Overview
Tlog 10 series data loggers can be widely used in each stage of the storage and cold chain logistics, such as refrigerated containers/ trucks, cooler bags, cooling cabinets, medical cabinets, freezers, and laboratories. The loggers feature an LCD screen and two buttons design. They support various start and stop modes, multiple threshold settings, two storage modes (stop when full & cyclic record) and PDF report automatically generated for users to check data without using software.
- USB Port
- LCD Screen
- Button
- Internal Sensor
- External Sensor
Model Selection
Model | Tlog 10 | Tlog 10E | Tlog 10H | Tlog 10 EH |
---|---|---|---|---|
Type | Internal Temperature | External Temperature | Internal Temperature and | |
Humidity | External Temperature and Humidity | |||
Measurement Range | -30°C~7o°c | |||
-22°F ~ 158°F | -40°F ~ 185 °F | |||
-40°F ~ 185 °F | -30°c ~70°c |
-22°F ~ 158°F
O%RH ~ 100%RH| -40°C ~ 85°C
-40°F ~185°F
Sensor| Digital Temperature Sensor| Digital Temperature and Humidity Sensor
Accuracy| Temperature: +0.5°C (-20°C ~ 40°C); +0.9°F (-4°F ~ 104°F)
1.0°C (-50°C ~ 85°C); +1.8°F (-58°F ~ 185 °F)
+3%RH (25°C: 20%RH ~ 80%RH), +S%RH (others)
Specifications
-
Resolution: Temperature: 0.1°C/0.1°F; Humidity: 0.1%RH
-
Memory: 32,000 points (MAX)
-
Logging Interval: 10 seconds ~ 24 hours
-
Start Mode: Press button or use software
-
Stop Mode: Press button, use software, or auto stop
-
Alarm Threshold: Configurable;
- Temperature: up to 3 high limits and 2 low limits;
- Humidity: 1 high limit and 1 low limit
-
Alarm Type: Single, cumulative
-
Alarm Delay: 10 seconds ~ 24 hours
-
Data Interface: USB port
-
Report Type: PDF data report
-
Battery: 3.0V disposable lithium battery CR2450
2 years for storage and use (25°C:10 minutes -
Battery Life: jooging interval and can last 180 days)
-
Protection Level: |P65
-
External probe length: 1.2m
-
Dimensions: 97mmx43mmx12.5mm (LxWxH)
Operation
Install Software
Please download and install the free ElitechLog software (macOS and Windows) from www.elitechlog.com/softwares.
Configure Parameters
First connect the data logger to the computer USB port, wait until the USB icon shows on the LCD, then configure via:
ElitechLog software:
- If you do not need to change the default parameters (in Appendix); please click Quick Reset under Summary menu to synchronize local time before usage;
- If you need to change the parameters, please click the Parameter menu, enter your preferred values, and click the Save Parameter button to complete the configuration.
Warning! For first time user or after battery replacement:
To avoid time or time zone errors, please make sure you click Quick Reset or
Save Porometer before usage to sync and configure your local time into the
logger.
Start Logging
Press Button:
Press and hold the left button for 5 seconds until the icon shows on the
LCD, indicating the logger starts logging.
Auto Start:
Immediate Start: The logger starts loggine after plueged out from the
computer.
Timed Start: The logger starts counting after removed from the computer,
and will start logging automatically after the set date/time.
Note: If the icon keeps flashing, it means the logger configured
Mark Events
Double click the left button to mark current temperature and time, up to 10 groups. After events marked, the LCD will display (Mark), Currently marked groups and (SUC),
Stop Logging
Press Button*: Press and hold the right button for S seconds until the icon
shows on the LCD, indicating the logger stops logging.
Auto Stop: When the recorded points reach the maximum memory, the
logger will stop automatically.
Use Software: Open ElitechLog software, click Summary menu, and
Stop Logging button.
Note:* Stop via Press Button is the default. If set as disabled, this
function will be invalid, please open ElitechLog software and click Stop
Logging button to step it.
**The Auto Stop function will be disabled automatically if you enabled the
Circular Logging.
Download Data
Connect the data logger to your computer USB port, wait until the USB icon
shows on the LCD, then download data:
Without ElitechLog Software: Simply find and open the removable storage
device ElitechLog, save the auto generated PDF report to your computer for
viewing.
With ElltechLog software: After the logger auto-upload its data to the ElitechLog software, click Export and select your preferred file format to export. If data failed to auto-upload, please manually click Download and then repeat above operation.
Re-use the Logger
To reuse a logger, please stopped it first.Then connect it to your computer
and use ElitechLog software to save or export the data.
Next, reconfigure the logger by repeating the operations in 2.
*Configure Parameters.** After finished, follow 3. Start Logging to restart
the logger for new logging.
Re-use the Logger
To reuse a logger, please stopped it first. Then connect it to your computer
and use ElitechLog software to save or export the data.
Next, reconfigure the logger by repeating the operations in 2.
Configure Parameters*. After finished, follow 3. Start Logging to restart the
logger for new logging.
Warning! * To make space for new loggings, all previous logging data
inside the logger will be deleted after re-configuration.
If you forgot to save/export data, please try to locate the logger in the
History menu of ElitechLog software.