Boston Scientific 4591 ACUITY Spiral Implantable Lead User Manual
- June 13, 2024
- Boston Scientific
Table of Contents
- Boston Scientific 4591 ACUITY Spiral Implantable Lead
- Product Information
- Product Usage Instructions
- Overview
- INFORMATION FOR USE
- ADVERSE EVENTS
- DEVICE FEATURES
- LEAD EVALUATION
- IMPLANTATION
- EVALUATING LEAD PERFORMANCE
- SPECIFICATIONS
- References
- Read User Manual Online (PDF format)
- Download This Manual (PDF format)
Boston Scientific 4591 ACUITY Spiral Implantable Lead
Product Information
The ACUITY Spiral Implantable Lead is a transvenous lead designed for chronic, left-ventricular pacing and sensing via the coronary veins. It is compatible with a compatible pulse generator and is available in three models: 4591, 4592, and 4593. The lead features a steroid collar spiral fixation, an electrode marker band, and an IS-1 unipolar connector.
Product Usage Instructions
Before using the ACUITY Spiral Implantable Lead, please read and follow the instructions provided in the lead manual. It is important to also refer to the applicable pulse generator physician’s manual and instructions for use on any implant accessories or tools.
Intended Audience
This product is intended for use by professionals who are trained or
experienced in device implant and/or follow-up procedures.
Indications and Usage
The ACUITY Spiral Implantable Lead is indicated for chronic, left-ventricular
pacing and sensing via the coronary veins when used in conjunction with a
compatible pulse generator.
Contraindications
The use of the ACUITY Spiral Lead is contraindicated in patients with a
hypersensitivity to a maximum single dose of 0.56 mg dexamethasone acetate
drug.
Warnings
It is important to observe the following warnings to ensure correct lead
implantation and to prevent lead damage, dislodgment, or harm to the patient:
- Refer to the indicated in the manual for specific warnings relevant to each warning listed.
- IS-1 refers to the international standard ISO 5841-3:2000.
Precautions
There are no specific precautions mentioned in the provided text extract.
Overview
ACUITY™ Spiral Lead
- REF: 4591/4592/4593
INFORMATION FOR USE
Device Description
Boston Scientific ACUITY™ Spiral coronary venous pace/sense leads, Models 4591/4592/4593, provide chronic left ventricular unipolar pacing and unipolar sensing. The leads have an over-the-wire design with an IS-11 unipolar connector and are steroid-eluting distal to the electrode. The lead is anchored with spiral fixation and the electrode is IROX™-coated (iridium oxide). Placement is achieved by inserting the lead through the coronary sinus and placing it into a branch of the cardiac veins. The ACUITY Spiral lead is used in conjunction with a compatible pulse generator.
Related Information
Instructions in the lead manual should be used in conjunction with other
resource material, including the applicable pulse generator physician’s manual
and instructions for use on any implant accessories or tools.
Intended Audience
This literature is intended for use by professionals trained or experienced in
device implant and/or follow-up procedures.
Indications and Usage
The ACUITY Spiral coronary venous, steroid-eluting, single-electrode
pace/sense leads, Models 4591/4592/4593, are transvenous leads intended for
chronic, left-ventricular pacing and sensing via the coronary veins when used
in conjunction with a compatible pulse generator.
Contraindications
Use of the ACUITY Spiral lead is contraindicated in patients with a
hypersensitivity to a maximum single dose of 0.56 mg dexamethasone acetate
drug.
Warnings
In the following list of warnings are indicated for those warnings that are
specific to other areas of the manual. Refer to the indicated for information
relevant to the warning. Failure to observe these warnings could result in
incorrect lead implantation, lead damage/dislodgment, or harm to the patient.
- Labeling knowledge. Read this manual thoroughly before implanting the lead to avoid damage to the system. Such damage can result in injury to or death of the patient.
- For single patient use only. Do not reuse, reprocess, or resterilize. Reuse, reprocessing, or resterilization may compromise the structural integrity of the device and/or lead to device failure which, in turn, may result in patient injury, illness, or death. Reuse, reprocessing, or resterilization may also create a risk of contamination of the device and/or cause patient infection or cross-infection, including, but not limited to, the transmission of infectious disease(s) from one patient to another. Contamination of the device may lead to injury, illness, or death of the patient.
- Use of right ventricular lead. When using a right ventricular (RV) pace/sense lead in conjunction with the ACUITY Spiral lead, it is recommended that a polyurethane-insulated RV lead be used. Failure to observe this warning could result in insulation damage of the RV lead, which can cause periodic or continual loss of pacing, or sensing, or both.
- Lead fracture. Lead fracture, dislodgment, abrasion, or an incomplete connection can cause a periodic or continual loss of pacing or sensing or both.
- Battery-powered equipment. The use of battery-powered equipment is recommended during lead implantation and testing to protect against fibrillation that might be caused by leakage currents.
