Alcolizer LE5 Alcohol Tester User Guide

June 13, 2024
Alcolizer

Alcolizer LE5 Alcohol Tester

Alcolizer LE5 Alcohol Tester

Passive Testing

  1. It is recommended the subject not consume food, drink or smoke 15 minutes prior to taking a test.
    Symbol
    Press the Test key “ ” to turn the LE5 Alcohol Tester on.
    Ensure the test type is set to “Passive Std” (no mouthpiece).
    A “press passive test” message will display
    Passive Testing

  2. Press the Test key – the screen will display “Please Wait” then “Ready for passive Test”.
    Hold the instrument 50mm – 100mm in front of the subject’s mouth so that the subject is breathing or talking directly into the device’s inlet.
    Passive Testing

  3. If using pressure activated passive, instruct the donor to blow into the device until you tell them to stop. The device will take a sample automatically. If using button activated passive, instruct the subject to count from 20 onwards in a one second rhythm until you tell them to stop.
    While they are counting, press the Test key. You will hear 3 clicks.
    Tell the subject to stop blowing or counting after the 3rd click.
    Passive Testing

  4. The device will analyse, emit a long beep, and then display a text based result.
    A green screen with “Pass No Alcohol” text indicates alcohol is not detected.
    A red screen with “Warning Alcohol Detected” text indicates alcohol is detected.
    In the latter case, the device automatically switches to a mouthpiece “Standard” test displaying “press for active test”.
    Passive Testing

Passive Testing Mouthpiece Testing

  1. The subject is not to consume food, drink or smoke 15 minutes prior to taking a test.
    Symbol

    Press the Test key “ ” to turn the LE5 Alcohol Tester on. Ensure the test type is set to “Standard” (with mouthpiece).
    A “press to test” message will display
    Passive Testing Mouthpiece Testing
    Attach an *Alkalized approved mouthpiece. Press the Test key – the screen will display “Please Wait” then “Blow”.
    Hold the breathalyzer in front of the subject’s mouth so that the subject can blow into the mouthpiece.
    Passive Testing Mouthpiece Testing
    Instruct the subject to take a deep breath and blow gently into the mouthpiece until you tell them to stop.
    You will hear a sequence of beeps, a click then a long beep to indicate the test is complete.
    Tell the subject to stop blowing.
    The device will analyze and then display a colour coded numerical Brace result.
    Passive Testing Mouthpiece Testing

  2. A green screen with “0.000g/210L” BrAC indicates that alcohol is not detected.
    A yellow screen with numerical BrAC indicates that alcohol is present but under the preconfigured threshold BrAC level.
    A red screen with numerical BrAC indicates that alcohol is detected and is over locally accepted BrAC levels.
    Passive Testing Mouthpiece Testing

Customer Support

World-class Alcohol and Drug testing solutions
Phone: 1300 789 908
Email:sales@alcolizer.com
www.alcolizer.com

  • Operate an Alcolizer LE5 Alcohol Tester in accordance with applicable regulations, policies and standards.
  • Completion of Alcolizer Certified training is recommended prior to use.
  • *Australian Standard certified for Active (Standard) tests when Alcolizer approved mouthpieces are utilised.
    Information correct as at: 30/06/2022 V400.02.300622

Documents / Resources

| Alcolizer LE5 Alcohol Tester [pdf] User Guide
LE5 Alcohol Tester, LE5, Alcohol Tester, Tester
---|---

Read User Manual Online (PDF format)

Read User Manual Online (PDF format)  >>

Download This Manual (PDF format)

Download this manual  >>

Related Manuals