NATIONAL INSTRUMENTS NI USB-6002 Low Cost DAQ USB Device User Manual

June 13, 2024
NATIONAL INSTRUMENTS

NATIONAL INSTRUMENTS logo NI USB-6002 Low Cost DAQ USB Device
User Manual

NI USB-6002 Low Cost DAQ USB Device

Bridging the gap between the manufacturer and your legacy test system.
1-800-915-6216
www.apexwaves.com
sales@apexwaves.com
All trademarks, brands, and brand names are the property of their respective owners.

COMPREHENSIVE SERVICES
We offer competitive repair and calibration services, as well as easily accessible documentation and free downloadable resources.
SELL YOUR SURPLUS
We buy new, used, decommissioned, and surplus parts from every NI series.
We work out the best solution to suit your individual needs.
Sell For Cash Get Credit Receive a Trade-In Deal

OBSOLETE NI HARDWARE IN STOCK & READY TO SHIP
We stock New, New Surplus, Refurbished, and Reconditioned NI Hardware.

Request a Quote CLICK HERE USB-6289

SPECIFICATIONS

NI USB-6002
Low-Cost DAQ USB Device
The following specifications are typical at 25 °C, unless otherwise noted. For more information about the NI USB-6002, refer to the NI USB-6001/6002/6003 User Guide available at ni.com/manuals.

Analog Input
Number of channels
Differential ………………………………………………………….. 4
Single-ended ………………………………………………………….. 8
ADC resolution……………………………………………………….16-bit
Maximum sample rate (aggregate)………………………………..50 kS/s
Converter type ……………………………….. Successive approximation
AI FIFO ………………………. 2,047 samples
Trigger sources ……………………………………. Software, PFI 0, PFI 1
Input range ………………………….. ±10 V
Working voltage ………………………….. ±10 V

Overvoltage protection
Powered-on…………………………………………………………..±30 V
Powered-off ………………………………………………………….. ±20 V
Input impedance ………………………………………………. >1 GΩ
Input bias current ………………………………………………. ±200 pA

Absolute accuracy

Typical at full scale …………………………………… 6 mV
Maximum over temperature, full scale …………………………………… 26 mV
System noise …………………………………… 0.4 mVrms
DNL …………………………………… 16-bit, no missing codes
INL …………………………………… ±1.8 LSB
CMRR …………………………………… 56 dB (DC to 5 kHz)
Bandwidth …………………………………… 300 kHz

Analog Output

Analog outputs …………………………………….. 2
DAC resolution…………………………………….. 16-bit
Output range …………………………………….. ±10 V
Maximum update rate…………………………………….. 5 kS/s simultaneous per channel, hardwaretimed
AO FIFO …………………………………….. 2,047 samples
Trigger sources …………………………………….. Software, PFI 0, PFI 1
Output current drive …………………………………….. ±5 mA
Short circuit current …………………………………….. ±11 mA
Slew rate …………………………………….. 3 V/μs
Output impedance …………………………………….. 0.2 Ω

Absolute accuracy (no load)
Typical at full scale ……………………………………8.6 mV
Maximum over temperature, full scale ………………………… 32 mV
DNL………………………………………….16-bit, no missing codes
INL…………………………………………………±4 LSB
Power-on state…………………………….. 0 V
Startup glitch ……………………………. -7 V for 10 µs

Timebase

Note The following specifications apply to the sampling accuracy for hardwaretimed analog input and analog output.
Timebase frequency …………………………………………………………………. 80 MHz
Timebase accuracy …………………………………………………………………. ±100 ppm
Timing resolution …………………………………………………………………. 12.5 ns

Digital I/O

13 digital lines

Port 0…………………………………………………………..8 lines
Port 1 …………………………………………………………..4 lines
Port 2 …………………………………………………………..1 lines

Function
P0.<0..7> ……………………………………….Static digital input/output
P1.0 ………………………………………………..Static digital input/output
P1.1/PFI 1 ………………………………………. Static digital input/output, counter source or digital trigger
PY <2.3 > ……………………………………… Otatic digital inpul/outpul
P2.O/PFT 0…………………………………… digital inpul/outpul, counter source or digital trigger
Direction control……………………. bach channel individually programmable as input or output
Output driver type……………………. bach channel individually programmable as open collector or active drive
Absolute maximum voltage range…………………-0.3 V to 5. 5 V with respect to D GND
Pull-down resistor ……………………. 47.5 KQ to D GND
POWeP-ON Stale ……………………. LIPUL

