This operation manual is for the Mode VB-M003 microscope. We recommend that you carefully review this manual prior to operating this device in order to optimize safety and performance, and to become fully familiar with its use. Keep this operation manual in an easily accessible place for future reference. WARNING: To prevent fire and electrical shock, do not expose unit to rain or moisture. CAUTION: No user-serviceable parts inside. Refer servicing to qualified service personnel.

1. Before use

1-1. Operation
  1. As the microscope is a precision instrument, always use it with care. Avoid dropping or shaking during operation. Make sure the microscope is placed on a flat surface.

  2. Do not put the microscope in direct contact with the sun. Do not expose the microscope to high temperature, damp, or dusty environment.

  3. When moving the microscope, you should use both hands, typically one hand under the base and one hand on the arm (stand) of the microscope and set it down carefully.
    CAUTION: It will damage the microscope if you hold the microscope either by the stage, focusing knob or head.

  4. Be sure the microscope is unplugged before replacing the bulb or fuse and wait until the bulb has cooled down.

  5. This microscope has a built in voltage transformer and can be used with wide voltage range from 110 to 240V.

  6. Only use the power cord supplied with the microscope.

1-2 Maintenance
  1. Wipe the lens gently with a soft lens tissue. After every use with oil or fingerprints clean the lens surfaces with lens paper or lens paper moistened with a little rubbing alcohol. If you do not have lens paper make sure to use a lint free towel or cloth.
    CAUTION: Alcohol is flammable. Do not place this chemical near fire or fire causing source. You should use this chemical in a ventilated place.

  2. Don’t use organic solution to wipe the surfaces of the other components.

  3. If the microscope comes in contact with a liquid, then you should turn off the power immediately and wipe it dry.

  4. Never disassemble or service the microscope yourself.

  5. After using, cover the microscope with a dust cover.

2. Diagram

VB-M003 Microscope Diagram

Vabiooth VB-M003 Constant Temperature Stage Trinocular Instruction Manual -
Diagram

How to set the temperature?(Only for constant temperature version) Turn on the constant temperature heating button , press and hold for 3 seconds and then the font will flash. Then press the thermostat key “+” “-” to adjust the temperature, the adjustable range of temperature is 0-50 degree.

3. Assembly

3-1 Install the Objectives

Install the objective into the microscope nosepiece from the lowest magnification to the highest, in a clockwise direction.

3-2 Insert the Eyepieces

(1) Remove the eyepiece tube covers. (2) Insert the eyepiece into the tube completely.

3-3 Install plug

Insert one side of the plug into the back socket of the microscope.

4. Operation

4-1 Turn on
  1. Insert the plug into an outlet and turn on.
  2. Adjust the light adjustment knob until you get the suitable brightness. (See fig.1)
4-2 Set the specimen slide
  1. Push the arm of the specimen holder back.
  2. Insert the slide with the cover glass facing up and allow the arm of the specimen holder forward. (See fig.2)
  3. Rotate the X and Y-axis knobs on the mechanical stage to move the specimen to the center (the center of the objective).
4-3 Adjust focus
  1. Move the objective 4X into the light path.
  2. Rotate the coarse-focusing-knob until the image appears.
  3. Turn the fine-focusing-knob to make the specimen clearer. (See fig.3)
    • When using the 4X and 10X objective, open the aperture diaphragm and field diaphragm to the widest position.
4-4 Adjust tension of the focus

If the course focus knob is difficult to turn when focusing or the stage drifts downwards by itself, you can solve the problem by adjusting the tension adjustment ring. Turn clockwise to tighten the controls or counterclockwise to loosen the controls (See fig.4)

Use the filter
Filter can make the background softer. There is a filters: Frosted. (You also place the Polarizer in this position when using the optional simple polarizing set)

Vabiooth VB-M003 Constant Temperature Stage Trinocular Instruction Manual -
Fig.4

4-4Adjust diaphragm

The diaphragm can prevent other light from entering and strengthen the image by limiting the beam of light entering the condenser. When the image of the diaphragm is just on the edge of the field of view, you can obtain the clearest image.

