PeakTech 5305 PH Meter Instruction Manual
- June 4, 2024
- PeakTech
Table of Contents
5305 PH Meter
Instruction Manual
Safety precautions
This device complies with EU regulations 2014/30 / EU
(electromagnetic compatibility) and 2011/65/EU (RoHS) as specified in the
addendum 2014/32 / EU (CE mark).
Damage caused by failure to observe the following information is excluded from
claims of any kind.
For the operational safety of the device, the following safety instructions for operating the device must be observed.
- Read these instructions carefully and make them accessible to subsequent users.
- Use this device only within its scope and specifications.
- Take all necessary safety precautions, especially when using acids.
- Wear personal safety equipment, gloves, and safety glasses when handling chemical products.
- Never operate the device if it is not completely closed.
- Do not operate the device near strong magnetic fields (motors, transformers, etc.)
- Avoid strong vibrations of the device Before starting the operation, the device should be stabilized to the ambient temperature (Important when transporting cold to warm rooms and vice versa)
- Do not make any technical changes to the device
- Only qualified service technicians may open the device, as well as perform maintenance and repair work
Cleaning the cabinet
Clean only with a damp soft cloth and a commercially available mild household
cleanser. Ensure that no water gets inside the equipment to prevent possible
shorts and damage to the equipment.
Introduction
The PeakTech 5305 is used for pH measurement of various liquids.
Thanks to the removable pH probe, it is possible to thoroughly clean and
remove residues of the liquid from the device and the probe after each
measurement. With the adjusting screw in the battery compartment, you are able
to calibrate the pH meter with the enclosed buffer liquids (a screwdriver is
included in the scope of delivery). To protect the measuring electrode, a
protective cap is attached to the end of the P 5305 (this is removed during
the measurement).
- Digital pH meter with illuminated display
- Large, 3½-digit LCD display
- Can be calibrated using a potentiometer (pH7)
- Unscrewable measuring electrode (dry stored)
- Waterproofed case
- Electrode protective cap
- Easy switching on / off by dip switches
Controls
Measuring operation
The pH meter is used to determine the pH values of various liquids.
The display, which has a backlight, and the intuitive handling of the device
make it possible to carry out the measurement of the pH measured value quickly
and are user-friendly.
4.1 Carrying out a pH measurement
To carry out the pH measurement, the protective cap that protects the
electrode of the probe must be removed. Then the device must be switched on
with the on / off switch. The backlight is permanently 0on during operation
to emphasize the measured value.
If the pH meter has not been used for a longer period of time, has been in
operation for a longer period of time, or if it has been used in environments
in which severe external influences can occur, calibration of the device is
recommended (see point 4.2).
Before the measurement, clean the probe with distilled water in order to rule
out possible falsifications of the measured value. Dry the probe carefully
with a cloth. Now the pH meter probe can be immersed in the liquid to be
measured. Make sure that only the lowest 4 cm of the probe are immersed in the
liquid.
After completing the measurement, remove the pH meter from the liquid and
clean the probe again with distilled water.
The PeakTech 5305 is able to measure the temperature of the liquid during the pH measurement. The display of the temperature unit can be changed by pressing the Temp / Cal button (° C / ° F). The temperature display can also be updated manually by pressing the Temp / Cal button for 5 seconds. If no further measurements are to be taken, put the protective cap of the electrode on the end of the probe again to protect it from dirt or damage.
4.2 Calibration of the pH meter.
The calibration of the pH meter is recommended if exact measured values are
required, the device has not been used for a long time, or if the device has
been used for a longer period of time.
To calibrate the device, a buffer solution is required which has a pH of 7.01
at room temperature.
To calibrate the area, remove the protective cap from the probe, fill the
buffer solution in a container and immerse the lower 4 cm of the probe in the
buffer solution.
Now move the pH meter carefully and slowly in the solution for a short time.
After a short time, the device settles to a measured value. When this has been
done, the value can be adjusted to exactly 7.01 using the calibration screw.
To check the pH meter, another buffer solution with a pH of 4.01 is included
in the scope of delivery.
After the calibration is done, clean the probe with distilled water and dry
the probe with a cloth.
Battery replacement
The PeakTech 5305 is a battery-operated measuring device. After a certain
period of use of the device, it may happen that the capacity of the batteries
is used up and new batteries have to be inserted. This becomes visible when
the display gets darker or looks a little blurry.
To change the batteries, make sure that the device is switched off.
Remove the upper screw cap of the device and remove the batteries from the
battery compartment. Now insert the new batteries of the appropriate polarity
(which is shown in the housing) into the battery compartment. Make sure to use
the correct batteries that the measuring device requires (4 x 1.5V AG13 button
cells).
After changing the battery, put the screw cap on the device and screw it back
on.
Specifications
Display| 3 1/2-digit, 18mm
LCD-display max 1999
---|---
Functions| PH:
0 bis 14 pH| Temperature:
0°C — 50°C
Resolution| 0,01 pH| 0,1°C
Accuracy| ± 0,01 pH| ± 1 °C
Sampling rate| 2 — 3 / Sec.
Calibration| External with a potentiometer for pH7
Operation temperature| 0°C…+50°C (32°F…122°C); < 80% RH
Power Supply| 4 x 1,5V Batteries (AG-13)
Dimensions (WxHxD)| 188 x 35 x 35 mm
Weight| 80 g
Accessories| Manual, 4 x 1,5V Batteries (AG-13),
Calibration buffer solution ph 7
Notification about the Battery Regulation
The delivery of many devices includes batteries, which for example serve to operate the remote control. There also could be batteries or accumulators built into the device itself. In connection with the sale of these batteries or accumulators, we are obliged under the Battery Regulations to notify our customers of the following:
Please dispose of old batteries at a council collection point or return them to a local shop at no cost. The disposal of domestic refuse is strictly forbidden according to the Battery Regulations. You can return used batteries obtained from us at no charge at the address on the east side of this manual or by posting with sufficient stamps. Contaminated batteries shall be marked with a symbol consisting of a crossed-out refuse bin and the chemical symbol (Cd, Hg or Pb) of the heavy metal which is responsible for the classification as a pollutant:
- “Cd” means cadmium.
- “Hg” means mercury.
- “Pb” stands for lead.
All rights, also for translation, reprinting, and copy of this manual or parts
are reserved.
Reproduction of all kinds (photocopy, microfilm, or other) only by written
permission of the publisher.
This manual considers the latest technical knowledge. Technical changings
which are in the interest of progress are reserved.
We herewith confirm that the units are calibrated by the factory according to
the specifications as per the technical specifications.
We recommend calibrating the unit again, after 1 year.