MEAN WELL RPS-45 Series 45W Reliable Green Medical Power Supply Owner’s Manual

June 13, 2024
MEAN WELL

MEAN WELL RPS-45 Series 45W Reliable Green Medical Power Supply

MEAN-WELL-RPS-45-Series-45W-Reliable-Green-Medical-Power-Supply-
PRODUCT

Product Information

  • Product Name: 45W Reliable Green Medical Power Supply
  • Series: RPS-45
  • User’s Manual: EN 60601-1 IEC 60601-1
  • Features: Highly reliable, green PCB type, high power density, accepts 80~264VAC input, various output voltages between 3.3V and 48V, working efficiency up to 91%, extremely low no load power consumption below 0.1W, suitable for Class(no FG) system design, extremely low leakage current less than 100A, conforms to international medical regulations (2*MOPP) and EMC BS EN/EN55011
  • Applications: Oral irrigator, Hemodialysis machine, Medical computer monitors, Sleep apnea devices
  • GTIN CODE: MW Search

Product Usage Instructions

To use the 45W Reliable Green Medical Power Supply (RPS-45), follow these instructions:

  1. Ensure the input voltage is within the range of 80~264VAC.
  2. Select the appropriate output voltage based on your requirements. The RPS-45 offers various output voltages between 3.3V and 48V.
  3. Connect the power supply to your device, ensuring proper polarity.
  4. Ensure the working temperature is within the range of -30°C to +70°C. Refer to the Derating Curve for temperature adjustments.
  5. Maintain a working humidity between 20% and 90% RH, non-condensing.
  6. Store the power supply in a temperature range of -40°C to +85°C with a humidity range of 10% to 95% RH, non-condensing.
  7. Ensure the operating altitude does not exceed 4000 meters.
  8. Follow the safety standards and regulations specified, including IEC60601-1, TUV BS EN/EN60601-1, EAC TP TC 004, UL ANSI / AAMI ES60601-1, and CAN/CSA-C22.2 No. 60601-1:14 – Edition 3.

Note:
For more detailed information and specifications, please refer to the user manual provided.

Features

  • 3″×2″ compact size
  • Medical safety approved (2 × MOPP) according to ANSI/AAMI ES60601-1 and IEC/BS EN/EN60601-1
  • Suitable for BF application with appropriate system consideration
  • Cooling by free air convection
  • EMI class B for class II configuration
  • No load power consumption<0.1W
  • Extremely low leakage current
  • Protections: Short circuit / Overload / Over voltage
  • Lifetime > 50K hours
  • Operating altitude up to 4000 meters
  • 3 years warranty

Applications

  • Oral irrigator
  • Hemodialysis machine
  • Medical computer monitors
  • Sleep apnea devices

GTIN CODE

Description

RPS-45 is a 45W highly reliable green PCB type medical power supply with a high power density on the 3″ by 2″ footprint. It accepts 80~264VAC input and offers various output voltages between 3.3V and 48V. The working efficiency is up to 91% and the extremely low no load power consumption is down below 0.1W. RPS-45 is able to be used for Class I (no FG) system design. The extremely low leakage current is less than 100MA. In addition, it conforms to international medical regulations (2*MOPP) and EMC BS EN/EN55011, perfectly fitting all kinds of BF rated “patient contact” medical system equipment.

