HT ITALIA HT9014 AC true RMS Clamp Meter User Manual

June 4, 2024
HT ITALIA

HT ITALIA HT9014 AC true RMS Clamp Meter

HT-ITALIA-HT9014-AC-true-RMS-Clamp-Meter-product-image

 SAFETY PRECAUTIONS AND PROCEDURES

The word “meter” in this manual means generically both the model HT9014 and the model HT9015 except notation specifically indicated. The meter complies with IEC/EN61010-1. For your own safety and in order to avoid damaging the instrument, you’re recommended to keep to the instructions contained in this manual and read carefully all the notes preceded by the symbol .

Take extreme care for the following conditions while measuring:

  • Do not measure voltage or current in humid or wet environment.
  • Do not use the meter in presence of explosive gas (material), combustible gas (material), steam or dust.
  • Insulate yourself from the object to be tested.
  • Do not touch exposed metal (conductive) parts such as test lead ends, sockets, fixing objects, circuits, etc.
  • If you detect anomalies of testing end (metal part) and attachment of the meter such as breakages, deformations, foreign substances, no display, etc., do not take any measurement.
  • Measuring voltage over 20V as it might cause human body electricity conduction.

The followings symbols are used on meter and in this manual:

  • Caution: refer to the instruction manual. An incorrect use may damage the tester or its components
  • High Voltage ranger: electrical shock risk
  • Double insulated instrument
  • AC Voltage or Current
  • DC Voltage or Current
  • Ground reference
PRELIMINARY
  • This apparatus has been designed for use in an environment of pollution degree 2. Indoor use.
  • It measures CURRENT and VOLTAGE on CAT IV 600V and CAT III 1000V plants. For measurement categories please see § 1.4.
  • You must comply with the usual safety regulations aimed at protecting you against the dangerous electric current and protecting the instrument against an incorrect operation.
  • Only the leads supplied with the instrument guarantee compliance with the safety standards. They must be in good conditions and they must be replaced, if necessary, with an identical model.
  • Do not test or connect to any circuit whose voltage or current exceeds the specified overload protection.
  •  Make sure that the batteries are installed correctly.
  • Before connecting the test probes to the installation, check that the function selector is positioned on the required measurement.
  •  Make sure that the LCD and the range indicator show the same as the function desired.
DURING USE

Always keep to the instructions contained in this manual.

CAUTION Non compliance with the CAUTIONs and/or the instructions may damage the tester and/or its components or injure the operator.

  • Before changing the switch position, take off the clamp jaw from the tested conductor or the electrical circuit in order to avoid any accident
  • When the clamp is connected to the circuits to be tested, never touch unused terminals
  • When testing resistors, do not add voltage. Although there is a protection circuit, excessive voltage would cause malfunctioning
  • Before measuring current, remove the voltage-resistance test leads
  • When measuring current, any strong current close to the clamp jaw will affect the accuracy
  • When measuring current, always put the tested conductor in the middle of the clamp jaw in order to obtain a more accurate reading
  • If the reading value or the sign indication remains unchanged during the measurement, check if the HOLD function is active
 AFTER USE
  • Once the measurements are completed, turn the rotary switch to OFF.
  • If you expect not to use the clamp for a long time, remove the battery.
MEASURING (OVERVOLTAGE) CATEGORIES DEFINITIONS

The norm EN 61010-1: Safety requirements for electrical equipment for measurement, control and laboratory use, Part 1: General requirements, defines what measuring category, usually called overvoltage category, is. On § 6.7.4: Measuring circuits, it says: (OMISSIS) circuits are divided into the following measurement categories:

  • Measurement category IV is for measurements performed at the source of the low-voltage installation.
    Examples are electricity meters and measurements on primary overcurrent protection devices and ripple control units.

  • Measurement category III is for measurements performed in the building installation. Examples are measurements on distribution boards, circuit breakers, wiring, including cables, bus-bars, junction boxes, switches, socket-outlets in the fixed installation, and equipment for industrial use and some other equipment, for example, stationary motors with permanent connection to fixed installation.