- Line-powered equipment used in the vicinity of the patient must be properly grounded.
- The lead connector must be insulated from any leakage currents that could arise from line-powered equipment.
- Use the corresponding finishing wire length. When using a finishing wire accessory kit, use the corresponding finishing wire model for the lead length. If the wrong finishing wire is used, the finishing wire tip may extend out of the distal end of the lead or not stabilize the lead properly.
- Excessive flexing. The lead is not designed to tolerate excessive flexing, bending, tension, or injection pressure. This could cause structural weakness, conductor discontinuity, or lead dislodgment.
- Magnetic Resonance Imaging (MRI) exposure. Do not expose a patient to the MRI environment. Strong electromagnetic fields in the MRI environment may interfere with the pulse generator and lead system and cause injury to the patient.
- Diathermy exposure. Patients with implanted leads should not receive diathermy treatment. Shortwave or microwave diathermy can cause tissue damage and injure the patient.
- Do not kink leads. Do not kink, twist, or braid the lead terminal with other leads, as doing so could cause lead insulation abrasion or conductor damage.
Precautions
In the following list of cautions are indicated for those cautions that are specific to other areas of the manual. Refer to the indicated for information relevant to the caution. Failure to observe these cautions could result in incorrect lead implantation, lead damage/dislodgment, or harm to the patient.
Sterilization and Handling
- If a package is damaged. The lead and accessories are sterilized with ethylene oxide gas (EO) before final packaging. When they are received, they are sterile, provided the container is intact. If the packaging is wet, punctured, opened or otherwise damaged, return the device to Boston Scientific at the address on the back cover of this manual.
- Storage temperature. Store at 25ºC (77ºF). Excursions permitted between 15º-30ºC (59º-86ºF). Transportation spikes permitted up to 50ºC (122ºF).
- Use by date. Implant the lead before or on the USE BY date on the package label because this date reflects a validated shelf life. For example, if the date is January 1, do not implant on or after January 2.
- Lead compatibility. Prior to implantation of this lead, confirm lead/pulse generator compatibility by calling Technical Services at the telephone number on the back cover of this manual.
- Dexamethasone acetate. It has not been determined whether the warnings, precautions, or complications usually associated with injectable dexamethasone acetate apply to the use of a low-concentration, highly localized, controlled-release device. For a listing of potentially adverse effects, refer to the Physicians’ Desk Reference.
- Defibrillating equipment. Defibrillating equipment should be kept nearby for immediate use during the implantation procedure.
Lead Evaluation and Implant
- Vein pick. The vein pick is not intended either for puncturing the vein or for dissecting tissue during a cutdown procedure.
- Avoid using unauthorized delivery tools. Do not use unauthorized delivery tools (e.g., stylet) to deliver the ACUITY Spiral lead.
- Remove the finishing wire. The finishing wire MUST BE REMOVED before connecting the lead to the pulse generator.
- Suture Sleeve. Do not suture directly over the lead body, as this may cause structural damage. Use the suture sleeve to secure the lead at the venous entry site.
- Do not wipe or immerse the distal lead tip in fluid prior to implant. Such treatment will reduce the amount of steroid available when the lead is implanted.
- Chronic repositioning. Optimum threshold performance might not be achieved if the lead is chronically repositioned because the steroid can be depleted.
- Protect from surface contamination. The conductor insulation is silicone rubber, which can attract particulate matter, and therefore must always be protected from surface contamination.
- Do not insert under the medial one-third region of the clavicle (subclavian puncture). When attempting to implant the lead via a subclavian puncture, do not insert the lead under the medial one-third region of the clavicle. Damage or chronic dislodgment of the lead is possible if the lead is implanted in this manner. If implantation via the subclavian vein is desired, the lead must enter the subclavian vein near the lateral border of the first rib and must avoid penetrating the subclavius muscle. It is important to observe these implant precautions to avoid clavicle/first rib damage or chronic dislodgment of the lead. It has been established in the literature that lead fracture can be caused by lead entrapment in such soft tissue structures as the subclavius muscle, costocoracoid ligament, or the costoclavicular ligament.
- Implant risks. Risks associated with this procedure are similar to any other catheterization procedure in the coronary sinus. Some patients can have a physical intolerance to different types of contrast agents. If this is known in advance, the physician should select an appropriate agent.
- Contrast medium. The type, amount, and rate of injection of the contrast medium must be determined by the physician’s medical judgment regarding the adequacy of the venogram obtained.
- Balloon catheter use. At the physician’s discretion, an occlusion balloon catheter may be used to identify the distal cardiac vein. For further instructions, see the literature accompanying the balloon catheter.
- Guidewire prolapse. Use fluoroscopy to verify the guide wire does not prolapse and catch on the distal tip of the lead. If this occurs, slowly extend the wire beyond the distal tip to free the guide wire and then retract it to reestablish movement of the guide wire.
