Table of Contents
Quickie Sedeo Pro Advanced Power Wheelchair
Product Information
This user manual provides instructions and information for the
usage of a wheelchair. The product is manufactured by Sunrise
Medical (US) LLC and the customer service department can be
contacted at 2842 N. Business Park Avenue Fresno, CA 93727 USA or
by calling 800-333-4000. The product
can be registered at
www.SunriseMedical.com/register.
The authorized dealer of the wheelchair is the best resource for
questions regarding chair safety, use, and maintenance. The manual
also provides space to fill in the dealer’s information including
their name, address, and telephone number.
The wheelchair is not made with Natural Rubber Latex, ensuring
that individuals with latex allergies can safely use the
product.
When the wheelchair reaches the end of its life, it should be
taken to an approved collection or recycling point designated by
the local or state government. The product is manufactured using a
variety of materials, most of which are fully recyclable. Proper
disposal and recycling of the product will help conserve natural
resources and protect the environment.
Product Usage Instructions
User Information
This section provides general information for the user of the
wheelchair. It includes details about this user manual and
instructions for obtaining further information.
Using the Seating System
This section provides instructions for using the seating system
of the wheelchair. It covers topics such as adjusting the
controller, setting laterals, using lap belts or seating
positioning belts, and setting the height of the chest strap
(Anterior tilt).
It is recommended to read the entire manual before using the
wheelchair and to save it for future reference.
Quickie® SEDEO® PRO/PRO ADVANCED
Supplier: User:
This manual must be given to the user of this wheelchair. Before using this wheelchair read this entire manual and save for future reference.
Owner’s Manual
Quickie® SEDEO® PRO/PRO ADVANCED
Distribuidor: Usuario:
Este manual debe ser entregado el usuario de esta silla de ruedas. Antes de usar esta silla de ruedas, lea este manual en su totalidad y guárdelo para futura referencia.
Cada una de las sillas se envía con un manual de instrucciones en inglés. El manual en español y francés está disponible en formato PDF en nuestra página en Internet: www.SunriseMedical.com. Ingrese a la página del producto específi co para descargar el manual, o comuníquese con el proveedor autorizado de Sunrise Medical.
Manual de Instrucciones
Quickie® SEDEO® PRO/PRO ADVANCED
Au Fournisseur: Ce manuel doit être remis à l’utilisateur de ce
fauteuil roulant.
Utilisateur:
Avant d’utiliser ce fauteuil roulant, lisez l’intégralité
de manuel et conserver pour référence future.
Chaque fauteuil est livré avec un manuel d’instructions en anglais. Les versions en espagnol et en français sont à votre disposition en format PDF sur le site : www.SunriseMedical.com. Veuillez vous rendre à la page de votre produit pour télécharger le manuel dans la langue souhaitée, ou contactez un fournisseur agréé Sunrise Medical.
Manuel d ‘instructions
P/N 248020 Rev. G
QUICKIE® SEDEO® PRO/PRO ADVANCED
Seating Manual
Instructions for Use
Quickie® SEDEO® PRO/PRO ADVANCED
I. INTRODUCTION
SUNRISE MEDICAL LISTENS Thank you for choosing a Quickie® wheelchair. We want to hear your questions or comments about this manual, the safety and reliability of your chair, and the service you receive from your Sunrise dealer. Please feel free to write or call us at the address and telephone number below:
SUNRISE MEDICAL (US) LLC Customer Service Department 2842 N. Business Park Avenue Fresno, CA 93727 USA 800-333-4000
Be sure to register your heelchair, and let us know if you change your address. This will allow us to keep you up to date with information about safety, new products, and options to increase your use and enjoyment of the wheelchair.
www.SunriseMedical.com/register
You can also register your product at: www.SunriseMedical.com/register
FOR ANSWERS TO YOUR QUESTIONS Your authorized dealer knows your wheelchair best, and can answer most of your questions about chair safety, use, and maintenance.
For future reference, please fill in the following:
Dealer: __
Address: _____
Telephone: __
Serial #: ____ Date/Purchased: ___
ADDITIONAL INFORMATION YOU SHOULD KNOW No component of this chair was made with Natural Rubber Latex.
DISPOSAL AND RECYCLING INFORMATION When this product reaches the end of its
life, please take it to an approved collection or recycling point designated
by your local or state government. This product is manufactured using a
variety of materials.
Your product should not be disposed of as ordinary household waste. You should
dispose of your product properly, according to local laws and regulations.
Most materials that are used in the construction of this product are fully
recyclable. The separate collection and recycling of your product at the time
of disposal will help conserve natural resources and ensure that it is
disposed in a manner that protects the environment.
Ensure you are the legal owner of the product prior to arranging for the
product disposal in accordance with the above recommendations.
248020 Rev. G
2
II. TABLE OF CONTENTS
Quickie® SEDEO® PRO/PRO ADVANCED
I. INTRODUCTION……………………………………………………………………………2
5.11 Pre-set fixed/swing away controller……………………………………22
II. TABLE OF CONTENTS………………………………………………………………….3
5.12 Pre-set laterals…………………………………………………………………………23
5.13 Lap belt/seating positioning belt ………………………………………….24
1.0 USER INFORMATION…………………………………………………………………….4
5.14 Setting the height of the chest strap (Anterior tilt)………..25
1.1 This user manual……………………………………………………………………..4
1.2 For further information…………………………………………………………4 6.0 USING THE SEATING SYSTEM……………………………………………………26
1.3 Symbols used in this manual…………………………………………………4
6.1 Getting in and out of the wheelchair (Transferring) ………26
6.2 Preparing for forward transfer…………………………………………….26
2.0 SAFETY……………………………………………………………………………………………….5
6.3 Preparing for side transfer…………………………………………………….28
2.1 Symbols and labels used on the product…………………………..5
6.4 Manual recline backrest…………………………………………………………28
2.2 Safety: Temperature ……………………………………………………………….5
6.5 Manually elevating leg rests knee angle………………………………28
2.3 Safety: Moving parts……………………………………………………………….5
6.6 Headrest……………………………………………………………………………………29
2.4 Safety: Choking hazard ………………………………………………………….5
6.7 Swing-away controller……………………………………………………………29
2.5 Safety: General warnings……………………………………………………….6
6.8 Laterals ……………………………………………………………………………………..29
2.6 Safety: Accessory trays ………………………………………………………….6
6.9 Push handle height………………………………………………………………….29
2.7 Safety: Stability…………………………………………………………………………6
6.10 Electric seating function ………………………………………………………..30
6.11 Seat memory position……………………………………………………………30
3.0 GENERAL DESCRIPTION
6.12 Retrieving a memory seat position………………………………………30
3.1 Maximum user weight……………………………………………………………7
6.13 Return to drive mode ……………………………………………………………30
6.14 Saving a new position to memory ………………………………………31
4.0 WARNINGS FOR COMPONENTS AND OPTIONS…………….8
6.15 Delete custom programmed memory……………………………….31
4.1 Seat depth adjustment…………………………………………………………..8
6.16 Return to drive mode ……………………………………………………………31
4.2 Armrests …………………………………………………………………………………..8
6.17 Curves……………………………………………………………………………………….32
4.3 Cushions……………………………………………………………………………………8
6.18 Breaking and emergency stop ……………………………………………..32
4.4 Fasteners…………………………………………………………………………………..8
6.19 Driving on a slope…………………………………………………………………..32
4.5 Footplate and footrests…………………………………………………………8
6.20 Anterior tilt function ……………………………………………………………..33
4.6 Positioning belts………………………………………………………………………8
6.21 Using the knee fixation………………………………………………………….33
4.7 Push handles…………………………………………………………………………….8
6.22 Using a chest strap …………………………………………………………………33
4.8 Seating systems ……………………………………………………………………….8
4.9 Upholstery fabric…………………………………………………………………….8 7.0 POWERED SEATING OPTIONS…………………………………………………34
4.10 Wiring and connectors………………………………………………………….9
7.1 Powered seating options……………………………………………………….34
4.11 Assignable controls…………………………………………………………………9
4.12 Power seating safety ………………………………………………………………9 8.0 REDUCING THE SEATING SYSTEM SIZE FOR
4.13 Power recline …………………………………………………………………………..9
TRANSPORTATION PURPOSES…………………………………………………35
4.14 Power leg rest………………………………………………………………………….9
8.1 Flip down bracket quick release ………………………………………….35
4.15 Lift, tilt, anterior tilt, or combo……………………………………………9
8.2 Swing away leg rests……………………………………………………………….35
4.16 C-ME® lift………………………………………………………………………………….10
8.3 Single post removable armrest…………………………………………….35
4.17 On/Off switch………………………………………………………………………….10
4.18 Suspension springs………………………………………………………………….10 9.0 MAINTENANCE AND CLEANING……………………………………………36
4.19 Anterior tilt position………………………………………………………………10
9.1 General maintenance……………………………………………………………..36
4.20 Swing away joystick arm………………………………………………………..10
9.2 Cleaning your seating system……………………………………………….36
4.21 Transit………………………………………………………………………………………..11
9.3 Hygiene measures when being re-used……………………………..37
9.4 General cleaning instructions……………………………………………….37
5.0 SETTING UP THE SEATING SYSTEM ……………………………………….12
9.5 Cleaning accessories ………………………………………………………………37
5.1 Pre-set backrest frame depth ………………………………………………12
9.6 Cleaning controls ……………………………………………………………………37
5.2 Pre-set armrests ……………………………………………………………………..13
9.7 Storage………………………………………………………………………………………37
5.3 Pre-set backrest………………………………………………………………………14
9.8 Maintenance and inspection schedule………………………………..37
5.4 Pre-set flip back armrests/single post removeable
armrests ……………………………………………………………………………………14 10.0 TECHNICAL SPECIFICATIONS: APPLICABLE NORMS /
5.5 Pre-set reclining flip back armrests …………………………………….16
STANDARDS …………………………………………………………………………………….38
5.6 Pre-set fixed centermount leg rest (short and long)………17
5.7 Pre-set Sedeo Seating fixed centermount leg rest ………….18 11.0 SUNRISE LIMITED WARRANTY…………………………………………………39
5.8 Pre-set power centermount leg rest………………………………….18
5.9 Pre-set swing away leg rest…………………………………………………..20
5.10 Pre-set headrest………………………………………………………………………21
3
248020 Rev. G
Quickie® SEDEO® PRO/PRO ADVANCED
1.0 USER INFORMATION
1.0 User information Congratulations on choosing a Sunrise Medical wheelchair.
Our high quality mobility products are designed to enhance your independence
and make your everyday life easier. As a part of our ongoing product
improvement initiative, Sunrise Medical reserves the right to change
specifications and design without notice. Further, not all features and
options offered are compatible with all configurations of the wheelchair. All
dimensions are approximate and may be subject to change. The intended lifetime
of the wheelchair is five (5) years. Please DO NOT use or fit any third party
components to the wheelchair unless they are officially approved by Sunrise
Medical.
1.1 This user manual This user manual will help you to use and maintain your
seating system safely. This seating manual is a supplement to Sunrise
Medical’s general wheelchair user manual: · The general wheelchair user manual
· The user manual for the seating system (this manual) · The user manual for
the controller · The user manual for the battery charger When necessary, this
seating user manual refers to other manuals as shown below:
General manual: Refers to the general user manual for the wheelchair.
Controller: Refers to the user manual for the controller.
Battery charger: Refers to the user manual for the battery charger.
Do not use your wheelchair until this entire manual (all booklets) has been
read and understood! If one of the user manuals was not included with your
wheelchair, please contact your Sunrise Medical authorized dealer immediately.
