Features

  • 8″x2″ compact size

  • Medical safety approved (2 x MOPP) according to
    ANSI/AAMI ES60601-1 and IEC/BS EN/EN60601-1

  • Suitable for BF application with appropriate system consideration

  • Cooling by free air convection

  • EMI class B for class II configuration

  • No load power consumption<0.1W

  • Extremely low leakage current

  • Protections: Short circuit / Overload / Over voltage

  • Operating altitude up to 4000 meters

  • 3 years warranty

Applications

  • Oralirrigator
  • Hemodialysis machine
  • Medical computer monitors
  • Sleep apnea devices

GTIN CODE

MW Search: https://www.meanwell.com/serviceGTIN.aspx

Description

RPS-65is a 65W highly reliable green PCB type medical power supply with a high power density on the 3″ by 2” footprint. It accepts 80~264VAC input and offers various output voltages between 3.3V and 48V.
The working efficiency is up to 91% and the extremely low no load power consumption is down below 0.1W. RPS-65is able to be used for Class Il (no FG) system design. The extremely low leakage currentis less than 100 ¢#A. In addition, it conforms to international medical regulations (2*MOPP) and EMC BS EN/EN55011, perfectly fitting all kinds of BF rated “patient contact” medical system equipment.

Model Encoding

MEAN WELL RPS-65 Series 65W Reliable Green Medical Power Supply -
fig1

SPECIFICATION

ORDER NO.| RPS653. 3| RPS655| RPS657. 5| RPS6512| RPS6515| RPS6524| RPS6548
---|---|---|---|---|---|---|---

OUTPUT

| DC VOLTAGE| 3.3V| 5V| 7.5V| 12V| 15V| 24V| 48V
RATED CURRENT| 10A| 10A| 8A| 5.42A| 4.34A| 2.71A| 1.36A
CURRENT RANGE| 0 ~ 11A| 0 ~ 11A| 0 ~ 8.8A| 0 ~ 5.96A| 0 ~ 4.77A| 0 ~ 2.98A| 0 ~ 1.49A
RATED POWER| 33W| 50W| 60W| 65W| 65.1W| 65W| 65.3W
PEAK LOAD(10sec.)| 36.3W| 55W| 66W| 71.5W| 71.6W| 71.5W| 71.5W
RIPPLE & NOISE (max.) **Note.2| 80mVp-p| 80mVp-p| 80mVp-p| 120mVp-p| 120mVp-p| 120mVp-p| 150mVp-p
VOLTAGE ADJ.RANGE| 2.9~3.6V| 4.7~5.5V| 7.12~8.3V| 11.4~13.2V| 13.5~16.5V| 22.8~27.6V| 45.6~52.8V
VOLTAGE TOLERANCE ****Note.3| ±2.0%| ±2.0%| ±2.0%| ±2.0%| ±1.0%| ±1.0%| ±1.0%
LINE REGULATION| ±0.5%| ±0.5%| ±0.5%| ±0.5%| ±0.5%| ±0.5%| ±0.5%
LOAD REGULATION| ±2.0%| ±2.0%| ±2.0%| ±2.0%| ±1.0%| ±1.0%| ±1.0%
SETUP, RISE TIME| 500ms, 30ms / 230VAC          500ms, 30ms / 115VAC at full load
HOLD UP TIME (Typ.)**| 30ms / 230VAC          12ms / 115VAC at full load

INPUT

| VOLTAGE RANGE Note. 4| 80 ~ 264VAC
FREQUENCY RANGE| 47 ~ 63Hz
EFFICIENCY (Typ.)| 80%| 84%| 85%| 88%| 89%| 90%| 91%
AC CURRENT (Typ.)| 1.5A / 115VAC         1A / 230VAC
INRUSH CURRENT (Typ.)| СOLD STAR 30A/115VAC  50A/230VAC
LEAKAGE CURRENT(max.) Note.5| Touch current< 100μA/264VAC

PROTECTION

|

OVERLOAD

| 115 ~ 150% rated output power
Protection type : Hiccup mode, recovers automatically after fault condition is removed

OVER VOLTAGE

| 3.8~4.5V| 5.7~6.8V| 8.6~11.3V| 13.8~16.2V| 17.2~20.3V| 27.6~32.4V| 55.2~64.8V
Protection type : Shut down o/p voltage, re-power on to recover