- Guidewire retraction. If the guide wire cannot be retracted, withdraw the lead/guide wire assembly through the guiding catheter. Remove the guide wire through the distal tip of the lead and reintroduce the lead using a new guide wire. Follow the positioning procedures discussed in this manual.
- Flushing a clotted lead. Flushing a clotted lead can compromise lead integrity. If clotting is suspected, remove the lead from the body and soak the lead in heparinized saline. Insert a guide wire into either the terminal or distal tip of the lead and advance the wire to clear clotting. If unsuccessful, use a new lead.
- Applying tools to the distal end of the lead. Applying tools to the distal end of the lead may result in lead damage.
- Kinking the finishing wire. Do not kink the finishing wire in the lead. Kinking the finishing wire could lock it in the lead or damage the conductor coil.
- Remove the finishing wire. If the finishing wire cannot be retracted from the lead, withdraw the lead and finishing wire together. Do not implant with the finishing wire inside the lead.
- Strain relief. When implanting the lead via a subclavian puncture, allow slack in the lead between the suture sleeve and the venous entry site. This will help minimize flexing at the suture sleeve and interaction with the clavicle/first rib region.
- Avoid too tight ligature. When ligating the vein, avoid too tight a ligature. A tight ligature might damage the lead insulation or sever the vein. Avoid dislodging the lead tip during the stabilizing procedure.
- Do not bend the lead near the lead-header interface. Insert the lead terminal straight into the lead port. Do not bend the lead near the lead-header interface. Improper insertion can cause insulation or connector damage.
- Connecting the lead. Ensure that the lead terminal for the ACUITY Spiral is connected to the LV IS-1 port of the pulse generator.
- Explanted leads. Return all explanted leads to Boston Scientific.
- Minimize dissection. To minimize the possibility of dissection, it is recommended that a guide wire be used when advancing the guiding catheter through the venous system, right atrium, or coronary sinus.
- Prevent renal failure. To prevent renal failure associated with the use of contrast media, consider the patient’s renal function prior to the implant procedure to determine the type, amount, and rate of injection of the contrast medium while performing a venogram.
ADVERSE EVENTS
Potential Adverse Events
Based on the literature and on pulse generator and/or lead implant experience,
the following list includes possible adverse events associated with the
implantation of products described in this literature:
- Acceleration of arrhythmias
- Adverse reaction to procedure (e.g., bradycardia, general, respiratory, hypotension)
- Air embolism
- Allergic reaction
- Bleeding
- Cardiac tamponade
- Chronic nerve damage
- Conductor coil fracture
- Coronary venous spasm
- Death
- Elevated thresholds
- Erosion/extrusion
- Extracardiac stimulation (e.g., phrenic, diaphragm, chest wall)
- Fibrotic tissue formation (e.g., keloid formation)
- Fluid accumulation
- Formation of hematomas or cysts
- Heart block
- Inappropriate therapy (e.g., shocks, ATP, pacing)
- Incomplete lead connection with pulse generator
- Infection
- Lead displacement/dislodgment
- Lead fracture
- Lead insulation breakage or abrasion
- Lead tip deformation and/or breakage
- Local tissue reaction
- Muscle and nerve stimulation
- Myocardial trauma (e.g., cardiac perforation, irritability, injury)
- Myopotential sensing
- Oversensing/understanding
- Pacemaker-mediated tachycardia
- Pericardial rub, effusion
- Pneumothorax/hemothorax
- Random component failures
- Shunting current or insulating myocardium during defibrillation with internal or external paddles
- Thrombosis/thromboembolism
- Valve damage
- Venous occlusion
- Venous trauma (e.g., perforation, dissection, erosion)
In addition to the implantation of an implantable cardioverter defibrillator and/or pacemaker lead system, possible adverse events associated with the implantation of a coronary venous lead system are listed below in alphabetical order:
- Allergic reaction to contrast media
- Breakage/failure of implant tools
- Coronary venous occlusion
- Coronary venous trauma (e.g., perforation, dissection, erosion)
- Prolonged exposure to fluoroscopic radiation
- Renal failure from contrast media used to visualize coronary veins
Warranty
- A limited warranty certificate for the lead is available. For a copy, contact Boston Scientific using the information on the back cover.
- Refer to the Contraindications, Warnings, Precautions, and Adverse Events sections of this manual for information concerning the performance of this device.
DEVICE FEATURES
Detailed Device Description
Features of the ACUITY Spiral lead include the following:
- Over-The-Wire Lead Design: The lead design consists of an open-lumen conductor coil that tracks over a 0.36 mm (0.014 in) diameter guide wire.
- Steroid: The silicone rubber collar near the electrode contains a nominal dose of 0.45 mg dexamethasone acetate. Upon exposure to body fluids, the steroid elutes from the lead to help reduce tissue inflammation response at the distal electrode.
- Ring Electrode with IROX Coating: The IROX coated ring electrode provides a pacing and sensing surface in the coronary venous system.
