MEAN WELL RPS-30 Series RPS-30-3.3 Power Supply Instruction Manual

June 13, 2024
MEAN WELL

**MEAN WELL RPS-30 Series RPS-30-3.3 Power Supply Instruction Manual

**

Features

  • 3″ x2 compact size
  • Medical safety approved (2 x MOPP) according to ANSI/AAMI ES60601-1 and IEC/BS EN/EN60601-1
  • Suitable for BF application with appropriate system consideration
  • Cooling by free air convection
  • EMI class B for class II configuration
  • No load power consumption<0.1W
  • Extremely low leakage current
  • Protections: Short circuit/Overload / Over voltage Lifetime > 105K hours
  • Operating altitude up to 4000 meters 3 years warranty

Description

RPS-30is a 30W highly reliable green PCB type medical power supply with a high power density on the 3″ by 2” footprint. It accepts 80~264VAC input and offers various output voltages between 3.3V and 48V.

The working efficiency is up to 92% and the extremely low no load power consumption is down below 0.1W. RPS-30is able to be used for Class Il (no FG) system design. The extremely low leakage current is less than 80 #A. In addition, it conforms to international medical regulations (2*MOPP) and EMC BS EN/EN55011, perfectly fitting all kinds of BF rated “patient contact” medical system equipment.

Applications

  • Oral irrigator
  • Hemodialysis machine
  • Medical computer monitors
  • Sleep apnea devices

GTIN CODE

MW Search: hitps://www.meanviell.comiserviceGTIN.aspx

Model Encoding

SPECIFICATION

ORDER NO.| RPS303. 3| RPS305| RPS307. 5| RPS3012| RPS3015| RPS3024| RPS3048
---|---|---|---|---|---|---|---
OUTPUT| DC VOLTAGE| 3.3V| 5V| 7.5V| 12V| 15V| 24V| 48V
RATED CURRENT| 6A| 6A| 4A| 2.5A| 2A| 1.25A| 0.625A
CURRENT RANGE| 0 ~ 6.6A| 0 ~ 6.6A| 0 ~ 4.4A| 0 ~ 2.75A| 0 ~ 2.2A| 0 ~ 1.375A| 0 ~ 0.687A
RATED POWER| 19.8W| 30W| 30W| 30W| 30W| 30W| 30W
PEAK LOAD(10sec.) Note.2| 21.8W| 33W| 33W| 33W| 33W| 33W| 33W
RIPPLE & NOISE (max.) **Note.3| 80mVp-p| 80mVp-p| 80mVp-p| 100mVp-p| 100mVp-p| 150mVp-p| 150mVp-p
VOLTAGE ADJ.RANGE| 3.1~3.6V| 4.7~5.5V| 7.12~8.3V| 11.4~13.2V| 13.5~16.5V| 22.8~27.6V| 45.6~52.8V
VOLTAGE TOLERANCE| ±2.0%| ±2.0%| ±2.0%| ±2.0%| ±2.0%| ±1.0%| ±1.0%
LINE REGULATION| ±0.5%| ±0.5%| ±0.5%| ±0.5%| ±0.5%| ±0.5%| ±0.5%
LOAD REGULATION| ±1.0%| ±1.0%| ±1.0%| ±1.0%| ±1.0%| ±1.0%| ±1.0%
SETUP, RISE TIME| 200ms, 30ms / 230VAC           200ms, 30ms / 115VAC at full load
HOLD UP TIME (Typ.)| 30ms / 230VAC          16ms / 115VAC at full load
INPUT| VOLTAGE RANGE Note. 5| 80 ~ 264VAC
FREQUENCY RANGE| 47 ~ 63Hz
EFFICIENCY (Typ.)| 80%| 82%| 84%| 88%| 89%| 89.5%| 92%
AC CURRENT (Typ.)| 1A / 115VAC         0.5A / 230VAC
INRUSH CURRENT (Typ.)| СOLD STAR 30A/115VAC  60A/230VAC
LEAKAGE CURRENT(max.) Note.6| Touch current< 80μA/264VAC
PROTECTION| OVERLOAD| 115 ~ 150% rated output power
Protection type : Hiccup mode, recovers automatically after fault condition is removed
OVER VOLTAGE| 3.8~5V| 5.7~6.8V| 8.6~11.3V| 13.8~16.2V| 17.2~20.3V| 28.4~32.4V| 55.2~64.8V
Protection type : Shut down o/p voltage, re-power on to recover
ENVIRONMENT| WORKING TEMP.| -30 ~ +70℃ (Refer to “Derating Curve”)
WORKING HUMIDITY| 20% ~ 90% RH non-condensing
STORAGE TEMP., HUMIDITY| -40 ~ +85℃, 10 ~ 95% RH non-condensing
TEMP. COEFFICIENT| ±0.03% / ℃ (0 ~ 50℃)
VIBRATION| 10 ~ 500Hz, 2G 10min./1cycle, period for 60min. each along X, Y, Z axes
OPERATING ALTITUDE Note.7| 4000 meters
SAFETY & **EMC (Note. 8)| SAFETY STANDARDS| IEC60601-1, TUV BS EN/EN60601-1, EAC TP TC 004, UL ANSI / AAMI ES60601-1 (3.1 version),CAN/CSA-C22.2 No. 60601-1:14 – Edition 3 approved; Design refer to BS EN/EN60335-1
ISOLATION LEVEL| Primary-Secondary: 2xMOPP
WITHSTAND VOLTAGE| I/P-O/P: 4KVAC
ISOLATION RESISTANCE| I/P-O/P:100M Ohms / 500VDC / 25℃/ 70% RH
EMC EMISSION| Parameter| Standard| Test Level / Note
Conducted emission| BS EN/EN55011 (CISPR11)| Class B
Radiated emission| BS EN/EN55011 (CISPR11)| Class B
Harmonic current| BS EN/EN61000-3-2| Class A
Voltage flicker| BS EN/EN61000-3-3| —–
EMC IMMUNITY| BS EN/EN55035, BS EN/EN60601-1-2
Parameter| Standard| Test Level / Note
ESD| BS EN/EN61000-4-2| Level 4, 15KV air ; Level 4, 8KV contact
RF field susceptibility| BS EN/EN61000-4-3| Level 3, 10V/m( 80MHz~2.7GHz )Table 9, 9~28V/m( 385MHz~5.78GHz )
EFT bursts| BS EN/EN61000-4-4| Level 3, 2KV
Surge susceptibility| BS EN/EN61000-4-5| Level 4, 2KV/Line-Line
Conducted susceptibility| BS EN/EN61000-4-6| Level 3, 10V
Magnetic field immunity| BS EN/EN61000-4-8| Level 4, 30A/m
Voltage dip, interruption| BS EN/EN61000-4-11| 100% dip 1 periods, 30% dip 25 periods,100% interruptions 250 periods
OTHERS| MTBF| 3550.0K hrs min.          Telcordia SR-332 (Bellcore) ; 628.7K hrs min.                MIL-HDBK-217F (25℃)
DIMENSION (LWH)| 76.250.824mm or 3″ 2″ 0.945″ inch
PACKING| 0.09Kg; 120pcs/11.8Kg/0.94CUFT
NOTE|

