HACH AS950 All Weather Refrigerated Sampler Instruction Manual

June 14, 2024
HACH

HACH AS950 All Weather Refrigerated Sampler Instruction Manual
All Weather Refrigerated Sampler

Specifications

Specifications are subject to change without notice.
Refrigerated Sampler and All Weather Refrigerated Sampler (AWRS)

Specification Refrigerated Sampler AWRS
Dimensions (W x D x H)1 61 x 61 x 112 cm (24 x 24 x44 in.) 76 x 81 x 130 cm

(30 x 32 x 51 in.)
Weight| 63.3 kg (140 lb)| 86 kg (190 lb)
Power requirements (includes compressor)| 115 VAC, 60 Hz, 3.3 A(18 locked rotor amps)230 VAC, 50 Hz, 1.7 A (9 lockedrotor amps)| 115 VAC, 60 Hz, 4.2 A or 6.4 A withcontroller compartment heater230 VAC, 50 Hz, 2.7 A or 4.1 A withcontroller compartment heater
Overload protection| Thermal overload relay, opens at 100 to 110 °C (212 to 230 °F)| 115 VAC: 7.5 A circuit breaker230 VAC: 5.0 A circuit breaker
Compressor| 1/6 HP| 1/5 HP115 VAC: 115 °C (239 °F) thermaloverload protector, 7.1 locked rotor amps230 VAC: 120 °C (248 °F) thermaloverload protector, 7.6 A peak start current
Operating temperature| 0 to 50 °C (32 to 122 °F)| 0 to 50 °C (32 to 122 °F); with ACbattery backup: 0 to 40 ºC (32 to 104 ºF); with controller compartment heater: –40 to 50 °C (–40 to 122 °F);with controller compartment heater and AC battery backup: –15 to 40 ºC (5 to104 ºF)

1 Refer to Figure 1 on page 6 for the sampler dimensions.

Specification Refrigerated Sampler AWRS
Storage temperature –30 to 60 °C (–22 to 140 °F) –30 to 60 °C (–22 to 140

°F)
Relative humidity| 0 to 95%| 0 to 95%
Installation category, pollution degree| II, 2| II, 2
Protection class| I| I
Temperature control| 4 (±0.8) °C (39 (±1.5) °F) inambient temperatures at a maximum of 50 °C (120 °F)| 4 (±0.8) °C (39 (±1.5) °F)2
Enclosure| 22-gauge steel (optional stainless steel) with vinyl laminate overcoat| IP24, low-density polyethylene with UV inhibitor
Sample bottle capacity| Single bottle: 10 L (2.5 gal) glass or polyethylene, or 21 L (5.5 gal) polyethylene
Multiple bottles: two 10 L (2.5 gal) polyethylene and/or glass, four 10 L(2.5 gal) polyethylene and/or glass, eight 2.3 L (0.6 gal) polyethylene and/or1.9 L (0.5 gal) glass, twelve 2 L (0.5 gal) polyethylene (AWRS only), twenty- four 1 L (0.3 gal) polyethylene and/or 350 mL (12 oz.) glass
Certifications| AC power supply: cETLus, CE

AS950 controller

Specification Details
Dimensions (W x H x D) 31.1 x 18.9 x 26.4 cm (12.3 x 7.4 x 10.4 in.)
Weight 4.6 kg (10 lb) maximum
Enclosure PC/ABS blend, NEMA 6, IP68, corrosion and ice resistant
Pollution degree/installation category 3, II
Protection class II
Display ¼ VGA, color
Power requirements Refrigerated sampler: 15 VDC supplied by a power supply

(115 to 220 VAC, 50/60Hz); AWRS: 15 VDC supplied by an integral power supply
Overload protection| 7 A, DC line fuse for the pump
Operating temperature| 0 to 50 °C (32 to 122 °F); Refrigerated sampler: 0 to 49 °C (32 to 120 °F); AWRS with controller compartment heater: –40 to 50 °C (–40 to 122 °F); AWRS with controller compartment heater and AC battery backup: –15 to 40 °C (5 to 104 °F)
Storage temperature| Refrigerated sampler: –40 to 60 °C (–40 to 140 °F); AWRS: –30 to60 °C (–22 to 140 °F)

2 Radio frequency interference in the 30 to 50 MHz range can cause a maximum temperature change of 1.3 °C (34.3 °F). Adjust the set point temperature to 2 to 10 °C (35.6 to 50 °F) to correct this interference.

