MTTS Beluga V1 Resuscitator
Company Information
MEDICAL TECHNOLOGY TRANSFER AND SERVICES Co., LTD
House No. 26, Alley 41, An Duong Vuong Street, Tay Ho District, Hanoi City,
Vietnam
- Tel: +84 24 3766 6521
- Fax: +84 24 3718 8050
- Email: assistance@mtts-asia.com
- www.mtts-asia.com
PROMETEO SRL Via Mantova 9 38122 Trento Italy
Introduction
- WARNING – a warning statement refers to the conditions when the possibility of injury to the patient or user exists if a procedure is not followed correctly.
- NOTE – a note statement provides additional information intended to clarify points, procedures or instructions.
- This user manual is intended for health care professionals.
- The Beluga Resuscitator is to be operated by qualified personnel only.
This manual, accessory Directions for Use, all precautionary information, and specifications should be read before use.
Beluga Resuscitator is a standalone T-piece resuscitator intended to provide positive pressure ventilation during resuscitation of infants. The device contains a built-in pump to generate positive end-expiratory pressure (PEEP) and Peak Inspiratory Pressure (PIP) without necessity of any external air sources. An integrated blender and oxygen monitoring function also allows oxygen-rich (>21% O2) air with accurate Fraction of inspired oxygen (FiO2) to be safely delivered to the patient.
Device Description
Overview
Device Description
Control Panel
Beluga Resuscitator functions are controlled by touch screen.
1 | Power Button | Touch and hold this button to turn off the device |
---|---|---|
2 | Realtime Pressure Graph | Plots measured pressure in realtime The graph will |
be expand in look screen
3| Timer| Displays Procedure Timers
4| Status Bar| Displays Infomation about locking, battery charging conditions
5| Pressure Measurement| Displays Measured Pressure
6| FiO2 Measurement| Displays Measured FiO2
7| Setting Button| Touch this button to enter Setting screen
8| Lock Button| Touch this button to lock the screen
9| Up and Down Buttons| Touch and hold these buttons to adjust a setpoint. The
setpoint can be selected by touching but 10, 11, or 12
10| PIP Setpoint| Touch this button to select PIP setpoint
---|---|---
11| FiO2 Setpoint| Touch this button to select FiO2 setpoint
12| Ventilation Rate Setpoint| Touch this button to select Ventilation rate
setpoint
Devices starts with PIP setpoint of 0 cmH2O. When selected, it is
automatically set to 20 cmH2O.
When PIP setpoint step size is set greater than 1, touching and holding the
Up/Down buttons will only increase/decrease the value once.
Warnings
- The Beluga Resuscitator is to be operated by qualified personnel only. This manual, accessory Directions for Use, all precautionary information, and specifications should be read before use.
- Explosion hazard. Do not use the Beluga Resuscitator in the presence of flammable anesthetics or other flammable substance in combination with air, oxygen-enriched environments, or nitrous oxide.
- Do not modify this equipment without authorization of the manufacturer.
- Using the device outside of the specified ambient temperature range (19°C to 37°C) or humidity range (30%RH to 90%RH) can compromise performance.
- Do not operate the device if any of the components appear to be damaged or broken. Damaged or broken components should be discarded and replaced.
- Do not position the device in a location that reduces its ventilation.
- Do not block any of the air vents on the back of the device.
- The presence of electro-surgical devices, shortwave or microwave equipment near this device could lead to electrical interference that would negatively affect the function of the device.
- Using tubing or other breathing circuit components without cleaning or disin-fection between patients can result in infections.
- The device should only be used with tubing, cables and connectors specified in this user manual. Use of components not specified in this user manual can result in serious injury to the patient or operator or damage to the device.
Setup
- Attach the T-piece circuit, gas hose
- Attach the T-piece circuit to the mixed gas outlet in the front of the device.
- Attach the compressed oxygen source to the CGA 1240 oxygen inlet.
- Plug in and power up
- The connector plugs into the side of the device and then directly into the power outlet in the wall.
- Press the Power Button to turn the device ON.
Only use T-piece circuit and power supply unit provided by MTTS.
The gas sources must provide pressure within the range of 10 to 60 psi (0.7 to
4.1 bar). If the gas sources are outside of this range, the device may not be
able to produce the desired %FiO2.
Attaching the compressed oxygen will not waste any oxygen. The machine will
not use any compressed oxygen until you increase the FiO2 setpoint.
Operation
Start the device
Touch PIP Setpoint to start the device.
The device will start immediately with default PIP Setpoint of 20 cmH2O.