  • Measurement category II is for measurements performed on circuits directly connected to the low voltage installation..
    Examples are measurements on household appliances, portable tools and similar equipment..

  • Measurement category I is for measurements performed on circuits not directly connected to MAINS.
    Examples are measurements on circuits not derived from MAINS, and specially protected (internal) MAINS-derived circuits. In the latter case, transient stresses are variable; for that reason, the norm requires that the transient withstand capability of the equipment is made known to the user.

GENERAL DESCRIPTION

The meter can perform the herewith measurements:

  • DC and AC TRMS Voltage up to 1000V
  • DC and AC TRMS Current up to 600A
  • Resistance and continuity test with buzzer
  • Capacitance
  • Frequency with test lead and jaws
  • Duty cycle
  • Diode test
  • Temperature with type K probe
  • AC voltage detection with and without contact with integrated sensor

Each parameter can be selected by rotating the 7 positions switch. To facilitate the hold function the HOLD key is available. There are also the following keys: “MODE “MAX/MIN”, “Hz%“, “PK/REL” and “ ”. For their use please see § 4.2. The selected quantity appears on a LCD display with indication of measurement units and functions. An analogical barograph it is also available

TRMS AND MEAN VALUE MEASURING INSTRUMENTS

Safety testers for alternate parameters are divided into two big families:

  • MEAN VALUE instruments: instruments which measure only the value of the wave at the fundamental frequency (50 or 60 Hz)
  • TRUE ROOT MEAN SQUARE instruments, also defined as TRMS: instruments which measure the true root mean square value of the quantity under test.

In presence of a perfectly sinusoidal wave, both families provide identical results. In presence of distorted waves the readings are different. Mean value instruments provide only the value of the fundamental wave while True RMS instruments provide the value of the entire wave, including harmonics (within the passband of the instrument). Accordingly, if the same quantity is measured with both kinds of instruments, the measured values are identical only if the wave is purely sinusoidal. Should it be distorted, True RMS instruments provide higher values than medium value instruments.

TRUE ROOT MEAN SQUARE VALUE AND CREST FACTOR DEFINITIONS

The current effective value is defined as follows: “In an interval of time equivalent to a period, an alternate current with effective value having an intensity of 1A, by passing on a resistor, disperses the same energy which would be dispersed in the same period of time by a direct current having an intensity of 1A”. From this definition comes the numerical expression: G= ∫ The effective value is indicated as RMS (root mean square). The Crest Factor is defined as the ratio between the Peak Value of a signal and its
effective value: CF (G)=  This value varies according to the waveform of the signal, for a purely sinusoidal wave it’s worth 2 =1.41. In presence of distortions the Crest Factor assumes higher values as long as the wave distortion is higher.

PREPARATION FOR USE

INITIAL

The tester has been checked from a mechanical and electrical point of view before shipment.
Every care has been taken to make sure that the instrument reaches you in perfect conditions.
However, it’s advisable to make a rapid check in order to detect any damage which may have occurred in transit. Should this be the case, enter immediately the usual claims with the carrier.
Make sure that all the accessories listed in § 6.3 are contained in the package. In case of discrepancies contact Your dealer.
In case of returning of the tester please keep to the instructions given in § 7.

POWER SUPPLY

The instrument is battery supplied. One battery 9V NEDA 1604 IEC 6F22 JIS 006P is included in the package. Battery life: about 200 hours. When batteries are low, the symbol
“ ” appears on the display. Replace them immediately, following the instructions given in § 5.2
The instrument disposes of the Auto Power OFF function (not disable) consisting in an automatic switching off 15 minutes of idleness.

CALIBRATION

The tester complies with the technical features listed in this manual. Its performances are guaranteed for one year.

STORAGE

In order to guarantee the accuracy of the measurements, after a period of storage under extreme environmental condition, wait for the necessary time so that the tester returns to normal measuring conditions (see environmental specifications, § 6.2.1).

