MTTS Colibri Phototherapy
Introduction
Phototherapy is one of the simplest and safest methods to treat jaundice in
neonatal intensive care. Untreated jaundice leads to an excessive level of
bilirubin in the blood which may potentially result in brain damage or death.
What does phototherapy do?
Reduces the bilirubin level in the blood through Photo-oxidation
Configurational isomerization Structural isomerization Phototherapy starts
medical treatment immediately after it’s turned on. It emits very little heat
because of the LED light source, so does not cause dehydration. Phototherapy
with Colibri Colibri Phototherapy equips you to deliver high-intensity
phototherapy to treat this potentially hazardous condition.
Device Description
Overview
Control Panel
1| On/Off Button| Press this button to turn Colibri Phototherapy display and
functions on and off.
---|---|---
2| Therapy Mode Button| Press this button to switch between standard and
intensive therapy mode.
3| Display| Shows therapy mode and treatment time.
Warnings
-
Using the device outside of the specified ambient temperature range
(10-35°C) or humidity range (0-90%RH) can compromise performance. -
Do not operate the device if any of the components appear to be damaged or broken. Damaged or broken components should be discarded and replaced.
-
Do not position the device in a location that impairs its ventilation.
-
Do not cover the device with blankets during operation.
-
The presence of electro-surgical devices, shortwaves, or microwave equipment near this device could lead to electrical interference that negatively affects the function of the device.
-
Do not modify this equipment without the authorization of the manufacturer.
-
LED lifetime warning: The blue LEDs in Colibri Phototherapy are rated to provide effective treatment for 60,000 hours.
If the Colibri Phototherapy has been used for over 60,000 hours, after turning ON the display will show the “total device hours” for 3 seconds and ‘Error 4’ message after that. The error message will continue to appear every second during treatment.
Note: The estimated 60,000 hour life span is equal to a full 7 years of non- stop use. After this period of time the LED phototherapy lights will continue to function, however treatment will be significantly less effective and Colibri Phototherapy should be replaced.
Setting up Phototherapy
Step 1 – Remove components from the box
Box contains:
- Stand
- Base
- canopy set
Step 2 – Stand assembly preparation
- Remove the protection label.
- 2-3. Unscrew slowly height adjustment knob.
While unscrewing the knob secure inner pole with hand in order to prevent uncontrolled releasing. 4. Remove height adjustment knob.
Step 3 – Base assembly
- 1. Insert the stand into the base.
- Screw in 3 bolts into the thread hole.
- 3-4. Connect and secure DC connector.
- Tighten the bolts using wrench (size 13).
Step 4 – Stand assembly
- Insert stand support ring.
- Screw in the height adjustment knob.
Step 5 – Canopy set assembly
- Pull out a plastic bracket with a DC connector from inside of the stand.
- Connect it with the canopy set wire.
- 3-4. Insert the connected wire back into the stand and cover it with the plastic bracket.
- Attach canopy set to the stand and secure it with screws.
2 people may be required to perform operations 2-4.
Step 6 – Plug in and power up
The connector plugs into the back of the Colibri Phototherapy base and then
directly into the power outlet in the wall.
- Only use power cords and connectors provided by MTTS.
- Colibri Phototherapy should only be used with power outlets that have a protective earth (ground) pin.
Using Phototherapy
Starting and stopping the therapy
- Plug in Colibri Phototherapy to the power outlet. Turn on the device by pressing the “On/Off Button”. The display shows the “total device hours”. This will be displayed for 3 seconds.
- The blue phototherapy lights will now start to shine in STANDARD MODE. The display will start counting the “treatment time” in STANDARD MODE. This example shows 1 minute of STANDARD therapy.
- The phototherapy treatment will continue until the doctor decides therapy is no longer necessary and the device can be switched off by pressing the “On/Off Button”. This example shows 72 hours and 30 minutes of STANDARD therapy. This example shows the OFF message.
Selecting therapy mode
Colibri Phototherapy can be operated in 2 modes:
- STANDARD: ~33 µW/cm2/nm
- INTENSIVE: ~57 µW/cm2/nm
- To select the mode press the “Therapy Mode Button” Colibri Phototherapy starts with STANDARD MODE once it is turned on. Pressing the “Therapy Mode Button” will switch it to INTENSIVE MODE. STANDARD MODE is indicated by the letter “n” on the display (“n” for “normal”) INTENSIVE MODE is indicated by the letter “b” on the display (“b” for “boost”) The display shows the “total device hours of INTENSIVE MODE”. This will be displayed for 3 seconds.