Adjust parameters
-
PIP
Select PIP setpoint circle, then touch and hold Up/Down button to adjust PIP .
The red horizontal level will move up/down according to the change. -
FiO2
Select FiO2 Setpoint circle, then touch Up/Down button to adjust FiO2. -
Ventilation Rate (VR)
Select VR Setpoint circle, then touch Up/Down button to adjust VR.
Ventilation rate is used by metronome only.
Metronome can be disabled in Setting panel.
Resuscitate patient
- Fit T-piece to resuscitation mask and place over the baby’s mouth and/or nose.
- B. Resuscitate by placing and removing thumb over the PEEP cap to allow inspiration and expiration.
Lock/unlock the screen
The screen can be locked by touch and hold the lock button. On the locked screen, all setpoints will disappear and the graph will expand. To unlock the screen, touch and hold the unlock button.
Setting screen
Setting screen allows some features of the device to be configured. To enter Setting screen, touch Setting button on the unlocked screen. To return the unlocked screen, touch the Return button on the Setting screen.
Following features are available on the Setting screen:
- Calibrate oxygen sensor
- Goto the Device information screen
- Enable/disable metronome
- Lock/unlock autorotate function
- Adjust PIP step size
- Adjust alarm volume
The Setting screen and available features may vary depending on the soft-ware version.
Calibrate oxygen sensor
To calibrate oxygen sensor, touch the ‘O2 Calib’ button on the Setting screen. The process will take 1-2 minutes to complete. After calibration, the FiO2 reading should be 21%.
Disconnect oxygen source before calibration.
Oxygen sensor must be calibrated at least every 3 months. It is recommended to
calibrate the sensor before each use.
Turning off the device
The device can be turned off by Power button either on the screen or in front of the device. Press and hold the button for 1.5 second.
Alarm system
General
List of alarms:
Alarm | Cause | Effect |
---|---|---|
HIGH PRESSURE | Measured pressure higher than the PIP setpoint by 4 cmH2O |
Device adjusts pressure by itself
Pressure >14 cmH2O for longer than 5 seconds| No effect
HIGH FiO2| Measured FiO2 higher than the setpoint by 6 %FiO2| Device controls
oxygen valve to regulate FiO2
PRESSURE SENSOR FAILURE| Pressure sensor disconnected or faulty| No effect
OXYGEN SENSOR FAILURE| Oxygen sensor disconnected or faulty| Oxygen valve
closed
LOW BATTERY| Battery remaining capacity is low| No effect
Alarm system test
All the alarms listed above are conisdered MEDIUM PRIORITY (ref. IEC
60601-1-8).
In order to verify the operation of auditory and visual alarms, perform the
following procedure:
- On the unlocked screen, set PIP to 20 cmH2O
- Occlude the mixed gas outlet for 10 seconds.
- Check pressure audible alarm sounds
Cleaning
- Comply with hospital, local and national guidelines for product cleaning frequencies.
- Ensure oxygen supply is turned off and disconnected from Beluga Resuscitator before performing cleaning procedures.
- Cleaning shall be performed at ambient condictions.
- Before cleaning, remove and discard all used disposable products using recommended method of disposal.
- Dust all surfaces with clean damp soft cloth.
- Clean all plastic surfaces with mild detergent solution (maximum 2% in wa-ter).
- Dry all the surfaces after cleaning with a clean soft cloth or paper towel.
- Ensure that no part of Beluga Resuscitator is immersed in any cleaning liquid or cleaning solution.
- Do not use abrasive cleaning solutions.
- Ensure all Beluga Resuscitator parts and accessories are checked before returning the device to service.
The reusable resuscitation circuit includes a number of parts that must be completely disassembled and sterilized before use.
These parts include:
Component | Recommended sterialization method |
---|---|
Silicone tube | Autoclaving |
PEEP valve | Low-temperature plasma sterilization OR soak disinfection with |
1:500 chlorine-containing disinfectant
Mask| Autoclaving
Maintenance
Only qualified personnel should carry out service and maintenance procedures.
- After the maintenance is completed, ensure the equipment is functioning correctly in accordance with the published performance specifications.
- Ensure only approved replacement parts are used during service and maintenance procedures.
- Please contact an authorized MTTS representative for further assistance with any servicing or maintenance requirement.
The Beluga Resuscitator includes a number of parts that must be replaced or serviced during the lifetime of the device.
These parts include:
Component | Approx. Duration of Use | Maintenance Information |
---|---|---|
Air filter | 6 months | Clean or replace. |
Oxygen sensor | 18 months | Replace if it is faulty or FiO2 control is not |
accurate after calibration.