Model Encoding

SPECIFICATION

ORDER NO.| RPS453. 3| RPS455| RPS457. 5| RPS4512| RPS4515| RPS4524| RPS4548
---|---|---|---|---|---|---|---






OUTPUT

| DC VOLTAGE| 3.3V| 5V| 7.5V| 12V| 15V| 24V| 48V
RATED CURRENT| 8A| 8A| 5.4A| 3.8A| 3A| 1.9A| 0.94A
CURRENT RANGE| 0 ~ 8.8A| 0 ~ 8.8A| 0 ~ 5.95A| 0 ~ 4.18A| 0 ~ 3.3A| 0 ~ 2.1A| 0 ~ 1.03A
RATED POWER| 26.4W| 40W| 40.5W| 45.6W| 45W| 45.6W| 45.1W
PEAK LOAD(10sec.) **Note.2| 29W| 44W| 44.6W| 50.2W| 49.5W| 50.2W| 49.4W
RIPPLE & NOISE (max.) Note.3| 60mVp-p| 60mVp-p| 80mVp-p| 100mVp-p| 100mVp-p| 120mVp-p| 120mVp-p
VOLTAGE ADJ.RANGE| 3.1~3.6V| 4.7~5.5V| 7.12~8.3V| 11.4~13.2V| 13.5~16.5V| 22.8~27.6V| 45.6~52.8V
VOLTAGE TOLERANCE Note.4| ±2.0%| ±2.0%| ±2.0%| ±2.0%| ±1.0%| ±1.0%| ±1.0%
LINE REGULATION| ±0.5%| ±0.5%| ±0.5%| ±0.5%| ±0.5%| ±0.5%| ±0.5%
LOAD REGULATION| ±2.0%| ±2.0%| ±2.0%| ±2.0%| ±1.0%| ±1.0%| ±1.0%
SETUP, RISE TIME| 500ms, 30ms / 230VAC         500ms, 30ms / 115VAC at full load
HOLD UP TIME (Typ.)**| 30ms / 230VAC         16ms / 115VAC at full load




INPUT

| VOLTAGE RANGE **Note.5| 80 ~ 264VAC
FREQUENCY RANGE| 47 ~ 63Hz
EFFICIENCY (Typ.)| 80.5%| 83%| 85%| 88%| 89%| 90%| 91%
AC CURRENT (Typ.)| 1.2A / 115VAC         1A / 230VAC
INRUSH CURRENT (Typ.)| СOLD STAR 30A/115VAC 60A/230VAC
LEAKAGE CURRENT(max.) Note.6**| Touch current< 100μA/264VAC




PROTECTION

| ****

OVERLOAD

| 115 ~ 150% rated output power
Protection type : Hiccup mode, recovers automatically after fault condition is removed


OVER VOLTAGE

| 3.8~5V| 5.7~6.8V| 8.6~11.3V| 13.8~16.2V| 17.2~20.3V| 28.4~32.4V| 55.2~64.8V
Protection type : Shut down o/p voltage, re-power on to recover




ENVIRONMENT

| WORKING TEMP.| -30 ~ +70℃ (Refer to “Derating Curve”)
WORKING HUMIDITY| 20% ~ 90% RH non-condensing
STORAGE TEMP., HUMIDITY| -40 ~ +85℃, 10 ~ 95% RH non-condensing
TEMP. COEFFICIENT| ±0.03% / ℃ (0 ~ 50℃)
VIBRATION| 10 ~ 500Hz, 2G 10min./1cycle, period for 60min. each along X, Y, Z axes
OPERATING ALTITUDE Note.7| 4000 meters










SAFETY & EMC

(Note. 8)

| SAFETY STANDARDS| IEC60601-1, TUV BS EN/EN60601-1, EAC TP TC 004,UL ANSI / AAMI ES60601-1 (3.1 version),

CAN/CSA-C22.2 No. 60601-1:14 – Edition 3 approved;Design refer to BS EN/EN60335-1

ISOLATION LEVEL| Primary-Secondary: 2xMOPP
WITHSTAND VOLTAGE| I/P-O/P: 4KVAC
ISOLATION RESISTANCE| I/P-O/P:100M Ohms / 500VDC / 25℃/ 70% RH



EMC EMISSION

| Parameter| Standard| Test Level / Note
Conducted emission| BS EN/EN55011 (CISPR11)| Class B
Radiated emission| BS EN/EN55011 (CISPR11)| Class B
Harmonic current| BS EN/EN61000-3-2| Class A
Voltage flicker| BS EN/EN61000-3-3| —–