- The blue phototherapy lights will now start to shine in INTENSIVE MODE. The display will start counting the “treatment time” in INTENSIVE MODE. This example shows 1 minute of INTENSIVE therapy.
- The phototherapy treatment will continue until the doctor decides therapy is no longer necessary and the device can be switched off by pressing the “On/Off Button”. Pressing the “Therapy Mode Button” will switch back the device to STANDARD MODE and restart the treatment counter. This example shows 13 hours and 20 minutes of INTENSIVE therapy. If there is a power interruption, when the power is restored the device will return to the mode it was in before the interruption of power occurred.
Height adjustment
Colibri Phototherapy can be moved up and down in order to select preferred dis-tance from the light source to the patient. The height of the device is between 115 and 140 cm.
- Recommended distance from the light source to the patient is 30-40 cm.
- Exercise caution when adjusting the height of the Colibri Phototherapy.
Battery backup (optional)
Battery Backup
The battery backup allows operating the device without an AC power supply:
- 3 hours in STANDARD MODE
- 2 hours in INTENSIVE MODE
The backup activates automatically as soon as the AC power supply is off. To check the battery charge level and obtain additional information, follow the steps below:
Low battery alarm When the lowest cell voltage reaches 3V, the device informs about low batteries. “Lo bAt’ and audible alarm lasts 10 seconds after which the devices turns off by it-self.
Safety Information
- Exercise caution when positioning the Colibri Phototherapy around the infant bed.
- Always lock the caster wheels of the Colibri Phototherapy before starting the treatment.
- In the event of a power outage, eye coverings should be left on the patient whenever they are under Colibri Phototherapy to protect the patient’s eyes in case the power returns.
- Always cover the patient’s eyes with a shield or patch whenever the patient’s eyes can be exposed to Colibri Phototherapy light. Periodically, you may want to ensure that the patient’s eyes are protected and infection free.
- Patients’ bilirubin values should be measured regularly while undergoing phototherapy treatment.
- To accurately assess the patient’s skin color, turn off the light when checking the patient’s condition.
- Regularly monitor the patient’s temperature and fluid status.
- Colibri Phototherapy should be disposed of separately from normal household waste so that components can be recycled.
- DO NOT disassemble any part of the Colibri Phototherapy. None of the parts of the device are designated to be repaired or replaced by service personnel. In case of any problems, contact the nearest MTTS representative.
- DO NOT use non-standard power cords with Colibri Phototherapy. Using non-standard power cords can result in damaging the device.
- DO NOT place Colibri Phototherapy in direct sunlight.
- Colibri Phototherapy should be used only by appropriately trained personnel and under the direction of qualified medical personnel familiar with currently known risks and benefits of infant phototherapy equipment use.
- Patients adjacent to the Colibri Phototherapy may need to be protected with protective shields or eye protection.
- The operator may experience some effects during prolonged exposure to the area irradiated by the Colibri Phototherapy.*
- Blue light can hinder clinical observations by masking skin color changes, such as cyanosis.
- An infant’s water balance may be influenced by phototherapy. Regular feeding times may need to occur more often in order to prevent dehydration.
- Varying ambient conditions (ambient temperature and/or different radiation sources) may negatively affect the patient. Refer to your hospital phototherapy policy and procedure regarding appropriate ambient conditions.
- Bilirubin photoisomers may be toxic.
- Do not use Colibri Phototherapy if it appears that any part of the device is not functioning properly or if any parts appear damaged. Contact MTTS.
- In case of any malfunction DO NOT attempt to service this machine.
- Colibri Photothrapy should only be serviced by a qualified MTTS technician.
- Due to photo effects, drugs and infusion liquids should not be stored in the radiation area.
- Do not use the in the presence of potentially combustible gases including oxygen, nitrous oxide, or other anesthetics.
- Sensitive individuals may experience headache, nausea, or mild vertigo if he/she stays too long in the illuminated area. Using Colibri Phototherapy in a well-lighted area or wearing glasses with yellow lenses can alleviate potential effects. Guard Dog Bones glasses (p/n 413BB) are recommended and are available online at www.safetyglasses.com.
Cleaning and disinfection
- If visibly soiled, clean the affected parts with soap and water using a soft cloth. Then wipe with a wet cloth and dry with a soft cloth. For difficult areas use a standard soft-bristled brush.
- To loosen heavy, dried-on soil, you may first need to saturate the spot with water.