Specification Details
Storage/operating humidity 100% condensing
Pump Peristaltic high speed with spring-mounted Nylatron rollers
Pump enclosure Polycarbonate cover
Pump tubing 9.5 mm ID x 15.9 OD mm (3/8-in. ID x 5/8-in. OD) silicone
Pump tubing life 20,000 sample cycles with: 1 L (0.3 gal) sample volume, 1

rinse, 6- minute pacing interval, 4.9 m (16 ft) of 3/8-in. intake tube, 4.6 m(15 ft) of vertical lift, 21 °C (70 °F) sample temperature
Vertical sample lift| 8.5 m (28 ft) for 8.8 m (29 ft) maximum of 3/8-in. vinyl intake tube at sea level at 20 to 25 °C (68 to 77 °F)
Pump flow rate| 4.8 L/min (1.25 gpm) at 1 m (3 ft) vertical lift with 3/8-in. intake tube typical
Sample volume| Programmable in 10-mL (0.34 oz) increments from 10 to 10,000 mL (3.38 oz to 2.6 gal)
Sample volume repeatability (typical)| ±5% of 200 mL sample volume with: 4.6 m (15 ft) vertical lift, 4.9 m (16 ft) of 3/8-in. vinyl intake tube, single bottle, full bottle shut-off at room temperature and 1524 m (5000 ft) elevation
Sample volume accuracy (typical)| ±5% of 200 mL sample volume with: 4.6 m (15 ft) vertical lift, 4.9 m (16 ft) of 3/8-in. vinyl intake tube, single bottle, full bottle shut-off at room temperature and 1524 m (5000 ft) elevation
Sampling modes| Pacing: Fixed Time, Fixed Flow, Variable Time, Variable Flow, EventDistribution: Samples per bottle, bottles per sample and time based (switching)
Run modes| Continuous or non-continuous
Transfer velocity (typical)| 0.9 m/s (2.9 ft/s) with: 4.6 m (15 ft) vertical lift, 4.9 m (16 ft) of 3/8-in. vinyl intake tubing, 21 °C (70 °F) and 1524 m (5000 ft) elevation
Liquid detector| Ultrasonic. Body: Ultem® NSF ANSI standard 51 approved, USP Class VI compliant. Contacting liquid detector or optional non- contact liquid detector
Air purge| An air purge is done automatically before and after each sample. The sampler automatically compensates for different intake tube lengths.
Tubing| Intake tubing: 1.0 to 30.0 m (3.0 to 99 ft) length, ¼-in. or 3/8-in. ID vinyl or 3/8-in. ID Teflon™-lined polyethylene with protective outer cover (black or clear)
Wetted materials| Stainless steel, polyethylene, Teflon, Ultem, silicone
Memory| Sample history: 4000 records; Data log: 325,000 records; Event log: 2000 records
Communications| USB and optional RS485 (Modbus)
Electrical connections| Power, auxiliary, optional sensors (2x), USB, distributor arm, optional rain gauge, thermal (AWRS only)
Analog outputs| AUX port: none; optional IO9000 module : Three 0/4–20 mA outputs to supply the recorded measurements (e.g., level, velocity, flow and pH) to external instruments
Specification| Details
---|---
Analog inputs| AUX port: One 0/4–20 mA input for flow pacing; optional IO9000 module : Two 0/4–20 mA inputs to receive measurements from external instruments (e.g., third-party ultrasonic level)
Digital outputs| AUX port: none; optional IO9000 module : Four low voltage, contact closure outputs that each supply a digital signal for an alarm event
Relays| AUX port: none; optional IO9000 module : Four relays controlled by alarm events
Certifications| CE, cETLus

**Figure 1 Refrigerated sampler and AWRS dimensions

**Dimensions 
Dimensions 
General information

In no event will the manufacturer be liable for direct, indirect, special, incidental or consequential damages resulting from any defect or omission in this manual. The manufacturer reserves the right to make changes in this manual and the products it describes at any time, without notice or obligation. Revised editions are found on the manufacturer’s website.