1.2 For further information Please contact your Sunrise Medical authorized
dealer if you have any questions regarding the use, maintenance, or safety of
your wheelchair. In case there is no authorized dealer in your area or you
have any questions, contact Sunrise Medical either in writing or by telephone.
For information about product safety notices and product recalls, go to
www.SunriseMedical.com
1.3 Symbols used in this manual
DANGER!
Potential risk of serious injury or death
WARNING!
Potential risk of injury
CAUTION!
Potential damage to equipment
NOTE:
General user advice
Not following these instructions may result in physical injury, damage to the
product, or damage to the environment!
If you are visually impaired, this document can be viewed
in PDF format at www.SunriseMedical.com
248020 Rev. G
4
2.0 SAFETY
Quickie® SEDEO® PRO/PRO ADVANCED
Follow the instructions carefully next to these warning symbols! Not paying careful attention to these instructions could result in physical injury or damage to the wheelchair or the environment. Wherever possible, safety information is provided in the relevant chapter.
2.1 Symbols and labels used on the product The signs, symbols, and instructions affixed to the wheelchair comprise part of the safety facilities. They must never be covered or removed. They must remain present and clearly legible throughout the entire lifespan of the wheelchair.
Replace or repair all illegible or damaged signs, symbols, and instructions immediately. Please contact your Sunrise Medical authorized dealer for assistance.
Fig. 2.1: Fig. 2.2: Fig. 2.3:
Fig. 2.4:
Warning Danger of finger entrapment (pinch point). Warning Maximum user weight for the wheelchair. Warning Ensure that the seat interface thumb screws are screwed in tight before using your wheelchair, especially on a slope. Warning Do not drive your wheelchair on a slope with the backrest reclined.
2.2 Safety: Temperature
WARNING!
· Avoid physical contact with the wheelchair’s motors at all times. Motors are
continuously in motion during use and can reach high temperatures. After use,
the motors will cool down slowly. Physical contact could cause burns. Allow
the motors after using at least 30 minutes to cool down.
· If you do not use the wheelchair, ensure that it is not exposed to direct
sunlight for lengthy periods of time. Certain parts of the wheelchair, such as
the seat, the back, and the armrests can become hot if they have been exposed
to full sunlight for too long. This may cause burns or allergic reactions to
the skin.
2.3 Safety: Moving parts
DANGER!
A wheelchair has moving and rotating parts. Contact with moving parts may
result in serious physical injury or damage to the wheelchair. Contact with
the moving parts of the wheelchair should be avoided (Fig. 2.5).
· Wheels (turning and caster) · Powered tilt in space adjustment · Powered
seat lift option · Powered recline adjustment · Swing-away bracket of the
controller · Power elevating leg rest
2.4 Safety: Choking hazard
DANGER!
This mobility aid uses small parts which under certain circumstances may
present a choking hazard to young children.
Fig. 2.1 Fig. 2.2
Fig. 2.3 Fig. 2.4
Fig. 2.5
5
248020 Rev. G
Quickie® SEDEO® PRO/PRO ADVANCED
2.0 SAFETY
2.5 Safety: General warnings
DANGER!
1. Always ensure that you are able to operate all controls from a comfortable
position. Paying attention to your posture is essential to ensure your
continued comfort and well being.
2. Always ensure that you can be seen clearly. Use the lights, especially if
you intend using your seating system in poor light.
3. This seating system has been set up to match the needs of a particular
user. If used by another user, then it may need to be adjusted.
4. Do not let children or others use your seating system. 5. Avoid spilling
liquids on to the controller or its connections, especially
hot, sticky, sugary, or corrosive liquids. Such contamination may compromise
the safe operation of the control system and damage the seating system
upholstery.
WARNING!
DO NOT use any wheelchair that has been involved in a motor vehicle accident.
A sudden stop and/or collision may structurally damage your wheelchair. There
may have been a change to the structure of the chair and/or damaged or broken
some of the components. Wheelchairs involved in sudden stops should be
inspected for possible failures in frame and/or components. Frame damage may
be represented by but not limited to: visual cracks, dents, metal distortion,
bends, or damage to the seating mounting. If the chair no longer drives
straight, it could be damaged. If the wheelchair has been involved in an
accident, discontinue use immediately and contact your Sunrise Medical
authorized dealer for a thorough inspection. If the damage is questionable or
if there is concern regarding the condition of the chair, Sunrise Medical
recommends replacement of the chair.
Wheelchairs involved in collisions should be replaced.
NOTE: the warranty is void on any wheelchair that is involved in a collision.
2.5.1 User and Attendants
DANGER!
1. When transferring the user in or out of the seating system, never use the
footplate as a platform.
2. When transferring the user in or out of the seating system, never use the
armrest as a means of support.
3. Do not lift this seating system by any parts that are removable. Doing so
may result in damage to the seating system or injury to the user.
4. Never try to stop or slow this seating system with the wheel locks (if
applicable). Wheel locks are designed to prevent the system from moving when
stopped. They are not brakes.
5. Always keep your hands and body parts clear of moving parts to help
prevent injuries.
2.5.2 User
DANGER!
1. Every seating system is different. 2. Take the time to learn the feel of
this system before you begin riding. 3. Be aware that you must develop your
own methods for the safe use of
this system that are best suited to your level of function and ability. 4.
Have someone help you practice bending, reaching, and transferring until
you learn how to do them safely. 5. Never try a new maneuver on your own
unless you are sure it is safe. 6. Get to know the areas where you plan to use
your system. Look for
hazards and learn how to avoid them. 7. Always wear a positioning belt. 8. Do
not use a seating system if the joystick does not spring back to the
neutral position. 9. Do not use your seating system if the joystick boot is
ripped or damaged. 10. Do not try to tip this seating system without an
attendant present. 11. Do not allow additional riders; the system may become
unstable.
2.5.3 Attendants
DANGER!
Ensure you heed all warnings and follow all instructions in each section of
this manual. Warnings that apply to the user also apply to you.
1. Do not stand or sit on any portion of this seating system. 2. You need to
work with the user, and the user’s doctor or health care
provider, to develop safe methods best suited to your abilities and those of
the user. 3. To manually push the system you must release the motor brakes. ·
Do not engage or disengage motor brakes unless power to the
system is off. · Ensure you have full control over the system when you release
the
motor brakes. When you do so the system will not have brakes. · Ensure the
system is on level ground before you release the motor
brakes. 4. Check to ensure push handle grips will not rotate or slip off.
2.5.4 Attendant control
DANGER!
Ensure you always have comfortable access to the controls while the chair is
moving and ensure that the controller is fixed securely to the chair. 1. Set
the speed of the attendant control to a speed that you can comfort-
ably follow. 2. Always turn off the power to the controller when leaving the
user in the
chair.
2.6 Safety: Accessory Trays
2.6.1 Vent tray
WARNING!
Using a vent tray will affect the stability and overall weight of your
wheelchair. 1. Ensure that the batteries and the ventilator are securely
fastened to the
tray before use. 2. Familiarize yourself with the increased size of the
wheelchair before
driving to prevent potential collisions. 3. When activating the recline
backrest ensure that the area around the
vent tray is clear. 4. Be aware that the vent tray assembly can cause a finger
trap (pinch point)
hazard when the backrest is reclined. 5. The vent tray is not designed to hold
anything other than the vent unit
and its batteries. It is not designed to hold an oxygen bottle, etc. Use of
these items may cause your wheelchair to become unstable. 6. The vent tray
must be mounted and maintained by a Sunrise Medical authorized dealer.
2.6.2 Swing away tray
DANGER!
1. The maximum load allowed for the tray is 5.5 lbs (2.5kg). 2. Do not
overload the tray. This could cause the tray to break or could
cause the chair to become unstable. 3. Do not leave lit cigarettes or other
heat sources on the tray as this could
cause the tray to deform and mark. 4. Ensure that all extremities and clothing
are clear when positioning the
tray for use.
2.7 Safety: Stability
DANGER!
1. Removing the batteries may cause the chair to become unstable causing it
to tip backwards.
2. Only remove the batteries while the chair is on the ground. 3. Do not
remove the batteries while the chair is on a lift table or on any
other elevated surface.
248020 Rev. G
6
3.0 GENERAL DESCRIPTION
General description This wheelchair is equipped with the Sedeo Pro/Pro Advanced seating system. The Sedeo Pro/Pro Advanced seating system provides excellent support for the head, torso, arms, and legs. Optimal body support ensures maximum independence for the wheelchair user.
The wheelchair must be adjusted by a qualified specialist. Incorrect settings may have negative consequences for the user.
Note: Seat depth adjustment may require reprogramming to prevent potential mechanical interferences. See section 5.1 for instructions.
Main Components (Fig.3.1)
1. Headrest 2. Joystick 3. Armrest 4. Seat cushion 5. Leg rest 6. Backrest cushion
3.1 Maximum user weight The overall maximum user weight shall not exceed:
SEDEO Pro/Pro Advanced
– 300 lbs (136 kg)
SEDEO Pro/Pro Advanced w/C-ME® up to 3mph
– 300 lbs (136 kg)
SEDEO Pro/Pro Advanced Heavy Duty with recline – 350 lbs (159 kg)
SEDEO Pro Heavy Duty without recline
– 450 lbs (204 kg)
WARNING!
Exceeding the weight capacities could lead to injury and/or permanent damage
to the equipment.
NOTE — Exceeding the weight limit will void the warranty.
Quickie® SEDEO® PRO/PRO ADVANCED
1 2 6
3 4
5
Fig. 3.1
7
248020 Rev. G
Quickie® SEDEO® PRO/PRO ADVANCED
4.0 WARNINGS FOR COMPONENTS AND OPTIONS
WARNING!
If you use parts or make changes not authorized by Sunrise Medical it may
create a safety hazard and will void the warranty.
4.1 Seat depth adjustment
WARNING!
Seat depth adjustment may require reprogramming to prevent potential
mechanical interferences.
4.2 Armrests
DANGER!
1. Armrests will not bear the weight of the user. 2. Never lift this system
by its armrests. They may come loose or break. 3. When transferring the user
in or out of the seating system, never use the
armrest as a means of support.
4.3 Cushions
WARNING!
1. Standard foam cushions and other body supports are not designed for the
relief of pressure.
2. If you suffer from pressure sores, or if you are at risk that they will
occur, you may need a special seating system or a device that will help
control your posture. Consult your health care provider to find out if you
need such a device for your comfort and health.
4.4 Fasteners
WARNING!
Many of the screws, bolts, and nuts on this system are special high-strength
fasteners. Use of improper fasteners may cause your system to fail. 1. Only
use fasteners provided by Sunrise Medical. 2. If fasteners become loose,
tighten them immediately. 3. Do not over- or under-tighten fasteners as this
can cause damage to your
system or its components. 4. Always tighten to specified torque. Specified
torque values are listed in
Section 9. 5. Do not reuse fasteners.
4.5 Footplate and footrests
DANGER!
1. At the lowest point, footrests should be at least 2-in (51mm) off the
ground. If set too LOW, they may “hang up” on obstacles you can expect to find
in normal use. This may cause the system to stop suddenly and tip forward.
2. To avoid a trip or fall when you transfer: · Ensure your feet do not “hang
up” or get caught in the space between the footrests. · Avoid putting weight
on the footrests, as the system may tip forward. · Remove or swing the
footrests out of the way, if possible.
3. The footrest should always be in the down position when operating the
system.
4. Never lift this system by the footrests. Footrests detach and will not
bear the weight of this system. Lift this system only by non-detachable parts
of the main frame.
4.6 Positioning belts
DANGER!
