PCE INSTRUMENTS PCE-423N Hot Wire Anemometer User Manual
- June 14, 2024
- PCE Instruments
Table of Contents
User Manual
PCE-423N Heat Wire Anemometer User manuals in various languageshttp://www
.pce-instruments.com
Safety notes
Please read this manual carefully and completely before you use the device for
the first time. The device may only be used by qualified personnel and
repaired by PCE Instruments personnel.
Damage or injuries caused by non-observance of the manual are excluded from
our liability and not covered by our warranty.
- The device must only be used as described in this instruction manual. If used otherwise, this can cause dangerous situations for the user and damage to the meter.
- The instrument may only be used if the environmental conditions (temperature, relative humidity, …) are within the ranges stated in the technical specifications. Do not expose the device to extreme temperatures, direct sunlight, extreme humidity or moisture.
- Do not expose the device to shocks or strong vibrations.
- The case should only be opened by qualified PCE Instruments personnel.
- Never use the instrument when your hands are wet.
- You must not make any technical changes to the device.
- The appliance should only be cleaned with a damp cloth. Use only pH-neutral cleaner, no abrasives or solvents.
- The device must only be used with accessories from PCE Instruments or equivalent.
- Before each use, inspect the case for visible damage. If any damage is visible, do not use the device.
- Do not use the instrument in explosive atmospheres.
- The measurement range as stated in the specifications must not be exceeded under any circumstances.
- Non-observance of the safety notes can cause damage to the device and injuries to the user.
- Do not touch the hot wire.
We do not assume liability for printing errors or any other mistakes in this
manual.
We expressly point to our general guarantee terms which can be found in our
general terms of business.
Specifications
Velocity
Measurement range| 0.00 … 30.00 m/s 0.0 … 5905.5 ft/min
Resolution| 0.01 m/s 0.1 ft/min
Accuracy the higher value applies| ±0.10 m/s or ±5 % of rdg. at 0 … 5 m/s
±0.30 m/s or ±5 % of rdg. at 5 … 30 m/s
±19.6 ft/min or ±5 % of rdg. at 0 … 984.2 ft/min ±59.0 ft/min or ±5 % of rdg.
at 984.2 … 5905.5 ft/min
Flow rate
Measurement range| 0.0 … 339120.0 m3 /h 0.0 … 99598.0 CFM
Resolution| 0.1 m3 /h 0.1 CFM
Temperature
Measurement range| 0 … 50 °C
32 … 122 °F
Resolution| 0.1 °C
0.1 °F
Accuracy| ±1 °C ±1.8 °F
Further specification
Telescopic probe length| 210 … 426 mm / 8.2 … 16.7″
Telescopic probe diameter| 12 mm / 0.47″
Memory space| 10,000 measuring points
File format| CSV
Interface| micro USB
Bracket| 3/8″ tripod bracket
Display| 2″ LCD, 176 x 220 pixels
Measurement rate| 1 Hz
Power supply (rechargeable battery)| internal rechargeable 1000 mAh battery
Power supply (external)| 5 V DC, 1 A
Operating conditions| 0 … 50 °C (32 … 122 °F), <85 % RH, non-condensing
Storage conditions| -10 … 60 °C (14 … 140 °F), <85 % RH, non-condensing
Dimensions
(without telescopic probe)| 145 x 50 x 35 mm / 5.7 x 1.9 x 1.3″
Weight| 156 g / 0.34 lbs
Delivery scope
1 x hot wire anemometer PCE-423N
1 x micro USB cable
1 x user manual
Device description
No. | Meaning / function |
---|---|
1 | flow sensor, temperature sensor |
2 | display |
3 | left function key |
4 | right function key |
5 | on/off key |
6 | Enter-V/F key, toggle key (velocity. flow rate) |
7 | back key |
8 | air slits |
9 | tripod bracket |
10 | micro USB connector |
Switching the meter on and off
To switch the meter on or off, press and hold the key.
Measurement preparation
Before making a measurement, slide the probe guard down. Switch on the device. Point the opening of the sensor in the direction of the flow to be measured. When you have finished your measurement, push the guard back up so that the sensor system is not damaged.
Measuring mode
7.1 Main display
After the meter has been switched on, the velocity measurement is carried out
immediately. At the same time, the current temperature is measured and
displayed. To select between the velocity and the flow rate measurement, press
and release the enter-V/F key once. The “V” on the display stands for
velocity and the “F” for flow. The flow rate depends on the velocity and the
set area. You can set the area under 8.2 Flow rate setting to set the area.
Note: Saving individual measured values is not possible in normal
measuring mode:
7.2 Average measurement (time-dependent)
To display the average, press the right function key in measuring mode to
select . The display now shows for the average reading (time-dependent) and .
Now press the left function key to select . The current velocity, the average,
the highest and the lowest measured value are displayed here. Press the left
function key to save the displayed values . Press the right function key to
start or pause the current average measurement . To repeat the average
measurement, go to normal measurement mode with the back key and start the
average measurement again.
7.3 Average measurement (different points)
To determine the average of different measuring points, press the right
function key in measuring mode.
Press the right function key again to select . To save the displayed values,
now press the left function key . The values are saved once per keystroke. To
include the current measuring point in the average measurement and select the
next one, press the right function key . To repeat the average measurement, go
back to normal measuring mode with the back key and start the average
measurement again.
Settings
To make settings, press the left function key in measuring mode.
Use the function keys to select between the individual functions displayed.
At the same time, you can change selected parameters with the function keys.
To open a setting, press the enter-V/F key. Press the back key to go back one
level. Press the enter-V/F key to confirm your entry.