Device information screen
To better understand the device performance, device information should be
checked before opening the device. From Setting screen, touch the Info button
to see this in-formation.
Air filter and oxygen sensor replacement
To access air filter and oxygen sensor, open the back hatch with a Philips head screwdriver. Gently take the filter/sensor out and replace with the new one.
Specifications
Performance Specifications
Control Settings
- Set PIP Default Default
- Range 0 – 60 cmH2O
- Set FiO2
- Default 21 %FiO2
- Range 21 – 100 %FiO2
- Set Ventilation rate
- Default Disabled
- Range 30 – 60 BPM
Displays
- Measured pressure display
- Accuracy ±1 cmH2O
- Range 0 – 100 cmH2O
Physical Specifications
- Dimensions (HxWxD) overall 20 cm x 22cm x 20 cm 2.5 kg
- Medium Priority Audible 3 pulse burst, followed by 2.5 second delay
- Medium Priority Visual Yellow flashing at 0.5 Hz
- Alarm Volume Adjustable
- Internal pump
- Type Blower, 12VDC
Electrical Specifications
- Power characteristics 60W, 100-240VAC, 47/63Hz
- Off-the-shelf external power supply
- International medical safety approvals
- (ANSI/AAMI/EN 60601-1, UL/TUV)
- Class I construction is standard (ground required)
- 100k hours MTBF
- Energy Star Efficiency Level V compliant
- RoHS 2 compliant
- Over voltage / over current protection
- Battery Type 11.1V Li-ion
- Capacity 6 cells, 6000mAh
Environmental Specifications
- Operating
- Ambient temperature +18°C to +35°C
- Humidity: 0% to 90% RH non condensing
- Atmospheric Pressure: 70-106kPa
- Transport and storage
- Ambient temperature 0°C to +50°C
- Humidity: 0% to 90% RH non condensing
- Atmospheric Pressure: 70-106kPa
- Exclusions Exclusions
Standards for Reference
- EN ISO 13485:2016
- EN ISO 14971:2012
- EN 60601-1:2006/A1:2013
- EN 60601-1-2:2015
- EN 60601-1-6:2010
- EN 60601-1-8:2007/A11:2017
- MEDDEV 2.12-1 Rev. 8
- MEDDEV. 2.7.1 Rev. 4
- MEDDEV. 2.12-2 Rev. 2
Explanation of Symbols
- Refer to User Manual before operating this device.
- Do not disassemble Beluga Resuscitator unless you are an MTTS trained technician or have been instructed to by qualified personnel.
- Keep out of direct sunlight.
- This device should be disposed of separately from normal household waste so that components can be recycled.
- Device manufacturer.
- Date of manufacture.
- Serial Number
- Power supply polarity
- ![Uolfin-N6RAI-LWS230W1-Z8-1-Light-Polished-Gold-and-Bell-Glass-Wall-Sconce-08]](https://manuals.plus/wp-content/uploads/2022/12/Uolfin-N6RAI-LWS230W1-Z8-1-Light-Polished-Gold-and-Bell-Glass-Wall-Sconce-08.png)12 Volt, Direct Current, 5 Ampere
- CE Marking with Notified Body Number
- Name and address of European Authorized Representative
Warranty Policy
General Terms
This MTTS Limited Warranty gives you, the customer, express limited warranty
rights from MTTS, the manufacturer for the duration specified on the Warranty
Card.
Please refer to the MTTS Website for an extensive description of your limited
warranty entitlements. In addi-tion, you may also have other legal rights
under applica-ble law or special written agreement with MTTS.
MTTS MAKES NO OTHER EXPRESS WARRANTY OR CONDITION WHETHER WRITTEN OR ORAL AND
MTTS EXPRESSLY DISCLAIMS ALL WARRANTIES AND CON-DITIONS NOT STATED IN THIS
LIMITED WARRANTY. TO THE EXTENT ALLOWED BY LOCAL LAW OF jURIS-DICTIONS OUTSIDE
VIETNAM, MTTS DISCLAIMS ALL IMPLIED WARRANTIES OR CONDITIONS, INCLUD-ING ANY
IMPLIED WARRANTIES OF MERCHANTABIL-ITY AND FITNESS FOR A PARTICULAR PURPOSE.
