Starkey ITC R and ITE R Hearing Aids User Manual
- June 13, 2024
- Starkey
Table of Contents
For charging ITC R & ITE R hearing aids
Your Custom Charger
User Manual
ITC R and ITE R Hearing Aids
Chargers included:
Starkey Custom Charger
Starkey Premium Custom Charger
Congrats!
The power to hear better is yours.
You’ve got a Custom Charger.
Smart choice.
Now you can get the most out of every day with the power to charge hearing
aids you can count on.
After all – when you hear better, you live better.
Go ahead, have a read.
And get ready to experience the power of better hearing.
What’s included
Chargers covered in this user manual (select yours):
-
Custom Charger
-
Premium Custom Charger*
-
Premium Custom Charger contains an on-board, rechargeable lithium-ion battery that is not removable.
Get to know your charger
- Meet your charger
- Use your charger
- Care for your charger
Meet your charger
Overview
-
Cleaning tool (included accessory)
-
Charging posts
-
Hearing aid indicator lights
-
Charger indicator lights
(Premium Custom Charger only) -
USB-C port
-
Power cord & adapter
Use your charger
It is recommended that you fully charge your hearing aids before using them.
To charge your hearing aids:
-
Plug your charger into an external power source, like a wall outlet.
NOTE: On the Premium Custom Charger only, the back charger indicator lights will turn on for a few seconds when your charger is connected.
Reminder – Custom Charger requires an external power source to charge hearing aids (e.g., a wall outlet).
Premium Custom Charger lets you charge hearing aids cordless or plugged into an external power source (e.g., a wall outlet). -
Place your hearing aids with the charging contacts facing down on the charging posts:
• The right front hearing aid indicator light flashes green for two seconds when you insert the right hearing aid.
• The left front hearing aid indicator light flashes green for two seconds when you insert the left hearing aid.
(If the indicator lights do not turn on, see the Troubleshooting guide on p. 16.)
• Your hearing aids will turn off and start charging. Charger can be open or closed during this. -
Check the charging status of your hearing aids using the hearing aid indicator lights on the front of the charger:
Front hearing aid indicator lights| Charging status
---|---
| Charging
| Fully charged
| Fault state*
Your hearing aids will fully charge in 3.5 hours.
- Take your hearing aids from the charger. They will turn on – ready to wear.
Before using your hearing aids for the first time, please see Your Custom Rechargeable Hearing Aid User Manual.
To charge your Premium Custom Charger:
- Plug your charger into an external power source, like a wall outlet.
- Check the charger indicator lights on the back of your charger to confirm the charging status of your charger.
Number of solid green charger indicator lights | Charger battery level |
---|---|
Up to 25% charged | |
Up to 50% charged | |
Up to 75% charged | |
100% charged |
Your charger will fully charge in under 3 hours.
Care for your charger
Cleaning tips
Keep your charger clean. This helps prevent damage from moisture and debris:
-
Use the included cleaning tool to clear debris from the charging posts.
This is where you insert your hearing aids for charging. -
Do not use water, cleaning fluids or solvents to clean the charging contacts. These can damage your charger.
-
Try to keep the lid closed to prevent buildup of dust and debris.
Storage tips
Store your charger in a clean and dry area. This helps preserve the life of
your charger and rechargeable hearing aid batteries:
- Choose a dresser or shelf in a room other than a bathroom or kitchen.
- Be sure the storage location is cool to avoid damage. Skip areas with extreme heat like a windowsill or hot car.
- It’s safe to leave your charger plugged in while wearing your hearing aids. But if you won’t be using it for an extended period (over a week), remove the power cord from your charger and/or external power source.
Troubleshooting guide
Problems with your charger? Try these tips. If issues continue, contact your hearing care professional.
Symptom | Possiblecauses | Solutions |
---|
Indicator lights do not turn on when hearing aids are placed on the charging
post.| Incorrect hearing aid orientation.| Reposition your hearing aids on the
charging posts.
Custom Charger:
Not plugged into an external power source (e.g., a wall outlet).
Premium Custom Charger:
Depleted onboard battery| Custom Charger:
Connect the power supply cord to your charger and plug into an external power
source (e.g., a wall outlet). The front hearing aid indicator lights will
turn on for a few seconds to indicate connection to the power source.
If they do not, contact your hearing care professional.
Premium Custom Charger: Connect the power cord to your charger and plug it
into the wall. The back charger indicator lights will turn on for a few
seconds to indicate connection to the power source. If they do not, contact
your hearing care professional.
Red blinking hearing aid indicator lights.| Fault occurred.| Remove your
hearing aids from the charging posts, wait until the front hearing aid
indicator lights turn off, then re-insert.
If the red blinking persists, contact your hearing care professional.