Digital Input
Input voltage range (powered on) …………………………………. 0 to 5 V
Input voltage range (powered off) ………………………………. 0 to 3.3 V
Input voltage protection …………………………… ±20 V on two lines per port (maximum of five lines for all ports) for up to 24 hours

Caution Do not leave a voltage above 3.3 V connected on any DIO line for extended periods of time when the device is powered off. This may lead to long term reliability issues.

Minimum VIH …………………………………………………………………. 2.3 V
Maximum VIL …………………………………………………………………. 0.8 V

Maximum input leakage current
At 3.3 V ………………………………………………………….. 0.8 mA
At 5 V ………………………………………………………….. 4.5 mA

Digital Output (Active Drive)

Maximum VOL (4 mA) ……………… 0.7 V
Maximum VOL (1 mA) …………………………… 0.2 V
Minimum VOH (4 mA) …………………………… 2.1 V
Minimum VOH (1 mA) …………………………… 2.8 V
Maximum VOH …………………………… 3.6 V
Maximum output current per line …………………. ±4 mA

Digital Output (Open Collector)
Maximum VOL (4 mA) ………………………….. 0.8 V
Maximum VOL (1 mA) ……………………………..0.2 V

Note Minimum VOH dependent on user-provided pull-up resistor and voltage source. Recommended pull-up resistor is 1 kΩ.

Using a 1 kΩ pull-up resistor and 5 V voltage source:
Minimum VOH …………………………… 3.5 V
Typical VOH ………………………. 4.5 V

Maximum output (sinking) current per line ……………….. -4 mA
Maximum pull-up voltage …………………………. 5 V
Maximum leakage current
At 3.3 V …………………………… 0.8 mA
At 5 V ……………………………. 4.5 mA

Counter

Number of counter………………………….. 1
Resolution………………………….. 32 – bit

Counter measurements…………………………..Ege Counting, rising or falling
Counter direction…………………………..Count UP
Counter SOULCE…………………………..PP 0 or PFI 1
Maximum input frequency………………………….. 5 MHz
Minimum high pulse width ………………………….. 100 ns
Minimum low pulse width ………………………….. 100 ns

+5 V Power Source

Output voltage +5 V, ±3%
Maximum current 150 mA
Overcurrent protection 200 mA
Short circuit current 50 mA
Overvoltage protection ±20 V

Bus Interface

USB specification USB Full Speed
USB bus speed 12 Mb/s

Physical Characteristics

Dimensions
Without screw terminal connector plugs ………………………… 75.4 mm × 86.2 mm × 23.6 mm, (2.97 in. × 3.40 in. × 0.93 in.)
With screw terminal connector plugs ………………………… 93.2 mm × 86.2 mm × 23.6 mm, (3.67 in. × 3.40 in. × 0.93 in.)NATIONAL INSTRUMENTS NI USB 6002 Low Cost
DAQ USB Device - Dimensions

Weight
Without screw terminal connector plugs ——- 83 g (2.93 oz)
With screw terminal connector plugs——- 105 g (3.70 oz)
I/O connectors: USB Micro-B receptacle, (1)
16-position screw terminal plugs
Screw-terminal wiring ——- 1.31 to 0.08 mm² (16 to 28 AWG)
Torque for screw terminals ……………………………….. 0.22 – 0.25 N · m (2.0 – 2.2 lb. · in.)
If you need to clean the module, wipe it with a dry towel.

Environmental

Temperature (IEC 60068-2-1 and IEC 60068-2-2)

Operating 0 to 45 °C
Storage -40 to 85 °C

Humidity (IEC 60068-2-56)

Operating 5 to 95% RH, noncondensing
Storage 5 to 90% RH, noncondensing

Pollution Degree (IEC 60664)………………………….2
Maximum altitude ……………………………. 2,000 m
Indoor use only.

