MEAN WELL RPS-200 Series 200W Reliable Green Medical Power Supply User Guide

June 13, 2024
MEAN WELL

MEAN WELL RPS-200 Series 200W Reliable Green Medical Power Supply User Guide
MEAN WELL RPS-200 Series 200W Reliable Green Medical Power
Supply

Features

  • • 4″x2″ compact size
  • Medical safety approved (2 x MOPP) according to ANSI/AAMI ES60601-1 and IEC/BS EN/EN60601-1
  • Suitable for BF application with appropriate system consideration
  • 140W convention, 200W force air
  • EMI Conduction for Class B Radiation for Class B with FG(Class I) and Class A without FG(Class II)
  • No load power consumption<0.5W
  • Extremely low leakage current
  • 12V/0.5A fan supply
  • Protections: Short circuit/Overload / Over voltage / Over temperature
  • Lifetime 65K hours
  • Operating altitude up to 5000 meters
  • 3 years warranty

Applications

  • Oral irrigator
  • Hemodialysis machine
  • Medical monitors
  • Sleep apnea devices
  • Pumps machine
  • Electric bed

TIN CODE

MW Search: https://www.meanwell.com/serviceGTIN.aspx

Description

RPS-200is a 200W highly reliable green PCB type medical power supply with a high power density (21.9W/ir) on the 4” by 2″ footprint. It accepts 80~264VAC input and offers various output voltages between 12V and 48V.
The working efficiency is up to 95% and the extremely low no load power consumption is down below 0.5W. RPS-200 is able to be used for both Class I (with FG) and Class 1l (no FG) system design. The extremely low leakage current s less than 130 ¢A. In addition, it conforms to the international medical regulations (2*”MOPP) and EMC BS EN/EN55011, perfectly fitting all kinds of BF rated “patient contact” medical system equipment.

Model Encoding

Model Encoding

Type Description Note
Blank PCB Type In stock
C Enclosed casing Type In stock

SPECIFICATION

MODEL| RPS-200-12 □| RPS-200-15 □| RPS-200-24 □| RPS-200-27 □| RPS-200-48
---|---|---|---|---|---
OUTPUT| DC VOLTAGE| 12V| 15V| 24V| 27V| 48V
CURRENT| 10CFM| 16.7A| 13.4A| 8.4A| 7.5A| 4.2A
Convection| 11.7A| 9.4A| 5.9A| 5.3A| 3A
RATED POWER| 10CFM| 200.4W| 201W| 201.6W| 202.5W| 201.6W
Convection| 140.4W| 141W| 141.6W| 143.1W| 144W
RIPPLE & NOISE (max.) Note.2| 100mVp-p| 100mVp-p| 120mVp-p| 120mVp-p| 120mVp-p
VOLTAGE ADJ. RANGE| 11.4~12.6V| 14.3~15.8V| 22.8~25.2V| 25.6 ~ 28.4V| 45.6 ~50.4V
VOLTAGE TOLERANCE Note.3| ±2.0%| ±2.0%| ±1.0%| ±1.0%| ±1.0%
LINE REGULATION| ±0.5%| ±0.5%| ±0.5%| ±0.5%| ±0.5%
LOAD REGULATION| ±1.0%| ±1.0%| ±1.0%| ±1.0%| ±1.0%
SETUP, RISE TIME| 700ms, 30ms/230VAC           700ms, 30ms/115VAC at full load
HOLD UP TIME (Typ.)| 16ms/230VAC          16ms/115VAC at full load
INPUT| VOLTAGE RANGE **Note.4| 80 ~ 264VAC          113 ~ 370VDC
FREQUENCY RANGE| 47 ~ 63Hz
POWER FACTOR| PF>0.94/230VAC PF>0.98/115VAC at full load
EFFICIENCY (Typ.)| 93%| 93.5%| 94%| 94%| 95%
AC CURRENT (Typ.)| 2A/115VAC         1A/230VAC
INRUSH CURRENT (Typ.)| COLD START 30A/115VAC            60A/230VAC
LEAKAGE ****CURRENT (max.)Note.5| Earth leakage current < 130μA/264VAC , Touch current < 40μA/264VAC
PROTECTION| OVERLOAD| 110 ~ 140% rated output power
Protection type : Hiccup mode, recovers automatically after fault condition is removed
OVER VOLTAGE| 13.2 ~ 15.6V| 16.5 ~ 19.5V| 26.4 ~ 31.2V| 29.7 ~ 35V| 52.8 ~ 62.4V
Protection type : Shut down o/p voltage, re-power on to recover
OVER TEMPERATURE| Protection type : Shut down o/p voltage, re-power on to recover
FUNCTION| FAN SUPPLY| 12V@0.5A for driving a fan ; tolerance +15% ~ -15% at main output 20% rated current (10CFM)
ENVIRONMENT| WORKING TEMP.| -30 ~ +70℃ (Refer to “Derating Curve”)
WORKING HUMIDITY| 20 ~ 90% RH non-condensing
STORAGE TEMP., HUMIDITY| -40 ~ +85℃, 10 ~ 95% RH non-condensing
TEMP. COEFFICIENT| ±0.03%/℃ (0 ~ 50℃)
VIBRATION| 10 ~ 500Hz, 2G 10min./1cycle, 60min. each along X, Y, Z axes
OPERATING ALTITUDE Note.6| 5000 meters
SAFETY & **EMC (Note 7)| SAFETY STANDARDS| IEC60601-1, TUV BS EN/EN60601-1, EAC TP TC 004,UL ANSI / AAMI ES60601-1 (3.1 version),CAN/CSA-C22.2 No. 60601-1:14 – Edition 3 approved;  Design refer to BS EN/EN60335-1
ISOLATION RESISTANCE| Primary-Secondary:  2xMOPP,  Primary-Earth:1xMOPP, Secondary-Earth:1xMOPP
WITHSTAND VOLTAGE| I/P-O/P:4KVAC I/P-FG:2KVAC  O/P-FG:1.5KVAC
ISOLATION **RESISTANCE| I/P-O/P, I/P-FG:100M Ohms / 500VDC / 25℃/ 70% RH
EMC EMISSION| Parameter| Standard| Test Level / Note
Conducted emission| BS EN/EN55011 (CISPR11)| Class B
Radiated emission| BS EN/EN55011 (CISPR11)| Class A (for Class II);Class B (for Class I)
Harmonic current| BS EN/EN61000-3-2| Class A
Voltage flicker| BS EN/EN61000-3-3| —–
EMC IMMUNITY| BS EN/EN55035, BS EN/EN60601-1-2
Parameter| Standard| Test Level / Note
ESD| BS EN/EN61000-4-2| Level 4, 15KV air ; Level 4, 8KV contact
RF field susceptibility| BS EN/EN61000-4-3| Level 3, 10V/m( 80MHz~2.7GHz )Table 9, 9~28V/m( 385MHz~5.78GHz )
EFT bursts| BS EN/EN61000-4-4| Level 3, 2KV
Surge susceptibility| BS EN/EN61000-4-5| Level 4, 4KV/Line-FG ; 2KV/Line-Line
Conducted  susceptibility| BS EN/EN61000-4-6| Level 3, 10V
Magnetic field immunity| BS EN/EN61000-4-8| Level 4, 30A/m
Voltage dip, interruption| BS EN/EN61000-4-11| 100% dip 1 periods, 30% dip 25 periods,100% interruptions 250 periods
OTHERS| MTBF| 2669.7K hrs min.     Telcordia SR-332 (Bellcore) ; 500.3K hrs min.     MIL-HDBK-217F (25℃)
DIMENSION (LWH)| PCB:101.650.829mm or 4″2″1.14″inch ; Enclosed type:103.46240mm or 4.07″2.44″1.57″inch
PACKING| PCB:0.19Kg; 72pcs/14.7Kg/0.84CUFT ; Enclosed type:0.3Kg; 60pcs/19Kg/1.06CUFT
NOTE**|

