NICHIRYO Nichimate Stepper Volume Adjustable Repetitive Pipetter Instructions

October 30, 2023
NICHIRYO

NICHIRYO Nichimate Stepper Volume Adjustable Repetitive Pipetter

CAUTION

To use this product safely and prevent it from damaging your body or your property, carefully read the following instructions and fully understand the features and operating procedures.

  1. Operate the pipette at a constant speed. Do not treat it roughly, otherwise, the liquid may splash.
  2. During operation, do not bring your face close to the nozzle etc., otherwise, an accident may occur.

FEATURES

  • 40 different volumes of liquid can be dispensed.
  • Quick and easy syringe replacement.
  • Ergonomic design and thumb knob position, allows handling without fatigue.
  • High accuracy dispensing is possible and reagent loss is minimized.
  • Quick and easy volume adjustment via a 5 stage click stop mechanism.

Fig.1
Adjustable volume No. Point mark Push button Marker space

  1. Push button (nylon)
  2. Volume adjustment knob (ABS)
  3. Body cap (ABS)
  4. Body (ABS)
  5. Piston lever (ABS)
  6. Handle (nylon)
  7. Connector (nylon)
  8. Syringe (Polypropylene)
  9. Piston (Polypropylene)

OPERATION

  1. Push the piston lever to its lowest position.
  2. Select an appropriate syringe and mount it to the body.
    • Mounting the syringe: Push the piston until it hits the syringe bottom (This step is unnecessary if the piston already hits the bottom).
    • Raise the handle (Fig 3).
    • Firmly insert the syringe flange into the groove at the lower part of the body, where a piece of plate spring is secured.
    • At the same time, insert the piston head into the connector (Fig 3) (Fig4).
    • Tighten the handle and fix the piston head (Fig 5).
    • When using either 25mL or 50mL syringe, use an adapter. (Fig 6).
  3. By turning the volume adjustment knob, align the No which corresponds to the volume to be dispensed with the appropriate point on the push button. For selection of adjustable volume No, see the attached table of dispensing volumes. To ensure free movement of the syringe, the piston lever may be moved back and forth 2 or 3 times to a distance of approximately 1cm.
  4. Put the nozzle of the syringe into the liquid to be dispensed and push up the piston lever as far as it can go, so that the syringe is filled with liquid. To obtain better precision, it is recommended to minimize the volume of the air in the syringe. Let the syringe suck in the liquid slowly.
  5. After suction, slowly withdraw the nozzle of the syringe from the liquid. This will ensure that liquid does not adhere to the nozzle surface. Should any liquid remain on the nozzle, wipe it off with a tissue paper before dispensing the liquid.
  6. Place the nozzle or tip of the syringe so that it touches the inside wall of the receptacle and depress the thumb knob as far as it will go. The desired volume of liquid will be delivered into the receptacle. Do not use the first aliquot.
  7. After finishing dispensing, the liquid remaining in the syringe can be ejected by pressing the piston lever down to the lowest level. Replace the syringe after ejecting the liquid from the syringe.
  8. Replace the syringe by reversing the syringe mounting sequence.

CAUTION
To use this product safely and prevent it form damaging your body or your property, carefully read the following instructions and fully understand the features and operating procedures.

  1. Operate the pipette at a constant speed. Do not treat it roughly, otherwise, the liquid may splash.
  2. During operation, do not bring your face close to the nozzle etc., otherwise, an accident may occur.

FEATURES

  • 40 different volumes of liquid can be dispensed.
  • Quick and easy syringe replacement.
  • Ergonomic design and thumb knob position, allows handling without fatigue.
  • High accuracy dispensing is possible and reagent loss is minimized.
  • Quick and easy volume adjustment via a 5 stage click stop mechanism.

**

**

  1. Push button (nylon)
  2. Volume adjustment knob (ABS)
  3. Body cap (ABS)
  4. Body (ABS)
  5. Piston lever (ABS)
  6. Handle (nylon)
  7. Connector (nylon)
  8. Syringe (Polypropylene)
  9. Piston (Polypropylene)

Mounting the Syringe

Attachments for 25mL and 50mL Syringes.