Safety information
The manufacturer is not responsible for any damages due to misapplication or misuse of this product including, without limitation, direct, incidental and consequential damages, and disclaims such damages to the full extent permitted under applicable law. The user is soley responsible to identify critical application risks and install appropriate mechanisms to protect processes during a possible equipment malfunction.

Please read this entire manual before unpacking, setting up or operating this equipment. Pay attention to all danger and caution statements. Failure to do so could result in serious injury to the operator or damage to the equipment.

Make sure that the protection provided by this equipment is not impaired. Do not use or install this equipment in any manner other than that specified in this manual.

Use of hazard information

D A N G E R
Indicates a potentially or imminently hazardous situation which, if not avoided, will result in death or serious injury.

W A R N I N G
Indicates a potentially or imminently hazardous situation which, if not avoided, could result in death or serious injury.

C A U T I O N
Indicates a potentially hazardous situation that may result in minor or moderate injury.

N O T I C E
Indicates a situation which, if not avoided, may cause damage to the instrument. Information that requires special emphasis.

Precautionary labels
Read all labels and tags attached to the instrument. Personal injury or damage to the instrument could occur if not observed. A symbol on the instrument is referenced in the manual with a precautionary statement.

This is the safety alert symbol. Obey all safety messages that follow this symbol to avoid potential injury. If on the instrument, refer to the instruction manual for operation or safety information.
This symbol indicates that a risk of electrical shock and/or electrocution exists.
This symbol indicates that a risk of fire is present.
This symbol indicates that the marked item can be hot and should not be touched without care.

Prohibited IconThis symbol indicates that the item is to be protected from fluid entry.
Prohibited IconThis symbol indicates that the marked item should not be touched.
Prohibited IconThis symbol indicates a potential pinch hazard.
This symbol indicates that the object is heavy.
SymbolThis symbol indicates that the marked item requires a protective earth connection. If the instrument is not supplied with a ground plug on a cord, make the protective earth connection to the protective conductor terminal.
Electrical equipment marked with this symbol may not be disposed of in European domestic or public disposal systems. Return old or end-of-life equipment to the manufacturer for disposal at no charge to the user.

Compliance and certification

C A U T I O N
This equipment is not intended for use in residential environments and may not provide adequate protection to radio reception in such environments

Canadian Radio Interference-Causing Equipment Regulation, ICES-003, Class A:
Supporting test records reside with the manufacturer.
This Class A digital apparatus meets all requirements of the Canadian Interference-Causing Equipment Regulations.

FCC Part 15, Class “A” Limits
Supporting test records reside with the manufacturer. The device complies with Part 15 of the FCC Rules. Operation is subject to the following conditions:

  1. The equipment may not cause harmful interference.
  2. The equipment must accept any interference received, including interference that may cause undesired operation.

Changes or modifications to this equipment not expressly approved by the party responsible for compliance could void the user’s authority to operate the equipment. This equipment has been tested and found to comply with the limits for a Class A digital device, pursuant to Part 15 of the FCC rules. These limits are designed to provide reasonable protection against harmful interference when the equipment is operated in a commercial environment. This equipment generates, uses and can radiate radio frequency energy and, if not installed and used in accordance with the instruction manual, may cause harmful interference to radio communications. Operation of this equipment in a residential area is likely to cause harmful interference, in which case the user will be required to correct the interference at their expense. The following techniques can be used to reduce interference problems:

  1. Disconnect the equipment from its power source to verify that it is or is not the source of the interference.
  2. If the equipment is connected to the same outlet as the device experiencing interference, connect the equipment to a different outlet.
  3. Move the equipment away from the device receiving the interference.
  4. Reposition the receiving antenna for the device receiving the interference.
  5. Try combinations of the above.

Product overview
D A N G E R
Chemical or biological hazards. If this instrument is used to monitor a treatment process and/or chemical feed system for which there are regulatory limits and monitoringrequirements related to public health, public safety, food or beverage manufacture or processing, it is the responsibility of the user of this instrument to know and abide by any applicable regulation and to have sufficient and appropriate mechanisms in place for compliance with applicable regulations in the event of malfunction of the instrument.