- When cleaning the device disconnect the power cord.
Irradiance measurement
Note: To ensure that the device is functioning properly, regularly
monitor Colibri Phototherapy in the following way.
Measurement instructions:
To test the irradiance of Colibri Phototherapy, measure using a light meter
specifically designed to measure wavelengths for phototherapy.
A stabilization time of 2 hours of continuous operation is required for
accurate measuring of the irradiance. When measuring the irradiance, prop up
the light meter 30-40 cm from the top of the canopies to achieve the
irradiance seen in the pictures below. Measure the irradiance of Colibri
Phototherapy in each of the sections seen in the pictures below. The lights
will decay gradually over time, reaching approximately 70% of their original
irradiance after the Colibri Phototherapy device has been used for 60,000
hours. After 60,000 hours of use, the light source will start to degrade
rapidly. It is recommended that the device be replaced after 60,000 hours of
use. If you must keep using the device after 60,000 hours of use, you should
increase the frequency at which you measure the irradiance of the device
.
Specifications
Performance Specifications
Standards for Reference
- EN ISO 13485:2016
- EN ISO 14971:2012
- EN ISO 15223-1:2016
- EN1041:2008
- EN 60601-1:2006 / A1:2013
- EN 60601-1-2:2015
- EN 60601-1-6:2010
- IEC 60601-2-50:2009
- MEDDEV 2.12-1 Rev. 8
- ISO 2244:2000
- MEDDEV. 2.7.1 Rev.4
- MEDDEV. 2.12-2 Rev.2
- ROHS 2015/863/EU
- WEEE 2012/19/EU
Explanation of Symbols
- This statement provides important information or highlights information that may be easily overlooked.
- This statement is a warning. Not following this statement could result in injury to the patient or operator, or damage to the device.
- CE Marking with Notified Body Number.
- Do not disassemble the device unless you are an MTTS trained technician or have been instructed to by qualified personnel (exclud-ing the removal of the bassinet).
- Refer to the user manual before operating this device.
- Refer to the user manual before operating this device.
- This device contains electronic parts. Do not dispose of it with normal waste. Dispose of it according to local guidelines for disposal of eletronics. Dispose of according to the WEEE directive in the European Union.
- Medical device
- Unique identification number
- A website where a user can obtain additional
- information on the medical product
- Model number
- Catalog number
- Name and address of European Authorized Representative.
- Device manufacturer.
- Date of manufacture.
- Serial number.
- An eye patch should always be placed on the patient before commencing treatment.
- Keep out of direct sunlight.
- Liquid ingress protection – level 1 (dripping water).
- Country of manufacture
Warranty Policy
General Terms
This MTTS Limited Warranty gives you, the customer, express limited
warranty rights from MTTS, the manu- facturer for the duration specified on
the Warranty Card. Please refer to the MTTS Website for an extensive de-
scription of your limited warranty entitlements. In addi- tion, you may also
have other legal rights under applica- ble law or special written agreement
with MTTS MTTS MAKES NO OTHER EXPRESS WARRANTY OR CONDITION WHETHER WRITTEN OR
ORAL AND MTTS EXPRESSLY DISCLAIMS ALL WARRANTIES AND CON- DITIONS NOT STATED
IN THIS LIMITED WARRANTY. TO THE EXTENT ALLOWED BY LOCAL LAW OF JURIS-
DICTIONS OUTSIDE VIETNAM, MTTS DISCLAIMS ALL IMPLIED WARRANTIES OR CONDITIONS,
INCLUDING ANY IMPLIED WARRANTIES OF MERCHANTABIL- ITY AND FITNESS FOR A
PARTICULAR PURPOSE. FOR ALL TRANSACTIONS OCCURRING IN VIETNAM ANY IMPLIED
WARRANTY OR CONDITION OF MERCHANT- ABILITY, SATISFACTORY QUALITY, OR FITNESS
FOR A PARTICULAR PURPOSE IS LIMITED TO THE DURATION OF THE EXPRESS WARRANTY
SET FORTH ABOVE. SOME COUNTRIES DO NOT ALLOW A LIMITATION ON HOW LONG AN
IMPLIED WARRANTY LASTS OR THE EXCLU- SION OR LIMITATION OF INCIDENTAL OR
CONSEQUEN- TIAL DAMAGES FOR CONSUMER PRODUCTS. IN SUCH COUNTRIES, SOME
EXCLUSIONS OR LIMITATIONS OF THIS LIMITED WARRANTY MAY NOT APPLY TO YOU. FOR
CONSUMER TRANSACTIONS. THE LIMITED WARRANTY TERMS CONTAINED IN THIS STATEMENT,
EXCEPT TO THE EXTENT LAWFULLY PERMITTED, DO NOT EXCLUDE, OR RESTRICT. OR