- Pace/Sense Configurations: The ACUITY Spiral lead offers various pace/sense configurations depending upon the programming options of a compatible device. Refer to the pulse generator manual for instructions.
- Distal Tip: The distal tip is protected by silicone rubber. This protection allows for atraumatic lead advancement through the coronary venous system.
- Spiral Fixation: The distal portion of the lead provides fixation after guide wire removal. The lead is anchored in position by removing the guide wire and allowing the distal tip to assume a spiral shape that lodges in the coronary venous system.
- Lead Body: The diameter of the distal lead body (working profile) is 4.1F (1.37 mm), (0.054 in). The diameter of the proximal lead body is 4.5F (1.5 mm), (0.059 in). The lead body consists of a single conductor coil that provides one pathway. The conductor coil is sheathed in silicone rubber tubing, which is subsequently sheathed in polyurethane tubing.
- IS-1 Unipolar Connector: The industry standard connector can be used in conjunction with a compatible cardiac device that accepts the IS-1 connector.
LEAD EVALUATION
Implant Information
Proper surgical procedures and techniques are the responsibility of the
medical professional. The described implant procedures are furnished for
informational purposes only. Each physician must apply the information in
these instructions according to professional medical training and experience.
The ACUITY Spiral lead is not designed, sold, or intended for use except as indicated.
Items Included
Items packaged include the following:
- (1) ACUITY Spiral Lead
- (1) Wire Guide
- (1) Vein Pick
- Literature Packet
WARNING:
Instructions in the lead manual should be used in conjunction with other
resource material, including the applicable pulse generator physician’s manual
and instructions for use on any implant accessories or tools.
Additional Implant Tools
The following is a list of devices used for implanting the lead, but not
packaged with the lead:
- Outer guiding catheter: An 8F removable outer guiding catheter with a minimum inner diameter of 2.21 mm (0.087-in) or greater, that is intended for accessing the coronary venous system
- Tools for advancing the guiding catheter to the right atrium and cannulating the coronary sinus:
- Guidewire, 0.81–0.97-mm (0.032–0.038-in) diameter (optional), that is intended for use in the coronary venous vasculature
- An inner guiding catheter, 6F (removable inner guiding catheter (optional) with a minimum inner diameter of 1.73 mm (0.068-in), that is intended for accessing the coronary venous system
- Deflectable tip mapping catheter, 6F (2-mm) (0.078-in) diameter (optional), that is intended for use in the coronary sinus ostium
- Guidewire, 0.36-mm (0.014-in) diameter, that is intended for use in the coronary venous system
- Finishing wire, designed to stabilize the positioned lead in the venous system during guiding catheter removal
WARNING:
When using a finishing wire accessory kit, use the corresponding finishing
wire model for the lead length. If the wrong length finishing wire is used,
the finishing wire tip may extend out of the distal end of the lead or not
stabilize the lead properly. See Table 1 for the available finishing wires to
be used with ACUITY Spiral.
Table 1. Available Finishing Wires for use with ACUITY Spiral
Finishing Wire | Finishing Wire Model Numbers and Lengths |
---|
FINISHING WIRE™ Universal
| 6004 (80 cm)
6005 (90 cm)
6007 (100 cm)
FINISHING WIRE™ SUPPORTRAK™
| 6667 (80 cm)
6668 (90 cm)
6669 (100 cm)
CAUTION:
The finishing wire MUST BE REMOVED before connecting the lead to the pulse
generator.
- Standard occlusion balloon, 6F (2-mm) (0.078-in) diameter, (optional), that is used to obtain venograms by occluding the coronary sinus
- Implant accessories
Opening Instructions
The outer package and sterile tray should be opened under clean conditions. To
ensure sterility, the sealed inner sterile tray must be opened using accepted
aseptic technique by scrubbed, masked personnel. The sterile tray is opened by
peeling back the cover.
Sterilization
CAUTION:
The lead and accessories are sterilized with ethylene oxide gas (EO) before
final packaging. When they are received, they are sterile, provided the
container is intact. If the packaging is wet, punctured, opened or otherwise
damaged, return the device to Boston Scientific at the address on the back
cover of this manual.
Storage
Store at 25ºC (77ºF). Excursions permitted between 15º-30ºC (59º-86ºF).
Transportation spikes permitted up to 50ºC (122ºF).
Surgical Preparation
Instrumentation for heart monitoring, imaging (fluoroscopy), external
defibrillation, and pacing threshold and sensitivity measurements should be
available during implantation. The sterile field should be large enough to
accommodate the use of the guide wires. Sterile duplicates of all implantable
items should also be available for use if accidental damage or contamination
occurs. Always isolate the patient from potentially hazardous leakage current
when using electrical instrumentation.
Nominal lengths of the leads are as follows:
Model | 4591 | 4592 | 4593 |
---|---|---|---|
Length | 80 cm | 90 cm | 100 cm |
Selection of the lead length appropriate to the patient’s cardiac anatomy is a matter of medical judgment.