The positioning belt is predominately used to support your posture. It can
also be used to limit slipping and/or sliding that you might experience when
the system is in motion. The positioning belt is not a transit rated safety
belt and should not be used in the place of a seat belt while being
transported in a motor vehicle. Improper use of positioning belts may cause
severe injury or death. If you use a positioning belt, ensure you follow the
recommendations in this section:
1. Ensure the rider does not slide down in the seating system seat. If this
occurs, the rider may suffer chest compression or suffocate due to pressure
from the belts.
2. The belts must be snug, but must not be so tight that they interfere with
breathing. You should be able to slide your open hand, flat, between the belt
and the rider.
3. A pelvic wedge or a similar device can help keep the rider from sliding
down in the seat. Consult with the rider’s doctor or health care provider to
find out if the rider needs such a device.
4. Use positioning belts only with a rider who can cooperate. Ensure the
rider or attendant can easily remove the belts in case of emergency.
5. NEVER use positioning belts: a. As a patient restraint. A restraint
requires a doctor’s order. b. On a rider who is comatose or agitated. c. As a
motor vehicle restraint. In an accident or sudden stop the rider may be thrown
from the system. Seating system positioning belts will not prevent this, and
further injury may result from the belts or straps.
If you fail to heed these warnings, damage to your system, a fall, tip-over,
or loss of control may occur and cause severe injury to the rider or others.
4.7 Push handles
DANGER!
1. Push handles provide secure points for an attendant to propel and control
the system. This helps to prevent a fall or tip-over.
2. Check to ensure push handle grips will not rotate or slip off.
4.8 Seating systems
DANGER!
1. Use of a seating system not approved by Sunrise Medical may affect or
interfere with other parts of the system.
2. Never change the seating system unless you consult your Sunrise Medical
authorized dealer first.
3. Never raise your seating system to a height of more than 19-in (482mm)
(measuring from the front of the seat pan to the floor) with all actuators in
their home position.
4.9 Upholstery fabric
DANGER!
1. Replace worn or torn fabric of the seat sling and/or seat back as soon as
possible. If you fail to do so, the seat may fail and cause you to fall. Worn
or torn fabrics may increase the potential for a fire hazard.
2. Fabric will deteriorate with age and use. Look for fraying, thin spots, or
stretching of fabric at rivet holes. Replace fabric as required.
3. Be aware that washing may reduce flame retardation qualities of the
fabric.
248020 Rev. G
8
Quickie® SEDEO® PRO/PRO ADVANCED
4.0 WARNINGS FOR COMPONENTS AND OPTIONS
4.10 Wiring and connectors
WARNING!
Never pull on cables directly. This can result in wires breaking inside the
connector or harness. To remove a plug or connector, always grasp the plug or
connector itself.
4.11 Assignable controls
DANGER!
It is possible to assign a function to an input control (buttons, jacks) to
your seating system. If controls have been assigned to perform a dual or
alternate function, ensure you know what each control operates. Please contact
your Sunrise Medical authorized dealer if you do not receive this information.
Failure to do so may result in damage and/or injury.
4.12 Power seating safety
DANGER!
1. Operation a. While driving this seating system in the reclined, elevated,
or tilted position use caution. b. Do not attempt to operate any power seating
option while positioned on an incline. c. Do not attempt to operate any power
seating option while this seating system is in motion. d. Do not attempt to
operate the recline, lift, or tilt power seating options around children. e.
Through programming it is possible to reverse the direction of all powered
seating functions. Ensure you know which direction your seat is going to move
before operating. Please discuss with your Sunrise Medical authorized dealer
which direction the power seating functions have been programmed.
2. Reduced Speed Driving (Creep and/or C-Me) This power system is designed to
automatically reduce the maximum driving speed to a creep and/or C-Me mode
when certain power seating limits are reached. Depending on the configuration
of the chair there might be multiple creep and/or C-Me speeds. Please discuss
the multiple creep and/or C-Me speeds with your Sunrise Medical authorized
dealer.
Before operating your chair in creep and/or C-Me mode, read all warnings and
guidelines related to this feature. They will help in understanding how to
reduce the risk of a fall or tip-over and how to identify environmental
conditions that could affect your safety when driving while the chair is in
creep and/or C-Me speed.
a. When driving indoors or outdoors, in creep mode, it should only be used on
flat, level, and hardscape surfaces.
b. Driving in creep mode should never be attempted on uneven terrain like:
gravel, grass, bumpy surfaces, uneven sidewalks, or hills.
c. This feature may be equipped with a system that allows the power chair to
drive up to 3 mph when the seat is elevated. This feature is called C-Me.
d. NEVER raise the seat when the chair is being operated in the proximity of
any of the terrain types listed above, or on any uneven surface.
3. Pinch points Pinch points may occur when operating the powered seating
options on this system. Ensure all hands and body parts are clear of all power
seating components that may result in pinch points before operating them.
4. Tilting power base with push handles Do not attempt to tilt power
wheelchairs by pulling down on the push handles. Power wheelchairs have heavy
bases. As such, attempting to tilt a power seating system to traverse
obstacles can damage components in the seating system and/or modular
actuators.
5. Remain clear during power accessory operation Please stay clear of any
powered actuator while the component is in motion. Sunrise Medical power
mechanisms move over large ranges of travel. Users should be aware of their
surroundings while components are in motion.
6. Maximum user weight Do not exceed the weight capacity for your specific
system/seating system configuration. Doing so could result in premature
failure or injury.
7. Latch mode Use caution while operating any power seating function in
latched mode. In latch mode, powered seating will not stop until a reverse
command is performed or the end of travel is met. Please consult with your
Sunrise Medical authorized dealer to receive training before using latch mode.
Failure to do so may result in potential serious injury.
4.13 Power recline
DANGER!
Pinch Point Avoid putting hands or fingers near the power recline mechanism
while in operation.
Power Recline Weight Capacity The maximum user weight capacity for the power
recline is up to 300 lbs (136 kg), 350 lbs (158 kg) for heavy duty chair.
Exceeding this weight capacity could lead to injury and/or permanent damage to
the equipment.
4.14 Power leg rest
DANGER!
Leg rest Weight Capacity The power leg rest accessory has a maximum user
weight capacity of up to 450 lbs (204 kg) depending on your system
configuration. Never exceed this capacity while operating the powered
actuator.
Anterior Tilt If equipped with anterior tilt the power leg rest accessory has
a maximum user weight capacity of up to 300 lbs (136 kg). Never exceed this
capacity while operating the powered actuators or wheelchair.
4.15 Lift, Tilt, Anterior Tilt, or Combo
DANGER!
Pinch Point Avoid putting hands, fingers, clothing items, or letting children
near the seating system when the Tilt/Lift mechanism is in operation.
Weight Capacity The weight capacity is a maximum user weight of up to 300 lbs
(136 kg), 450 lbs (204 kg) for heavy duty tilt. Never exceed this capacity
while operating the powered actuator.
9
248020 Rev. G
Quickie® SEDEO® PRO
4.0 WARNINGS FOR COMPONENTS AND OPTIONS
4.16 C-Me® Lift
DANGER!
The C-ME accessory has a maximum user weight capacity of 300 lbs (136 kg) with
a maximum speed of 3 mph at full lift. Never exceed this capacity while
operating the lift.
WARNING!
Exceeding these weight capacities could lead to injury and/or permanent damage
to the equipment.
Note: Exceeding the weight limit will void the warranty.
4.17 On/Off switch
WARNING!
1. Only use the ON/OFF switch to stop the chair in an emergency. This will
result in an abrupt stop, and may cause you to fall.
2. To slow your chair to a stop, return the joystick to neutral.
4.18 Suspension springs
WARNING!
The suspension springs are set to factory defaults when the chair is
manufactured. These springs should not need to be adjusted and/or detached
from the chair. Adjusting these springs could cause instability.
For the safety of the user and stability of the chair:
1. NEVER use a spanner wrench to adjust the springs on this chair. 2. DO NOT
remove the suspension springs. 3. If the springs need replacing, contact
Sunrise Medical Customer Service
for parts and service information.
4.19 Anterior tilt position
WARNING!
1. Moving toward an anterior tilt position and operation in a anterior tilt
position must only be done on a flat surface. When the chair is in anterior
tilt position, never reach out from the chair, due to the higher tipping risk.
2. The operation of the anterior tilt function (and other predefined seating
positions) will cause multiple actuators to operate simultaneously.
3. If the seating system operates unintentionally or erratically immediately
switch off the system with the On/Off button.
4.20 Swing away joystick arm
DANGER!
Before adjusting the swing away arm, switch off the controller to avoid
accidental displacement of the joystick which would cause unwanted movement of
your wheelchair. Be aware that the width of your chair may have increased and
you may not be able to go through certain obstacles.
DANGER!
1. Only operate the wheelchair at a low and maneuverable speed when the
parallel swing-away is in use. (Fig. 4.1)
2. When transferring to and from the wheelchair, do not use the remote as a
means of support.
3. Do not hang any items on or over the parallel swing-away remote assembly
as this could damage the swing-away mechanism.
Note: hanging items on/near the joystick may cause unwanted movement and
potential of serious injury or death!
Fig. 4.1
248020 Rev. G
10
Quickie® SEQDuEiOck®ieP®RSOED/PEROO®APDRVOANCED
4.0 WARNINGS FOR COMPONENTS AND OPTIONS
4.21 Transit Important Transit notes for the chair:
DANGER!
1. If possible and feasible, the rider should transfer to the Original
Equipment Manufacturer (OEM) vehicle seat and use the vehicle restraint.
2. If the seating system is used for transport, it must be used in a forward-
facing seating position during travel in a motor vehicle.
3. The recommended clear zones for wheelchair seated occupants restrained by
both pelvic and upper-torso belt(s) and only by a pelvic belt are shown in the
diagrams (Fig. 4.2 and 4.3) and described below. Frontal Clear Zones (FCZ)
need to be larger when upper torso belt(s) are not used. The Rear Clear Zone
of 16-in (40.5 cm) is measured from the rearmost point on an occupant’s head.
The Front Clear Zone is measured from the front most point of an occupant’s
head and is 26-in (66 cm) with pelvic and upper-torso belt(s) and 37-in (94
cm) with only a pelvic belt. The frontal clear zone may not be achievable for
wheelchair-seated drivers.
4. The estimated seated height (HHT) from the ground or floor to the top of
the wheelchair-seated occupant’s head ranges from approximately 47-in (119 cm)
for a small adult female to about 61-in (155 cm) for a tall adult male (Fig.
4.3).
5. The rider must not weigh more than the weight capacity of the wheelchair.
6. Attach WTORS to securement points in accordance with the manufacturer’s
instructions and SAE J2249. 7. Ensure hip guides are mounted and are locked in
place. 8. Only use Wheelchair Tie-down and Occupant Restraint Systems (WTORS)
which meet the require-
ments of SAE J2249 Recommended Practice-Wheelchair Tie-down and Occupant
Restraint Systems For Use in Motor Vehicles. Do not use WTORS designed to rely
on the wheelchair structure to transfer occupant restraint loads to the
vehicle. 9. In order to reduce the potential of injury to vehicle occupants,
wheelchair mounted accessories, such as trays and respiratory equipment,
should be removed from the seating and wheelchair and secured separately. 10.
Postural supports and positioning devices should not be relied on for occupant
restraint. 11. Do not alter or substitute wheelchair frame parts, components,
or seating. 12. Sudden stops or impacts can structurally damage your system.
Chairs involved in such incidents should be replaced.