C A U T I O N
Fire hazard. This product is not designed for use with flammable liquids

The sampler collects liquid samples at specified intervals and keeps the samples in a refrigerated cabinet. Use the sampler for a wide variety of aqueous sample applications and also with toxic pollutants and suspended solids. Refer to Figure 2

Lockable cabinet door (AWRS only)
Push the round button in the center of the latch to open the door. Flip the latch to close the door tight. Two keys are supplied for the door lock. Over time, it may be necessary to tighten the adjustment screw on the door latch.

Controller compartment heater (AWRS only)
The controller compartment heater is a factory-installed option. The heater prevents liquid from freezing in the tubing, extends the life of the tubing and pump components and prevents the collection of ice and snow on the cover.

Figure 2 Refrigerated sampler and AWRS
Product overview

1 Refrigerated sampler 6 Controller 11 Controller cover
2 Power supply 7 Refrigerated cabinet 12 Compartment heater

option
3 Pump| 8 Cover latch| 13 Door latch
4 Liquid detector| 9 AWRS| 14 Bottle tray
5 Non-contacting liquid detector| 10 Access cover

Product components

W A R N I N G
Personal injury hazard. Instruments or components are heavy. Use assistance to install or move.

The instrument weighs a maximum of 86 kg (190 lb). Do not try to unpack or move the instrument without sufficient equipment and people to do it safely. Use correct lifting procedures to prevent injury. Make sure that all used equipment is rated for the load, for example, a hand truck must be rated for a minimum of 90 kg (198 lb). Do not move the sampler when filled sample bottles are in the refrigerated cabinet.

Make sure that all components have been received. Refer to Figure 3. If any items are missing or damaged, contact the manufacturer or a sales representative immediately.

Figure 3 Sampler components
Product overview

1 Optional cover 5 All Weather Refrigerated Sampler (AWRS)
2 Refrigerated sampler 6 Intake tubing, vinyl or teflon-lined
3 Components for single-bottle option 7 Strainer
4 Components for multiple-bottle option 8 Tubing coupler3

**Installation

**

D A N G E R
Multiple hazards. Only qualified personnel must conduct the tasks described in this section of the document.

Site installation guidelines
D A N G E R
Explosion hazard. The instrument is not approved for installation in hazardous locations.

This instrument is rated for an altitude of 2000 m (6562 ft) maximum. Although the use of thisequipment above the 2000 m altitude does not show any substantial safety concern, the manufacturer recommends that users with concerns contact technical support.

  • Only install the refrigerated sampler in an indoor location. Install the AWRS in an indoor or outdoor location.
  • Make sure that the temperature at the location is in the specification range. Refer to Specifications on page 3.
  • Install the sampler on a level surface. Adjust the sampler feet to make the sampler level. Refer to Figure 1 on page 6 for the sampler dimensions.
  • Use the installed anchor brackets and user-supplied 3 /8-in. bolts for the AWRS. Refer to Figure 4.
  • Plumb a drain tube to the ½ in.-14 NPT female connector on the bottom of the sampler.

Figure 4 AWRS anchor bracket locations with mounting dimensions
Dimensions 

  1. Optional anchor brackets
  2. Anchor brackets (2x)

Prepare the sampler

Clean the sample bottles
Clean the sample bottles and caps with a brush, water and a mild detergent. Flush the sample bottles with fresh water followed by a distilled water rinse.

Install a single bottle
When a single bottle is used to collect one composite sample, do the steps that follow. When multiple bottles are used, refer to Install multiple bottles on page 13. When the bottle is full, the full bottle shut-off stops the sampling program. Install the sample bottle as shown in Figure 5.

Figure 5 Single bottle installation
Installation

Install multiple bottles
When multiple bottles are installed, a distributor arm moves the sample tube over each bottle. Sample collection automatically stops when the specified number of samples are collected.

  1. Assemble the sample bottles as shown in Figure 6. For eight or more bottles, make sure that the first bottle is near the bottle one indicator in the clockwise direction.
  2. Put the bottle assembly in the sampler. For eight or more bottles, align the wires in the slots in the bottom tray.