Lead Accessories
The following items are packaged in the lead tray and are also available from
Boston Scientific as accessory items:
Vein Pick
- The vein pick is a sterile, disposable, nontoxic, plastic device designed to assist with placement of the guiding catheter into the vein.
- To use the vein pick during a cutdown procedure, isolate and open the selected vein using an appropriate instrument. Introduce the point of the vein pick via this incision into the lumen of the vein. With the point of the vein pick facing in the direction of the desired guiding catheter passage, gently raise and tilt the pick. Pass the guiding catheter under the vein pick and into the vein.
CAUTION:
The vein pick is not intended either for puncturing the vein or for dissecting
tissue during a cutdown procedure.
Wire Guide
The wire guide is intended to ease insertion of a guide wire into the lumen at
the terminal of the lead (Figure 1).
Suture Sleeve
The suture sleeve is an adjustable, tubular reinforcement positioned over the
outer lead insulation. It is designed to secure and protect the lead at the
venous entry site after lead placement. Using a suture sleeve reduces the
possibility of structural damage caused by suturing directly over the lead
body.
CAUTION:
Do not suture directly over the lead body, as this may cause structural
damage. Use the suture sleeve to secure the lead at the venous entry site.
Handling the Lead
Observe the following when handling the lead:
WARNING:
The lead is not designed to tolerate excessive flexing, bending, tension, or
injection pressure. This could cause structural weakness, conductor
discontinuity, or lead dislodgment.
CAUTIONS:
- Do not wipe or immerse the distal lead tip in fluid prior to implant. Such treatment will reduce the amount of steroid available when the lead is implanted.
- Optimum threshold performance might not be achieved if the lead is chronically repositioned because the steroid can be depleted.
- The conductor insulation is silicone rubber, which can attract particulate matter, and therefore must always be protected from surface contamination.
IMPLANTATION
Inserting the Lead
The lead may be inserted using one of the following two methods:
Via cut down through the left or right cephalic vein.
- Only one incision over the deltopectoral groove is required to insert the guiding catheter through the cephalic vein. The endocardial lead is inserted into the right or left cephalic vein in the deltopectoral groove.
- The vein pick packaged with this lead can be used during a cutdown procedure to aid insertion of the guiding catheter into the vein. Before inserting the guiding catheter, see the section, “Lead Accessories” for instructions on using the vein pick.
- Percutaneously or via cutdown through the subclavian vein or internal jugular vein—typically the left subclavian or right internal jugular vein.
- A subclavian introducer set is available from Boston Scientific for use during percutaneous lead insertion.
CAUTION:
When attempting to implant the lead via a subclavian puncture, do not insert
the lead under the medial one-third region of the clavicle. Damage or chronic
dislodgment to the lead is possible if the lead is implanted in this manner.
If implantation via the subclavian vein is desired, the lead must enter the
subclavian vein near the lateral border of the first rib and must avoid
penetrating the subclavius muscle. It is important to observe these implant
precautions to avoid clavicle/first rib damage or chronic dislodgment to the
lead. It has been established in the literature that lead fracture can be
caused by lead entrapment in such soft tissue structures as the subclavius
muscle, costocoracoid ligament, or the costoclavicular ligament.2
Leads placed by percutaneous subclavian venipuncture should enter the subclavian vein, where it passes over the first rib (rather than more medially), to avoid entrapment by the subclavius muscle or ligamentous structures associated with the narrow costoclavicular region.3 It is recommended to introduce the lead into the subclavian vein near the lateral border of the first rib.
The syringe should be positioned directly above and parallel to the axillary vein to reduce the chance that the needle will contact the axillary or subclavian arteries or the brachial plexus. Use of fluoroscopy is helpful in locating the first rib and in guiding the needle. The steps below explain how to identify the skin entry point and define the course of the needle toward the subclavian vein where it crosses the first rib.
- Referring to Figure 2, identify points St (sternal angle) and Cp (coracoid process).
- Visually draw a line between St and Cp, and divide the segment into thirds. The needle should pierce the skin at the junction of the middle and lateral thirds, directly above the axillary vein (point Ax).
- Place an index finger on the clavicle at the junction of the medial and middle thirds (point V), beneath which point the subclavian vein should be located.
- Press a thumb against the index finger and project one or two centimeters below the clavicle to shield the subclavius muscle from the needle (when hypertrophy of the pectoralis muscle is apparent, the thumb should project about two centimeters below the clavicle because the subclavius muscle should be hypertrophied as well) (Figure 3).
- Feel with the thumb the pressure from the passage of the needle through the superficial fascia; direct the needle deep into the tissues toward the subclavian vein and the underlying first rib. Fluoroscopic guidance will reduce the chance that the needle would pass below the first rib and into the lung.