FOR ALL TRANSACTIONS OCCURRING IN VIETNAM ANY IMPLIED WARRANTY OR CONDITION OF
MERCHANT-ABILITY, SATISFACTORY qUALITY, OR FITNESS FOR A PARTICULAR PURPOSE IS
LIMITED TO THE DURATION OF THE EXPRESS WARRANTY SET FORTH ABOVE. SOME
COUNTRIES DO NOT ALLOW A LIMITATION ON HOW LONG AN IMPLIED WARRANTY LASTS OR
THE EXCLU-SION OR LIMITATION OF INCIDENTAL OR CONSEqUEN-TIAL DAMAGES FOR
CONSUMER PRODUCTS. IN SUCH COUNTRIES, SOME EXCLUSIONS OR LIMITATIONS OF THIS
LIMITED WARRANTY MAY NOT APPLY TO YOU. FOR CONSUMER TRANSACTION, THE LIMITED
WARRANTY TERMS CONTAINED IN THIS STATEMENT, EXCEPT TO THE EXTENT LAWFULLY
PERMITTED, DO NOT EXCLUDE, RESTRICT, OR MODIFY BUT ARE IN ADDITION TO THE
MANDATORY STATUTORY RIGHTS APPLICABLE TO THE SALE OF THIS PRODUCT TO YOU.
This Limited Warranty is applicable in all countries and may be enforced in
any country or region where MTTS or its authorized service providers offer
warranty service for the same product model number (subject to the terms and
conditions set forth in this Limited Warranty) Under this Limited Warranty,
products purchased in one country or region may be transferred to another
country or region where MTTS or its authorized service providers offer
warranty service for the same product model number. Warranty terms, service
availability, and service response times may vary from country or region to
country or region. Standard warranty service response time is subject to
change due to local parts availability. If so, your MTTS authorized service
provider can provide you with details. MTTS will not alter form, fit, or
function of this MTTS product to make it operate in a country for which it was
never intended to function for legal or regulatory reasons. MTTS is not
responsible for any tariffs or duties that may be incurred in transferring the
products. MTTS guarantees that the product that you have purchased or leased
from MTTS is free from defects in materials or workmanship under normal use
during Limited Warranty Period. The Limited Warranty Period starts on the date
of purchase or lease from MTTS, or from the date MTTS completes installation.
Your dated sales or delivery receipt, showing the date of purchase of the
product, is your proof of the purchase or lease date. You may be required to
provide proof or purchase or lease as a condition of receiving warranty
service. You are entitled to hardware warranty service according to the terms
and conditions of this document if a repair to your MTTS product is required
within the Limited Warranty Period. Unless otherwise stated, and to the extent
permitted by local law, new MTTS product may be manufactured using new
materials and used materials equivalent to new in performance and reliability.
MTTS may repair or replace MTTS products
- with new or previously used products or parts equivalent to new in performance and reliability, or
- with equivalent products to an original product that has been discontinued.
Replacement parts are warranted to be free from defects in material or workmanship for ninety (90) days or, for the reminder of Limited Warranty Period of the MTTS product they are replacing or in which they are installed, whichever is longer
MTTS will, at its sole discretion, repair or replace any components or product that manifests a defect in materials or workmanship during the Limited Warranty Period. All component parts removed under this Limited Warranty become the property of MTTS. In the unlikely event that your MTTS product has recurring failures, MTTS at its sole discretion, may elect to provide you with
- a replacement unit selected by MTTS that is the same or equivalent to your MTTS product in performance or
- to give you a refund of your purchase price or lease payments (less interests) instead of a replacement.
This is your exclusive remedy for defective products.
Exclusions
MTTS DOES NOT WARRANT THAT THE OPERATION OF THIS PRODUCT WILL BE UNINTERRUPTED
OR ERROR- FREE. MTTS IS NOT RESPONSIBLE FOR DAMAGE THAT OCCURS AS RESULT OF
YOUR FAILURE TO FOLLOW THE INSTRUCTIONS INTENDED FOR THE PRODUCT.
This Limited Warranty does not apply to expendable or consumable parts and
does not extend to any product from which the serial number has been removed
or that gas been damaged or rendered defective
- as a result of accident, misuse, abuse, contamination, improper or inadequate maintenance or calibration (if required) or other external causes;
- by operation outside the usage parameters stated in the user documentation shipped with the product;
- by software, interfacing, parts or supplies not supplier by MTTS;
- improper site preparation or maintenance;
- virus infection;
- loss or damage in transit ; or
- by modification or service by anyone other than
- MTTS personnel,
- an MTTS authorized service provider, or
- your own installation of end-user replaceable MTTS or MTTS approved parts if available for your MTTS product in the servicing country or region.
