ARROW EU-25552-HP Pressure Injectable Two-Lumen PICC Instruction Manual
- June 4, 2024
- Arrow
Table of Contents
EU-25552-HP Pressure Injectable Two-Lumen PICC
Instruction Manual
Pressure Injectable Two-Lumen PICC
Contents:
1: Two Lumen Catheter: 5 Fr. (1.81 mm OD) x 55 cm, Pressure Injectable, TPort
Connector, Blue FlexTip®, Contamination Guard and Placement Wire
1: GlideThruTM Peel-Away Sheath: 5 Fr. x 2-3/4″ (7 cm) Radiopaque over 5 Fr.
Dilator
1: Spring-Wire Guide, Nitinol, Marked: .018″ (0.46 mm) dia. x 17-3/4″ (45 cm)
(Straight Soft Tip on One End – Straight Stiff Tip on Other) with Arrow
Advancer
1: Introducer Needle: Echogenic 21 Ga. x 2-3/4″ (7 cm) TW 1: Syringe: 10 mL
Luer-Lock
1: SecondSiteTM Adjustable Hub: Fastener
1: SecondSiteTM Adjustable Hub: Catheter Clamp
1: SharpsAway® II Locking Disposal Cup
1: Catheter Trimmer
1: Safety Scalpel: #11
1: Patient ID Card
1: Patient Information Booklet
1: Chart Sticker
1: Paper Tape Measure
1: Tourniquet
1: Dressing: GRIP-LOK®1 Catheter Fixation Device
2: Valve: MicroCLAVE®2 Neutral Displacement Connector
1A registered trademark of TIDI.
2A registered trademark of ICU Medical, Inc.
All components are CE 2797 unless otherwise noted.
Warning: Read all package insert warnings, precautions, and instructions
prior to use. Failure to do so may result in severe patient injury or death.
www.teleflex.com/IFU Not made with natural
rubber latex. Fluid path components are non-pyrogenic.
Contraindications: The Pressure Injectable PICC is contraindicated
wherever there is the presence of device-related infections or the presence of
thrombosis in the intended insertion vessel or catheter pathway. A clinical
assessment of the patient must be completed to ensure no contraindications
exist.
Lumen| Priming Volume*
(mL)| Gravity Flow Rate
(mL/hr)| Pump Flow Rate
(mL/hr)| MAX Pressure Injection
Flow Rate** (mL/sec)
---|---|---|---|---
Distal (18 Ga.)
Proximal (18 Ga.)| 0.48
0.51| 356
356| 2934
3317| 4
4
- Priming volumes are approximate and are done without accessories. Flow rate values are approximate and are determined using deionized water at 100 cm head height. Pump flow rates are determined at a maximum pump pressure of 10 PSIG and represent approximate flow capabilities.
Pressure injection flow rates are determined at the injector pressure setting of 300 psi maximum using media of 11.8 centipoise viscosity, with 152 cm pressure tubing.
Pressure Injectable Two-Lumen PICC**
EU Authorized Representative and Importer:
Teleflex Medical IDA Business and
Technology Park Dublin Road, Athlone
2797 Co. Westmeath, Ireland
Arrow International LLC
A subsidiary of Teleflex Incorporated
3015 Carrington Mill Blvd.
Morrisville, NC 27560 USA
www.teleflex.com
LBL056484 R00 (2021-06)
References
Read User Manual Online (PDF format)
Read User Manual Online (PDF format) >>