PRODUCT OPERATION

  1. Push the piston lever to its lowest position.
  2. Select an appropriate syringe and mount it to the body.
  3. Mounting the syringe:
  4. Push the piston until it hits the syringe bottom (This step is unnecessary if the piston already hits the bottom). Raise the handle (Fig 3).
  5. Firmly insert the syringe flange into the groove at the lower part of the body, where a piece of plate spring is secured. At the same time, insert the piston head into the connector (Fig 3) (Fig4).
  6. Tighten the handle and fix the piston head (Fig 5).
  7. When using either 25mL or 50mL syringe, use an adapter. (Fig 6).
  8. By turning the volume adjustment knob, align the No which corresponds to the volume to be dispensed with the appropriate Δ point on the push button. For selection of adjustable volume No, see the attached table of dispensing volumes. To ensure free movement of the syringe, the piston lever may be moved back and forth 2 or 3 times to a distance of approximately 1cm.
  9. Put the nozzle of the syringe into the liquid to be dispensed and push up the piston lever as far as it can go, so that the syringe is filled with liquid. To obtain better precision, it is recommended to minimize the volume of the air in the syringe. Let the syringe suck in the liquid slowly.
  10. After suction, slowly withdraw the nozzle of the syringe from the liquid. This will ensure that liquid does not adhere to the nozzle surface. Should any liquid remain on the nozzle, wipe it off with a tissue paper before dispensing the liquid.
  11. Place the nozzle or tip of the syringe so that it touches the inside wall of the receptacle and depress the thumb knob as far as it will go. The desired volume of liquid will be delivered into the receptacle.
  12. Do not use the first aliquot.
  13. After finishing dispensing, the liquid remaining in the syringe can be ejected by pressing the piston lever down to the lowest level. Replace the syringe after ejecting the liquid from the syringe.
  14. Replace the syringe by reversing the syringe mounting sequence.

PRECAUTIONS

  1. Never disassemble the body.
  2. Upon dispensing, do not use the first aliquot.
  3. After suction of liquid, small air bubbles may occur around the piston. They, however, do not affect the dispensing accuracy.
  4. Be sure to mount BMT-UT tips on the nozzle of the 0.05mL syringe.
  5. When the body has been stained, wipe the stain out with adequate cleaning materials, such as neutral detergent. Never use inappropriate materials such as alcohol and thinner.

DISPENSING VOLUMES AND MAXIMUM NUMBER OF DISPENSES

Syringe                                          No. 1 2 3 4 5
NS-V 0.05mL 1μL 2μL 3μL 4μL
NS-S 0.50mL 10μL 20μL 30μL 40μL
NS-M 1.25mL 25μL 50μL 75μL 100μL
NS-N 2.50mL 50μL 100μL 150μL 200μL
NS-L 5.00mL 100μL 200μL 300μL 400μL
NS-X 12.50mL 200μL 500μL 750μL 1000μL
NS-Y 25.00mL 0.5mL 1.0mL 1.5mL 2.0mL
NS-Z 50.00mL 1.0mL 2.0mL 3.0mL 4.0mL
Maximum number of Dispenses 48 23 15 11 8

Note: 0.05mL syringe (NS-V) requires the use of BMT-UT tips.

ACCURACY AND PRECISION

Syringe Volume Accuracy (systematic error) Precision (random error)
NS-V 0.05mL 1~5μL ±5.0%
NS-S 0.50mL 10~50μL ±1.7%
NS-M 1.25mL 25~125μL ±1.6%
NS-N 2.50mL 50~250μL ±1.2%
NS-L 5.00mL 100~500μL ±2.0%
NS-X 12.50mL 200~1250μL ±1.1%
NS-Y 25.00mL 0.5~2.5mL ±1.2%
NS-Z 50.00mL 1.0~5.0mL ±1.5%

Data on the 0.05mL syringe represents data obtained when using the BMT-UT tip. NB: The accuracy and precision of NS-V depends much on the operator’s skill. Measurement condition: distilled water temperature 20℃.

ORDERING INFORMATION

Cat No. Description Quantity Cat No. Description Quantity
00-NSTP Main Body 1 00-NS-Y 25mL syringe 50/box
00-NS-V 0.05mL syringe (with tips) 100/box 00-NS-Z 50mL syringe 25/box
00-NS-S 0.5mL syringe 100/box 00-NS-ADPY Attachment for 25mL Syringes
(00-NS-Y) 1
00-NS-M 1.25mL syringe 100/box
00-NS-N 2.5mL syringe 100/box 00-NS-ADPZ Attachment for 50mL Syringes
(00-NS-Z) 1
00-NS-L 5.0mL syringe 100/box
00-NS-X 12.5mL syringe 100/box 00-NS-STD Stand to hold two dispensers 1

Note: Please contact us in case of the compatibility question of previous adapter(00-NS-ADP).

Read User Manual Online (PDF format)

Read User Manual Online (PDF format)  >>

Download This Manual (PDF format)

Download this manual  >>

Related Manuals