EMC IMMUNITY

| BS EN/EN55035, BS EN/EN60601-1-2
Parameter| Standard| Test Level / Note
ESD| BS EN/EN61000-4-2| Level 4, 15KV air ; Level 4, 8KV contact
RF field susceptibility| BS EN/EN61000-4-3| Level 3, 10V/m( 80MHz~2.7GHz )

Table 9, 9~28V/m( 385MHz~5.78GHz )

EFT bursts| BS EN/EN61000-4-4| Level 3, 2KV
Surge susceptibility| BS EN/EN61000-4-5| Level 4, 2KV/Line-Line
Conducted susceptibility| BS EN/EN61000-4-6| Level 3, 10V
Magnetic field immunity| BS EN/EN61000-4-8| Level 4, 30A/m


Voltage dip, interruption

| BS EN/EN61000-4-11| 100% dip 1 periods, 30% dip 25 periods,

100% interruptions 250 periods


OTHERS

| MTBF| 3334.3K hrs min.      Telcordia SR-332 (Bellcore) ; 726.2K hrs min.      MIL-HDBK-217F (25℃)
DIMENSION (LWH)| 76.250.824mm or 3″ 2″ 0.945″ inch
PACKING| 0.11Kg; 120pcs/14.2Kg/0.94CUFT






NOTE

| 1. All parameters NOT specially mentioned are measured at 230VAC input, rated load and 25℃ of ambient temperature.

2.  33% Duty cycle maximum within every 30 seconds. Average output power should not exceed the rated power.

3.  Ripple & noise are measured at 20MHz of bandwidth by using a 12″ twisted pair-wire terminated with a 0.1μf & 47μf parallel capacitor.

4.  Tolerance : includes set up tolerance, line regulation and load regulation.

5.  Derating may be needed under low input voltages. Please check the derating curve for more details.

6.  Touch current was measured from primary input to DC output.

7.  The ambient temperature derating of 3.5℃/1000m with fanless models and of 5℃/1000m with fan models for operating altitude higher than 2000m(6500ft).

8.  The power supply is considered a component which will be installed into a final equipment. “All the EMC tests are been executed by mounting the unit on a 360mm*360mm metal plate with 1mm of thickness.” The final equipment must be re-confirmed that it still meets EMC directives. For guidance on how to perform these EMC tests, please refer to “EMI testing of component power supplies.”

(as available on http://www.meanwell.com)

※ Product Liability Disclaimer:For detailed information, please refer to https://www.meanwell.com/serviceDisclaimer.aspx

 | Arrow.com.| File Name:RPS-45-SPEC  2022-09-20

Block Diagram

MEAN-WELL-RPS-45-Series-45W-Reliable-Green-Medical-Power-Supply-fig-
\(2\)

Derating Curve

MEAN-WELL-RPS-45-Series-45W-Reliable-Green-Medical-Power-Supply-fig-
\(3\)

Static Characteristics

MEAN-WELL-RPS-45-Series-45W-Reliable-Green-Medical-Power-Supply-fig-
\(4\)

Mechanical Specification

Top View

MEAN-WELL-RPS-45-Series-45W-Reliable-Green-Medical-Power-Supply-fig-
\(5\)

Side View

MEAN-WELL-RPS-45-Series-45W-Reliable-Green-Medical-Power-Supply-fig-
\(6\)

AC Input Connector (CN1) : JST B3P-VH or equivalent

Pin No. Assignment Mating Housing Terminal
1 AC/N

JST VHR

or equivalent

|

JST SVH-21T-P1.1

or equivalent

2| No Pin
3| AC/L

DC Output Connector (CN2) : JST B4P-VH or equivalent

Pin No. Assignment Mating Housing Terminal
1 +V

JST VHR

or equivalent

|

JST  SVH-21T-P1.1

or equivalent

2| +V
3| -V
4| -V

Installation Manual

Please refer to: http://www.meanwell.com/manual.html.

Downloaded from Arrow.com.