  1. All parameters NOT specially mentioned are measured at 230VAC input, rated load and 25 of ambient temperature.
  2. Ripple & noise are measured at 20MHz of bandwidth by using a 12″ twisted pair-wire terminated with a 0.1μf & 47μf parallel capacitor.
  3. Tolerance : includes set up tolerance, line regulation and load regulation.
  4. Derating may be needed under low input voltages. Please check the derating curve for more details.5. Touch current was measured from primary input to DC output.
  5. The ambient temperature derating of 3.5℃/1000m with fanless models and of 5℃/1000m with fan models for operating altitude higher than 2000m(6500ft).
  6. The power supply is considered a component which will be installed into a final equipment. All the EMC tests are been executed by mounting the unit on a 360mm*360mm metal plate with 1mm of thickness. The final equipment must be re-confirmed that it still meets EMC directives. For guidance on how to perform these EMC tests, please refer to “EMI testing of component power supplies.”(as available on http://www.meanwell.com)

w.comPr.oduct Liability Disclaimer:For detailed information, please refer to https://www.meanwell.com/serviceDisclaimer.aspx

Block Diagram

fosc: 65KHz

Block Diagram

Derating Curve

AMBIENT TEMPERATURE (C)
Derating Curve

Output Derating VS Input Voltage

INPUT VOLTAGE (VAC) 60Hz
Output Derating

Mechanical Specification

RPS-200 (PCB Type)

Product Dimensions
Product Dimensions

RPS-200-C (Enclosed Type)

Product Dimensions
Product Dimensions
Product Dimensions

AC Input Connector (CN1) : JST B3P-VH or equivalent

Pin No. Assignment Mating Housing Terminal
1 AC/L JST VHRor equivalent JST SVH-21T-P1.1or equivalent
2 No Pin
3 AC/N

DC Output Connector (CN2): JST B6P-VH or equivalent

Pin No. Assignment Mating Housing Terminal
1,2,3 +V JST VHRor equivalent JST SVH-21T-P1.1or equivalent
4,5,6 -V

FAN Connector(CN101) : JST B2B-PH-K-S or equivalent

Pin No. Assignment Mating Housing Terminal
1 +12V JST PHR-2or equivalent JST SPH-002T-P0.5Sor equivalent
2 DC COM

Note

  1. The FAN supply is designed to serve as the source of the additive external fan for the cooling of the power supply, enabling the full load delivery and assuring the best life span of the product. Please do not use this FAN supply to rive other devices.
  2. The PCB type(Blank type)EMI Conduction for Class B. Radiation for Class B with FG(Class I ) and Class A without FG(Class II)
  3. The enclosed type(-C type) model is not suitable for the configuration within a Class II (no FG) system but is suggested to used within a Class I (with FG) system.

Installation Manual

Please referto: http://www.meanwell.com/manual.htm|

Downloaded from ATFOW.COM!

File Name:RPS-200-SPEC 2022-09-20

ANSI/AAMI ES60601-1 BS EN/EN60601-1 IEC60601-1 TPTC004