Figure 6 Multiple bottle installation
Installation

1| Retainer for 24 1-L poly bottles| 4| Bottle tray for 8 to 24 bottles| 7| Front of sampler
---|---|---|---|---|---
2| Retainer for 24 350-mL glass bottles| 5| Slot for bottle tray| 8| Removable tray (AWRS only)
3| Bottle one indicator| 6| Floor of refrigerated sampler| 9 Insert (refrigerated sampler only)

Plumb the sampler
Install the intake tube in the middle of the sample stream (not near the surface or bottom) to make sure that a representative sample is collected.

  1. For a sampler with the standard liquid detector, connect the tubing to the sampler as shown in Figure 7.
    Note: When Teflon-lined tubing is used, use the tubing connection kit for Teflon-lined PE tubing.

  2. For a sampler with the optional non-contacting liquid detector, connect the tubing to the sampleras shown in Figure 8.
    Note: When Teflon-lined tubing is used, use the tubing connection kit for Teflon-lined PE tubing.

  3. Install the intake tube and strainer in the main stream of the sample source where the water is turbulent and well-mixed. Refer to Figure 9.

    • Make the intake tube as short as possible. Refer to Specifications on page 3 for the minimum intake tubing length.

    • Keep the intake tube at a maximum vertical slope so that the tube drains completely between samples.
      Note: If a vertical slope is not possible or if the tube is pressurized, disable the liquid detector. Calibrate the sample volume manually.

    • Make sure that the intake tube is not pinched

Figure 7 Plumbing—Standard liquid detector
Connection
Figure 8 Plumbing—Non-contacting liquid detector
Connection
Figure 9 Site installation
Connection

  1. Strainer
  2. Vertical lift
  3. Mounting surface

Electrical installation

Connect the sampler to power

D A N G E R
Electrocution hazard. If this equipment is used outdoors or in potentially wet locations, a Ground Fault Circuit Interrupt (GFCI/GFI) device must be used for connecting the equipment to its main power source.

D A N G E R
Fire hazard. Install a 15 A circuit breaker in the power line. A circuit breaker can be the local power disconnect, if located in close proximity to the equipment.

D A N G E R
Electrocution hazard. Protective Earth Ground (PE) connection is required.

W A R N I N G
Electrocution hazard. Make sure that there is easy access to the local power disconnect

Connect the power cord on the AWRS or both power cords on the refrigerated sampler. The refrigerator starts after a 5-minute delay. Use a power line filter or connect the power cord for the controller to a different branch circuit to decrease the possibility of electrical transients.

Controller connections

W A R N I N G
Electrical shock hazard. Externally connected equipment must have an applicable country safety standard assessment.

Figure 10 shows the electrical connectors on the controller

Figure 10 Controller connections
Controller connections 

1 Sensor 2 port (optional) 5 USB connector
2 Thermal unit port (AWRS only) 6 Rain gauge/RS485 port

(optional)
3| Power supply port| 7| Auxilliary I/O port
4| Sensor 1 port (optional)| 8| Distributor arm/Full bottle shut- off port

Connect a Sigma 950 or FL900
If sample pacing is flow based, supply the controller with a flow input signal (pulse or 4–20 mA). Connect a Sigma 950 or an FL900 Flow Logger to the AUX I/O port.

As an alternative, connect a flow sensor to a sensor port. Refer to Connect a sensor on page 19. Item to collect: Multi-purpose auxiliary full cable, 7 pin

  1. Connect one end of the cable to the flow meter. Refer to the flow meter documentation.
  2. Connect the other end of the cable to the AUX I/O port on the controller.

Connect a non-Hach flow meter
To connect a non-Hach flow meter to the AUX I/O port, do the steps that follow. Items to collect: Multi-purpose auxiliary half cable, 7 pin

  1. Connect one end of the cable to the AUX I/O port on the controller.
  2. Connect the other end of the cable to the flow meter. Refer to Figure 11 and Table 1.