  1. All parameters NOT specially mentioned are measured at 230VAC input, rated load and 25℃ of ambient temperature.
  2.  33% Duty cycle maximum within every 30 seconds. Average output power should not exceed the rated power.
  3. Ripple & noise are measured at 20MHz of bandwidth by using a 12″ twisted pair-wire terminated with a 0.1μf & 47μf parallel capacitor.
  4. Tolerance : includes set up tolerance, line regulation and load regulation.
  5. Derating may be needed under low input voltages. Please check the derating curve for more details.
  6. Touch current was measured from primary input to DC output.
  7. The ambient temperature derating of 3.5℃/1000m with fanless models and of 5℃/1000m with fan models for operating altitude higher than 2000m(6500ft).
  8. The power supply is considered a component which will be installed into a final equipment. The final equipment must be re-confirmed that it still meets EMC directives. For guidance on how to perform these EMC tests, please refer to “EMI testing of component power supplies.”

(as available on http://www.meanwell.com)※ Product Liability Disclaimer For detailed information, please refer to https://www.meanwell.com/serviceDisclaimer.aspx

| Arrow.com.| File Name:R PS-30 -SPEC __2022-09-20

Block Diagram

Derating Curve

Static Characteristics

Mechanical Specification

AC Input Connector (CN1): JST B3P-VH or equivalent

Pin No. Assignment Mating Housing Terminal
1 AC/N JST VHRor equivalent JST SVH-21T-P1.1or equivalent
2 No Pin
3 AC/L

DC Output Connector (CN2) : JST B2P-VH or equivalent

Pin No. Assignment Mating Housing Terminal
1 +V JST VHRor equivalent JST SVH-21T-P1.1or equivalent
2 -V

Installation Manual

Please referto : http://www.meanwell.com/manual.htm|

ANSI/AAMI ES60601-1 BS EN/EN60601-1 IEC60601-1 TPTC004

References

Read User Manual Online (PDF format)

Read User Manual Online (PDF format)  >>

Download This Manual (PDF format)

Download this manual  >>