Positioning the Lead
Positioning the lead includes the following steps:
- Insert a guiding catheter into the ostium of the coronary sinus to provide a path for lead placement.
- Obtain a venogram to visualize the coronary venous system.
- Place the lead through the guiding catheter in the coronary venous system by advancing the lead over a guide wire.
Referring to Figure 4, the lead is introduced into the coronary venous system through the ostium of the coronary sinus and advanced into its tributaries. The coronary sinus and its tributaries include the great cardiac vein, middle cardiac vein, left posterior vein, and left marginal vein. All cardiac veins are potential sites for implantation of the ACUITY Spiral lead. Variability in patient anatomy may preclude placement in one or more of the suggested sites.
Note:
It is recommended that a venogram be performed to determine the patient’s
cardiac anatomy. Any preexisting condition of the patient, e.g., coronary
stent or coronary artery bypass graft (CABG), should be taken into
consideration while using proper medical judgment to determine the best lead
implant site.
Inserting the Guiding Catheter
Recommended methods for finding the coronary ostium include but are not
limited to the following: a) placing a guide wire 0.81–0.97 mm
(0.032–0.038 in.) diameter in the ostium first and then following the guide
wire with the guiding catheter or b) inserting a 6F (2 mm) (0.078 in.)
diameter (or smaller) fixed curve or deflectable tip mapping catheter through
the guiding catheter and then into the ostium.
Obtaining a Venogram
CAUTION:
Risks associated with this procedure are similar to any other catheterization
procedure in the coronary sinus. Some patients can have a physical intolerance
to different types of contrast agents. If this is known in advance, the
physician should select an appropriate agent.
Once the guiding catheter is in place and while under fluoroscopy, inject a small amount of contrast medium into the coronary sinus to confirm proper placement of the guiding catheter tip in the coronary sinus. The contrast agent will flow out of the coronary sinus. Once the position is confirmed, use a minimum amount of contrast to identify the coronary sinus branch vein. Save the acquired venogram for future reference of the venous anatomy.
CAUTIONS:
- The type, amount, and rate of injection of the contrast medium must be determined by the physician’s medical judgment regarding the adequacy of the venogram obtained.
- At the physician’s discretion, an occlusion balloon catheter may be used to identify the distal cardiac veins. For further instructions, see literature accompanying the balloon catheter.
Inserting the Lead Into the Guiding Catheter
The ACUITY Spiral lead can be delivered through the guiding catheter used to
cannulate the coronary sinus after the venogram has been obtained.
Alternatively, the ACUITY Spiral lead can be delivered through a secondary
inner catheter that has been introduced through the cannulation catheter for
the purpose of sub-selecting a branch vein.
Note:
The inner catheter must be removable over the lead and must have a minimum
inner diameter of 1.73 mm (0.068 in).
Placing the Lead
The following section describes two preferred methods for the ACUITY Spiral
lead placement over a guide wire after the guiding catheter has been
positioned in the coronary sinus and a venogram has been obtained.
Notes:
- The guiding catheter serves as a conduit for the delivery of implantable coronary venous leads and can help protect the ACUITY Spiral lead during the placement of other leads.
- It is recommended to flush the guide wire’s protective hoop and the inner lumen of the guide catheter with heparinized saline before and during guide wire use.
- To prevent blood from clotting in the lead, it is recommended to flush the inner lumen of the lead with heparinized saline before and during use.
- Position the guide catheter tip as close as possible to the origin of the target branch vein.
- The physician should consider the venous anatomy of the patient when selecting the appropriate guide wire for lead delivery. Guide wires with varying distal stiffness will straighten the spiral fixation to varying degrees. Guide wires with more distal support will provide the greatest amount of spiral straightening.
- Under fluoroscopy confirm that the marker band, proximal to the spiral fixation, remains within the branch vein.
Method A
- Insert the 0.36-mm (0.014-in) diameter guide wire into the guiding catheter and advance the tip of the wire through the coronary sinus to the desired position within the venous system.
- Insert the proximal end of the guide wire into the distal opening of the lead. While inserting the guide wire, carefully straighten the helix to prevent perforating the lead or damaging the conductor coil.
- While holding the guide wire in place, advance the lead over the wire to the desired lead position.
Method B
- Insert the floppy tip of the 0.36-mm (0.014-in) diameter guide wire into the terminal pin of the lead. Extend at least 3 cm of the guide wire beyond the distal tip of the lead to ensure the guide wire slides easily through the lumen and to straighten the spiral fixation of the lead.
- Insert the lead/guide wire assembly into the guiding catheter. Under fluoroscopy, advance the lead until the tip of the lead is even with, but does not extend beyond the tip of the guide catheter. Advance the guide wire through the coronary sinus to the desired position within the venous system.
- While holding the guide wire in place, advance the lead over the wire to the desired lead position.
When the lead is in the desired target branch vein, advance the lead to a distal location within that branch. Remove the guide wire while applying gentle forward pressure on the lead until the spiral fixation engages.