ENVIRONMENT

| WORKING TEMP.| -30 ~ +70℃ (Refer to “Derating Curve”)
WORKING HUMIDITY| 20% ~ 90% RH non-condensing
STORAGE TEMP., HUMIDITY| -40 ~ +85℃, 10 ~ 95% RH non-condensing
TEMP. COEFFICIENT| ±0.03% / ℃ (0 ~ 50℃)
VIBRATION| 10 ~ 500Hz, 2G 10min./1cycle, period for 60min. each along X, Y, Z axes
OPERATING ALTITUDE Note.6| 4000 meters

SAFETY & **EMC (Note. 7)**

|

SAFETY STANDARDS

| IEC60601-1, TUV BS EN/EN60601-1, EAC TP TC 004,UL ANSI / AAMI ES60601-1 (3.1 version),

CAN/CSA-C22.2 No. 60601-1:14 – Edition 3 approved; Design refer to BS EN/EN60335-1

ISOLATION LEVEL| Primary-Secondary: 2xMOPP
WITHSTAND VOLTAGE| I/P-O/P: 4KVAC
ISOLATION RESISTANCE| I/P-O/P:100M Ohms / 500VDC / 25℃/ 70% RH

EMC EMISSION

| Parameter| Standard| Test Level / Note
Conducted emission| BS EN/EN55011 (CISPR11)| Class B
Radiated emission| BS EN/EN55011 (CISPR11)| Class B
Harmonic current| BS EN/EN61000-3-2| Class A
Voltage flicker| BS EN/EN61000-3-3| —–

EMC IMMUNITY

| BS EN/EN55035,BS EN/EN60601-1-2
Parameter| Standard| Test Level / Note
ESD| BS EN/EN61000-4-2| Level 4, 15KV air ; Level 4, 8KV contact
RF field susceptibility| BS EN/EN61000-4-3| Level 3, 10V/m( 80MHz~2.7GHz )
Table 9, 9~28V/m( 385MHz~5.78GHz )
EFT bursts| BS EN/EN61000-4-4| Level 3, 2KV
Surge susceptibility| BS EN/EN61000-4-5| Level 4, 2KV/Line-Line
Conducted susceptibility| BS EN/EN61000-4-6| Level 3, 10V
Magnetic field immunity| BS EN/EN61000-4-8| Level 4, 30A/m
Voltage dip, interruption| BS EN/EN61000-4-11| 100% dip 1 periods, 30% dip 25 periods,
100% interruptions 250 periods

OTHERS

| MTBF| 3334.3K hrs min.       Telcordia SR-332 (Bellcore) ; 959.1K hrs min.        MIL-HDBK-217F (25℃)
DIMENSION (LWH)| 76.250.824mm or 3″ 2″ 0.945″ inch
PACKING| 0.11Kg; 120pcs/14.2Kg/0.94CUFT
NO| 1. All parameters NOT specially mentioned are measured at 230VAC input, rated load and 25℃ of ambient temperature.
2.  Ripple & noise are measured at 20MHz of bandwidth by using a 12″ twisted pair-wire terminated with a 0.1μf & 47μf parallel capacitor.
3.  Tolerance : includes set up tolerance, line regulation and load regulation.
4.  Derating may be needed under low input voltages. Please check the derating curve for more details.
5.  Touch current was measured from primary input to DC output.
6.  The ambient temperature derating of 3.5℃/1000m with fanless models and of 5℃/1000m with fan models for operating altitude higher than 2000m(6500ft).
7.  The power supply is considered a component which will be installed into a final equipment. The final equipment must be re-confirmed that it still meets EMC directives. For guidance on how to perform these EMC tests, please refer to “EMI testing of component power supplies.” (as available on http://www.meanwell.com)
※ Product Liability Disclaimer:For detailed information, please refer to https://www.meanwell.com/serviceDisclaimer.aspx

65W Reliable Green Medical Power Supply

  • Block DiagramMEAN WELL RPS-65 Series 65W Reliable Green Medical Power Supply - Block Diagram
  • Derating CurveMEAN WELL RPS-65 Series 65W Reliable Green Medical Power Supply - Derating Curve
  • Static Characteristics

Mechanical Specification

MEAN WELL RPS-65 Series 65W Reliable Green Medical Power Supply - Mechanical
Specification

AC Input Connector (CN1) : JST B3P-VH or equivalent

Pin No. Assignment Mating Housing Terminal
1 AC/N JST VHR or equivalent JST SVH-21T-P1.1 or equivalent
2 No Pin
3 AC/L

DC Output Connector (CN2) : JST B4P-VH or equivalent

Pin No. Assignment Mating Housing Terminal
1 +V JST VHR or equivalent JST SVH-21T-P1.1 or equivalent
2 +V
3 -V
4 -V