Data Logger RC-51 User Manual
Product Overview
This data loggeris mainly used to record the temperature of food, pharmaceuticals and chemicals, etc.in the storage and transportation. It is especially applicable to container transportation of temperature sensi ~tive goods by sea, air and road for large export-oriented enterprises and global chain enterprises.
Specification
Size: 131 (Length) * 24 (Diameter) mm
Technical paramet
Temperature measuring range: -30°C~70°C
Resolution: 0.1C
Sensor: Built-in NTC thermistor
Temperature accuracy: 05°C (-20°C~40°C); +1°C (others)
Record capacity: 32000 points (MAX)
Alarm type: continuous, cumulative
Alarm setting: no alarm, upper/lower limit alarm, multiple alarms
Record interval: 10 sec ~ 24 hour continuously set
Data interface: USB
Report type: Al format doc
Power supply: single-use lithium battery 3.6V (replaceable)
Battery life: at least 12 months at 25°C with 15 min record interval
Use the data logger for the first time
Download the data management software from the link below.
http://www.e-elitech.com/xiazaizhongxin/
Install the software first. Insert data logger to computer USB port and
install the drive software according to the prompt information.Open the
software; the data logger will automatically upload information after
connected to the computer. View information and save the configuration to
calibrate the time.
Configure parameters
Refer to the data management software instruction for details.
When connected to USB, the data logger displays Figure 19.
Start the data logger
There are three modes to start it—instant-on, manual start, and timing start
Instant-on: After parameter configuration, the data logger starts recording
immediately when it disconnects to USB.
Manual start: After parameter configuration, press and hold the button for 5
seconds to start the data logger. In this mode, it has start delay function If
this function is enabled, the data logger will not record data immediately
after start-up but start recording after the set delay time elapses.
Timing start: After parameter configuration and disconnection with USB, the
data logger starts recording when it reaches the set time.
View data temporarily
If you need to view simple statistical information, you may directly press the
button to turn page and check. The LCD screen can display MKT,average value,
Max value and Min value.
If you need detailed information, please connect the data logger to computer
USB. After a few min (in 3 min), the data will be saved in the USB disk of the
data logger in Al format report. You can open it by Al or PDF reader.
Moreover, you can connect the data logger to a computer and analyze the data
vertically and horizontally by the data management software.
Stop the data logger
There are several modes to stop it—manual stop,over – Max-record – capacity
stop (enable/disable manual stop), stop via software Manual stop: When the
data logger is recording in this mode, youmay press and hold the button for 5
second to stop it. You can also use the software tostopit. If the record
capacity reaches the Max value (32000 points) and the data logger is not
stopped manually. The data logger will save the data circularly by deleting
the initial data. (It keeps the statistical in formationof the whole
transportation process)
Note: When the record capacity exceeds the Max capacity (32000points) in the
manual mode, the data logger can continue recording the temperature state of
the whole transportation process but only keep the detail of the last 32000
points. Please use the “manual stop” mode with cautionif you have demand of
tracing back the detail of the whole process.
Over-Max-record-capacity stop (enable manual stop): In this mode, you can stop
the data logger by hand or via software, or it wil stop automatically when the
record data reaches the Max capacity (32000 points).
Over-Max-record-capacity stop (disable manual stop): In this mode, it will
stop automatically when the record data reaches the Max capacity (32000
points), or you stop it via software.
Stop via software: You can stop the data logger via software in any mode.
View data
Connect the data logger to the computer via USB and then view the data.
View Al report: Open the USB disk to view the exported Al report.
View report via the data management software: Open the software and import the
data, the software will display the configurationinfo and record data
Display menu instruction
The data logger displays different pages based on the settings. Below is the
Max display info. If you do not set relative info, it will not appear in page
turning.
Menu 1: Start delay time or the remaining time of timing start (Hr: Min.
10Sec).
See Figure 1,2 (This page s displayed only in start delay or timing start
status)Menu 2:
Current temperature. See Fig. 3, 4 (Static »indicates itis
recording.)Menu
3: Current record points. See Fig.5 (Static = indicates the current record
points exceed the Max capacity and the data logger recorded circularly) Menu
4: Current record interval. See Fig 6 (e.g. if the digit N following the
decimal point represents N*10 sec. Fig.6 shows the record interval i set to 12
min 50 sec)) Menu 5 MKT value. See Fig7 (Static indicates it stops recording)
Menu 6: Average temperature value. See Fig8 Menu 7: Max temperature value. See
Fig.9 Menu & Min temperature value. See Fig.10
Menu 9,10,11: Set upper limit of temperature. See Fig.11,1213Menu 12,13: Set lower limit of
temperature. See Fig.14,15
Content of Al report
The Al document varies based on the set alarm types.
Whenitissetto “noalarm” , there is no alarm info on the top right corer of the
first page or color mark among data.
Whenitissetto “alarm” , relative alarm info appearsin the alarm info column
based on the selected alarms. Over high temperature data is in red. Over low
temperature data is in blue. Normal data is in black Ifalarm cases occur,
there will be marked as alarm status on top right comerof the first page,
otherwise, itis in normal status.
Finish view
Exit the data logger after viewing the report
Product diagram
1 | USB port |
---|---|
2 | LCD screen |
3 | Button |
4 | Transparent cap |
5 | Battery compartment |
Replace the battery
Step 1. Rotate the transparent cap and remove it in the direction shown in
Fig.20.Step 2. Press
the snap to remove the compartment. See Fig 21Step 3. Remove the battery compartment. See
Fig.22Step 4.
Install and replace the battery. See Fig.23Step 5. Adjust the button and the intemal light pipe to
the same side, snap the compartment shut. See Fig.24Step 6. Rotate the transparent cap to install
it in the direction shown in Fig.25 Notice:
Please replace the battery after shutting down the data logger.If not, it
causes time disorder.
After replacing the battery, you need to configure parameters to calibrate the
time.
Standard configuration
1 piece of RC-51 temperature data logger
1 piece of user manual
Adds: No.1 Huangshan Rd, Tongshan Economic Development Zone,
Xuzhou, Jiangsu, China
Tel: 0516-86306508
Fax: 4008875666-982200
Hotline: 400-067-5966
URL: www.e-elitech.com
ISO9001:2008 1S014001:2004 OHSAS18001:2011 ISO/TS16949:2009
V1.0
References
Read User Manual Online (PDF format)
Read User Manual Online (PDF format) >>