Alcolizer g-210L BrAC AlcoCONNECT Toolbox User Manual
- June 13, 2024
- Alcolizer
Table of Contents
Alcolizer g-210L BrAC AlcoCONNECT Toolbox User Manual
Introduction
AlcoCONNECT Toolbox assists in the analysis of data downloaded and software update from the following Alcolizer devices. Not all features apply to all devices.
- HH3 Alcohol Tester
- HH4 Alcohol Tester
- LE5 Alcohol Tester
- LE5 Druglizer
The AlcoCONNECT Toolbox software can be downloaded from https://www.alcolizer.com/software-updates/
Installing AlcoCONNECT Toolbox Software
AlcoCONNECT Toolbox is downloaded as a zip file, so you may need to extract the Windows installer package. Install the program. Installation is quick, there is no set up needed or any successful notification.
Following a successful installation, an AlcoCONNECT Toolbox icon will appear on the Desktop. If it does not, then search for AlcoCONNECT in your program list.
Figure 1 AlcoCONNECT Toolbox Shortcut
Connecting Alcolizer device to AlcoCONNECT Toolbox
Once installed onto your computer, use AlcoCONNECT Toolbox as follows:
- Connect the Alcolizer device to the computer via the USB cable.
- Double click the AlcoCONNECT Toolbox icon.
- Turn the Alcolizer device on.
A successful connection between the Alcolizer Device and the computer/AlcoCONNECT
Toolbox will be shown by a green bar on the Read Event Log window.
Figure 2 Green bar indicates a device is connected
Configuring AlcoCONNECT Toolbox
Language
Access Language Settings by clicking on File, then Language in the menu and
then choose the language you would like to use
Figure 3 Customise the language used in AlcoCONNECT Toolbox
Log Settings
Access Log Settings by clicking on File, then Options in the menu.
Record Filter
When the Log Settings window appears, the Record Filter tab will selected by
default. You can filter the data by record type and column. Check or uncheck
boxes as required to customise the log output format.
Figure 4 – Log Settings Window
Default Save Options
In the Log Settings window click on the Save Options tab.
You can customise the:
- You can customise the:
- Path
- Click on Browse and choose the required directory
- File Name Pattern –
- Click on Show Tokens and a list of tokens are displayed with description and sample.
- You can add a file name pattern (Token) to the file name pattern by double clicking on the row.
- You can add spaces or characters between the tokens in the file name pattern field.
- Token options include
- [MMSerial] – Main module serial number
- [SMSerial] – Sample module serial number
- [MMFWV] – Main module firmware version
- [SMFWV] – Sample module firmware version
- [DateTime0-3] – date and time in various formats
- File Format – options include
- CSV
- XLSX
- XML
Figure 5 Set default Save Options
Firmware Upgrade Location
In the Log Settings window click on the Firmware Upgrade Location tab. Click
on Browse andchoose the directory where you will save any firmware files
provided by Alcolizer. Click on OK.
Logs
Reading Logs
Click the Read Event Log button and the Event Log will then be displayed. You
can also click on Log Records in the menu and choose Read Event Log.
Figure 6 – Read Event Log Button
Event Logs can be saved for future analysis.
-
Click the “Read Event Log” button to populate the event log with records.
-
Click on File > Save Event Log.
-
Choose where you want to save the log.
-
Click the “Save as type:” drop down menu and select required format (Options are CSV, XML or xlsx).
Figure 7 – Save as type drop down options
-
Click the “Save” button, and the file will save in the selected format/location.
Figure 8 Breathalyser Event Log
Figure 9 Druglizer Event Log
Navigating around the Event Log Page
Refer to the below table and Figure 5 for a description of the Event Log page.
Figure
Reference| Icon, Button or
Heading| Description
---|---|---
1| Read Event Log Icon| Select to download records from the Alcolizer device.
2| File Icon| Select for:Save Event Log — save as a .csv or xml file to a file
location in computer or elsewhere.Language – Choose a language if the
Alcolizer device is configured for multiple languages.Options – Set Log
Settings.
3| Log Records| Activates the Event Log application. Same function as the Read
Event Log icon.
4| Set Access| Used by authorised persons to access certainAIcoCONNECT Toolbox
functionality. Select the icon after inputting correct access code.
5| Sample Serial Number| The serial number of the sample module.
6| Record Number| The number of each record.
7| Event Type| The type of test performed or event. E.g. breath test type
(passive, standard), test timeout or refusal etc.
8| Date and Time| The data and time of the event.
9| Result Qualitative| Passive test result.
10| True Result| Mouthpiece test result.
11| Serial Number Main Serial No| The serial number of the Alcolizer device.
12| Drug Name| The name of the drug tested
13| Drug Test Result| The drug test result
14| Cartridge ID| The production ID of the cartridge batch
15| Cartridge Expiry| Expiry date of the test cartridge
NOTE Firmware version display is now displayed in a new format X.Y.0 this is the equivalent to previous format X.Y
Saving Logs
When you have read event logs, you can save them for future analysis. See
Default Save Options to see how to configure how logs are saved by
default.