Important Transit notes for Occupant restraints:
13. Attach occupant restraints in accordance with the manufacturer’s
instructions and SAE J2249. 14. Both pelvic and upper torso restraint belts
must be used to restrain the occupant to reduce the possi-
bility of head and chest impacts with the vehicle components and serious risk
of injury to the user and other vehicle occupants (Fig. 4.4) 15. The upper
torso restraint belt (A) should be mounted to the vehicle (B) pillar. Failure
to do so will increase the risk of serious abdominal injuries to the user
(Fig. 4.4). 16. A head restraint suitable for transportation must be fitted
and suitably positioned at all times during transportation. 17. The pelvic
restraint belt (C) must be worn low across the front of the pelvis so that the
angle of the pelvic belt is within the preferred zone of 30 to 75 degrees to
the horizontal (Fig. 4.4). 18. A steeper (greater) angle within the preferred
zone is desirable closer to, but not exceeding, 75 degrees (Fig. 4.6). 19. The
upper torso restraint belt (A) must fit over the shoulder and across the chest
(Fig. 4.4). 20. Restraint belts must be adjusted as tightly as possible
consistent with user comfort, and should not constrict breathing (Fig. 4.5).
Fig.4.2 Fig.4.3 Fig.4.4 Fig.4.5
Fig.4.6
11
248020 Rev. G
Quickie® SEDEO® PRO/PRO ADVANCED
5.0 SETTING UP THE SEATING SYSTEM
5.0 Setting up the seating system The average wheelchair does not exist. This
is why Sunrise Medical wheelchairs can be pre-set according to specific needs
of the user.
Pre-settings are carried out once, the first time the wheelchair is going to
be used. They must be executed by a qualified specialist, unless explicitly
stated otherwise.
Tool and calibration marks A M5 Allen key is supplied with the seating system.
This can be used by a Sunrise Medical authorized dealer for adjusting all the
seating system settings.
CAUTION!
There are calibration marks on various parts of the seating system. The
factory settings are indicated.
5.1 Pre-set backrest frame depth (Fig. 5.1 and 5.2) 1. Loosen the seat pan
bolts (A and B) by one turn. 2. Adjust the seat back (C) to the desired
position. 3. Tighten the seat pan bolts (A and B). See section 9 for specified
torque values.
CAUTION!
1. When a bigger seat depth is required, ensure that the harnesses are
lengthened before the mechanical adjustment takes place.
2. There are 2 sizes seat rail available: 15.75-in – 20.50-in (400-520 mm),
and 19-in – 22-in (480-560 mm). Choose the correct seat rail to match the
adjusted seat depth.
3. Mount extra seat depth mounting plate for seat depth 19-in – 22-in for
more stability of the long seat rail (check with Sunrise Medical Customer
Service for COG mounting location). (Fig. 5.3)
4. For Anterior Tilt only: there are two programming sets, one for seat depth
17-in to 19-in and one for seat depth 20-in to 22-in. If adjusting seat depth
from one range to the other, new program is required to avoid potential
mechanical interferences. Contact Sunrise Technical Support to receive a new
program.
DANGER!
We strongly recommend that you contact your Sunrise Medical authorized dealer
to adjust the seat depth, because adjusting the seat depth will affect the
center of gravity.
Fig.5.1 Fig.5.2
Fig.5.3
248020 Rev. G
12
5.0 SETTING UP THE SEATING SYSTEM
Quickie® SEDEO® PRO/PRO ADVANCED
5.2 Pre-set armrests
Flip back armrests / Single post removable armrests width (Fig. 5.4) Fine
adjustments can be done with the armrest only. For larger adjustments also
pre-set the seat plate width.
1. Loosen (A) by one turn. 2. Adjust (B) to the desired position. 3. Tighten
(A). See section 9 for specified torque values.
CAUTION!
Please check that the new armrest position does not interfere with any seat
positioning.
Reclining flip back armrest width (Fig. 5.5) 1. Loosen (A) by one turn. 2.
Adjust (B) to the desired position. 3. Tighten (A). See section 9 for
specified torque values.
CAUTION!
Ensure that the new armrest position does not interfere with any seat
positioning.
Seat plate width (Fig. 5.6 – Fig. 5.8) Lift the seat cushion off the hook and
loop strips to access the seat plates.
For standard seat pans:
1. Loosen (A) by one turn. 2. Adjust (B) to the desired position. 3. Tighten
(A). See section 9 for specified torque values.
For HD seat pans: 1. Loosen (A) and (F) by one turn. 2. Adjust (B) to the
desired position. (Fig. 5.6) 3. Tighten (A) and (F). See section 9 for
specified torque values.
If adjusting seat width to 22-in or less: 4. Remove screws (C), then slide
seat rails (D) out of the mounting channel. 5. When the rail is removed,
remove the support brackets (E) and set aside. 6. Attach again the rail (D)
and add screws (C). 7. Adjust seat width per steps 1-3.
The seat width is indicated by the scale that runs alongside each slot (B).
Fig.5.4 Fig.5.5 Fig.5.6
Fig.5.7
Fig.5.8
13
248020 Rev. G
Quickie® SEDEO® PRO/PRO ADVANCED
5.3 Pre-set backrest Backrest angle (Fig. 5.9) 1. Turn (A) in order to adjust
(B). Backrest height with fixed back or reclining back (Fig. 5.10) 1. Loosen
(A) by one turn. 2. Adjust (B) to the desired position. 3. Tighten (A). See
section 9 for specified torque values.
5.0 SETTING UP THE SEATING SYSTEM
BA
1
B
1 M5 Fig.5.9
5.4 Pre-set flip back armrests / single post removable armrests
Armrest height (Fig. 5.11) 1. Loosen (A) by one turn. 2. Adjust (B) to the
desired position. 3. Tighten (A). See section 9 for specified torque values.
Ensure that the arms are supported well, without the shoulders being pushed
upwards.
Armrest depth (Fig. 5.12) 1. Loosen (A) by one turn. 2. Adjust (B) to the
desired position. 3. Tighten (A). See section 9 for specified torque values.
Position the flip back armrest further back in order to obtain more space for
sidewards transfers. Position the armrest further to the front to obtain more
support for forward transfers.
Armrest pad depth (Fig. 5.13) 1. Loosen (A) by one turn. 2. Adjust (B) to the
desired position. 3. Tighten (A). See section 9 for specified torque values.
2
A
1
3
M5
1
M5
3
Fig.5.10
B
A
2 Fig.5.11
1 3B
A
2
1 M5
2 M5
Fig.5.12
B
2
248020 Rev. G
11 M5
A
3
3 M5 Fig.5.13
14
5.0 SETTING UP THE SEATING SYSTEM
Quickie® SEDEO® PRO/PRO ADVANCED
Armrest pad width To remove: (Fig. 5.14 and 5.15) 1. Loosen screws (A). 2.
Remove armpad (B). 3. Loosen screws (C) by one turn. 4. Remove mounting plate
(D).
To position: (Fig. 5.16 and 5.17) 5. Slide slot (E or F) onto clamp nuts (G).
CAUTION!
Ensure that the slot (E or F) and clamp nuts (G) are fully engaged. 6. Adjust
the depth of (D) and tighten screws
(C). See section 9 for specified torque values.
To attach again: (Fig. 5.14) 7. Attach again the armpad (B) with screws (A).
See section 9 for specified torque values.
Armrest angle (Fig. 5.18) 1. Loosen (A) by one turn. 2. Adjust (B) to the
desired position. 3. Tighten (A). See section 9 for specified
torque values.
Side guard height (Fig. 5.19) 1. Loosen (A) by one turn. 2. Adjust (B) to the
desired position. 3. Tighten (A). See section 9 for specified
torque values.
B 2
A
1 M5
Fig.5.14
GD
5 DA
B
2
BE CF Fig.5.16
A 1
2 3
B
Fig.5.18
DF
4 CE 3 M5
Fig.5.15
DA
CE
6 M5
Fig.5.17
A
1 3
Fig.5.19
15
248020 Rev. G
Quickie® SEDEO® PRO/PRO ADVANCED
5.0 SETTING UP THE SEATING SYSTEM
5.5 Pre-set reclining flip back armrests
Armrest height Fine or gross adjustment, it is a two-part process:
Part 1 (Fig. 5.20) 1. Loosen (A) in order to adjust (B).
Part 2 (Fig. 5.21) 2. Loosen (A) in order to adjust (B). 3. Retighten screws
(A-Fig. 5.20 and A-Fig. 5.21). See section 9 for specified torque values.
Armrest pad depth (Fig. 5.22) 1. Loosen both screws (A) by one turn. 2. Adjust
(B) to the desired position. 3. Tighten (A). See section 9 for specified
torque values.
Armrest width adjustment (Fig. 5.23 and 5.24) 1. Using 8mm hex tool, loosen
both screws (A) by one turn. 2. Adjust (B) in or out to the desired position.
3. Tighten (A). See section 9 for specified torque values.
WARNING!
All screws should be checked periodically for tightness. A loose arm pad could
slip down when pressed. See maintenance schedule in section 9.
B
A Fig.5.20 B
A Fig.5.21
B
A Fig.5.22
B
248020 Rev. G
A Fig.5.23
B
B Fig.5.24
16
5.0 SETTING UP THE SEATING SYSTEM
Quickie® SEDEO® PRO/PRO ADVANCED
5.6 Pre-set fixed center mount leg rest (short and long)
Removing the assembly (Fig. 5.25) 1. Remove the 3 screws (A) and lower the
footrest assembly from the seat. (Use a 5mm hex key)
Lower leg length (Fig. 5.26) 1. Remove the 6 securing screws (B). (Use a 5mm
hex key) 2. Move footplate up or down to desired height. 3. Attach again
screws (B) through adjustment holes and tighten the hardware. See section 9
for specified
torque values.
Calf pad height adjustment (Fig. 5.27) 1. Remove the 2 securing screws (C).
(Use a 5mm hex key) 2. Move calf pads to desired height. 3. Tighten screws (C)
through adjustment holes. See section 9 for specified torque values.
Fig.5.25
Fig.5.26
Fig.5.27
17
248020 Rev. G
Quickie® SEDEO® PRO/PRO ADVANCED
5.0 SETTING UP THE SEATING SYSTEM
5.7 Pre-set Sedeo Seating fixed center mount leg rest
Removing the assembly (Fig. 5.28) 1. Remove the two screws (A). (Use a 5mm hex
key) 2. Fold forward the neoprene cover (C). 3. By hand, remove the two
spacers (B). 4. Remove one screw (D) and lower the footrest assembly from the
seat.
Lower leg length (Fig. 5.29) 1. Loosen the two screws per footplate (D) and
adjust to the desired height. (Use a 5mm hex key) 2. When finished adjusting,
tighten screws (D). (Use a 5mm hex key) See section 9 for specified torque
values.
Calf pad height adjustment (Fig. 5.29) 1. Loosen the two screws (A) per calf
pad and adjust calf pads (C) to desired height. (Use a 5mm hex
key) 2. When finished adjusting, tighten screws (A). (Use a 5mm hex key) See
section 9 for specified torque
values.
Fig.5.28
5.8 Pre-set power center mount leg rest
DANGER!
1. Be aware of your environment to ensure you do not injure your legs when
leg rests are extended. 2. Always ensure that the leg rests or foot plates do
not come into contact with the casters before
driving the wheelchair. 3. Leg rests are not to be used for lifting or
carrying the wheelchair under any circumstances. 4. As with all moving parts,
be careful not to trap your fingers. (Fig. 5.30)
Removing the assembly (Fig. 5.31) 1. Remove the three screws (A) and lower the
footrest assembly from the seat. (Use a 5mm hex key) Note: Neoprene cover not
shown for clarity.
