ARKTEK YBC-5 Passive Vaccine Storage Device User Manual

June 14, 2024
ARKTEK

ARKTEK logo YBC-5
Passive Vaccine Storage Device
Maintenance Manual

YBC-5 Passive Vaccine Storage Device

INTRODUCTION

Purpose of Manual
This manual will help you to properly maintain the Arktek YBC-5 device and, if needed, to resolve any issues that might occur.
Device Features

  • Advanced materials and superior vacuum insulation
  • Easy to use and maintain (simple, safe, robust and reliable)
  • Keeps vaccines safely between 0° and 10 °C for at least 35 days in an ambient temperature range of 15 ° to 43 °C with no external power source.
  • Real-time display of the internal temperature.
  • Visual and audible alarm for out-of-range temperature (below 0 °C).
  • Built-in disposable battery,5 years no need to change.
  • Lid lock for secure storage.

Device Information

ITEM SPEC
Model identifier Arktek alpha-numeric designation YBC-5

Ambient Temperature
Expected conditions of surrounding environment of use| 15° C to 43° C
Vaccine storage temperature range Internal temperature in vaccine payload area| 0°C to 10°C
Capacity Volume of internal payload space| 5.4 L
Unit Dimension (OD x H) Outside diameter and height| φ528 x 747 mm
Net Weight Unloaded| 26.5 Kg
Gross Weight Fully loaded with packages| 32.5 Kg

Safety precautions

  • Device is designed to store vaccine. Do not use for other purposes (e.g,, do not use for storing liquid nitrogen).
  • Device is designed for stationary use. Do not as a portable-transport cold box.
  • Some parts may be flammable (e.g., nylon ropes). Keep away from open flame and flammable or explosive items.
  • Keep device away from acid, alkali, and other corrosive substances.
  • Do not allow children to play on or near device.
  • Only allow authorized personnel to perform repairs or maintenance. Do not attempt to disassemble, repair, or modify the unit on your own.
  • Do not dispose of the device on your own, even if it is severely damaged. Contact a technical supervisor to perfom an authorized disposal.

MAINTENANCE & CLEANING

General instructions
The exterior of the device can be cleaned as needed. The interior of the device should be cleaned during the process of exchanging ice blocks. Do not clean the interior when it is in use for vaccine storage, as doing so will cause unnecessary temperature exchange and thus reduce vaccine hold time.
To clean the exterior of the device
You can clean the exterior of the device when it is loaded with vaccines, however, do not shake or tilt the device. You must avoid damaging the vaccine vials.

  1. Use soap and water to wash the device with a hose, sponge, or cloth. Do not use abrasives or cleaning powders.
    Note: The device’s electronics are can tolerate some moisture. However, do not submerge the device.

  2. Dry the device with a lint-free towel or allow it to air dry.

To clean the interior of the device:
To clean the interior of the Arktek, take these steps:

  1. Remove the cap.

  2. Remove the vaccine stacks. Follow established cold chain vaccine handling practices with any removed vaccines.

  3. If needed, clean the vaccine cup-stack containers with soap and water and allow them to dry.

  4. Remove and set aside the ice blocks.

  5. Use mild soap and water or a 1% chlorine bleach solution to clean the inside of the device with a hose, sponge, or cloth. Do not scrub the interior with abrasives or cleaning powders.
    Caution: Take care of the wiring that is affixed on the interior of the device; these wires are for the sensors that monitor the internal temperature of the device.
    Caution: The device should not be submerged in water.

  6. Thoroughly wipe down the interior of the device with a towel. The interior of the device can be air-dried if time allows. You can also turn the device upside down as needed to allow it to drain.

  7. Leave the cap off of the device during storage for next use. It will remain tethered to the device.

TROUBLESHOOTING

Note: If, for trouble shooting, you need to open the cap or change the internal parts during the use cycle, you should remove everything (vaccine stacks and ice blocks) from the device and follow established protocols for cold chain handling of vaccines so as to ensure that they remain potent and effective. Please record all observed issues and attempted solutions in the device log.