OPERATING INSTRUCTIONS

INSTRUMENT DESCRIPTION

Command description

HT-ITALIA-HT9014-AC-true-RMS-Clamp-Meter-01Fig. 1: Instrument description

LEGEND:

  1. Inductive clamp jaw
  2. LED for AC voltage detection
  3. Jaw trigger
  4. Function selector
  5. HOLD / key
  6. PK/REL key
  7. LCD display
  8.  MODE key
  9. Hz% key
  10. MAX/MIN key
  11. V VΩ CA PHz % Temp input jack
  12. COM input jack
  13.  Battery cover

Alignment marks
Put the conductor within the jaws on intersection of the indicated marks as much as possible (see Fig. 2) in order to meet the meter accuracy specifications.

LEGEND

  1. Alignment marks
  2. Conductor

Fig. 2: Alignment marks

FUNCTION KEY DESCRIPTION

H key
By pushing “H” key the parameter measured value is frozen on the display and the symbol ”HOLD” appears on it. This mode is disabled by pushing “H” key or moving the rotary switch.

key
Keep pressed the “H” key for the backlight activation. This light could help the operator reading the display while he’s measuring in dark sites. For sake of battery saving after 10 seconds the light is automatically switched off.

MODE key
By pushing MODE key the selection of double measured functions which are present at display is possible. In particular this key is active in Ω CAP position to select among resistance test, diode test, continuity test and capacitance test, and in Temp position to select between °C and °F in temperature test and in 60A , 600A position (only HT9015) to select between AC or DC current measurements.

MAX/MIN key
By pushing MAX/MIN key the parameter maximum and minimum values are measured. Both values are stored in memory and they can be shown cyclically when pressing “MIN MAX” key. The display shows the symbol linked to the selected function: “MAX” for maximum value, the symbol “MIN” for minimum value. This mode is activated on each measurement except for continuity test, diode test capacitance test, frequency test and duty cycle test. The barograph disappear in the functions where the MAX/MIN feature is active.
This mode is disabled keeping pressed MAX/MIN key or moving the rotary switch.

Hz% key
With rotary switch on , 60A∼, 600A∼ (only HT9014) and 60A , 600A (only HT9015) positions, the Hz% key permits to select the frequency test (Hz) or the duty cycle test (%)

PK/REL key
With rotary switch on , , CAP and DC current (only HT9015) the PK/REL key permits to perform the zeroing of value at display and a relative measurement of the parameter under test.
By pushing PK/REL key the instrument takes the actual parameter’s value as on offset and performs a relative measurement. The relative value is displayed, it’s calculated as actual value – offset value, the “REL” symbol is shown at display and the barograph is disabled.
This mode is disabled for resistance test, continuity test, AC current test, temperature test, frequency test, duty cycle test and diode test and with MAX/MIN mode activated. This mode is disabled pressing PK/REL key or moving the rotary switch
With rotary switch on 60A∼, 600A∼ (only HT9014) and 60A , 600A positions (only HT9015) the PK/REL key permits to activated the maximum peak measurement of AC current (calculated with response time <10ms). The “P” symbol is shown at display and the value is continuously updated by the meter.
This mode is disabled pressing PK/REL key or moving the rotary switch

FUNCTIONS OF ROTARY SWITCH DESCRIPTION

AC Voltage measurement

CAUTION Maximum input for AC Voltage measurements is 1000Vrms. Do not take any voltage measurement exceeding this limit in order not to risk electrical shock or damaging the tester

HT-ITALIA-HT9014-AC-true-RMS-Clamp-Meter-03 Fig. 3: Taking AC voltage measurements

  1. Approach the meter closest to AC source and note the turn on of red LED which is placed to the bottom of clamp jaws (see Fig. 1 – part 2) which detect the AC voltage
    Rotate the switch on position

  2. Insert the red test lead plug into VΩ CA PHz % Temp jack and the black test lead plug into COM jack (Fig. 3)

  3. Connect the two long ends of test leads to the desired circuit, then reading will be displayed

  4. When “O.L.” symbol is displayed, the result exceeds the maximum value that the instrument is capable of measuring

  5. For HOLD, MAX/MIN and REL features please refer to § 4.2

CAUTION

  • Due to the high input impedance is possible that the meter performs the zero of the display taking longer
  • The possible blinking values shown at display with open input terminals are not a problem of the meter and these values will not be added to the measured value.