While charging hearing aids, no indicator lights are lighting up.| Hearing
aids have finished charging, Premium Custom Charger has moved to “power save”
mode.| To trigger indicator lights, remove hearing aid from charging post for
three seconds, then re-insert it.
Hearing aid indicator lights will remain illuminated for 10 seconds.
Hearing aids whistling in the charger.| Incorrect hearing aid orientation.|
Reposition your hearing aids in the charging posts. Confirm that the charger
indicator lights begin to light up.
Charger is not plugged into an external ower source or a working external
power source.| Custom Charger:
Your charger may not be plugged into an adequate external power source (e.g.,
a wall outlet).
Test another wall outet or power supply to ensure charger is powered. If you
do not have access to an external power source, push and hold your hearing
aid’s user control for 3 seconds to turn off.
Premium Custom Charger: Your charger’s onboard battery may be fully
depleted. Apply external power to your charger.
If you do not have the power cord, push and hold your hearing aid’s user
control for 3 seconds to turn it off.
Additional guidance
- Intended use
- FCC information
- Regulatory information
Intended Use
These hearing aid chargers are intended to charge lithium-ion rechargeable
hearing aids, and are to be used with the patient’s selected hearing aid
solution. For information about your hearing aid, please consult the hearing
aid specific user manual.
Your charger is designed to operate in public and residential environments and
is designed to comply with international Electromagnetic Compatibility
emissions and immunity standards for medical devices. However, it is still
possible that you may experience interference caused by power line
disturbances, airport metal detectors, electromagnetic fields from other
medical devices, radio signals and electrostatic discharges.
If you use other medical devices or wear implantable medical devices such as
defibrillators or pacemakers and are concerned that your hearing instruments
might cause interference with your medical device, please contact your
physician or the manufacturer of your medical device for information about
the risk of disturbance.
Your charger should not be used during an MRI procedure or in a hyperbaric
chamber or other oxygen-rich environments.
Your charger is classified as a Type B applied part under the IEC 60601-1
medical device standard.
Your charger is not formally certified to operate in explosive atmospheres
such as may be found in coal mines or certain chemical factories.
Your charger should be stored and transported within the temperature, humidity, and pressure anges of -10°C (14°F) to +45°C (113°F),10%-95% H, and 70 kPa – 106 kPa (equivalent to altitudes rom 1,200 ft (380 m) below sea level to 10,000 ft 3,000 m).
The charging temperature range is between 10°C (50°F) and 40°C (104°F) and
between 10%-95% RH and 70 kPa-106 kPa.
CAUTION:
-
If the product is hot, do not touch it until cool.
-
If the product is not working, do not disassemble.
Due to a shock hazard, contact the manufacturer and send the device in for repair. -
Keep out of reach from children. Do not ingest anything in the package, including desiccant, cleaning tool, etc.
-
Any cords and AC adapters must be approved or listed by a Nationally Recognized Testing Laboratory.
Clinical benefit
The charger itself does not provide a direct clinical benefit.
An indirect clinical benefit is provided by enabling the parent device
(rechargeable hearing aids) to achieve their intended purpose.
We are required by regulations to provide the following warnings:
WARNING: Use of wireless charger directly next to other electronic
equipment should be avoided because it could result in improper performance.
If such use is necessary, note as to whether your hearing aids and the other
equipment are operating normally.
WARNING: Use of accessories, components or replacement parts other than
those provided by the manufacturer of your charger could result in increased
electromagnetic emissions and decreased electromagnetic immunity and could
result in degradation of performance.
WARNING: If Portable Radio Frequency communications equipment is used
closer than 30 cm (12 inches) from your charger, degradation of the
performance of your charger could result. If this occurs, move away from the
communications equipment.
DO NOT OPEN CHARGER, NO USERSERVICEABLE PARTS INSIDE
The Custom Charger and Premium Custom Charger have a rating of IP 22 per IEC
60529. This means the unit is protected against solid objects over 12 mm, such
as fingers and falling water drops, such as condensation.
Do not connect the charger to any devices not described in this user manual.
The service life of both the Custom Charger and Premium Custom Charger is 3
years.
The Custom Charger and Premium Custom Charger have been tested to, and have
passed, the following emissions and immunity tests:
- IEC 60601-1-2 radiated and conducted emissions requirements for a Group 1 Class B device as stated in CISPR 11.
- Harmonic distortion and voltage fluctuations affecting the power input source as stated in Table 2 of IEC 60601-1-2.
- RF radiated immunity at a field level of 10 V/m between 80 MHz and 2.7 GHz as well as higher field levels from communications devices as stated in Table 9 of IEC 60601-1-2.
- Immunity to power frequency magnetic fields at a field level of 30 A/m and proximity magnetic fields as stated in Table 11 of IEC 60601-1-2.
