CAN-45854-XPCN1A Pressure Injectable Arrowg+ard Blue Plus Four-Lumen CVC
Kit Instruction Manual
Contents
- 1: Arrowg+ard Blue Plus®1 Four-Lumen Catheter: 8.5 Fr. (3.0 mm OD) x 20 cm, Pressure Injectable
- 1: Spring-Wire Guide, Nitinol, Marked: .032″ (0.81 mm) dia. x 23-5/8″ (60 cm) (Straight Soft Tip on One End – “J” Tip on Other) with Arrow® GlideWheel™ Wire Advancer
- 1: Catheter: 18 Ga. x 2-1/2″ (6.35 cm) Radiopaque over 20 Ga. RW Introducer Needle
- 1: Injection Needle: Eclipse™2 25 Ga. x 1-1/2″ (3.81 cm) RW
- 1: Injection Needle: Eclipse™2 22 Ga. x 1-1/2″ (3.81 cm) RW
- 1: Pressure Transduction Probe
- 1: Introducer Needle: Echogenic 18 Ga. x 2-1/2″ (6.35 cm) XTW and 5 mL Arrow® Raulerson Spring-Wire Introduction Syringe
- 1: Syringe: 3 mL Luer-Lock
- 1: Tissue Dilator: 10 Fr. (3.5 mm) x 10.2 cm
- 1: SecondSite™ Adjustable Hub: Fastener
- 1: SecondSite™ Adjustable Hub: Catheter Clamp
- 1: SharpsAway® Disposal Cup 1: SharpsAway® II Locking Disposal Cup
- 1: Maximal Barrier Drape™ with 4″ fenestration
- 1: Towel
- 1: Needle Holder
- 1: Forceps
- 1: Safety Scalpel: #11
- 1: Checklist/CLIP Sheet
- 1: Sterile Procedure Sign
±Precaution: Contains Phthalate: DEHP.
Warning: The results of certain animal experiments have shown phthalates
to be potentially toxic to reproduction. Proceeding from the present state of
scientific knowledge, risks for male premature infants cannot be excluded in
the case of long-term exposure or application. Medical products containing
phthalates should be used only temporarily with pregnant women, nursing
mothers, babies and infants. Not made with natural rubber latex. Store below
25°C (77°F). Avoid excessive heat above 40°C (104°F).
Contraindications: The Arrowg+ard Blue Plus antimicrobial catheter is
contraindicated for patients with known hypersensitivity to chlorhexidine,
silver sulfadiazine and/or sulfa drugs.
Lumen| Priming Volume* (mL)| Gravity Flow Rate† (mL/hr)| MAX Pressure
Injection Flow Rate** (mL/sec)
---|---|---|---
Distal (16 Ga.)| 0.40| 2283| 5
Medial 1 (14 Ga.)| 0.50| 4316| 10
Medial 2 (18 Ga.)| 0.36| 1079| not rated
Proximal (18 Ga.)| 0.39| 1162| not rated
Priming volumes are approximate and are done without accessories. † Flow rate
values are approximate and are determined using deionized water at 100 cm head
height.
Pressure injection flow rates are determined using media of 11.8 centipoise
(cP) viscosity, with 152 cm administration set tubing.