Note: In some installations, it is necessary to connect external equipment to the Pulse input, Special output and/or Program Complete output with long cables. Since these are ground-referenced pulse interfaces, false signaling can be caused by transient ground differences between each end of the cable. High ground differentials are typical in heavy industrial environments. In such environments, it may be necessary to use third-party galvanic isolators (e.g., optocouplers) in line with the affected signal(s). For the Analog input, external ground isolation is typically not necessary because the 4–20 mA transmitter typically supplies isolation

Figure 11 Auxiliary connector
Auxiliary connector 
Table 1 Half cable wiring information

Pin Signal Color4 Description Rating
1 +12 VDCpoweroutput White Power supply positive output.Only use with pin
2. Battery power to the I/O module:12 VDC nominal; Power supply to theI/O

module: 15 at 1.0 A maximum.
2| Common| Blue| Negative return of powersupply. When the powersupply is used, pin 2 isconnected to earth ground5.

4 The wire color refers to the colors of multi-purpose cables (8528500 and 8528501).
5 All mains powered equipment that connects to the controller terminals must be NRTL listed.

Table 1 Half cable wiring information (continued)

Pin Signal Color 4 Description Rating
3 Pulse input or Analog input Orange This signal is a sample collection

trigger from the flow logger (pulse or4–20 mA) or a simple floating (dry) contact closure.| Pulse input —Reacts to a positive pulse with respect to pin 2. Termination (pulled low): pin 2 through a series1 kΩ resistor and 10 kΩ resistor. A7.5 zener diode is in parallel with the 10 kΩ resistor as a protection device. Analog input —Reacts to the analog signal that enters pin 3 and returns on pin 2. Input burden: 100 Ω plus 0.4 V; Input current (internal limit): 40 to50 mA maximum6Absolute maximum input: 0 to 15 VDC with respect to pin 2.Signal to make the input active: 5 to 15 V positive-going pulse7 with respect to pin 2, 50 millisecond minimum.
4| Liquid level input or Auxiliary control input| Black| Liquid level input —Start or continue the sampling program. A simple float level switch can supply input. Auxiliary control input — Start a sampler after the sampling program on another sampler ends. As an alternative, start a sampler when a trigger condition occurs. For example, when a high or low pH condition occurs, the sampling program starts.| Termination (pulled high): internal +5 V supply through an 11 kΩ resistance with a series 1 kΩ resistor and 7.5 V zener diode terminated to pin 2 for protection. Trigger: High to low voltage with a low pulse of 50 milliseconds minimum.Absolute maximum input: 0 to 15 VDC with respect to pin 2. Signal to make the input active: external logic signal with 5 to 15 VDC power source. The drive signal must be typically high. The external driver must be able to sink0.5 mA at 1 VDC maximum at the logic low level.A logic high signal from a driver with a power source of more than 7.5 V will source current into this input at the rate of: I = (V – 7.5)/1000 where I is the source current and V is the power supply voltage of the driving logic.Dry contact (switch) closure:50 millisecond minimum between pin 4 and pin 2. Contact resistance: 2 kΩ maximum. Contact current: 0.5 mA DC maximum

6 Long-term operation in this state voids the warranty.
7 Source impedance of the driving signal must be less than 5 kΩ.
4 The wire color refers to the colors of multi-purpose cables (8528500 and 8528501)

Table 1 Half cable wiring information (continued)

Pin Signal Color 4 Description Rating
5 Special output Red This output goes from 0 to+12 VDC with respect to pin

2 after each sample cycle. Refer to the Mode setting of the hardware settings for the AUX I/O port. Refer to the AS950 operations documentation.| This output has protection against short circuit currents to pin 2. External load current: 0.2 A maximumActive high output: 15 VDC nominal with AC power to the AS950 controller or a 12 VDC nominal with battery power to the AS950 controller.
6| Program Complete output| Green| Typical state: open circuit. This output goes to ground for 90 seconds at the end of the sampling program.Use this output to start another sampler or to signal an operator or data logger at the end of the sampling program.| This output is an open drain output with 18 V zener clamp diode for over- voltage protection. The output is active low with respect to pin 2.Absolute maximum ratings for the output transistor: sink current = 200 mA DC maximum; external pull-up voltage= 18 VDC maximum
7| Shield| Silver| The shield is a connection to earth ground when AC power is supplied to a sampler to control RF emissions and susceptibility to RF emissions.| The shield is not a safety ground. Do not use the shield as a current carrying conductor.The shield wire of cables that are connected to the AUX I/O port and are more than 3 m (10 ft) should connected to pin 7.Only connect the shield wire to earth ground at one end of the cable to prevent ground loop currents.