Fig.5.29
Fig.5.30 A
248020 Rev. G
18
Fig.5.31
5.0 SETTING UP THE SEATING SYSTEM
Quickie® SEDEO® PRO/PRO ADVANCED
Lower leg length (Fig. 5.32) 1. Loosen the two screws per footplate (D) and
adjust to the desired height. (Use a 5mm hex key) 2. When finished adjusting,
tighten screws (D). (Use a 5mm hex key) See section 9 for specified torque
values.
Calf pad height adjustment (Fig. 5.32) 1. Loosen two screws (A) per calf pad
and adjust calf pads (C) to desired height. (Use a 5mm hex key) 2. When
finished adjusting, tighten screws (A). (Use a 5mm hex key) See section 9 for
specified torque
values.
Ankle angle (Fig. 5.33) 1. Flip up the footplate (A). 2. Turn (B) in order to
adjust the ankle angle as desired. 3. Fold down the footplate (A). 4. Repeat
on the opposite side.
DANGER!
1. After adjustment and with the seat in its normal drive position, ensure
there is a gap of at least 2-in (51mm) between the floor and the footplate at
all times.
2. Perform this test with the user sitting in the chair loading the leg rest.
3. Using the chair with the leg rest set too low might result in personal
injury or damage to the
wheelchair and its surroundings.
Width adjustment (Fig. 5.34) 1. Remove (A) using a 6mm Allen wrench. 2. Move
(B) to the desired side of the footplate mount (C). 3. Attach again (A)
through items (B, D, and E). 4. Torque (A) to 13.5-14.9 Nm (120-132 in-lbs).
After the bolt is torqued, the footplate should still rotate
up and down but with 0.5-5 lbs of resistance.
A B
Fig.5.32
M4
2 1
Fig.5.33
Fig.5.34
19
248020 Rev. G
Quickie® SEDEO® PRO/PRO ADVANCED
5.0 SETTING UP THE SEATING SYSTEM
5.9 Pre-set swing away leg rest
Leg rest width (Fig. 5.35) 1. Loosen (A) by one turn. 2. Adjust (B) to the
desired position. 3. Tighten (A). See section 9 for specified torque values.
Lower leg length (Fig. 5.36) 1. Loosen (A) a few turns. 2. Adjust (B) to the
desired position. 3. Tighten (A). See section 9 for specified torque values.
CAUTION!
Ensure that there is at least 2-in (51mm) of free space under the footplate
for the wheelchair to drive over obstacles.
Ankle angle (Fig. 5.37) 1. Loosen (A) a few turns. 2. Adjust (B) to the
desired position. 3. Tighten (A). See section 9 for specified torque values.
Knee angle (Fig. 5.38) The position of the foot plates can be changed by
adjusting the knee angle. 1. Loosen (A) a few turns. 2. Adjust (B) to the
desired position. 3. Tighten (A). See section 9 for specified torque values.
Note: Change the height or knee angle if the footplate touches the caster
wheel.
CAUTION!
Check after setup and/or adjustment of wheelchair. After the initial setup by
the Sunrise Medical authorized dealer or after making any adjustments to the
wheelchair, a check to confirm that no interferences exist must be completed.
Interferences can happen (but not limited to) adjustments to the joystick,
armrest, specialty control mounts, leg rest, headrest, etc. After an
adjustment has been made, the movement of that device may collide or hinder
the operation of another component of the wheelchair. A check must be
performed prior to operation of the chair to ensure all moving pieces move
freely and through the entire range of motion without coming into contact with
another part of the wheelchair.
1
A
3
M5
1
M5
2
3
B
Fig.5.35
A 13
2
M5
1
B
M5
3
Fig.5.36
B2
M5 M5
B
A1 3 Fig.5.37
2
A1 3
Fig.5.38
248020 Rev. G
20
5.0 SETTING UP THE SEATING SYSTEM
Quickie® SEDEO® PRO/PRO ADVANCED
Calf pad height (Fig. 5.39) 1. Loosen (A) a few turns. 2. Adjust (D) to the
desired position. 3. Tighten (A). See section 9 for specified torque values.
Calf pad depth (Fig. 5.39) 4. Loosen (B) a few turns. 5. Adjust (D) to the
desired position. 6. Tighten (B). See section 9 for specified torque values.
Calf pad width (Fig. 5.39) 7. Loosen (C) on the back of (D) a few turns. 8.
Adjust (D) to the desired position. 9. Tighten (C). See section 9 for
specified torque values.
5.10 Pre-set headrest
DANGER!
1. If a headrest is part of the options selected for the user/chair, never
drive the chair without the headrest attached.
2. Never use the chair without a headrest if a person is transported with the
chair in a vehicle.
Headrest height (Fig. 5.40) 1. Loosen (A) a few turns. 2. Adjust (B) to the
desired position. 3. Tighten (A). See section 9 for specified torque values.
Fixed height setting (Fig. 5.41) 1. Loosen (C) by one turn. 2. Slide (D) as
far down as possible. 3. Tighten (C). See section 9 for specified torque
values.
Headrest position (Fig. 5.42) 1. Loosen (A) by one turn. 2. Adjust (B and C)
to the desired position. 3. Tighten (A). See section 9 for specified torque
values.
Note: Correct the headrest height and tilt the headrest cushion as required.
Headrest cushion width (Fig. 5.43) 1. Loosen (A) by one turn. 2. Adjust (B) to
the desired position. 3. Tighten (A). See section 9 for specified torque
values.
A13
D
2 5
B4 6
Fig.5.39 C7 9
8
Fig.5.40
Fig.5.41
Fig.5.42
Fig.5.43
21
248020 Rev. G
Quickie® SEDEO® PRO/PRO ADVANCED
5.11 Pre-set fixed / swing away controller Bracket depth of the controller
(Fig. 5.44) 1. Loosen (A) by one turn. 2. Adjust (B) to the desired position.
3. Tighten (A). See section 9 for specified torque values. Height of the
controller (1) (Fig. 5.45) 1. Loosen (A) by one turn. 2. Adjust (B) to the
desired position. 3. Tighten (A). See section 9 for specified torque values.
Height of the controller (2) (Fig. 5.46) 1. Loosen (C) by one turn. 2. Adjust
(D) to the desired position. 3. Tighten (C). See section 9 for specified
torque values.
5.0 SETTING UP THE SEATING SYSTEM
Fig.5.44
Fig.5.45
Fig.5.46
248020 Rev. G
22
5.0 SETTING UP THE SEATING SYSTEM
5.12 Pre-set laterals
Laterals width (Fig. 5.47) 1. Loosen (A) completely. 2. Adjust (B) to the
desired position. 3. Tighten (A). See section 9 for specified torque values.
Laterals height (Fig. 5.48) 1. Loosen (A) by one turn. 2. Adjust (B) to the
desired position. 3. Tighten (A). See section 9 for specified torque values.
Laterals depth (Fig. 5.49) 1. Loosen (A) by one turn. 2. Adjust (B) to the
desired position. 3. Tighten (A). See section 9 for specified torque values.
4. Loosen (C) by one turn. 5. Adjust (B) to the desired position. 6. Tighten
(C). See section 9 for specified torque values.
Laterals angle (Fig. 5.50) 1. Loosen (A) by one turn. 2. Adjust (B) to the
desired position. 3. Tighten (A). See section 9 for specified torque values.
4. Loosen (C) by one turn. 5. Adjust (B) to the desired position. 6. Tighten
(C). See section 9 for specified torque values.
Quickie® SEDEO® PRO/PRO ADVANCED
Fig.5.47 Fig.5.48
Fig.5.49
Fig.5.50
23
248020 Rev. G
Quickie® SEDEO® PRO/PRO ADVANCED
5.0 SETTING UP THE SEATING SYSTEM
5.13 Lap belt / seating positioning belt (Fig. 5.51)
DANGER!
1. This product is only to be used to position a single person in a
wheelchair. 2. Lap belts are not suitable for transit purposes. Approved
occupant restraint systems must be used. 3. Failure to heed these warnings may
result in severe injury or death. 4. Ensure that the carer or attendant is
trained in the correct operation of the belts. 5. Untrained carers or
attendants may lead to delays, if an emergency situation occurs.
Mounting a lap belt /seating positioning belt 1. Attach (A) to the seating
frame using provided screws on the left and on the right. (Fig. 5.52) 2.
Tighten (B). See section 9 for specified torque values.
Positioning a person with a lap belt 1. Adjust the lap belt to suit, leaving
no more than a hand’s width gap for comfort and safety. (Fig. 5.53) 2. The
hand clearance should be with the lap belt under normal tension and not allow
large gaps or
loops. 3. Generally, the lap belt should be fixed so that the belt sits at an
angle of approximately 45° (Fig. 5.54),
and when correctly adjusted should not allow user to slip down in the seat. 4.
Place the belt loosely across the seat with the opening end of the buckle
facing to the right for a
left-handed person and to the left for a right handed person. Pass the other
ends of the belt through the gap between the backrest posts and the backrest.
DANGER!
1. Always ensure that the lap belt is correctly secured and adjusted prior to
use. 2. Too loose a belt could cause the user to slip down and cause serious
injury. 3. Check lap belt and securing components at regular intervals for any
signs of fray or damage. Replace if
necessary. 4. When servicing, check for correct operation of the release
buckle and for any signs of wear on the
material or plastic brackets. 5. Regular checks/activities:
· As with all positioning components, there is a need for corrective
adjustments as the person changes their seating position over time.
· Check the belts regularly for correct fitting, to ensure the safety and
comfort of the user.
CAUTION!
1. Standard Sling: Lap strap movement is restricted by upholstery. 2.
Contoured Back: Use universal bottom bracket as per recline back method.
Fig.5.51 A
B Fig.5.52
Fig.5.53
Fig.5.54
248020 Rev. G
24
5.0 SETTING UP THE SEATING SYSTEM
Quickie® SEDEO® PRO/PRO ADVANCED
5.14 Setting the height of the chest strap (Anterior Tilt) (Fig. 5.55) NOTE: A 5mm Hexagon wrench (Allen Key) is required.
Positioning 1. Seat the user into the chair in the correct
seating position. 2. Loosen the Allen screw to be able to position
the chest strap at the desired height. 3. Tighten the Allen screws to a torque
value of
6.8-7.9 Nm [5 to 5.8 ft-lbs].
Fig. 5.55
DANGER!
1. To prevent choking hazards the chest strap may not be adjusted too high or
too low relative to the user’s body.
2. Ensure the rider does not slide in the wheelchair seat. If this occurs,
the rider may suffer chest compression or suffocate due to pressure from the
chest strap.
3. Never use the chest strap as a patient restraint. 4. Never use the chest
strap on a user who is comatose or agitated.
CAUTION!
Ensure the chest strap and/or positioning belt position does not interfere
with any other device attached to the wheelchair.
Adaptation adjustment 1. Seat the user into the wheelchair. 2. Feed the buckle
on the chest strap through the loop. (Fig. 5.56) 3. Pull the chest strap so
that it fully supports the user’s upper body firmly and comfortably. (Fig.
5.57) 4. Ensure that chest strap is not too tight. 5. Close the buckle until a
loud click is heard. (Fig. 5.58) 6. The D-Ring facilitates easy adjustment for
the user. (Fig. 5.59) 7. Ensure the chest strap fully supports the user and
that there is enough room to fit a hand’s width be-
tween the user and the support. (Fig. 5.60)
Safety: In addition to the chest strap latches, the chest strap incorporates
two further safety features that prevent the unintended opening of the belt.
- Each of the chest strap latches is secured with an overlaying safety hook
and loop strap. 2. In addition, on each side of the chest strap there is a
securing strap which is guided through the D-Ring,
adding another degree of safety.
