MTTS WM-WL-MAN-03 Wallaby Warmer User Guide
- June 12, 2024
- MTTS
Table of Contents
WM-WL-MAN-03 Wallaby Warmer
User Guide
Refer to User Manual for comprehensive information
Modes of operation
Touch MODE Selection panel to enter MODE Selection screen
BABY MODE
Set temperature: 34ºC – 37.5ºC
In BABY MODE, the warmer provides stable control of the baby’s skin
temperature by automatically adjusting the heater power to compensate for
varying physiological and environmental conditions.
PREWARM MODE
Set temperature: not available In PREWARM MODE, the warmer provides
general alarm free prewarming of the heater and environment above the warmer
mattress.
MANUAL MODE
Set heater power: 0-100%
In MANUAL MODE, the warmer provides user-adjustable heater power and the
option to monitor the baby’s skin temperature using skin sensor.
If the patient skin temperature is more than 1ºC above the set point, then the
heater is OFF and the output value displayed is 0 (zero).
Timers
APGAR
Alarm at 1, 5 and 10 minutes. Switches off after 60 minutes.
PROCEDURE TIMERS
Alarm at the end of preset period from 1 to 99 minutes.
Alarms
Alarm | Mode | Cause | Effect | Mute Button |
---|---|---|---|---|
HIGH SKIN TEMP | BABY | Measured skin temperature high- er than set temperature | ||
by 1oC | Heater power disabled | Silences audible alarm only | ||
MANUAL | Measured skin temperature higher 38oC | Heater power disabled | Silences |
audible alarm only
PREWARM| Not active| Not active| Not active
LOW SKIN TEMP| BABY| Measured skin temperature lower than set temperature by
1oC| Heater reduced to 25% of power| Cancels alarm and restores heater power
MANUAL| Not active| Not active| Not active
PREWARM| Not active| Not active| Not active
SENSOR DISCONNECT| BABY| Skin sensor unplugged or faulty| Heater reduced to
25% of power| Silences audible alarm only
MANUAL| Not active| Not active| Not active
PREWARM| Not active| Not active| Not active
POWER FAILURE| ALL| Power supply to Warmer failed| Warmer disabled| Silences
audible alarm only
ERROR| ALL| Software or hardware failure (refer to Service Manual for more
information on error codes)| Warmer disabled| Silences audible alarm only
When available, the mute alarm period is 10 min.
X-ray procedure
- Turn the main switch of the device off and silence the power failure alarm
- Wait until the device cools down
- Lift the overhead unit up
- Move overhead unit left or right
- After X-ray procedure is completed return the overhead unit to center position
- Make sure the overhead unit is locked in center postion
- Turn the device on
Turning off the Warmer
Press and hold the OFF symbol for 3 seconds
The action will activate POWER FAILURE alarm.