4-5 Adjust aperture diaphragm

The aperture diaphragm decides the numerical aperture of the illumination. If the N.A. of illumination is matching
with N.A. of the objective, you can obtain better resolution and the contrast. If the contrast is low, you can adjust the N.A. of the condenser to 70%-80% of the N.A. of the objective when using. Adjust the upper diaphragm ring until you see the image like above (See fig.7). Use of diaphragm and put the N.A. of condenser to 80% of the objective. e.g.: objective 40X (N.A. =0.65), put the aperture diaphragm to about 0.65X0.8=0.52. (See fig.8)

4-6 Use the oil objective
  1. Use the objective 40X to focus the specimen.

  2. Put a drop of oil on the specimen. (See fig.9)

  3. Rotate the nosepiece to place the 100X objective (oil) into the light path. Then use the fine focusing knob to focus.
    Make sure there is no air bubble in the oil.
    A. Take out the eyepieces to examine the air bubble. Open the aperture diaphragm and field diaphragm fully and observe the edge of the objective from the tube without the eyepiece.
    B. You can rotate nosepiece slightly and swing the oil objective some times to remove the air bubble.

  4. After using oil, clean the 100X lens with clean lens paper using rubbing alcohol if the oil does not come off easily.

5. Specification

5-1 VB-M003 Microscope general specifications

Vabiooth VB-M003 Constant Temperature Stage Trinocular Instruction Manual -
VB-M003 Microscope general specifications

6. Trouble shooting

Vabiooth VB-M003 Constant Temperature Stage Trinocular Instruction Manual -
Troubleshooting Vabiooth VB-M003 Constant Temperature Stage
Trinocular Instruction Manual - Troubleshooting

7. How to connect the monitor?

Vabiooth VB-M003 Constant Temperature Stage Trinocular Instruction Manual -
How to connect the monito

  1. Install and connect the monitor according to the picture above, then turn on the power button
  2. Select AV1 channel
  3. Monitor “M” can adjust BRIGHTNESS, CONTRAST, COLOUR, ZOOM TCON, LANGUAGE (“Vabiooth” YouTube channel has Instructional video)

8. How to take still photos and videos?

If you need to take still shots and videos, please contact Vabiooth@outlook.com, we will send you a install document and device to make the monitor get this function for free.
Instructional Video link: https://bit.ly/3FYu9aS

Vabiooth VB-M003 Constant Temperature Stage Trinocular Instruction Manual -
How to take still photos and videos

Vabiooth provides 18 months warranty and lifetime free service.
If you encounter any issues with the product, Please contact the Vabiooth team to promptly receive troubleshooting tips or a replacement.

Contact us

Vabiooth@outlook.com
+86 17306342950
Vabiooth (How-to videos, warranty details)

FAQ

  1. How to install the microscope?
    Please refer to our user manual. In addition, we have recorded an installation video for each machine. Please search for “Vabiooth” on youtube to watch.

  2. What are the options for eyepieces?
    To provide a better customer experience, all models’eyepieces have been upgraded to WF25X eyepieces, and if you need WFlOX, WF16X, 2X Teleconverter, Please contact us.

  3. How do I prevent dirt and grime from getting into the eyepieces and objectives?
    Remind the user to use the Vabiooth Microscope Cover to keep your microscope covered all the time when not in use. This will help keep the eyepieces and objectives from water and dust. If you need the microscope cover, please contact us.

  4. Why are your machines divided into so many versions?
    We want to improve the displays color restoration & clarity ability to show perfect imaging for different applications, our microscopes are mainly divided into blood version, constant temperature version, general version (Different versions use different chips). The blood version is used to observe blood cells. The constant temperature version is suitable for breeding and farms to artificial insemination {35-36 degrees is the most active temperature for sperm), the universal version is suitable for experimental research, veterinarians to observe pond organisms, parasites, and soil.