Safety

This product meets the requirements of the following standards of safety for electrical equipment for measurement, control, and laboratory use:

  • IEC 61010-1, EN 61010-1
  • UL 61010-1, CSA 61010-1

Note For UL and other safety certifications, refer to the product label or the Online Product Certification section.

Electromagnetic Compatibility

This product meets the requirements of the following EMC standards for sensitive electrical equipment for measurement, control, and laboratory use:

  • EN 61326-1 (IEC 61326-1): Class A emissions; Basic immunity
  • EN 55011 (CISPR 11): Group 1, Class A emissions
  • EN 55022 (CISPR 22): Class A emissions
  • EN 55024 (CISPR 24): Immunity
  • AS/NZS CISPR 11: Group 1, Class A emissions
  • AS/NZS CISPR 22: Class A emissions
  • FCC 47 CFR Part 15B: Class A emissions
  • ICES-001: Class A emissions

Note In the United States (per FCC 47 CFR), Class A equipment is intended for use in commercial, light-industrial, and heavy-industrial locations. In Europe, Canada, Australia, and New Zealand (per CISPR 11) Class A equipment is intended for use only in heavy-industrial locations.
Note Group 1 equipment (per CISPR 11) is any industrial, scientific, or medical equipment that does not intentionally generate radio frequency energy for the treatment of material or inspection/analysis purposes.
Note For EMC declarations and certifications, and additional information, refer to the Online Product Certification section.

CE Compliance ****

This product meets the essential requirements of applicable European Directives, as follows:

  • 2006/95/EC; Low-Voltage Directive (safety)
  • 2004/108/EC; Electromagnetic Compatibility Directive (EMC)

Online Product Certification
To obtain product certifications and the DoC for this product, visit ni.com/certification, search by model number or product line, and click the appropriate link in the Certification column.

Environmental Management

NI is committed to designing and manufacturing products in an environmentally responsible manner. NI recognizes that eliminating certain hazardous substances from our products is beneficial to the environment and to NI customers.
For additional environmental information, refer to the Minimize Our Environmental Impact web page at ni.com/environment. This page contains the environmental regulations and directives with which NI complies, as well as other environmental information not included in this document.

Waste Electrical and Electronic Equipment (WEEE)

EU Customers This symbol indicates that waste products should be disposed of separately from municipal household waste according to WEEE Directive 2002/96/EC of the European Parliament and the Council on waste electrical and electronic equipment (WEEE). All products at the end of their life cycle must be sent to a WEEE collection and recycling center. Proper WEEE disposal reduces environmental impact and the risk to human health due to potentially hazardous substances used in such equipment. Your cooperation in proper WEEE disposal will contribute to the effective usage of natural resources. For information about the available collection and recycling scheme in a particular country, go to ni.com/environment/weee.

Measures for the Control and Management of Pollution from Electronic Information Products (China RoHS)

National Instruments ni.com/environment/rohs_china. (For information about China RoHS compliance, go to ni.com/environment/rohs_china.)

Device Pinout

NATIONAL INSTRUMENTS NI USB 6002 Low Cost DAQ USB Device - Device
Pinout

Refer to the NI Trademarks and Logo Guidelines at ni.com/trademarks for information on National Instruments trademarks.
Other product and company names mentioned herein are trademarks or trade names of their respective companies. For patents covering National Instruments products/technology, refer to the appropriate location: Help»Patents in your software, the patents.txt file on your media, or the National Instruments Patent Notice at ni.com/patents. You can find information about end-user license agreements (EULAs) and third-party legal notices in the readme file for your NI product. Refer to the Export Compliance Information at ni.com/legal/export-compliance for the National Instruments global trade compliance policy and how to obtain relevant HTS codes, ECCNs, and other import/export data. NI MAKES NO EXPRESS OR IMPLIED WARRANTIES AS TO THE ACCURACY OF THE INFORMATION CONTAINED HEREIN AND SHALL NOT BE LIABLE FOR ANY ERRORS.
U.S. Government Customers: The data contained in this manual was developed at private expense and is subject to the applicable limited rights and restricted data rights as set forth in FAR 52.227-14, DFAR 252.227-7014, and DFAR 252.227-7015.

© 2014 National Instruments. All rights reserved.
374371A-01
NI USB-6002 Specifications
© National Instruments

References