CAUTIONS:
- Use fluoroscopy to verify the guide wire does not prolapse and catch on the distal tip of the lead. If this occurs, slowly extend the wire beyond the distal tip to free the guide wire and then retract it to reestablish movement of the guide wire.
- If the guide wire cannot be retracted, withdraw the lead/guide wire assembly through the guiding catheter. Remove the guide wire through the distal tip of the lead and reintroduce the lead using a new guide wire. Follow the positioning procedures previously discussed.
- Flushing a clotted lead can compromise lead integrity. If clotting is suspected, remove the lead from the body and soak the lead in heparinized saline. Insert a guide wire into either the terminal or distal tip of the lead and advance the wire to clear clotting. If unsuccessful, use a new lead.
- Applying tools to the distal end of the lead may result in lead damage.
EVALUATING LEAD PERFORMANCE
Evaluating Lead Position
Verify electrical performance of the lead using a pacing system analyzer or
similar monitor before attaching the lead to the pulse generator. Once the
lead is placed in the desired location, withdraw the guide wire tip into the
pacing lead so the spiral fixation is engaged. Perform the measurements for
voltage threshold (at 0.5 ms pulse width), R-wave amplitude, and pacing
impedance, using recommended values in Table 2.
Table 2. Recommended Threshold and Sensing Measurements
Ventricular Data
Voltage thresholda| < 2.5 V
R-wave amplitude| > 5.0 mV
Lead Impedance| 300- 2000 W
- Pulse width setting 0.5 ms.
See Figure 5 and Figure 6 for pacing system analyzer connections. Threshold measurements can be taken immediately after the lead is positioned and the spiral fixation is engaged.
Note:
The guide wire must be withdrawn so the spiral fixation is engaged when
performing lead evaluation.
Perform the lead evaluation process:
- Take measurements using one or more of the pacing and/or sensing configurations allowed by the pulse generator.
- If satisfactory measurements free of extra cardiac stimulation are not achieved in any available configuration, reposition the lead.
Repositioning the Lead
Recommended methods for repositioning the lead include:
-
Reposition the lead to a more proximal location within the branch vein. Repeat the lead evaluation process.
Notes:- Under fluoroscopy confirm that the marker band, proximal to the spiral fixation, remains within the branch vein.
- While pulling out the lead, keep the wire in place so that the bias shape is not damaged.
-
Reposition the lead to a new branch vein if measurements from method one are unsatisfactory.
Removing the Guiding Catheter
Once the lead is positioned, remove the guide wire from the lead. Next, remove
the finishing wire from its packaging and insert it into the lead according to
the manufacturer’s instructions. Peel away the introducer sheath, if used.
While holding the lead and finishing wire in place, remove the guiding
catheter using the method described in the guiding catheter instructions for
use. Using fluoroscopy, verify that the position of the lead tip does not
change during the removal of the guiding catheter. Hold the proximal end of
the lead near the venous entry site, disconnect the finishing wire from the
terminal pin and withdraw the finishing wire from the lead. Verify under
fluoroscopy that the lead has not moved.
Allow extra slack in the lead in the atrium for a strain relief to reduce the chance of dislodgment.
CAUTIONS:
- Do not kink the finishing wire in the lead. Kinking the finishing wire could lock it in the lead or damage the conductor coil.
- If the finishing wire cannot be retracted from the lead, withdraw the lead and finishing wire together. Do not implant with the finishing wire inside the lead.
Securing the Lead
After the lead is satisfactorily positioned, use the following steps to secure
the lead to the vein to achieve permanent hemostasis and lead stabilization.
Suture sleeve tie-down techniques can vary with the lead insertion technique
used. A suture sleeve is provided for this purpose.
Percutaneous Implant Technique
- Peel back the introducer sheath and slide the suture sleeve deep into the tissue (Figure 7).
- Using both grooves, ligate the suture sleeve and the lead to the fascia. For additional stability, the sleeve may be secured to the lead first before securing the sleeve to the fascia.
- Check the suture sleeve after tie-down to demonstrate stability and lack of slippage by grasping the suture sleeve with fingers and trying to move the lead in either direction.
CAUTION:
When implanting the lead via a subclavian puncture, allow slack in the lead
between the suture sleeve and the venous entry site. This will help minimize
flexing at the suture sleeve and interaction with the clavicle/first rib
region.
Venous Cutdown Technique
- Slide the suture sleeve into the vein past the distal groove. Ligate the vein around the suture sleeve to obtain hemostasis. Next, using the same groove, secure the lead and vein to the adjacent fascia (Figure 8).
- Using the proximal groove, secure the sleeve and the lead to the adjacent fascia. For additional stability, the sleeve may be secured to the lead first before securing the sleeve to the fascia.
- Check the suture sleeve after tie-down to demonstrate stability and lack of slippage by grasping the suture sleeve with fingers and trying to move the lead in either direction.