WARNING!
1. Never use the anterior tilt function without the knee fixation, chest strap, and positioning belt attached and safely locked.
2. Ensure the rider does not slide in the wheelchair seat. If this occurs, the rider may suffer chest compression or suffocate due to pressure from the supports.
3. Never use positioning supports as a patient restraint. 4. Never use positioning supports on a user who is comatose or agitated.
Fig. 5.56 Fig. 5.57 Fig. 5.58 Fig. 5.59
Fig. 5.60
25
248020 Rev. G
Quickie® SEDEO® PRO/PRO ADVANCED
6.0 USING THE SEATING SYSTEM
6.1 Getting in and out of the wheelchair (Transferring) Sunrise Medical recommends that you consult your health care provider for assistance in developing your personal front or side transfer technique to best suit your needs and avoid any personal injury.
DANGER!
To prevent a fall:
1. Ensure controller is switched off during transfers to avoid unintentional movement. If you fail to do so
you may touch the joystick and cause your system to move when you do not expect it.
2. Ensure that the free wheel switch has been set to “drive”. (
General manual)
3. Rotate the front casters so that they point forward.
4. Work with your health care provider to learn all safe methods for transfers.
5. Ensure motor brakes are engaged. This keeps the system from moving when you transfer.
6. It is dangerous to transfer on your own. It requires good balance and agility. Be aware that there is a
point during every transfer when the seating system seat is not below you.
7. Learn how to position your body and how to support yourself during a transfer.
8. Have someone help you until you are sure you can do a safe transfer on your own.
9. Do not stand on the footrest. Putting your full weight on the footrest may lead to wheelchair damage
and to hazardous situations, including the wheelchair tipping up.
10. Ensure your feet do not “hang up” or get caught in the space between the footrests.
6.2 Preparing for forward transfer
DANGER!
Move your wheelchair as close as possible to the seat you intend to transfer
to.
Swing-away leg rests (Fig. 6.1) 1. Fold up the foot supports (A). 2. Swing the
calf plates (B) to the side (if applicable). 3. If possible, push (C) to the
rear and swing leg rests (D) to the side.
Centermount foot plate (Fig. 6.2) 1. Flip up (A) so they are out of the way.
Lower the centermount foot plate (Fig. 6.3) 1. Ensure the area is level and
free of obstacles and hazards. 2. Turn the control system ON. 3. Select the
powered centermount leg rest. 4. Push the joystick forward until the
footplates rest fully on the ground. (Fig. 6.3) 5. Switch the control system
OFF before transferring.
B
2 1
A
C D 3
Fig.6.1
Fig.6.2
Fig.6.3
248020 Rev. G
26
6.0 USING THE SEATING SYSTEM
Remove the front knee fixation (Anterior Tilt configuration only)
To remove (Fig. 6.4, Fig. 6.5) 1. Press the Red lever, on the top of the front
knee fixation assembly. 2. Gently lift the assembly up and out of the
receiver.
To refit (Fig. 6.4, Fig. 6.5) 1. Press the Red lever, on the top of the front
knee fixation assembly. 2. Place the lower edge of the tube into the receiver.
3. Push home and release the button. 4. Ensure the assembly is fixed firmly in
place.
Lower the center mount footplate (
) R-Net Control System
1. Ensure the area is level and free of obstacles and hazards.
2. Turn the control system ON.
3. Select the powered center mount leg rest.
4. Push the joystick forward until the footplates rest fully on the ground.
5. Switch the control system OFF before transferring..
Quickie® SEDEO® PRO/PRO ADVANCED
Fig.6.4
Fig.6.5
27
248020 Rev. G
Quickie® SEDEO® PRO/PRO ADVANCED
6.0 USING THE SEATING SYSTEM
6.3 Preparing for side transfer
DANGER!
1. Move your system as close as you can to the seat you are transferring to.
If possible, use a transfer board.
2. Transfer as far back onto the seat surface as you can. This will reduce
the risk that you will miss the seat or fall.
CAUTION!
1. Some versions of the controller will need to be swung aside. 2. Remove the
tray before folding back the armrest. 3. Do not lean on the armrest when
folded back.
Flip back armrest (Fig. 6.6) 1. Press (A). 2. Fold (B) to the rear.
Single post removable armrest (Fig. 6.7) 1. Tip (A) downwards. 2. Remove (B).
Reclining flip back armrest (Fig. 6.8) 1. Flip (A) to the rear.
B
2
1
A
Fig.6.6
B
2
A
1
Fig.6.7
1
A
6.4 Manual recline backrest (Fig. 6.9) To manually recline backrest: 1. Pull
up lever (A). 2. Lean forward or backward to adjust (B) to the desired
position. Note: Keep pulling up on (A). 3. Release (A) when desired position
is obtained.
6.5 Manually elevating leg rests knee angle (Fig. 6.10) The angle of the
comfort leg rests can be adjusted independently of each other. 1. Push handle
(A) forward. 2. While sitting in the chair, lift your leg. 3. Release the
handle when in the desired position.
Fig.6.8
2
B A
1 3
Fig.6.9
A
248020 Rev. G
28
Fig.6.10
6.0 USING THE SEATING SYSTEM
6.6 Headrest Removing / positioning the headrest (Fig. 6.11) 1. Loosen knob
(A). 2. Slide (B) out of or into bracket (C). 3. Tighten (A).
Quickie® SEDEO® PRO/PRO ADVANCED
6.7 Swing-away controller The joystick is mounted with a mechanism that allows
the joystick to be locked in two forward positions. (Fig. 6.12)
Swing-away the controller bracket (Fig. 6.13) 1. Push (A) outward or inward.
2. To return to forward position, push the back of the joystick forward until
it locks into place (click). Note: Swinging back the controller bracket will
allow you to pull up close to a table. Full operation of the controller
remains possible.
Fig.6.11 Fig.6.12
6.8 Laterals (Fig. 6.14) 1. Press (A) to swing away (B).
6.9 Push handle height (Fig. 6.15) 1. Loosen (A) by one turn. 2. Slide (B)
into the desired position. 3. Tighten (A). See section 9 for specified torque
values.
DANGER!
Do not use the push handles to tip the whole chair to the rear (in order to go
up a sidewalk, for example). The push handle is not suitable for such use.
29
Fig.6.13
B A
1 Fig.6.14
Fig.6.15 248020 Rev. G
Quickie® SEDEO® PRO/PRO ADVANCED
6.0 USING THE SEATING SYSTEM
6.10 Electric seating function The seating system has an electric seating
function. This function is the seat position to which all the memory seat
positions (described in Section 7) come back to by pulling the joystick into
the backward direction when in the seating memory mode.
DANGER!
When the anterior tilt or supine function is used, the center of gravity is
moved away from the original safe position into a raised or retracted
position. This can increase the likelihood of tipping over. So only use this
seat function on a level floor and drive in the lowest speed. Not doing so
could lead to personal injury.
DANGER!
The use of knee fixation and chest straps are mandatory when using the
anterior tilt function.
WARNING!
1. Before using the anterior tilt function ensure that the chest belt is
secure. 2. Before using the anterior tilt function ensure that the front knee
fixation is in place. 3. Only use the anterior tilt function on a sound and
level surface.
6.11 Seat memory position The control system on the seating system has up to 6
predefined seat memory positions. Each ordered memory location has a set of
factory seat positioning presets: 1. Seated 2. Supine 3. Pressure Relief 4.
Feet Above Heart or Anterior Tilt 5. Transfer 6. C-ME® (if equipped) The seat
actuators are all working with a feedback system to track their positions and
movements.
This is important: 1. To reach the required memory position. 2. To prevent
mechanical clashes of the seat. 3. To operate the actuators especially towards
the memory positions in an ergonomic, smooth, and safe
way.
All ordered memory positions, if not suitable for the user, can be changed and
customized to the needs required. The software takes into account the range of
the available mechanical movements and also all safety aspects.
6.12 Retrieving a memory seat position (Fig. 6.16 to 6.18) 1. Press the “Mode”
button (A) one or more times until a seat icon appears in the LCD control
panel
display. 2. Move the joystick to the left or right to select a memory seat
position (Seating, Supine, Pressure Relief,
Feet Above Heart, Transfer, or C-ME®). A seat icon and the memory description
for the relevant position are shown in the LCD control panel display. 3. Move
and hold the joystick forwards. The seat adjusts to the factory preset
position. The movement can be stopped at any time and started again when
moving the Joystick forward. 4. In any of the 6 memory seat positions moving
the Joystick backwards brings the seat into the “Seating” memory position.
Once the seat has adjusted to the saved position, it stops moving.
NOTE:
1. Movement of the seat can be stopped at any time by releasing the joystick.
2. During a stop period, another memory position can be selected with a right
or left movement of the
Joystick. After selecting a new Memory position, moving the joystick forward
will operate the seat into the new selected position.
6.13 Return to drive mode Press the “Mode” button one or more times until a
standard display image with speed indicator appears in the LCD control panel
display.
Fig.6.16 Fig.6.17 Fig.6.18
248020 Rev. G
30
6.0 USING THE SEATING SYSTEM
Quickie® SEDEO® PRO/PRO ADVANCED
6.14 Saving a new position to memory
Fig.619
1. Remove the rear cover of the backrest and connect a switch with a 3.5mm Jack to the I6 harness
socket (A) on the CxSM (Complex Seating Module) (Fig. 6.19).
2. Press the “Mode” button one or more times until a seat icon appears in the LCD control panel display
(Fig. 6.20).
3. Move the joystick to the left or right to select a memory seat position (Seating, Supine, Pressure Relief,
Feet Above Heart, Transfer, or C-ME®) (Fig. 6.21). A seat icon and the memory description for the
relevant position are shown in the LCD control panel display (Fig. 6.20).
4. Move and hold the joystick forwards (Fig. 6.21). The seat adjusts to the factory preset position. The
movement can be stopped at any time and started again when moving the joystick forward.
5. Stop the seat movement where required and select with a left or right joystick movement the seat
actuator that needs to be changed.
6. Move the joystick forward or backwards until the required position is reached. Select other actuators
A
if required and move them where needed.
7. Move the joystick left or right until the factory memory position is reached and where the new posi-
tion shall be programmed in.
8. Press and hold the switch connected to the CxSM (Fig. 6.23) and pull the joystick backwards for 3
seconds until a beep can be heard.
9. Release the joystick and the button. Your new required position is now stored and the actuators will
always stop in this new programmed end position.
6.15 Delete custom programmed memory (Fig. 6.19 – 6.22) 1. Remove the rear
cover of the backrest and connect a switch with a 3.5mm Jack to the I6 harness
socket (A) on the CxSM (Fig. 6.19). 2. Press the “Mode” button one or more
times until a seat icon appears in the LCD control panel display
(Fig. 6.20). 3. Move the joystick to the left or right to select a memory seat
position (Seated, Pressure Relief, Supine,
Feet Above Heart, Transfer, or C-ME®) (Fig. 6.21). A seat icon and the memory
description for the relevant position are shown in the LCD control panel
display (Fig. 6.20). 4. Press and hold the switch connected to the CxSM (Fig.
6.22) and pull the joystick backwards for 6 seconds. After 3 seconds you will
hear the programming beep and after 6 seconds you will hear a change in the
beep sound. 5. As soon as you hear this new sound release the joystick and the
button. All memory positions are now set to the factory settings.
6.16 Return to drive mode Press the “Mode” button one or more times until a
standard display image with speed indicator appears in the LCD control panel
display.