Connect a sensor
To connect a sensor (e.g., pH or flow sensor) to a sensor port, refer to Figure 12.

Figure 12 Connect a sensor
Connect a sensor
4 The wire color refers to the colors of multi-purpose cables (8528500 and 8528501).

Startup

Set the instrument to on
The refrigerator starts after a 5-minute delay when power is supplied to the sampler. The refrigerator continues to operate when the controller is set to off or the power is removed from the controller.

Push the POWER key on the controller to set the controller to on.

To set the refrigerator to off, push the POWER key on the controller. Then, disconnect the power cord on the AWRS or the two power cords on the refrigerated sampler.

Preparation for use
Install the analyzer bottles and stir bar. Refer to the operations manual for the startup procedure

Maintenance

D A N G E R
 Multiple hazards. Only qualified personnel must conduct the tasks described in this se ction of the document .

D A N G E R
 Electrocution hazard. Remove power from the instrument before doing maintenance or service activities.

W A R N I N G
Biohazard exposure. Obey safe handling protocols during contact with sample bottles and sampler components.

W A R N I N G
Multiple hazards. The technician must make sure that the equipment operates safely and correctly after maintenance procedures.

N O T I C E
Do not disassemble the instrument for maintenance. If the internal components must be cleaned or repaired, contact the manufacturer.

Clean the instrument

C A U T I O N

Fire hazard. Do not use flammable agents to clean the instrument.

N O T I C E
Do not clean the controller compartment heater with liquids of any kind.

If water is not sufficient to clean the controller and the pump, disconnect the controller and move the controller away from the sampler. Allow sufficient time for the controller and pump to dry before theb parts are re-installed and put back into service.

Clean the sampler as follows:

  • Refrigerator—clean the condenser fins and coils as needed with a brush or vacuum.
  • Sampler cabinet and tray—clean the internal and external surfaces of the sampler cabinet with a damp cloth and mild detergent. Do not use abrasive cleaners or solvents.

Replace the desiccant
A desiccant cartridge in the controller absorbs moisture and prevents corrosion. Monitor the desiccant color through the window. Refer to Figure 13. Fresh desiccant is orange. When the color is green, replace the desiccant.

  1. Unscrew and remove the desiccant cartridge. Refer to Figure 13.
  2. Remove the plug and discard the spent desiccant.
  3. Fill the desiccant tube with fresh desiccant.
  4. Install the plug.
  5. Apply silicone grease to the O-ring.
  6. Install the desiccant tube in the controller.

Figure 13 Desiccant cartridge
Desiccant cartridge 

1 Plug 3 O-ring
2 Desiccant tube 4 Desiccant window

Pump maintenance

C A U T I O N
Setting Icon Pinch hazard. Remove power from the instrument before maintenance or service activities are done.

Replace the pump tubing

N O T I C E
Use of tubing other than that supplied by the manufacturer can cause excessive wear on mechanical parts and/or poor pump performance.

Examine the pump tubing for wear where the rollers rub against the tubing. Replace the tubing when the tubing shows signs of wear.

Pre-requisites:
Pump tubing—pre-cut or bulk 4.6 m or 15.2 m (15 ft or 50 ft)

  1. Disconnect the power to the controller.
  2. If the bulk tubing is used, cut the tubing and add alignment dots. Refer to Figure 14.
  3. Remove the pump tubing as shown in the illustrated steps that follow.
  4. Clean the silicone residue from the interior of the pump housing and from the rollers.
  5. Install the new pump tubing as shown in the illustrated steps that follow.

Figure 14 Pump tubing preparation
Pump tubing preparation 
Table

1 To intake tubing 7 Length for the refrigerated sampler
2 Alignment dots 8 Length for the AWRS
3 To fitting on sampler base 9 Length for refrigerated sampler and

controller with standard liquid detector
4 Length for controller with standard liquid detector| 10 Length for refrigerated sampler and controller with non-contacting liquid detector
5 Length for controller with optional non- contacting liquid detector| 11 Length for AWRS and controller with standard liquid detector
6 Length inside the pump| 12 Length for AWRS and controller with non- contacting liquid detector

Installation Instruction
Installation Instruction
Clean the rotor
Clean the rotor, pump tube tracks and pump housing with a mild detergent. Refer to Replace the pump tubing on page 21 and the illustrated steps that follow.
Installation Instruction

Replace the distributor arm tube
The distributor arm moves over each bottle during multiple bottle sampling. Replace the tube in the distributor arm when the tube is worn. Make sure that the correct tube is used for the correct distributor and distributor arm.