DC Voltage measurement
CAUTION
Maximum input for DC Voltage measurements is 1000V. Do not take any voltage measurement exceeding this limit in order not to risk electrical shock or damaging the teste.

HT-ITALIA-HT9014-AC-true-RMS-Clamp-Meter-4 Fig. 4: Taking DC voltage measurements

  1. Rotate the switch on position
  2. Insert the red test lead plug into VΩ CA PHz % Temp jack and the black test lead plug into COM jack (Fig. 4)
  3. Connect the two long ends of test leads to the desired circuit, then reading will be displayed
  4. When “O.L.” symbol is displayed, the result is over the maximum value that the instrument is able to measure
  5. For HOLD, MAX/MIN and REL features please refer to § 4.2

CAUTION

  1. Due to the high input impedance is possible that the meter performs the zero of the display taking longer.
  2. The possible blinking values shown at display with open input terminals are not a problem of the meter and these values will not be added to the measured value.

Resistance measurement
CAUTION: Before taking any in circuit resistance measurement, remove power from the circuit to be tested and discharge all the capacitors.

Fig. 5: Taking Resistance measurement

  1. Rotate the switch on Ω CAP position
  2.  Insert the red test lead plug into VΩ CA PHz % Temp jack and the black test lead plug into COM jack (Fig. 5)
  3. Connect the two long ends of test leads to the desired circuit, then reading will be displayed
  4. When “O.L.” symbol is displayed, the result under test exceeds the maximum value that the instrument is capable of measuring
  5. For HOLD and MAX/MIN features please refer to § 4.2

Continuity test and Diode test
CAUTION: Before taking any in circuit resistance measurement, remove power from the circuit to be tested and discharge all the capacitors.

HT-ITALIA-HT9014-AC-true-RMS-Clamp-Meter-6 Fig. 6: Taking Continuity test and Diode test

  1. Rotate the switch on Ω CAP position.
  2. Pushing MODE key and select continuity test. The symbol is shown at display.
  3. Insert the red test lead plug into VΩ CA PHz % Temp jack and the black test lead plug into COM jack and perform continuity test on the object under test (see Fig. 6 – left side). Buzzer emits sound if the measured resistance value is less about 60Ω.
  4. Pushing MODE key and select diode test. The “ ” symbol is shown at display.
  5. Connect the red test leads to the anode of diode on test and the black test lead on the cathode ones (see Fig. 6 – right side). Reverse position of test leads to reading reverse polarization voltage
  6. Displayed values within 0.4V and 0.7V (direct junction) and “OL” (reverse junction) are correspondent to a correct result. A “0mV” value means a shorted device while a “OL” indication in both side means a broken device. paragraph is disabled in diode test.

Capacitance measurement
CAUTION: When testing in-circuit capacitors, turn off the power of the circuit to be tested and discharge all the capacitors.

Fig. 7: Taking Capacitance measurement

  1. Rotate the switch on Ω CAP position
  2. Pushing MODE key and select capacitance test. The “nF” symbol is shown at display
  3. Insert the red test lead plug into VΩ CA PHz %Temp jack and the black test lead plug into COM jack (Fig. 7)
  4. Connect the two long ends of test leads to the desired circuit, then reading will be displayed. Bargraph is disabled in capacitance measurement
  5. When “O.L.” symbol is displayed, the result under test exceeds the maximum value that the instrument is capable of measuring
  6. For HOLD and REL features please refer to § 4.2

CAUTION: For readings of capacitance <40nF push PEAK/REL key before making test in order to improve the measured result.