Note:
If venous entry is made using a lead introducer, ligate the lead to the
adjacent fascia using the suture sleeve to prevent lead movement.
CAUTION:
When ligating the vein, avoid too tight a ligature. A tight ligature might
damage the lead insulation or sever the vein. Avoid dislodging the lead tip
during the stabilizing procedure.
Connection to a Pulse Generator
- Remove the finishing wire from the lead before connecting the lead to the pulse generator. A finishing wire left in the lead could cause (1) lead perforation or (2) myocardial or coronary venous perforation.
- When the lead is secured at the venous entry site, reverify position and threshold measurements and then connect the lead to the pulse generator using the procedure described in the applicable pulse generator physician’s manual.
WARNING:
Do not kink, twist, or braid the lead terminal with other leads, as doing so
could cause lead insulation abrasion or conductor damage.
CAUTIONS:
- Insert the lead terminal straight into the lead port. Do not bend the lead near the lead-header interface. Improper insertion can cause insulation or connector damage.
- Ensure that the lead terminal of the ACUITY Spiral lead is connected to the LV IS-1 port of the pulse generator.
Notes:
- If a lubricant is needed when connecting the lead to the pulse generator, sterile water is suggested.
- If the lead terminal will not be connected to a pulse generator at the time of lead implantation, the lead connector must be capped before closing the pocket incision. The IS-1 lead cap is designed specifically for this purpose. Place a suture around the lead cap to keep it in place.
Giving consideration to patient anatomy and pulse generator size and motion, gently coil any excess lead and place adjacent to the pulse generator. It is important to place the lead into the pocket in a manner that minimizes lead tension, twisting, sharp angles, and/or pressure.
Returning Explanted Products
CAUTION: Return all explanted leads to Boston Scientific.
Examination of explanted leads can provide information for continued improvement in system reliability. Use a Boston Scientific Returned Product Kit to properly package the lead and complete an Observation/Complication/Out- of-Service Report form. Send the form and kit to Boston Scientific at the address on the back of this manual.
Note:
Disposal of explanted devices is subject to local, state, and federal
regulations. Contact your sales representative or call the telephone number on
the back of the manual for a Returned Product Kit.
Symbols on Packaging
The following symbols may be used on leads packaging and labeling (Table 3).
Table 3. Symbols on packaging
SPECIFICATIONS
SPECIFICATIONS (Nominal)
Model and Length
| 4591 – 80 cm
4592 – 90 cm
4593 – 100 cm
---|---
Terminal compatibility| IS-1
Electrode configuration| Unipolar (single)
Compatibility| Pulse generators that accept IS-1 connectors
Insertion Diameter| 1.60 mm
Recommended introducer size| Determined by guiding catheter size
Recommended guiding catheter size
| Outer Catheter (cannulation catheter): 8F, with an inner diameter of 2.21 mm
(0.087 in) or greater
Inner Catheter (branch sub- selection catheter): 6F, with an inner diameter of
1.73 mm (0.068 in) or greater
Steroid| 0.45 mg dexamethasone acetate
Conductors: Type Material|
Quadfilar
MP35N™ with Tantalum Core
Electrode:|
Surface area| 5.2 mm2
Material| Platinum iridium substrate
Coating| IROX (iridium oxide) coating
Lead Body:|
Proximal body diameter| 4.5F (1.5 mm)
Distal body (working profile) diameter| 4.1F (1.37 mm)
Inside diameter| 0.022 in (0.56 mm)
Tip diameter| 2.6F (0.86 mm)
Insulation material| Silicone rubber, polyurethane 55D
Protective sleeve material| Polyurethane 55D
Terminal pin and ring material| Titanium
Fixation mechanism| 3 dimensional spiral
Location of Marker Band| 41 mm from the distal tip
---|---
Maximum lead conductor| 4591 – 71 W
resistance (ohms) from terminal| 4592 – 77 W
pin to distal electrode| 4593 – 82 W
CENELEC pacing impedance test resulta| 600 W
CENELEC sensing impedance test resulta| 765 W
The CENELEC pacing and sensing impedance test provides a standardized way to compare the performance of lead designs. The test result does not necessarily reflect clinical performance.
Contact Information
Boston Scientific
- 4100 Hamline Avenue North St. Paul, MN 55112-5798 USA
Guidant Europe NV/SA; Boston Scientific
- Green Square, Lambroekstraat 5D 1831 Diegem, Belgium.
- 1.800.CARDIAC (227.3422)
- +1.651.582.4000
Boston Scientific (Australia) Pty Ltd
- PO Box 332 BOTANY NSW 1455 Australia
- Free Phone 1800 676 133
- Free Fax 1800 836 666
© 2013 Boston Scientific Corporation or its affiliates. All Rights Reserved.
References
Read User Manual Online (PDF format)
Read User Manual Online (PDF format) >>