Fig.6.20
Fig.6.21
Fig.6.22
31
248020 Rev. G
Quickie® SEDEO® PRO/PRO ADVANCED
6.0 USING THE SEATING SYSTEM
6.17 Curves
DANGER!
Full speed turns should not be attempted. If you need to turn sharply you must
reduce your speed with the joystick or speed setting. This is particularly
important when traveling across or down a slope. Disregarding this advice
could lead to your wheelchair tipping over.
WARNING!
Use indicators when changing direction.
6.18 Breaking and emergency stop There are three ways to stop your wheelchair:
- Simplest and safest way to stop the wheelchair is to release the joystick.
This
will bring the chair to a halt in a controlled manner. 2. Pulling back the joystick will brake the chair abruptly with a fast stop. 3. Switching the control system off while the chair is in motion will also bring
the chair to a halt.
WARNING!
1. Switching the control system off is only to be used in an emergency situation as the stopping action is very abrupt.
2. Never try to stop or slow this seating system with the wheel locks (if applicable). Wheel locks are designed to prevent the system from moving when stopped. They are not brakes.
6.19 Driving on a slope Your wheelchair has been designed and tested to allow its use on slopes or gradients of up to 7.5° for 6 mph and 9° for 8 mph in standard configuration. However, you have the option of adjusting your position with the anterior tilt option, powered center mount leg rest, or a combination of these options.
DANGER!
Do not use the anterior tilt feature on slopes or uneven surfaces.
DANGER!
1. In certain circumstances your wheelchair could become unstable. 2. Before
attempting to climb or descend a slope or a curb, caution should be
taken when using weight shift options (e.g. powered recline or anterior tilt)
of the seat and/or your body for a counter balance weight. 3. To improve
stability lean forward when driving uphill, with the seat and back in an
upright position. 4. Alternatively sit in an upright position when traveling
in a forward, downhill direction and /or recline the seat backwards. 5. We
strongly recommend that you return the seat and back to their home positions
before attempting to climb or descend a slope. Failure to do this may cause
the wheelchair to become unstable. 6. If you are in any doubt about the
capabilities of your wheelchair on a slope, then do not attempt to drive up or
down the slope/curb; try to find an alternative route.
Gradients: ascents
WARNING!
1. When going uphill, keep the chair moving. 2. Steer by carefully moving the
joystick forwards making slight Left and Right
adjustments as you go. 3. If you have stopped on a hill, you should start
slowly. 4. If necessary, lean forward.
Gradients: descents On descents, it is important not to let the wheelchair
accelerate beyond its normal level of ground speed.
DANGER!
1. Anterior tilt, posterior tilt, elevate, and leg rest adjustments can
affect the wheelchair center of gravity. Under certain conditions this might
cause instability that could lead to unintentional tipping. It may be
necessary to carry out a risk assessment if the wheelchair is adjusted to
extremes.
2. Stopping distances on slopes can be significantly longer than on level
ground.
248020 Rev. G
32
6.0 USING THE SEATING SYSTEM
Quickie® SEDEO® PRO/PRO ADVANCED
6.20 Anterior tilt function
DANGER!
This power chair may be equipped with an anterior tilt function. This function
makes it possible for the user to tilt forward in the wheelchair.
WARNING!
1. Before using the anterior tilt function ensure that the front knee
fixation is in place.
2. Only use the anterior tilt function on a level surface.
6.21 Using the knee fixation
WARNING!
Always also use the knee fixation when operating the anterior tilt function.
DANGER!
Due to the risk of choking it is extremely hazardous to use the chest strap
without using the knee fixation.
6.22 Using a chest strap
WARNING!
1. The chest strap may only be assembled and adapted by a Sunrise Medical
authorized dealer.
2. Before operating the anterior tilt function, the chest strap and the
positioning belt must be buckled in accordance with these user instructions.
DANGER!
Due to the risk of choking it is extremely hazardous to use the chest strap
without using the knee fixation.
Usage:
DANGER!
1. This product is only to be used to position a single person in a
wheelchair. 2. It is not suitable for use as a transportation restraint. 3.
Failure to heed these warnings may result in severe injury or death. 4. An
unintended opening of the chest strap can result in the user falling
forward out of the chair. 5. If it is likely that the movements or cognitive
agility of the user could lead to
an unintended opening of the chest strap, a caregiver or attendant has to be
present at all times when the wheelchair is in use. 6. Ensure that the
caregiver or attendant is trained in the correct operation of the belts. 7.
Untrained caregivers or attendants may lead to delays, if an emergency
situation occurs.
33
248020 Rev. G
Quickie® SEDEO® PRO/PRO ADVANCED
7.0 POWERED SEATING OPTIONS
7.1 Powered seating options The wheelchair can be equipped with the following
powered options to adjust your seating position. These options can be operated
by the user without the use of tools.
Powered adjustable seating options (Fig. 7.1 – Fig. 7.3) How to operate these
powered seating options can be found in:
CONTROLLER 1. Powered tilt function 2. Powered seat lift 3. Powered backrest
recline 4. Powered elevating leg rests 5. Tilt forward function
DANGER!
1. Use of powered adjustable seating options may influence the stability of
the wheelchair. 2. Only use the powered adjustable seating options when the
wheelchair is on a flat surface! Do Not
attempt to operate any power seating option while positioned on an incline. 3.
Please stay clear of any powered actuator while the component is in motion.
Sunrise Medical power
mechanisms move over large ranges of travel. Users should be aware of their
surroundings while components are in motion. 4. While driving this seating
system in the reclined, elevated, or tilted position use caution. 5. Do not
attempt to operate any power seating option while this seating system is in
motion. 6. Do not attempt to operate the tilt power seating option around
children. 7. Through programing it is possible to reverse the direction of all
powered seating functions; ensure you know which direction your seat is going
to move before operating. Please discuss with your Sunrise Medical authorized
dealer which direction the power seating functions have been programed. 8.
Please be aware that the lift/tilt modules present a trap hazard (pinch
point). Ensure that during operation the lift/tilt modules are free from all
clothing, hands, feet, and other extremities to prevent injury. 9. Do not
drive on ramps or slopes with the seating in the elevated position. Before
attempting to climb or descend a slope, return the wheelchair to a standard
seated drive position, with the backrest upright. 10. The following order of
operations should be used when moving powered seating functions: a. To a
tilted and/or reclined position: Tilt, Recline, Legs. b. From a tilted and/or
reclined position: Tilt, Legs, Recline. Note: If equipped with a powered seat
elevator, always operate the powered seating functions with the seat elevator
in its lowest position. 11. This power system is designed to automatically
reduce the maximum driving speed to a creep and/ or C-Me mode when certain
power seating limits are reached. Depending on the configuration of the chair
there might be multiple creep and/or C-Me speeds. Talk to your Sunrise Medical
authorized dealer about creep and/or C-Me speeds for your chair.
1 m Powered adjustable seating option
Powered adjustable seating option
Fig. 7.1 Fig. 7.2
248020 Rev. G
Powered adjustable seating option
Fig. 7.3
34
8.0 REDUCING THE SEATING SYSTEM SIZE FOR TRANSPORTATION PURPOSES
Quickie® SEDEO® PRO/PRO ADVANCED
WARNING!
The wheelchair can be reduced in size for transportation without a passenger.
Any components of the seating system that can be detached easily should be
removed. Store away such components safely so that they cannot cause injury or
damage, or become damaged themselves.
8.1 Flip down backrest quick release (Fig. 8.1) Folding the backrest fully forward: 1. Loosen (A).
Do not remove (A).
CAUTION!
2. Fold (B) fully forward. Note: If necessary, first remove the back cushion and/or the armrest assemblies.
Folding the backrest fully forward with recline (Fig. 8.2) 1. Remove bolt and pin (A) to fold seat back forward. Do not discard the bolt and pin.
Fig. 8.1
8.2 Swing away leg rests (Fig. 8.3) Removing a swing-away leg rest 1. Push the
catch (A) to the rear in order to unlock the leg rest. 2. Swing the leg rest
outwards. 3. Remove the leg rest (B) from its bracket by lifting. Note: You
should hear an audible click when replacing the leg rest. This indicates that
the leg rest has locked into place.
8.3 Single post removable armrest Removing the armrest assembly (Fig. 8.4) 1.
Tip (A) downward. 2. Remove (B).
Positioning the armrest assembly (Fig. 8.5) 1. Slide (A) onto (B). 2. Push (C)
upward.
1 A B
2
B
Fig. 8.2 2
Fig 8.3
1
A
Fig 8.4
2 CB
AA 1
B
Fig 8.5
35
248020 Rev. G
Quickie® SEDEO® PRO/PRO ADVANCED
9.0 MAINTENANCE AND CLEANING
The wheelchair’s lifespan is dependent on it being well maintained.
For information concerning specific settings, maintenance, or repair work,
please contact your Sunrise Medical authorized dealer. Always ensure you
mention the model and year of manufacture and identification number provided
on the serial number label of the wheelchair when contacting your Sunrise
Medical authorized dealer.
CAUTION!
The wheelchair should be serviced by your Sunrise Medical authorized dealer
once a year or, in the case of intensive use, every six months. For a list of
approved authorized dealers in your area please contact Sunrise Medical
Service Center.
The contact details of your local Sunrise Medical Service Center can be found
on the inside front cover of this booklet.
Website addresses are on the back cover.
9.1 General maintenance 1. All fasteners should be checked quarterly for wear,
such as loose bolts
or broken components. 2. Loose fasteners should be re-tightened. See section 9
for specified
torque values. 3. Check all straps monthly for fraying, ripped seams, or other
indications
of excessive wear damage. Discontinue use if damage is found. 4. Chest straps
should be replaced at the first indication of damage and/or
excessive wear. 5. If a broken or loose component is found, discontinue use
immediately
and contact your Sunrise Medical authorized dealer for replacement. 6. Check
all hook and loop fastening straps for correct adhesion when
pressed together. 7. Ensure that any contamination, such as fluff, hair, etc
is removed from
the hook and loop straps. Such contamination may affect adhesion.
Daily checks Perform the daily check routine before driving as described
below.
Checking seating 1. Ensure that all cushions are in place. 2. Ensure that the
backrest is correctly fitted and adjusted. 3. Visually inspect the wheelchair
to ensure the leg rests, armrests, etc. are
correctly positioned and attached to the wheelchair and all fasteners are
sufficiently tightened. See section 9 for specified torque values. 4. Ensure
that the seat interface knobs are tight and secure.
Checking clothing on potential entrapment When operating the wheelchair,
ensure that your clothing does not hamper the wheelchair (e.g. too long).
Before use, always ensure your clothing or accessories do not come into
contact with the wheels or other moving and/ or rotating parts in which they
could become entangled.
Weekly checks Perform the weekly check routine before driving as described
below.
Checking connectors and cables 1. Ensure that all connectors are securely
mated. 2. Check the condition of all cables and connectors for damage.
Checking controller 1. Check the thin rubber gaiter or boot around the base of
the joystick
shaft for damage or splitting. Check visually only; do not handle the gaiter.
2. Ensure that all components of the control system are securely mounted.
Checking controls 1. Switch on the hand control Do the lights flash? This
signifies that there
is a fault in the electronic system. 2. Operate all of the electric options,
including lights and indicators (if
fitted) to ensure that they work correctly. 3. Drive the wheelchair in each of
the drive profiles to ensure the wheel-
chair performs as it did before. 248020 Rev. G
General weekly checks 1. Check tires for proper inflation level (if
applicable). 2. Check frame and base for foreign objects and/or obstructions.
3. Ensure the knobs that attach the seat frame to the front of the base are
not loose.
