NICHIRYO 8100 Repetitive Syringe Dispenser Instruction Manual

June 13, 2024
NICHIRYO

NICHIRYO 8100 Repetitive Syringe Dispenser

NICHIRYO 8100 Repetitive Syringe Dispenser

FEATURES

  • 27 different volumes of liquid can be dispensed.
  • Fast change syringes.
  • Light and slender design permits easy handling without fatigue.
  • Excellent accuracy and precision.
  • Autoclavable syringe.(60.0 mL and 3.0 mL syringes are not autoclavable.)

PARTS AND MATERIALS

FEATURES
FEATURES

  1. Thumb Knob (Nylon)
  2. Volume Adjustment Dial(Nylon)
  3. Handle Cap(ABS)
  4. Handle (ABS)
  5. Piston Lever (Aluminium)
  6. Syringe Holder (Polypropylene)
  7. Piston Holder (Polypropylene)
  8. Piston Flange (Polypropylene)
  9. Syringe Flange (Polypropylene)
  10. Syringe (Polypropylene)
  11. Piston (Polypropylene & Rubber)
    FEATURES

OPERATION

  1. Push the piston lever to its lowest position.

  2. Select the appropriate syringe. Move the piston of the syringe up and down several times to ensure smooth operation in the dispenser.

  3. Push the syringe flange and the piston flange firmly into the syringe holder and the piston holder respectively. (Fig.2)

  4. Select the volume to be dispensed by turning the volume adjustment dial to the appropriate number. (Fig.1)

  5. Place the nozzle of the syringe into the liquid to be dispensed.

  6. Pull the piston lever up so that the syringe is filled with liquid. It is recommended to minimize the volume of air in the syringe to obtain better accuracy and precision.(Fig.2)

  7. Wipe the nozzle of the syringe or tip with a tissue.

  8. If the syringe has been fully filled, the thumb knob should be depressed once, and the liquid expelled before using the dispenser. The dispenser will then deliver the number of dispenses in accordance with the table attached to the side of the handle.

  9. If the syringe is only partially filled, the dispenser can be used by pressing the thumb knob until it reaches the black line at the end of the syringe.
    To ensure that the dispenser maintains its accuracy, do not dispense liquid after the piston has passed the black line.

  10. Place the nozzle or tip of the syringe so that it is touching the inside wall of the receptacle and depress the thumb knob as far as it will go. The desired volume of liquid will then be delivered into the receptacle.

  11. After finishing dispensing, the liquid remaining in the syringe can be ejected by pushing the piston lever downwards.

ATTACHMENT OF 60.0 ML SYRINGE

  1. To mount the adapter or the syringe, tilt the adapter so that square cut on the center hole of the adapter faces upward, and insert the piston flange through the hole, keeping the adapter tilted. Mount adapter on the syringe by turning the adapter clockwise until it will turn no further.
  2. Attach the piston flange to the piston holder and the flange on the adapter to the syringe holder by pushing them against the holders, keeping the flat side of the piston flange faced against the holder.
    OPERATION

SPECIFICATIONS

Volume| Accuracy (systematic error)|

precision (random error)

---|---|---
SG-V2| (0.06mL)| ±2.5%| ≦2.5%
SG-S| (0.6mL)| ±1.5%| ≦1.0%
SG-M| (1.5mL)| ±1.2%| ≦1.1%
SG-N3| (3mL)| ±1.0%| ≦0.9%
SG-L| (6mL)| ±0.9%| ≦0.8%
SG-Y| (15mL)| ±0.8%| ≦0.8%
SG-Z| (60mL)| ±1.4%| ≦1.0%

The above specifications were acquired using water at a temperature of 20°C.
The SG-S and SG-M tests were done using syringes having pipette tips on the nozzles.

DISPENSING VOLUMES

Size of Syringe

| 0.06mL (SG-V2)| 0.6mL (SG-S)| 1.5mL (SG-M)| 3.0mL (SG-N3)| 6.0mL (SG-L)| 15.0mL (SG-Y)| 60.0mL (SG-Z)|

Maximum Number Dispense

---|---|---|---|---|---|---|---|---
Volume Adjust-ment Dial| 1| 1uL| 10uL| 25uL| 50uL| 100uL| 250uL| 1.0mL| 58
2| 2uL| 20uL| 50uL| 100uL| 200uL| 500uL| 2.0mL| 28
3| 3uL| 30uL| 75uL| 150uL| 300uL| 750uL| 3.0mL| 18
4| 4uL| 40uL| 100uL| 200uL| 400uL| 1000uL| 4.0mL| 13
5| 5uL| 50uL| 125uL| 250uL| 500uL| 1250uL| 5.0mL| 10

ORDERING INFORMATION

Cat. No.

| Description|

Quantity

---|---|---
00-8100| One instrument with one each of 6.0mL, 1.5mL and 0.6mL syringe| 1
00-SG-V2| 0.06mL syringe without tip| 100
00-SG-S| 0.6mL syringe with tip| 100
00-SG-M| 1.5mL syringe with tip| 100
00-SG-N3| 3.0mL syringe without tip| 100
00-SG-L| 6.0mL syringe without tip| 100
00-SG-Y| 15.0mL syringe without tip| 50
00-SG-Z| 60.0mL syringe without tip, but with one adapter to mount the syringe to the handle| 50
00-SG-ZA| Adapter for 60mL syringe| 1
00-8100S| Stand to hold two handles with syringes| 1

For repair service or information you may contact your local distributor.