- Click the “Read Event Log” button to populate the event log with records.
- Click on File > Save Event Log.
- Choose where you want to save the log.
- Click the “Save as type:” drop down menu and select required format (Options are CSV, XML or xlsx).
- Click the “Save” button, and the file will save in the selected format/location.
Figure 10 – Save as type drop down options
Firmware Updates (Druglizer Only)
If you have been provided a firmware update file from Alcolizer Technology, follow these steps to install it on your device. Currently this is only applicable to Druglizer devices. (According to Ang).
-
Connect your device to AlcoCONNECT Toolbox.
-
Click on Tools in the menu and then click on Firmware updates.
-
A Firmware Updater window will appear. Click either the small rectangular button on as circled below. Your computer will display a window for you to locate the firmware. Navigate to the folder you unzipped the firmware file into and select the provided file and click on Open.
-
The new firmware file version will be displayed next to the existing firmware. If AlcoCONNECT™ Toolbox recognises newer firmware it will allow an update and so in that instance, click the ‘Update Device Firmware’ button
-
A bar at the bottom of the window will display progress. The instrument will emit a beep to confirm install completion. If required remove and reinsert the Sample Module to help reset the instrument.
-
On your Alcolizer device, open the About screen and check the firmware (Software) Version. The versions displayed should be the same as indicated in the filenames.
Extra Features
To access the following extra features, enter the access code ‘00005000’ and
click on Set Access.
Set Time
A new Set Time button appears next to the Read Event Log button. A popup
window willappear that displays the Alcolizer device time and PC time. You can
choose to sync yourdevice to the PC time or set the date and time manually.
When entering a date manually:
- Enter the date in the first field in this format DD/MM/YY.
- Enter the time in the following format HH:MM:SS PM or HH:MM:SS AM.
- The Manual Setting button will be greyed out if an invalid date or time is entered.
- The incorrect value will be displayed in red text.
Figure 11 Set Alcolizer Device time manually or from the PC time
Configuration Editor
Click on Tools and then Device Config. Click on Read from device to read the
current device config. You will be able to change most but not all
configuration settings. Anythinghighlighted in dark grey cannot be modified.
You should understand what each configuration item changes before modifying
it. An example of changing a config item is explained below using the Code 0
item.
Figure 12 Configuration Editor
Code 0 Example
Code 0 is a code that can be entered on your Breathalyser devices that will be
linked with any breath tests. It can be used to define something about the use
of the device at the time of use, eg who is using it or where are they using
it etc.
Enable Code 0 (Area Code)
- Enter Code 0 in the search bar. This will filter the config items to anything that matches code 0.
- Set the value of configuration item “Code 0” to True. Check the Enable Upload checkbox and click on Write to device.
- A popup with the title Device Busy will appear. If the config upload is small, this only appears for a second.
- Restart the device.
- Enter the User Menu and the item Area Code should appear in the menu.
- Select it and you should be able to enter a value for this.
Figure 13 Enabling Code 0
Figure 14 Code 0 (Area Code) appears in User Menu
Change Code 0 Label
- Enter Code 0 in the search bar. This will filter the config items to anything that matches code 0.
- Set the value of configuration item “Code 0 Label” to Staff ID. Check the Enable Upload checkbox and click on Write to device.
- A popup with the title Device Busy will appear. If the config upload is small, this only appears for a second.
- Restart the device.
- Enter the User Menu and the item Staff ID should appear in the menu.
- Select it and you should be able to enter a value for this.
Figure 15 Changing Code 0 Label
Figure 16 Code 0 (Staff ID) appears in User Menu
Change Code Edit on Start Enable
- Enter Code 0 in the search bar. This will filter the config items to anything that matches code 0.
- Set the value of configuration item “Code 0 Edit on Start Enable” to 1. Check the Enable Upload checkbox and click on Write to device.
- A popup with the title Device Busy will appear. If the config upload is small, this only appears for a second.
- Restart the device.
- Observe that the code screen is now available at startup.
Figure 17 Changing Code 0 Edit on Start Enable
Figure 18 You are now prompted to enter Staff ID on device start up.
Upload a config file
If Alcolizer Technology has provided you with a config file to upload, then
follow these instructions.
- Click on Tools and then Device Config.
- Click on Read from device to read the current device config.
- Click on Open from file and choose the provided S5C config file.
- The changes will be applied to the listed config and each config item that will be uploaded will be checked.
- Click on Write to device to upload the config to your Alcolizer Device.
- A Device Busy pop up will appear and then disappear.
- Check your device to make sure the required config changes were made.
Clearing (Deleting) Event Logs on an Instrument
- Click Log Records > Clear Device Records.
- Click on OK to confirm you would like to clear all records from the Alcolizer device.
- Restart your Alcolizer device before performing any tests.
About Screen
The About screen gives you details about the version that is installed. You should always have the latest version installed. You can find the latest version on our website (www.alcolizer.com)
- To access the About screen, click on About and then About This Program in the top menu.
- Look for Version: HE-#.##.# to find the installed version number.
Read User Manual Online (PDF format)
Read User Manual Online (PDF format) >>