Problems

| Typical Causes|

Resolution

---|---|---
External components are damaged.| Improper handling or accidental collision.| Replace entire device.
Screen does not display anything.| 1.  Wire connection came off or wire broken.
2.  LCD display and/or main PCB assembly have failed.
3. Battery is dead.| 1.Reconnect or replace with new components.
2. Replace with new components.
3. Replace with a new battery.
The “Err” message displays for a long time.| 1. The connection of the data cable has come off.
2. The data cable is broken| 1. Reconnect.
2. Replace with a new one.
Visual and/or audible alarm.| The interior temperature is below 0°C.| Remove everything and reload with properly conditioned ice blocks.
The thermometer’s pointer does change position over time (when in the frozen ice block)| Thermometer is broken.| Replace with a new thermometer.
Interior temperature rises more than 2°C / day during the use cycle.| The vacuum Dewar is broken.| Stop using the device. Call the authorized customer
service engineer.

DETAILED PARTS-ASSEMBLY VIEW

ARKTEK YBC-5 Passive Vaccine Storage Device - Assembly

LIST OF PARTS

Item number in expanded view| Spare parts number| Spare parts description in English
---|---|---
1| 92703A207| Flat Head Torx Screw M38
2| YBC-5.8-1| Battery Cover
3| 92703A207| Flat Head Torx Screw M3X8
4| YBC-5.0-8| Dewar Top Deck
5| 91771A110| Stainless Flat Head Torx Screw M3X12
6| YBC-5.7-3| Call Top Plate
7| YBC-5.7-5| Main Seal
8| YBC-5.7-2| Glass Window Edges Masked
9| YBC-5.7-7| LCD Display
10| YBC-5.7-6| LCD Foam Backer
11| YBC-5.7-4| Display Bracket
12| 91772A073| Stainless Pan Head Screw M2
4
13| YBC-5.6-1| Main PCB O-Ring
14| YBC-5.7-1| Main PCB ASSY
15| 92115A211| Stainless Pan Head Screw ST38
16| 91500A537| Stainless Flat Head Torx Screw M6X12
17| YBC-5.6-1| Main PCB O-RING
18| 91400A107| Stainless Pan Head Screw ST3.5
9.5
19| YBC-5.6-2| Battery Clamp
20| YBC-5.6-5| Battery Clamp Foam
21| YBC-5.6-6| Battery
22| 91400A107| Stainless Pan Head Screw ST3.59.5
23| YBC-5.6-7| Loudspeaker
24| YBC-5.6-9| Snap Button-Female
25| YBC-5.2-4| Nylon Rope-2
26| YBC-5.6-8| Shoulder Screw M5.0X10,Shoulder: φ 6.35,Length : 38mm
27| 90945A760| Flat Washer, IDφ6.6,ODφ12,Thickness0.8mm
28| YBC-5.6-3| Latch
29| YBC-5.6.2| Lock Inner Bracket
30| 90945A760| Flat Washer, IDφ6.6,ODφ12,Thickness0.8mm
31| YBC-5.6-4| Collar
32| YBC-5.6.1| Lock Outer Bracket
33| 90065A077| Pan Head Screw ST3.5
9.5
34| 90945A760| Flat Washer, IDφ6.6,ODφ12,Thickness0.8mm
35| 92147A029| Split Lock Washer,φ7
36| 90107A011| Flat Washer,φ5
37| 91240A011| Hex Nut,M5
38| YBC-5.0-14| Flex Sensor Folded PCB Assy
39| YBC-5.0-9| Hall Sensor Cradle
40| YBC-5.2-5| Snap Button-male
41| YBC-5.2-4| Nylon Rope-1
42| YBC-5.2| Plug
43| YBC-5.3.2-1| Nylon Rope-3
44| YBC-5.3.2-2| Handle Washer
45| YBC-5.3-1| Vaccine Stack Top Shell
---|---|---
46| YBC-5.3.1-5| Foam Bumper Strip
47| YBC-5.4.1| Vaccine Cup Small
48| YBC-5.3.1| Vaccine Cup Large
49| YBC-5.3.3-1| Vaccine Cup Divider-Male
50| YBC-5.3.3-2| Vaccine Cup Divider-Female
51| YBC-5.5-1| Ice Block
52| YBC-5.5-5| Thermometer Plug
53| YBC-5.5-6| Thermometer
54| YBC-5.5-2| Ice Block Cap
55| 91771A539| Flat Head Screw M6X16
56| YBC-5.0-7| Vaccine Stack Plate Fixed Axis
57| YBC-5.0-5| Vaccine Stack Plate
58| YBC-5.0-10| Stop Compression Sleeve For Wire Rope
59| YBC-5.0-11| Wire Rope
60| YBC-5.1| Dewar