ARROW EU-25052-IR Pressure Injectable Two-Lumen PICC Instruction Manual
- June 4, 2024
- Arrow
Table of Contents
ARROW EU-25052-IR Pressure Injectable Two-Lumen PICC
Contents
- 1: Two Lumen Catheter: 5 Fr. (1.81 mm OD) x 50 cm, Pressure Injectable, Blue FlexTip®
- 1: GlideThruTM Peel-Away Sheath: 5 Fr. x 2-3/4″ (7 cm) Radiopaque over 5 Fr. Dilator
- 1: Spring-Wire Guide: .018″ (0.46 mm) dia. x 13-1/8″ (33 cm) (Straight Soft Tip on Both Ends)
- 1: Spring-Wire Guide, Marked: .018″ (0.46 mm) dia. x 51-3/16″ (130 cm) (Straight Soft Tip on One End – Straight Stiff Tip on Other)
- 1: Introducer Needle: Echogenic 21 Ga. x 2-3/4″ (7 cm) TW
- 1: Syringe: 10 mL Luer-Lock 1: SecondSiteTM Adjustable Hub: Fastener
- 1: SecondSiteTM Adjustable Hub: Catheter Clamp
- 1: SharpsAway® II Locking Disposal Cup 1: Catheter Trimmer
- 1: Safety Scalpel: #11 1: Patient ID Card 1: Patient Information Booklet
- 1: Chart Sticker 1: Paper Tape Measure 1: Tourniquet
- 1: Dressing: GRIP-LOK®1 Catheter Fixation Device
- 2: Valve: MicroCLAVE®2 Neutral Displacement Connector 1A registered trademark of TIDI.
- 2A registered trademark of ICU Medical, Inc. All components are CE 2797 unless otherwise noted.
Warning: Read all package insert warnings, precautions, and instructions
prior to use. Failure to do so may result in severe patient injury or death.
www.teleflex.com/IFU Not made with natural rubber latex. Fluid path components
are non-pyrogenic.
Contraindications: The Pressure Injectable PICC is contraindicated
wherever there is presence of device related infections or presence of
thrombosis in the intended insertion vessel or catheter pathway. Clinical
assessment of the patient must be completed to ensure no contraindications
exist.
| Priming
**Volume***
| Gravity Flow
Rate†
| Pump
Flow Rate††
| MAX Pressure Injection
Flow Rate**
---|---|---|---|---
Lumen| (mL)| (mL/hr)| (mL/hr)| (mL/sec)
Distal (18 Ga.)| 0.4| 356| 3106| 4
Proximal (18 Ga.)| 0.43| 396| 3490| 4
- Priming volumes are approximate and are done without accessories.
- Flow rate values are approximate and are determined using deionized water at 100 cm head height. Pump flow rates are determined at maximum pump pressure of 10 psig and represent approximate flow capabilities.
- Pressure injection flow rates are determined at the injector pressure setting of 300 psi maximum using media of 11.8 centipoise viscosity, with 152 cm pressure tubing.
EU Authorized Representative and Importer: Teleflex Medical IDA Business and Technology Park Dublin Road, Athlone Co. Westmeath, Ireland Arrow International LLC Subsidiary of Teleflex Incorporated 3015 Carrington Mill Blvd. Morrisville, NC 27560 USA\ LBL056812 R00 ( 2021-06)
Documents / Resources
|
ARROW EU-25052-IR Pressure Injectable Two-Lumen
PICC
[pdf] Instruction Manual
EU-25052-IR Pressure Injectable Two-Lumen PICC, EU-25052-IR, Pressure
Injectable Two-Lumen PICC, Injectable Two-Lumen PICC, Two-Lumen PICC, PICC
---|---
References
Read User Manual Online (PDF format)