Temperature measurement
CAUTION: Do not allow the temperature sensor to contact a surface that is energized above 30 V RMS or 60 V DC, such voltages pose a shock hazard.

HT-ITALIA-HT9014-AC-true-RMS-Clamp-Meter-08 Fig. 8: Taking Temperature measurement

  1. Rotate the switch on Temp position
  2. Pushing MODE key and select the kind of measure. “°C” or “°K” symbols are shown at display respectively for Celsius or Fahrenheit temperature measurements
  3. Insert the type K bead probe in VΩ CA PHz %Temp and COM inputs terminals using the standard adapter and observing the correct polarity (see Fig. 8). The temperature value is shown at display
  4. For HOLD and MAX/MIN features please refer to § 4.2

DC Current measurement (only HT9015)
CAUTION: Make sure that all the test leads are disconnected from the meter terminals for current measurement.

HT-ITALIA-HT9014-AC-true-RMS-Clamp-Meter-09 Fig. 9: Taking DC current measurements

  1. Rotate the switch on 60A or 600A position

  2. Pushing MODE key and select the kind of measure “DC”

  3. Pushing PK/REL key to perform the zeroing of residual magnetization current
    CAUTION: The zeroing operation of residual magnetization current is essential to obtain accurate results

  4. Put the conductor to be tested inside to the center of clamp jaw to perform accurate measurements. Consider the notches on jaws as reference (see Fig. 2). The reading will be displayed

  5. On DC current measurement the “-“ symbol at display means that the meter is connected in the opposite side compared with current direction (see Fig. 9)

  6. When “O.L.” symbol is displayed, the result exceeds the maximum value that the instrument is capable of measuring. Select an higher range

  7. For HOLD, MAX/MIN and REL features please refer to § 4.

AC Current measurement
CAUTION: Make sure that all the test leads are disconnected from the meter terminals for current measurement.

HT-ITALIA-HT9014-AC-true-RMS-Clamp-Meter-10 Fig. 9: Taking AC current measurements

  1. Rotate the switch on 60A∼ or 600A∼ (only HT9014) or 60A or 600A position (only HT9015)

  2. Pushing MODE key and select the kind of measure “AC” (only HT9015)

  3. Put the conductor to be tested inside to the center of clamp jaw to perform accurate measurements. Consider the marks on jaws as reference (see Fig. 2). The reading will be displayed

  4. When “O.L.” symbol is displayed, the result exceeds the maximum value that the
    instrument is capable of measuring. Select an higher range

  5. For HOLD, MAX/MIN and PEAK features please refer to § 4.2

CAUTION: The possible blinking values shown at display with open input terminals are not a problem of the meter and these values will not be added to the measured value

Frequency and Duty cycle measurement
CAUTION

  • On frequency test with test leads the maximum input for AC Voltage measurements is 1000Vrms. Do not take any voltage measurement exceeding this limit in order not to risk electrical shock or damage the tester
  • On frequency test with jaws make sure that all the test leads are disconnected from the meter terminals for current measurement.

HT-ITALIA-HT9014-AC-true-RMS-Clamp-Meter-11 Fig. 10: Taking frequency and duty cycle measurements

  1. Rotate the switch on position for frequency measurement with test leads or 60A or 600A positions (only HT9015) or 60A∼ or 600A∼ positions (only HT9014) for
    frequency measurements with jaws

  2. Pushing Hz% key the “Hz” symbol for frequency measurement or the “%” symbol for duty cycle measurement are shown at display

  3. Insert the red test lead plug into VΩ CA PHz %Temp jack and the black test lead plug into COM jack (see Fig. 10 – left side) for frequency measurement with test leads. Open the clamp and put the conductor to be tested in the middle of the clamp jaw (see Fig. 10 – right side) for frequency measurements with jaws, then reading will be displayed. Bargraph is disabled in frequency and duty cycle test