Setting | Meaning |
---|---|
Unit | Set the unit |
Area / Funnel | Enter the dimensions for the flow rate |
Auto-off | Set the automatic power off |
History | Read memory |
Brightness | Show brightness of the display |
Date / Time | Set date and time |
Calibration | Calibrate flow |
8.1 Set unit
To set the unit for the flow and the temperature, go to the Settings menu and
to the Unit menu item. Here, you can select between the temperature or the
flow rate. Then you can change the unit.
Function | Units |
---|---|
Flow rate | m/s, m³/h / ft/min, CFM |
Temperature | °C, °F |
8.2 Flow rate setting
To specify the dimensions for the flow rate, go to the settings and select the
menu item “Area/Funnel”. Here, you can specify four different dimensions for
square as well as round areas. In addition, a cylindrical area can be
specified. The last selected area is adopted in measuring mode.
8.3 Automatic power off
Via the automatic power off function, the meter can be set to switch itself
off after the set time. The following setting options are available here:
“NONE” (function deactivated), 2 minutes, 10 minutes, 1 hour.
8.4 View saved data
To view the saved data, go to the settings and select the item “History”. All
saved measurement data can be viewed there. It is not possible to delete or
edit the data on the meter itself.
Note: If no data has been saved, this item cannot be opened.
8.4.1 Transfer data to the computer
To transfer the measurement data to a computer, the meter must first be
switched off. Now connect the meter to a computer which recognises it as a
mass data storage device. The data can now be found in CSV data format.
8.5 Display brightness
To adjust the display brightness, go to the settings and open the item
“Brightness”. You can now use the function keys to adjust the brightness of
the display.
8.6 Date and time
To set the date and time, go to the settings and open the menu item “Data/Time
Setup”. Here, the date and time can be set in the format YYYY/MM/DD. You can
also choose between English and German style here. Use the function keys to
change the parameters as desired. Use the enter-V/F key to accept the value
and jump to the next item at the same time.
8.7 Calibration
To calibrate the flow rate, go to the “Calibration” item in the settings.
Herethe displayed measured value can be set using a factor (“Gain”).
Thedefault value is 1.00.
The coefficient range is between 0.5 and 2.
Contact
If you have any questions, suggestions or technical problems, please do not hesitate to contact us. You will find the relevant contact information at the end of this user manual.
Disposal
For the disposal of batteries in the EU, the 2006/66/EC directive of the
European Parliament applies. Due to the contained pollutants, batteries must
not be disposed of as household waste.
They must be given to collection points designed for that purpose.
In order to comply with the EU directive 2012/19/EU we take our devices back.
We either re-use them or give them to a recycling company which disposes of
the devices in line with law.
For countries outside the EU, batteries and devices should be disposed of in
accordance with your local waste regulations.
If you have any questions, please contact PCE Instruments.
PCE Instruments contact information
Germany
PCE Deutschland GmbH
Im Langel 26
D-59872 Meschede
Deutschland
Tel.: +49 (0) 2903 976 99 0
Fax: +49 (0) 2903 976 99 29
info@pce-instruments.com
www.pce-instruments.com/deutsch|
France
PCE Instruments France EURL
23, rue de Strasbourg
67250 Soultz-Sous-Forets
France
Téléphone: +33 (0) 972 3537 17
Numéro de fax: +33 (0) 972 3537 18
info@pce-france.fr
www.pce-instruments.com/french| Spain
PCE Ibérica S.L.
Calle Mayor, 53
02500 Tobarra (Albacete)
España
Tel. : +34 967 543 548
Fax: +34 967 543 542
info@pce-iberica.es
www.pce-instruments.com/espanol
---|---|---
United Kingdom
PCE Instruments UK Ltd
Unit 11 Southpoint Business Park
Ensign Way, Southampton
Hampshire
United Kingdom, SO31 4RF
Tel: +44 (0) 2380 98703 0
Fax: +44 (0) 2380 98703 9
info@pce-instruments.co.uk
www.pce-instruments.com/english|
Italy
PCE Italia s.r.l.
Via Pesciatina 878 / B-Interno 6
55010 Loc. Gragnano
Capannori (Lucca)
Italia
Telefono: +39 0583 975 114
Fax: +39 0583 974 824
info@pce-italia.it
www.pce-instruments.com/italiano|
Turkey
PCE Teknik Cihazları Ltd.Şti.
Halkalı Merkez Mah.
Pehlivan Sok. No.6/C
34303 Küçükçekmece – İstanbul
Türkiye
Tel: 0212 471 11 47
Faks: 0212 705 53 93
info@pce-cihazlari.com.tr
www.pce-instruments.com/turkish
The Netherlands
PCE Brookhuis B.V.
Institutenweg 15
7521 PH Enschede
Nederland
Telefoon: +31 (0)53 737 01 92
info@pcebenelux.nl
www.pce-instruments.com/dutch| United
States of America
PCE Americas Inc.
1201 Jupiter Park Drive, Suite 8
Jupiter / Palm Beach
33458 FL
USA
Tel: +1 561-320-9162
Fax: +1 561-320-9176
info@pce-americas.com
www.pce-instruments.com/us| Denmark
PCE Instruments Denmark ApS
Birk Centerpark 40
7400 Herning
Denmark
Tel.: +45 70 30 53 08
kontakt@pce-instruments.com
www.pce-instruments.com/dansk
© PCE Instruments
References
- France.fr : Explore France and its wonders - Explore France
- iberica.es
- Make an offer on the domain instruments.co.uk - Domains.co.uk
- Computer Instruments | Home
- Discover Italy: Official Tourism Website - Italia.it
- PCE Teknik Cihazlar Paz. Tic. Ltd.Şti. | PCE Instruments
- Industrial Measurement Products and Solutions | PCE Instruments
Read User Manual Online (PDF format)
Read User Manual Online (PDF format) >>