Note: The distributor tubing is not the same as the pump tubing. The pump tubing installed in the distributor assembly can damage the distributor. Also, samples can be missed because the distributor arm cannot move easily

  1. Remove the tube from the distributor arm and from the ceiling of the sampler cabinet.
  2. Insert the new tube into the distributor arm. Extend the tube past the end of the distributor arm 4.8 mm (3/16 in.) or 19 mm (3/4 in.) as shown in item 1 of Figure 15.
  3. Insert the other end of the tube into the fitting on the ceiling of the sampler cabinet.
  4. Complete the diagnostic test for the distributor to make sure that the operation is correct.

Figure 15 Distributor assembly
Distributor assembly 

1| Tube extension| 4 Distributor arm lengths:152.4 mm (6.0 in.),177.8 mm (7.0 in.) or190.8 mm (7.51 in.)| 7| Distributor tube
---|---|---|---|---
2| Nozzle| 5| Distributor motor
3| Distributor arm| 6| Shaft

Replace the power supply—refrigerated sampler
W A R N I N G
Fire hazard. Use only the external power supply that is specified for this instrument.

To replace the power supply for the refrigerated sampler, refer to Figure 16.
Figure 16 Power supply replacement
Power supply replacement

Disposal

D A N G E R
Child entrapment hazard. Remove the doors on the refrigerated cabinet before disposal.

C A U T I O N
Chemical exposure hazard. Dispose of chemicals and wastes in accordance with local, regional and national regulations.

Troubleshooting

General troubleshooting
Table 2 shows causes and corrective actions for several common problems

Problem Possible cause Solution
No instrument power Problem with the main power source. Make sure that AC

power gets to the electrical outlet.
Defective power supply (refrigerated sampler only)| Replace the power supply.
Defective controller| Contact technical support.

Table 2 Troubleshooting table (continued)

Problem Possible cause Solution
Sampler does not have sufficient lift. Strainer is not completely submerged.

Install the shallow depth strainer (2071 or 4652).
Intake tube has a leak.| Replace the intake tube.
Pump tube is worn.| Replace the pump tubing on page 21.
Pump roller assembly is worn.| Contact technical support.
Sample volume is not correct.| Incorrect volume calibration| Repeat the volume calibration.
Incorrect tube length is specified in the sampling program.| Make sure that the correct tube length is in the sampling program.
Intake tube does not purge completely.| Make sure that the intake tube is as vertical and as short as possible.
Strainer is not completely submerged.| Install the shallow depth strainer (2071 or 4652).
Worn pump tubing and/or roller assembly.| Replace the pump tube and/or roller assembly.
The liquid detector is disabled.| Turn the liquid detector on and complete a volume calibration.
Liquid detector does not operate properly.| Calibrate the liquid detector with the same liquid that is sampled.

Phone Icon
HACH COMPANY World Headquarters

P.O. Box 389, Loveland, CO 80539-0389 U.S.A.
Tel. 970-669-3050
800-227-4224 (U.S.A. only)
Fax 970-669-2932
[email protected]
www.hach.com

HACH LANGE GMBH
Willstätterstraße 11
D-40549 Düsseldorf, Germany
Tel. +49 (0) 2 11 52 88-320
Fax +49 (0) 2 11 52 88-210
[email protected]
www.de.hach.com

HACH LANGE Sàrl
6, route de Compois
1222 Vésenaz
SWITZERLAND
Tel. +41 22 594 6400
Fax +41 22 594 6499

© Hach Company/Hach Lange GmbH, 2015–2017, 2021.
All rights reserved. Printed in Germany

References

Read User Manual Online (PDF format)

Loading......

Download This Manual (PDF format)

Download this manual  >>

HACH User Manuals

Related Manuals