Quarterly checks Perform the quarterly check routine before driving as
described below. 1. Check all straps for fraying, ripped seams, or other
indications of ex-
cessive wear damage. Discontinue use if damage is found, and contact a Sunrise
Medical authorized dealer. 2. Check chair for any missing or loose nuts,
bolts, and/or fasteners. Contact a Sunrise Medical authorized dealer if found.
3. Check wheelchair for any broken components, unusual wear, or other
indicators of excessive wear. Discontinue use if damage is found, and contact
a Sunrise Medical authorized dealer.
WARNING!
1. If you are in any doubt about the performance requirements of your
wheelchair contact your Sunrise Medical authorized dealer.
2. After performing any maintenance or repairs on the wheelchair you must
ensure that it is functioning correctly before it is used.
3. A complete inspection, safety check, and service should be made by a
Sunrise Medical authorized dealer at least once per year.
4. All fasteners must be replaced like for like using the correct length,
tensile strength, and materials.
5. When replacing self-locking nuts, or nuts/studs secured with a thread
locking solution, ensure that a suitable thread locking solution is reapplied
to the new fastener.
6. Check all hook and loop fastening straps for correct adhesion when pressed
together.
7. Ensure that any contamination, such as fluff, hair, etc is removed from
the hook and loop straps. Such contamination may affect adhesion.
8. Do not reuse fasteners.
9.2 Cleaning your seating system The wheelchair should be wiped over once per
week with a slightly damp, not wet, cloth and any fluff or dust that has
accumulated around the actuators should be blown or dusted away.
CAUTION!
Ensure that you dry all parts of your wheelchair if it becomes wet or damp
after cleaning or if it is used in a wet or damp environment.
DANGER!
It is important that should the wheelchair be used by more than one person it
is cleaned thoroughly to ensure there is no cross contamination.
9.2.1 Cleaning and inspection of your seating
WARNING!
Tears, dents, wearing, or slackening of upholstery particularly near to metal
could result in poor posture or lower levels of comfort and pressure relief.
Cleaning instructions for seating Wash all parts of the covers with a gentle-
wash detergent at 100°F (38°C). You can spin-dry the covers, but do not dry
them in a dryer. Remove all parts of the covers independently of each other
and wash them separately. Take out the foam inlays prior to washing and close
the hook and loop fasteners!
Where upholstery cannot be removed, clean regularly to prevent build up or
soiling. Clean with a damp, soapy cloth. Disinfectants may be used in dilution
as specified by their manufacturer. Ensure surfaces are rinsed well with clean
water and dried thoroughly.
36
9.0 MAINTENANCE AND CLEANING
Quickie® SEDEO® PRO/PRO ADVANCED
9.3 Hygiene measures when being re-used Prior to the wheelchair being reused, it must be carefully prepared. All surfaces which come into contact with the user must be treated with a disinfectant.
11. The upholstery material must not be saturated, since the stain remover
might then change it.
12. Do not use a hose or a pressure washer to clean your chair.
To do this, you must use a suitable disinfectant for rapid alcohol-based
disinfection of medical products and medical devices.
Please take into account the manufacturer’s instructions for the disinfectant
you are using.
WARNING!
1. Always read the label on any commercial or domestic cleaning substances.
2. Always follow the instructions carefully.
CAUTION!
1. Do not use solvents, bleaches, abrasives, synthetic detergents, wax
polishes, or aerosols.
2. Disinfectants may be used in dilution as specified by the manufacturer. 3.
Ensure surfaces are rinsed with clean water and dried thoroughly.
9.4 General cleaning instructions
CAUTION!
1. You should remove stains on your seat as soon as possible. 2. After a long
period of use, you should clean the cover of your seat with
a standard commercially available dry foam cleaning agent. 3. You should
always clean the whole surface and not just individual spots
in order to avoid unattractive edges. The longer you wait the more difficult
it will be to remove them. 4. Avoid powerful rubbing with aqueous solutions.
This will roughen up cover fabric. 5. Dirt stains (eg. beer, blood, cola, red
wine, etc.) are best removed with the dry foam cleaning agents or with a mild
action detergent. 6. Please follow the respective instructions of use when
treating the covers. 7. The covers should be allowed to dry for at least 48
hours after cleaning with foam cleaning agents. 8. For grease stains (eg. ball
point pen, lipstick, chewing gum, etc.) use a proprietary stain remover.
Carefully rub the soiled location on the surface only with a clean cloth
impregnated with the stain remover. 9. Use only a very small quantity of stain
remover. 10. Test a small, inconspicuous area first.
9.5 Cleaning accessories 1. All parts/accessories such as swing away trays
should be cleaned with a
damp cloth. 2. All lateral supports, headrest, armrests, side guards, calf
pads, lap belts,
and knee pads should be cleaned with a damp cloth.
9.6 Cleaning controls Should the controller of your wheelchair become soiled
or dirty, it can be wiped with a damp cloth with a dilute disinfectant until
clean.
DANGER!
1. This is important should the wheelchair be used by more than one person to
ensure there is no cross contamination.
2. Ensure that wafer boards, joysticks (all variants), head arrays, and
switches (all variants) are cleaned with a mild disinfectant and a cleaning
cloth to avoid any cross contamination possibilities. Following removal from
chair regularly wash the sip and puff mouth piece and tube to maintain
cleanliness and functionality.
3. Ensure control is switched off before cleaning.
9.7 Storage To keep the upholstery and other parts of the wheelchair in good
condition, store in a dry environment as close to room temperature as
possible.
CAUTION!
1. Never store your wheelchair in direct sunlight or in a damp/outdoor
environment.
2. Direct sunlight might bleach upholstery/plastic parts and cause metal
parts to become hot.
9.8 Maintenance and inspection schedule
To be performed by:
End User / Caregiver
Sunrise Medical Authorized Dealer
Annually
Quarterly*
Weekly
Daily*
Maintenance and Inspection Schedule
Daily Check routine as described in Section 9.1
R
Weekly Check routine as described in Section 9.1
R
Clean wheelchair upholstery as described in Section 9.2
R
Quarterly Check routine as described in Section 9.1
R
Complete inspection, safety check, and service should be made by a Sunrise Medical authorized dealer
R
- Daily = Before each use of the wheelchair. Quarterly = Every three (3) months.
If fasteners are found to be loose a Sunrise Medical authorized dealer should tighten the fastener to the torque setting per size detailed below, unless stated otherwise in all appropriate owner’s manuals or instruction sheets (if applicable).
· M8 Fasteners = 13.5-14.9 Nm [120-132 in-lbs] · M5 Fasteners = 3.1-3.6 Nm [27-32 in-lbs]
· M6 Fasteners = 6.8-7.9 Nm [60-70 in-lbs]
· M4 Fasteners = 0.7-0.9 Nm [6-8 in-lbs]
37
248020 Rev. G
Quickie® SEDEO® PRO/PRO ADVANCED
10.0 TECHNICAL SPECIFICATIONS: APPLICABLE NORMS / STANDARDS
This product complies with the regulations and guidelines for medical aids and carries a CE symbol. The product meets the requirements and standards below. These are checked by independent institutions.
Standard EU guideline 93/42 EEC
Definition / Description Applicable as mentioned in Appendix 1
Maximum User Weight N/A
NEN-EN 12182:2014 Class B
“The requirements from EN12182 (2014) as mentioned in
450 lbs / 204 kg
NEN-EN 12184:
Assistive products for persons with disability – General require-
ments and test methods”
NEN-EN 12184:2014 Class B
Electrically powered wheelchairs, scooters and their chargers – 450 lbs / 204 kg Requirements and test methods
ISO 7176-8:2014
Requirements and test methods for impact, static and fatigue N/A strengths
ISO 7176-9:2009
Climate tests for electric wheelchairs
N/A
ISO 7176-14:2008
Requirements and test methods for control systems for elec- N/A tric wheelchairs
ISO 7176-16:2012
Requirements for resistance to ignition of upholstered parts N/A
ISO 7176-19:2008
Wheelchairs – Part 19: Wheeled mobility devices for use as seats in motor vehicles: The Q-series meet the crash test requirements as described in ISO 7176-19
450 lbs / 204 kg
ISO 7176-4:2008
Wheelchair Energy Consumption Theoretical Range
450 lbs / 204 kg
ANSI/RESNA WC-1:2009 Requirements and test methods for wheelchairs (including
N/A
scooters)
248020 Rev. G
38
11.0 SUNRISE LIMITED WARRARNTY
Quickie® SEDEO® PRO/PRO ADVANCED
THIS GUARANTEE DOES NOT AFFECT YOUR LEGAL RIGHTS IN ANY WAY.
Sunrise Medical* provides a guarantee, as set out in the warranty conditions,
for wheelchairs to its customers covering the following.
Warranty conditions
1. FOR ONE (1) YEAR – We warrant all Sunrise-made parts and components of
this wheelchair including: motors, gearboxes, and electronic components,
against defects in materials and workmanship for one year from the date of
first consumer purchase.
2. BATTERIES – The batteries are covered by a one (1) year warranty provided
through the original battery manufacturers.
3. LIMITATIONS 1. We do not warrant: a. Tires and tubes, upholstery, pads,
and push-handle grips. b. Damage from neglect, accident, misuse, or from
improper installation or repair. c. Products modified without Sunrise
Medical’s express written consent. d. Damage from exceeding the weight limit.
2. This warranty is VOID if the original chair serial number tag is removed or
altered. 3. This warranty applies in the USA and Canada only. Check with your
Sunrise Medical authorized dealer to find out if international warranties
apply. 4. This warranty is not transferable and only applies to the first
consumer purchase of this wheelchair through a Sunrise Medical authorized
dealer.
4. WHAT WE WILL DO – Our sole liability is to repair or replace covered
parts. This is the exclusive remedy for consequential damages.
5. WHAT AUTHORIZED DEALER MUST DO 1. Obtain from Sunrise Medical, while this
warranty is in effect, prior approval for return or repair of covered parts.
Contact Customer Service for a Returned Material Authorization (RMA) number
and information as to where to return the product. Upon contacting Customer
Service please provide Model number, Serial number, description of what is
required for service/repair, and any additional information as to reason for
service being performed to the wheelchair.
6. NOTICE TO CONSUMER 1. You must pay the cost of labor to repair, remove, or
install parts even if they are covered under warranty. 2. If allowed by law,
this warranty is in place of any other warranty (written or oral, express or
implied, including a warranty of merchantability, or fitness for a particular
purpose). 3. This warranty gives you certain legal rights. You may also have
other rights that vary from state to state or province to province.
7. This guarantee is subject to the law of the country in which the product
was purchased from Sunrise Medical.
- Denotes the Sunrise Medical facility from which the product was purchased.
ADDITIONAL WARRANTY INFORMATION FOR AUSTRALIA ONLY
1. For goods provided by Sunrise Medical Pty Ltd in Australia, our goods come with a guarantee by Sunrise Medical that cannot be excluded under the Australian Consumer Law.
2. You are entitled to a replacement or refund for a major failure and for compensation for any other reasonably foreseeable loss or damage.
3. You are also entitled to have the goods repaired or replaced if the goods fail to be of acceptable quality and the failure does not amount to a major failure.
4. The benefits to you given by this warranty are in addition to your other rights and remedies under a law in relation to the goods to which the warranty relates.
Record your serial number here for future reference:
__ (Refer to Base owner’s manual for serial number label location)
39
248020 Rev. G
www.SunriseMedical.com
Customer Service: 800.333.4000
In Canada 800.263.3390
Sunrise Medical 2842 N. Business Park Ave. Fresno, CA 93727 · USA ©2023
Sunrise Medical (US) LLC
P/N 248020 Rev. G