  4. When “O.L.” symbol is displayed, the result exceeds the maximum value that the instrument is capable of measuring

  5. For HOLD feature please refer to § 4.2

MAINTENANCE

GENERAL INFORMATIONS
  1. This digital clamp meter is a precision instrument. Whether in use or in storage, please do not exceed the specification requirements to avoid possible damages or dangers.
  2. Do not place this meter at high temperatures or humidity or expose it to direct sunlight.
  3. Be sure to turn off the meter after use. If you expect not to use the tester for a long time, remove the battery in order to avoid leakages of battery liquid that would damage the internal parts.
BATTERY REPLACEMENT

When “ ” appears on the display, replace the battery.

CAUTION
Only expert and trained technicians must perform this operation. Remove the test leads or the conductor under test before replacing the battery

  1. Rotate the switch on OFF.
  2. Remove the test leads or the objects to be tested.
  3. Remove the screw from the battery cover, and detach the battery cover from the bottom cover.
  4. Remove the battery
  5. Replace the battery with a new one same type (see § 6.1.2)
  6. Replace the battery cover and screw
  7. Use the appropriate battery disposal methods for Your area
CLEANING

For cleaning the instrument use a soft dry cloth. Never use a wet cloth, solvents or water, etc.

END OF LIFE

CAUTION: this symbol indicates that equipment its accessories and battery shall be subject to a separate collection and correct disposal.

TECHNICAL SPECIFICATIONS

CHARACTERISTICS

Accuracy is calculated as [% rdg + (number of dgt) x resolution]. It is referred to the following reference conditions: 18°C ÷ 28°C (65°F ÷ 83°F) with RH < 75%
DC Voltage (Autorange)

Range| Resolution| Accuracy| Input impedance| Overload protection
---|---|---|---|---
600.0mV| 0.01mV| ±(1.0%rdg + 3dgt)| 10MW| 1000VDC/ACrms
6.000V| 0.001V
60.00V| 0.01V
600.0V| 0.1V
1000V| 1V

The meter emits a continuous sound with VDC >1000V

AC Voltage TRMS (Autorange) – (instrument HT9014)

Range| Resolution| Accuracy| Input impedance| Bandwidh| Overload protection
---|---|---|---|---|---
6.000V| 0.001V| ±(1.0%rdg+10dgt) (50 ¸ 60Hz)

±(3.5%rdg+5dgt) (61 ¸ 400Hz)

| 10MW| 50-400Hz| 1000VDC/ACrms
60.00V| 0.01V
600.0V| 0.1V
1000V| 1V

Integrated sensor for AC voltage detection: LED turn on for phase-earth voltage > 100V, 50/60Hz
The meter emits a continuous sound on 1000V range for VAC>750V

AC Voltage TRMS (Autorange) – (instrument HT9015)

Range| Resolution| Accuracy| Input impedance| Bandwidh| Overload protection
---|---|---|---|---|---
6.000V| 0.001V| ±(1.0%rdg+10dgt) (50 ¸ 60Hz)

±(3.5%rdg+5dgt) (61 ¸ 400Hz)

| 10MW| **** 50-400Hz| 1000VDC/ACrms
60.00V| 0.01V| ±(1.0%rdg.+4dgt) (50 ¸ 60Hz)

±(3.5%rdg+5dgt) (61 ¸ 400Hz)

600.0V| 0.1V
1000V| 1V

Integrated sensor for AC voltage detection: LED turn on for phase-earth voltage > 100V, 50/60Hz
The meter emits a continuous sound on 1000V range for VAC>750V

Resistance and Continuity test (Autorange)

Range| Resolution| Accuracy| Buzzer| Overload protection
---|---|---|---|---
600.0W| 0.1W| ±(1.0%rdg+5dgt)| ≤ 60W| 600VDC/ACrms
6.000kW| 0.001kW
60.00kW| 0.01kW
600.0kW| 0.1kW
6.000MW| 0.001MW
60.00MW| 0.01MW| ±(2.0%rdg.+10dgt)