Limitation of Liability
IF YOUR MTTS PRODUCT FAILS TO WORK AS WARRANT-ED ABOVE, THE MAXIMUM LIABILITY
OF MTTS UNDER THIS LIMITED WARRANTY IS EXPRESSLY LIMITED TO THE LESSER OF THE
PRICE YOU HAVE PAID FOR THE PRODUCT OR THE COSTS OF REPAIR OR REPLACE-MENT OF
ANY HARDWARE COMPONENTS THAT MAL-FUNCTION IN CONDITIONS OF NORMAL USE.
EXCEPT AS INDICATED ABOVE, IN NO EVENT WILL MTTS BE LIABLE FOR ANY DAMAGES
CAUSED BY THE PROD-UCT OR THE FAILURE OF THE PRODUCT TO PERFORM, INCLUDING ANY
LOST PROFITS OR SAVINGS OR SPE-CIAL, INCIDENTAL, OR CONSEqUENTIAL DAMAGES.
MTTS IS NOT LIABLE FOR ANY CLAIM MADE BY A THIRD PARTY OR MADE BY YOU FOR A
THIRD PARTY.
THIS LIMITATION OF LIABILITY APPLIES WHETHER DAMAGES ARE SOUGHT, OR CLAIM
MADE, UNDER THIS LiMiTED WARRANTY OR AS A TORT CLAiM (iNCLUD-iNG NEGLiGENCE
AND STRiCT PRODUCT LiABiLiTY), A CONTRACT CLAIM, OR ANY OTHER CLAIM. THIS
LIMITA-TION LIABILITY CANNOT BE WAIVED OR AMENDED BY ANY PERSON. THIS
LIMITATION OF LIABILITY WILL BE EFFECTIVE EVEN IF YOU HAVE ADVISED MTTS OF THE
POSSIBILITY OF ANY SUCH DAMAGES. THIS LIMITATION OF LIABILITY, HOWEVER, WILL
NOT APPLY TO CLAIMS FOR PERSONAL INjURY.
THIS LIMITED WARRANTY GIVES YOU SPECIFIC LEGAL RIGHTS. YOU MAY ALSO HAVE OTHER
RIGHTS THAT MAY VARY FROM COUNTRY TO COUNTRY. YOU ARE ADVISED TO CONSULT
APPLICABLE COUNTRY LAWS FOR A FULL DETERMINATION OF YOUR RIGHTS.
Limited Warranty Period
The Limited Warranty Period for this MTTS product is a specified, fixed period commencing of the date of purchase and specified on the Warranty Card. The date on your sales receipt is the date of purchase unless MTTS or your reseller informs you otherwise in writing.
Customer Responsibilities
In order to avoid the risk of charges for issues not covered by your limited
warranty (issues that are not due to defects in materials and workmanship on
MTTS product), you will be asked to assist MTTS as follows:
-
Verify configurations, load most recent firmware, install software patches, run MTTS diagnostics and utilities;
-
Implement temporary procedures or workarounds provided by MTTS while MTTS works on permanent solution;
-
Cooperate with MTTS in attempting to resolve the problem using online chat, email, or telephone.
This may involve performing routine diagnostic procedures, installing additional software updates or patches; -
Perform additional tasks as defined within each type of warranty service provided by MTTS and any other actions that MTTS may reasonably request in order to best perform the warranty support
CUSTOMER IS RESPONSIBLE FOR DELIVERING THE PRODUCT (AND ALL COSTS INVOLVED) FROM HIS LO- CATION TO THE MTTS AUTHORIZED SERVICE POINT.
Contacting MTTS
If your MTTS product fails during the Limited Warranty Period and the
suggestions in the product documentation do not solve the problem, you can
receive support by doing one of the following:
- Locate and contact your nearest MTTS service provider via MTTS website:
- Call the Technical Support Centre:
- +84 24 3766 6521
Before calling MTTS or an MTTS authorized service provider please have the following information available:
- Product serial number and model name
- Applicable error messages
- Detailed questions
MEDICAL TECHNOLOGY TRANSFER AND SERVICES Co., LTD
House No. 26, Alley 41, An Duong Vuong Street, Tay Ho District, Hanoi City,
Vietnam
- Tel: +84 24 3766 6521
- Fax: +84 24 3718 8050
- Email: assistance@mtts-asia.com
- www.mtts-asia.com
PROMETEO SRL Via Mantova 9 38122 Trento Italy
References
Read User Manual Online (PDF format)
Read User Manual Online (PDF format) >>