- Immunity to ESD levels of +/- 8 kV conducted discharge and +/- 15 kV air discharge.
- Immunity to Electrical fast transients on the power input at a level of +/- 2 kV at a 100 Hz repetition rate.
- Immunity to Surges on the power input of +/- 1 kV line to line.
- Immunity to conducted disturbances induced by RF fields on the power input as stated in Table 6 of IEC 60601-1-2.
- Immunity to voltage dips and interruptions on the power input as stated in Table 6 of IEC 60601-1-2.
EU ADVERSE EVENT REPORTING
Any serious incident that has occurred in relation to your Starkey device
should be reported to your local Starkey representative and the Competent
Authority of the Member State in which you are established. A serious incident
is defined as any malfunction,
deterioration in the characteristics and/or performance of the device, or
inadequacy in the device user manual/labeling which could lead to the death or
serious deterioration in the state of health of the user, OR could do so upon
recurrence.
FCC information
These devices comply with part 15 of the FCC rules and with ISED Canada
license-exempt RSS standards. Operation is subject to the following two
conditions: (1) This device may not cause harmful interference, and (2) this
device must accept any interference received, including interference that may
cause undesired operation of the device.
NOTE: The manufacturer is not responsible for any radio or TV
interference caused by unauthorized modifications to this equipment.
Such modifications could void the user’s authority to operate the equipment.
Regulatory information
Starkey Labs Canada Co.
2476 Argentia Road, Suite 301
Mississauga, ON L5N 6M1
www.starkeycanada.ca
Starkey Laboratories (Germany) GmbH
Weg beim Jäger 218-222
22335 Hamburg
Germany
Waste from electronic equipment must be handled according to local regulations
Consult user manual
Keep dry
Instructions for Disposal of Old Electronics
Starkey Laboratories, Inc. encourages, the EU requires, and your local
community laws may require that your Premium Custom Charger be disposed of via
your local electronics recycling/disposal process.
The instructions below are provided for the benefit of disposal/recycling
personnel. Please include this manual when disposing of your charger.
FOR DISPOSAL/RECYCLING PERSONNEL ONLY
The Premium Custom Charger contains a lithium-ion polymer battery.
To remove the battery from the charger, follow one of two options:
Option 1
-
Insert pry bars between the plastic and the USB-C metal housing (top and bottom)
-
Pry apart to break the seam
-
Clip battery wires and dispose battery
Option 2 -
Using the style groove as a guide, sever the plastic case on all 4 sides with a dremel blade, ultrasonic knife, hot wire, or band saw blade
-
In each case, establish a limit guide so entrance over 3/32 inch beyond the outer shell is prevented
-
Clip battery wires and dispose battery
Regulatory information
Symbol | Symbol Meaning | Applicable Standard | Symbol Number |
---|---|---|---|
Manufacturer | BS EN ISO | ||
15223-1:2021 | 5.1.1 | ||
Authorized representative in the European Community | BS EN ISO | ||
15223-1:2021 | 5.1.2 | ||
Date of Manufacture | BS EN ISO | ||
15223-1:2021 | 5.1.3 | ||
Catalogue Number | BS EN ISO | ||
15223-1:2021 | 5.1.6 | ||
Serial Number | BS EN ISO | ||
15223-1:2021 | 5.1.7 | ||
Medical Device | BS EN ISO | ||
15223-1:2021 | 5.7.7 | ||
Keep Dry | BS EN ISO | ||
15223-1:2021 | 5.3.4 | ||
Temperature Limit | BS EN ISO | ||
152231:2021 | 5.3.7 | ||
Humidity Limitation | BS EN ISO | ||
15223-1:2021 | 5.3.8 | ||
Caution | BS EN ISO | ||
15223-1:2021 | 5.4.4 | ||
General warning sign | EC 60601-1, Reference no. Table D.2, Safety sign 2 |
ISO 7010- WC01
| Refer to instruction manual/booklet| EC 60601-1, Reference no. Table D.2,
Safety sign 10| ISO 7010-M002
| Collect Separately| DIRECTIVE 201 2/19/EU (VVEEE)| Annex IX
****| Class II equipment| IEC 60417
Reference no. Table D.1| Symbol 9 (IEC 60417- 5172)
| Direct current| IEC 60601-1 Reference no. Table D.1| IEC 60417-5031
| Recycling Symbol| European Parliament and Council
Directive 94/62/EC| Annex I-VII
Notes____.
Starkey is a trademark of Starkey Laboratories, Inc.
©2023 Starkey Laboratories, Inc. All Rights Reserved. 5/23 BKLT3164-01-EN-XX-
CA
Patent: www.starkey.com/patents
Read User Manual Online (PDF format)
Read User Manual Online (PDF format) >>