Compensation
- Insert a trimming tool (Slot width max. 2.7 mm) through the opening in the BNC plug.
- Adjust the compensation. Turn the adjustment screw until the oscilloscope shows an exact square wave (3).
For an accurate indication of the measuring signal, it is necessary to adjust the capacity of the probe to the input capacity of the oscilloscope. For this purpose, the BNC plug of the connecting lead is provided with a countersunk, slot-headed adjusting screw. For calibration, connect the probe to the oscilloscope and the tip to the reference calibration signal of the oscilloscope.
Measurement Categories
(according to EN 61010-031 : 2002)
CAT I: Applies to test objects that are not connected to the mains.
CAT II: Applies to measurements on equipment that is connected to the
mains or supplied from the mains without constituting a part of the mains
installation.
CAT III: Applies to measurements inside the house or building
installation.
CAT IV: Applies to measurements at the supply source of the installation.
Under compensated
Over compensated
Correctly compensated
Safety Hints
In cases where the reference connnection is not at earth potential (“floating
measurements”), this probe must be used only with insulated oscilloscopes
provided with touch-protected inputs.
Never use this probe to carry out measurements on sources whose potential to
earth exceeds 1000 V.
The measurable input voltage is frequency-dependent, also see voltage
limitation curve ill. 3.
This probe is designed for indoor use at altitudes below 2000 m and in an
environment with a pollution degree of 2 (according to IEC/EN 61010-031:
2002).
The protection given by this test probe can be impaired if it is not used in
the manner intended.
Supplied accessories (see also ill. 1)
- HC400: Push-on hook clip
- GS400: Push-on reference contact
- GM400: Reference lead with crocodile clip and fork-type plug for connecting to the shielded contact on the side of the probe (ill. 2, b)
- BA400: Push-on BNC adapter
- SCC: Set of colour clips (5 x 2 pcs.) to identify the probe connecting lead (ill. 2, c)
Description of the probe (see ill. 2)
A: Insulating sleeve over the shielded contact
B: Opening at side with shielded contact for connection of reference lead
GM 400
C: SCC colour clips fitted on connecting lead
D: BNC plug with integrated compensation
E: Opening to adjusting screw
Technical details Isoprobe IV – 100:1
Dividing ratio: 100:1
Frequency range: DC to 500 MHz
Input capacitance: 4.6 pF
Compensation range: 10 pF … 25 pF
(works setting: 15 pF)
Rise time: 0.9 ns
Input resistance: 100 MΩ ± 1 % Rated voltage (frequency-dependent, see
derating curve ill. 3)
Shield / earth: max. 300 V, CAT III
Probe tip / shield: max. 300 Vr.m.s.
Pollution degree: 2
Operating temperature: +5 °C … +40 °C
Atmospheric humidity: 80 % up to 31 °C, falling linearly to 50 % at 40 °C
Altitude: < 2000 m
Cable length: 1.20 m
Making a reference connection
The output signal always relates to a reference point. For optimum measuring
accuracy, the shield of the test probe should be connected to this reference
point by the shortest possible route.
Used symbols
Caution! Refer to accompanying documents.
COMPREHENSIVE SERVICES
We offer competitive repair and calibration services, as well as easily
accessible documentation and free downloadable resources.
SELL YOUR SURPLUS
We buy new, used, decommissioned, and surplus parts from every NI series. We
work out the best solution to suit your individual needs. Sell For Cash Get
Credit ‘ Receive a Trade-In Deal
OBSOLETE NI HARDWARE IN STOCK & READY TO SHIP
We stock New, New Surplus, Refurbished, and Reconditioned NI Hardware.
1-800-915-6216
www.apexwaves.com
Sales@apexwaves.com
References
- Electrical connectors and contacts by Stäubli Electrical Connectors AG
- SP500C Oscilloscope Probe NI | Apex Waves