MODIFY BUT ARE IN ADDITION TO THE MANDATORY STATUTORY RIGHTS APPLICABLE TO THE
SALE OF THIS PRODUCT TO YOU. This Limited Warranty is applicable in all
countries and may be enforced in any country or region where MTTS or its
authorized service providers offer warranty service for the same product model
number (subject to the terms and conditions set forth in this Limited
Warranty) Under this Limited Warranty, products purchased in one country or
region may be transferred to another countryor region where MTTS or its
authorized service provid- ers offer warranty service for the same product
model number. Warranty terms, service availability, and service response times
may vary from country or region to con- try or region. Standard warranty
service response time is subject to change due to local parts availability. If
so, your MTTS authorized service provider can provide you with details. MTTS
will not alter form, fit, or function of this MTTS product to make it operate
in a country for which it was never intended to function for legal or
regulatory reasons. MTTS is not responsible for any tariffs or duties that may
be incurred in transferring the products. MTTS quarantees that the product
that you have pur- chased or leased from MTTS is free from defects in ma-
terials or workmanship under normal use during Lim- ited Warranty Period. The
Limited Warranty Period starts on the date of purchase or lease from MTTS, or
fromthe date MTTS completes installation. Your dated sales or delivery
receipt, showing the date of purchase of the product, is your proof of the
purchase or lease date. You may be required to provide proof or purchase or
lease as a condition of receiving warranty service. You are entitled to
hardware warranty service according to the terms and conditions of this
document if a repair to your MTTS prod- uct is required within the Limited
Warranty Period. Unless otherwise stated, and to the extent permitted by local
law, new MTTS product may be manufactured using new materials and used
materials equivalent to new in performance and reliability. MTTS may repair or
replace MTTS products (al with new or previously used products or parts
equivalent to new in performance and reliability, or (b) with equivalent
products to an original product that has been discontinued. Replacement parts
are warrant- ed to be free from defects in material or workmanship for ninety
(90) days or, for the reminder of Limited Warranty Period of the MTTS product
they are replacing or in which they are installed, whichever is longer. MTTS
will, at its sole discretion, repair or replace any components or product that
manifests a defect in materials or workmanship during the Limited Warranty
Period. All component parts removed under this Limited Warranty become the
property of MTTS. In the unlikely event that your MTTS product has recurring
failures, MTTS at its sole discretion, may elect to provide you with (a) a
replacement unit selected by MTTS that is the same or equivalent to your MTTS
product in performance or (b) to give you a refund of your purchase price or
lease payments (fewer interests) instead of a replacement. This is your
exclusive remedy for defective products.
Exclusions
MTTS DOES NOT WARRANT THAT THE OPERATION OF THIS PRODUCT WILL BE
UNINTERRUPTED OR ERROR- FREE. MTTS IS NOT RESPONSIBLE FOR DAMAGE THAT OCCURS
AS A RESULT OF YOUR FAILURE TO FOLLOW THE INSTRUCTIONS INTENDED FOR THE
PRODUCT. This Limited Warranty does not apply to expendable or consumable
parts and does not extend to any product from which the serial number has been
removed or that gas has been damaged or rendered defective (a) as a result of
an accident, misuse, abuse, contamination, improper or inadequate maintenance
or calibration (if required or other external causes; (b) by operation outside
the usage parameters stated in the user documentation shipped with the
product; (c) by software, interfacing, parts or supplies not supplier by MTTS:
(d) improper site prepara- tion or maintenance; (el virus infection; (f) loss
or dam- age in transit; or (q) by modification or service by anyone other than
(il MTTS personnel, (ill an MTTS authorized service provider, or lil your own
installation of end-user replaceable MTTS or MTTS approved parts if available
for your MTTS product in the servicing country or region.
