DLX-45552-VPSC Pressure Injectable Arrowg Instructions

June 13, 2024
Arrow

DLX-45552-VPSC Pressure Injectable Arrowg

Product Information

  • Product Name: ADVANCE

  • Features: Antimicrobial, Antithrombogenic

  • Model Number: DLX-45552-VPSC

  • Catheter Length: 55 cm

  • Diameter: 5.5 Fr.

  • Lumen: 2 Fr.

  • Guide Wire Diameter: .018 inch (0.46 mm)
    spring-wire guide

  • Contents:

    • GlideThruTM Peel-Away Sheath: 5.5 Fr. x 2-3/4 (7 cm) Radiopaque
      over 5.5 Fr. Dilator

    • Spring-Wire Guide, Nitinol, Marked: .018 (0.46 mm) dia. x
      17-3/4 (45 cm) (Straight Soft Tip on One End – Straight Stiff Tip
      on Other) with Arrow Advancer

Product Usage Instructions

  1. Ensure that the catheter and accessories are sterile before
    use.

  2. Select the appropriate size of the ADVANCE catheter based on
    the patient’s requirements.

  3. Prepare the insertion site according to standard medical
    procedures.

  4. Peel away the GlideThruTM Peel-Away Sheath to expose the
    Dilator.

  5. Insert the Dilator into the desired location while gently
    advancing it.

  6. Once the Dilator is in place, remove it carefully, leaving the
    catheter in position.

  7. Thread the Spring-Wire Guide through the catheter lumen,
    ensuring that the Soft Tip is on the desired end.

  8. Advance the Spring-Wire Guide to the desired location using the
    Arrow Advancer.

  9. When the Spring-Wire Guide is in place, remove the Arrow
    Advancer and gently withdraw the Spring-Wire Guide, leaving the
    catheter in position.

  10. Secure the catheter in place using appropriate medical
    procedures.

Please consult the user manual for more detailed instructions
and safety precautions.

LBL071556 R00 (2023-05)

ADVANCE
Antimicrobial Antithrombogenic

DLX-45552-VPSC

5.5 2 Fr.

Lumen

55 cm catheter length

.018 inch dia. spring-wire guide

Pressure Injectable Arrowg+ard Blue Advance® Two-Lumen PICC pre-loaded with Arrow® VPS PrecisionTM Stylet
Contents:

1: Two Lumen TaperFree® Catheter with Arrowg+ard Blue Advance® Antimicrobial/Antithrombogenic Protection1: 5.5 Fr. (1.88 mm OD) x 55 cm, Pressure Injectable, Contamination Guard, Arrow® VPS PrecisionTM Stylet, Tuohy-Borst Adapter, Stiffening Stylet (0.36 mm dia. x 75 cm with straight distal tip) and T-Port Connector
1: GlideThruTM Peel-Away Sheath: 5.5 Fr. x 2-3/4″ (7 cm) Radiopaque over 5.5 Fr. Dilator
1: Spring-Wire Guide, Nitinol, Marked: .018″ (0.46 mm) dia. x 17-3/4″ (45 cm) (Straight Soft Tip on One End – Straight Stiff Tip on Other) with Arrow Advancer
1: Injection Needle: SafetyGlideTM2 25 Ga. x 5/8″ (1.60 cm) RW 1: Safety Introducer Needle: Echogenic 21 Ga. x 2-3/4″ (7 cm) TW 1: Blunt Fill Needle: 18 Ga. x 1-1/2″ (3.81 cm) 1: Introducer Needle: Echogenic 21 Ga. x 2-3/4″ (7 cm) TW 1: Syringe: 5 mL Luer-Lock 4: 10 mL Luer-Lock Pre-Filled Saline Syringe 1: 5 mL 1% Lidocaine HCl Solution and Alcohol Prep 1: 3 mL Applicator 2% CHG and 70% IPA ChloraPrep®3 One-Step Solution 2: 3 mL Applicator 2% CHG and 70% IPA ChloraPrep®3 One-Step Solution with
Hi-Lite OrangeTM Tint 2: Dust Cap: Non-Vented 1: SharpsAway® Disposal Cup 1: SharpsAway® II Locking Disposal Cup 1: Catheter Trimmer 1: Towel: 24″ x 36″ (61 cm x 91 cm) 1: Maximal Barrier DrapeTM with two 13 cm (5″) x 15 cm (6″) fenestrations 1: Safety Scalpel: #11 1: Sterile Procedure Sign

1: Patient ID Card 1: Patient Information Booklet 1: Medication Label: 1% Lidocaine 2: Paper Tape Measure 1: Tourniquet 1: Foam Electrodes (3 per pack) 10: Gauze Pad: 2″ x 2″ (5 cm x 5 cm) 10: Gauze Pad: 4″ x 4″ (10 cm x 10 cm) 1: Surgical Apparel: Impervious Gown 1: Dressing: Tegaderm®4 CHG 3-1/2″ x 4-1/2″ (8.5 cm x 11.5 cm) 1: Transducer Cover with Gel and Rubber Bands 1: Surgical Apparel: Mask with Eye Shield 1: Dressing: STATLOCK®5 PICC Plus Catheter Stabilization Device 1: Surgical Apparel: Bouffant Cap 1: Surgical Apparel: Mask 1: Tape: Steri-Strip®4
1Licensed under US Patent No. 7,329,412 and 11,219,706. 2A trademark of Becton, Dickinson and Company. 3A registered trademark of CareFusion or one of its subsidiaries. 4A registered trademark of 3M Company. 5A registered trademark of C. R. Bard, Inc.

Lumen

Priming Volume*
(mL)

Gravity Flow Rate
(mL/hr)

Pump Flow Rate
(mL/hr)

MAX Pressure Injection
Flow Rate** (mL/sec)

MR Safe Not made with natural rubber latex. Store between 20 ­ 25°C (68 ­ 77°F). Arrow International provides the enclosed medication label(s) for your convenience. Please ensure that these labels are applied to the correct syringe and corresponding medication.
Contraindications: The Pressure Injectable Arrowg+ard Blue Advance antimicrobial/antithrombogenic catheter is contraindicated: · for patients with known hypersensitivity to chlorhexidine · in the presence of device related infection in the intended insertion vessel or catheter pathway · in the presence of thrombosis in the intended insertion vessel or catheter pathway

Pink (18 Ga.)

0.56

350

3050

5

White (18 Ga.)

0.58

360

3100

5

  • Priming volumes are approximate and are done without accessories. Flow rate values are approximate and are determined using deionized water at 100 cm head height. Pump flow rates are determined at maximum pump pressure of 10 psig and represent approximate flow capabilities. ** Pressure injection flow rates are determined at the injector pressure setting of 300 psi maximum using media of 11.8 centipoise viscosity, with 152 cm pressure tubing.