xylem XA00288 EXO Calibration Worksheet Instruction Manual

June 15, 2024
xylem

xylem XA00288 EXO Calibration Worksheet Instruction Manual
xylem logo

This EXO Calibration Worksheet is designed to walk you through the calibration process for each smart sensor.
Please use the EXO User Manual and EXO University as resources for additional information and best practices.

Calibration Date Date of Last Cal Start Time End Time Performed By
Sonde Model Sonde Serial # Sonde Firmware Kor Version Connectors
Lubricated Batteries Replaced
Yes         / No Date:
Conductivity Sensor Serial # Sensor Firmware Temp Check Cond in Air
QC Score
Pass      / Fail
Standard Value Temperature Pre Cal Post Cal Lot#
Cal Point 1
Wiped Cell Constant (Range: 0.469/cm ± 0.05) Non-Wiped Cell Constant
(Range: 5.1/cm ± 10%)

Notes:
DO| Sensor Serial #| Sensor Firmware| Cap Serial #| Cap Replaced| | QC Score
| | | | Date:|
Barometric Pressure| Temperature| Pre Cal| Post Cal| Lot#| Expiration
Cal Point 1| | | | | |
Cal Point 2| | | | | |
DO Gain| | (Range: 0.85 – 1.15)|
Notes:
pH| Sensor Serial #| Sensor Firmware| Module Lot Code| Module Replaced| QC Score
| | | | Date:|
Standard Value| Temperature| Pre Cal| Post Cal| mV| Lot#| Expiration
Cal Point 1| | | | | | |
Cal Point 2| | | | | | |
Cal Point 3| | | | | | |
| 7 to 10 (negative Δ)| 7 to 4 (positive Δ)|
Delta Slope| | | (Range: 165−180 ± 50 mV)|
mV per Decade| | (Range: 55 – 60 pH/mV)|
Notes:
ORP| Sensor Serial #| Sensor Firmware| Module Lot Code| Module Replaced| QC Score
---|---|---|---|---|---
| | | | Date:|
Standard Value| Temperature| Pre Cal| Post Cal| Lot#| Expiration
Cal Point 1| | | | | |
| _
Use_ solution’s temperature chart to determine standard value.|
Notes:
Turbidity| Sensor Serial #| Sensor Firmware| | QC Score
| | |
Standard Value| Temperature| Pre Cal| Post Cal| Lot#| Expiration
Cal Point 1| | | | | |
Cal Point 2| | | | | |
Cal Point 3| | | | | |
Notes:
Total Algae| Sensor Serial #| Sensor Firmware| Standards Prepared| QC Score
| | | Date:|
Temp-Compensated Standard Value (RFU or µg/L)| Temperature| Pre Cal| Post Cal
Cal Point 1| | | |
Cal Point 2| | | |
Cal Point 1| | | |
Cal Point 2| | | |
_
Indicate_ PC or PE| *Use the table in the EXO Manual to determine temp- compensated values. Indicate RFU or µg/L.
Notes:
fDOM| Sensor Serial #| Sensor Firmware| Standards Prepared| QC Score
| | | Date:|
Temp-Compensated Standard Value (RFU or QSU)| Temperature| Pre Cal| Post Cal
Cal Point 1| | | |
Cal Point 2| | | |
| _
Use the table in the EXO Manual to determine temp-compensated values. Indicate RFU or QSU.
Notes:
NitraLED| Sensor Serial #| Sensor Firmware| | QC Score
| | |
Standard Value| Temperature| Pre Cal| Post Cal| Lot#| Expiration
Cal Point 1| | | | | |
Cal Point 2| | | | | |
| Unfiltered Turbidity| Filtered Turbidity| LED1 Coeff| LED2 Coeff| NOM Coeff
Site-Specific Correction| | | | |
|
(Range: <1 FNU)| (Range: -0.001 − -0.007)| (Range: -1.1 − -1.7)_
Notes:
Depth| Depth Offset| Pre Cal| Post Cal| QC Score
| | | |
Notes:
Nitrate ISE| Sensor Serial #| Sensor Firmware| Module Lot Code| Module Replaced| QC Score
---|---|---|---|---|---
| | | | Date:|
Standard Value| Temperature| Pre Cal| Post Cal| mV| Lot#| Expiration
Cal Point 1| | | | | | |
Cal Point 2| | | | | | |
Cal Point 3(Temp Standard)| | | | | | |
| 1 to 100 mg/L (negative Δ)|
Slope| | (Range: 90−130 mV)
Notes:
Ammonium ISE| Sensor Serial #| Sensor Firmware| Module Lot Code| Module Replaced| QC Score
| | | | Date:|
Standard Value| Temperature| Pre Cal| Post Cal| mV| Lot#| Expiration
Cal Point 1| | | | | | |
Cal Point 2| | | | | | |
Cal Point 3(Temp Standard)| | | | | | |
| 1 to 100 mg/L (positive Δ)|
Slope| | (Range: 90−130 mV)
Notes:
Chloride ISE| Sensor Serial #| SensorFirmware| Module LotCode| ModuleReplaced| Standards Prepared| QC Score
| | | | | Date:|
Standard Value| Temperature| Pre Cal| Post Cal| mV| Lot#| Expiration
Cal Point 1| | | | | | |
Cal Point 2| | | | | | |
Cal Point 3(Temp Standard)| | | | | | |
| 10 to 1000 mg/L (negative Δ)|
Slope| | (Range: 80−130 mV)
Notes:
Rhodamine| Sensor Serial #| Sensor Firmware| Standards Prepared| QC Score
| | | Date:|
Standard Value| Temperature| Pre Cal| Post Cal|
Cal Point 1| | | | |
Cal Point 2| | | | |
Cal Point 3| | | | |
Cal Point 4| | | | |
Notes:
Central Wiper| Sensor Serial #| Sensor Firmware| Wiper Brush Replaced| QC Score
| | | Date:|
Mode: Normal          / NitraLED| Pre Cal| Post Cal
| |
Notes:

Calibrate Your Sensors

Your data is only as good as your last calibration. Consult the User Manual for full calibration procedures. To reorder additional reagents, membranes and calibration solutions, you may reference the chart or link below:

pH Buffer : Available in different volumes
Conductivity Standard : Available in different volumes & concentrations
Turbidity Standard: Available in different concentrations
NitraLED Standard: Available in different concentrations
ORP Calibrator: Available in different volumes
Confidence Solution

Browse: Reagents, Membranes and Calibration Solutions

Additional Resources

Please reference the EXO User Manual or browse EXO University, our video library on topics such as calibration, maintenance, storage, and more!

User Manual Setting
EXO User Manual
YSI.com/EXO

YSI, a Xylem brand
1725 Brannum Lane
Yellow Springs, OH 45387
 +1.937.767.7241
 info@ysi.com
 YSI.com
YSI.com/EXO

References

Read User Manual Online (PDF format)

Read User Manual Online (PDF format)  >>

Download This Manual (PDF format)

Download this manual  >>

Related Manuals