Limitation of Liability
IF YOUR MTTS PRODUCT FAILS TO WORK AS WARRANT- ED ABOVE. THE MAXIMUM LIABILITY
OF MTTS UNDER THIS LIMITED WARRANTY IS EXPRESSLY LIMITED TO THE LESSER OF THE
PRICE YOU HAVE PAID FOR THE PRODUCT OR THE COSTS OF REPAIR OR REPLACE- MENT OF
ANY HARDWARE COMPONENTS THAT MAL- FUNCTION IN CONDITIONS OF NORMAL US EXCEPT
AS INDICATED ABOVE, IN NO EVENT WILL MTTS BE LIABLE FOR ANY DAMAGES CAUSED BY
THE PROD- UCT OR THE FAILURE OF THE PRODUCT TO PERFORM, INCLUDING ANY LOST
PROFITS OR SAVINGS OR SPE- CIAL, INCIDENTAL, OR CONSEQUENTIAL DAMAGES. MTTS IS
NOT LIABLE FOR ANY CLAIM MADE BY A THIRD PARTY OR MADE BY YOU FOR A THIRD
PARTY. THIS LIMITATION OF LIABILITY APPLIES WHETHER DAMAGES ARE SOUGHT, OR A
CLAIM MADE, UNDER THIS LIMITED WARRANTY OR AS A TORT CLAIM (INCLUDING
NEGLIGENCE AND STRICT PRODUCT LIABILITY. A CONTRACT CLAIM, OR ANY OTHER CLAIM.
THIS LIMITA- TION LIABILITY CANNOT BE WAIVED OR AMENDED BY ANY PERSON. THIS
LIMITATION OF LIABILITY WILL BE EFFECTIVE EVEN IF YOU HAVE ADVISED MTTS OF THE
POSSIBILITY OF ANY SUCH DAMAGES. THIS LIMITATION OF LIABILITY, HOWEVER, WILL
NOT APPLY TO CLAIMS FOR PERSONAL INJURY. THIS LIMITED WARRANTY GIVES YOU
SPECIFIC LEGAL RIGHTS. YOU MAY ALSO HAVE OTHER RIGHTS THAT MAY VARY FROM
COUNTRY TO COUNTRY. YOU ARE ADVISED TO CONSULT APPLICABLE COUNTRY LAWS FOR A
FULL DETERMINATION OF YOUR RIGHTS.
Limited Warranty Period
The Limited Warranty Period for this MTTS product is a specified, fixed period
commencing of the date of purchase and specified on the Warranty Card. The
date on your sales receipt is the date of purchase unless MTTS or your
reseller informs you otherwise in writing
Customer Responsibilities
In order to avoid the risk of charges for issues not covered by your
limited warranty issues that are not due to defects in materials and
workmanship on MTTS products), you will be asked to assist MTTS as follows:
Verify configurations, load most recent firmware, in- stall software patches,
run MTTS diagnostics and utilities; Implement temporary procedures or
workarounds provided by MTTS while MTTS works on a permanent solution;
Cooperate with MTTS in attempting to resolve the problem using online chat,
email, or telephone. This may involve performing routine diagnostic
procedures, installing additional software updates or patches; Performing
additional tasks as defined within each type of warranty service provided by
MTT, and any other actions that MTTS may reasonably request in order to best
perform the warranty support CUSTOMER RESPONSIBLE FOR DELIVERING THE PRODUCT
(AND ALL COSTS INVOLVED FROM HIS LO- CATION TO THE MTTS AUTHORIZED SERVICE
POINT.
Contacting MTTS
If your MTTS product fails during the Limited Warranty Period and the suggestions in the product documentation do not solve the problem, you can receive support by do- in one of the following:
- Locate and contact your nearest
- MTTS service pro-video via MTTS website: http://www.mtts-asia.com/support/.
- Call the Technical Support Center: +84 43 766 6521
- Before calling MTTS or an MTTS-authorized service pro- video please have the following information available:
- Product serial number and model name
- Applicable error messages
- Detailed questions
MEDICAL TECHNOLOGY TRANSFER AND SERVICES Co., LTD House No. 26, Alley 41, An Duong Vuong Street, Tay Ho District, Hanoi City, Vietnam
- Tel: +84 24 3766 6521
- Fax: +84 24 3718 8050
- Email: assistance@mtts-asia.com.
- www.mtts-asia.com.
- Logic s.r.l.
- Via Antonio Pigafetta 1
- 34147 Trieste, Italy
- Infant Phototherapy for Jaundice Treatment
- Colibri – V1
- 111720640961
- Effective date: November 2022
- Issue Number: 01 Version: 09 EN
- Issue Date: November 2022 www.mtts-asia.com.
- Tel: +84 24 3766 6521
References
Read User Manual Online (PDF format)
Read User Manual Online (PDF format) >>