Test current of continuity test: < 0.35mA

DC Current (only HT9015)

Range Resolution *Accuracy ()** Overload protection
60.00A 0.01A ±(2.2%rdg.+10dgt) 600AACrms
600.0A 0.1A ±(2.0%rdg+8dgt)

(*) Referred to cable inside to the center of clamp jaws
Position sensitivity: ±2.0%rdg

AC TRMS Current (instrument HT9014)

Range| Resolution| *Accuracy ()| Bandwidh| Overload protection**
---|---|---|---|---
60.00A| 0.01A| ±(2.8%rdg+12dgt)| 50 ¸ 60Hz| 600AACrms
600.0A| 0.1A| ±(2.8%rdg+8dgt)
60.00A| 0.01A| ±(4.5% rdg+10dgt)| 61 ¸ 400Hz
600.0A| 0.1A| ±(5.0%rdg+10dgt)

PEAK function: response time <10ms ; (*) Referred to cable inside to the center of clamp jaws
Position sensitivity: ±2.0%rdg

AC TRMS Current (instrument HT9015)

Range| Resolution| *Accuracy ()| Bandwidh| Overload protection**
---|---|---|---|---
60.00A| 0.01A| ±(2.2%rdg+12dgt)| 50 ¸ 60Hz| 600AACrms
600.0A| 0.1A| ±(2.2%rdg+8dgt)
60.00A| 0.01A| ±(3.5% rdg+12dgt)| 61 ¸ 400Hz
600.0A| 0.1A| ±(3.5%rdg+8dgt)

PEAK function: response time <10ms ; (*) Referred to cable inside to the center of clamp jaws
Position sensitivity: ±2.0%rdg

Capacitance (Auto range)

Range Resolution Accuracy Overload protection
40.00nF 0.01nF ±(3.5%rdg+40dgt) 600VDC/ACrms
400.0nF 0.1nF ±(2.5%rdg+5dgt)
4.000mF 0.001mF
40.00mF 0.01mF
400.0mF 0.1mF
4000mF 1mF ±(5.0%rdg+5dgt)

Diode test

Range Test current Open voltage
**** 0.9mA typical 2.8VDC

Frequency with test leads (Auto range)

Range| Resolution| Accuracy| Sensitivity| Overload protection
---|---|---|---|---
99.99Hz| 0.01Hz| ****

±(1.0%rdg+5dgt)

| ****

³15Vrms

| ****

600VDC/ACrms

999.9Hz| 0.1Hz
9.999kHz| 0.001kHz
60.00kHz| 0.01kHz

Frequency with jaws (Auto range)

Range| Resolution| Accuracy| Sensitivity| Overload protection
---|---|---|---|---
99.99Hz| 0.01Hz| ±(1.0%rdg+5dgt)| ³10A (60A)

³50A(600A)

| 600AACrms
999.9Hz| 0.1Hz
9.999kHz| 0.001kHz

Duty Cycle (Auto range)

Range Resolution Sensitivity Accuracy
0.5% ¸ 99.0% 0.1% ³15Vrms ±(1.2%rdg+2dgt)

Temperature with type K probe (Auto range)

Range Resolution *Accuracy ()** Overload protection
-20.0 ¸ 400.0°C 0.1°C ±(2.0%rdg +3°C) 600VDC/ACrms
400 ¸ 760°C 1°C ±(2.0%rdg +5°C)
-4.0 ¸ 752.0°F 0.1°F ±(2.0%rdg +6°F)
752 ¸ 1400°F 1°F ±(2.0%rdg +9°F)

(*) Accuracy of type K probe not considered

Safety

  • Comply with: IEC/EN 61010-1
  • Insulation: double insulation
  • Pollution degree: 2
  • Max height of use: 2000m (6562 ft)
  • Installation category: CAT IV 600V, CAT III 1000V to ground

General data

  • Mechanical characteristics
  • Dimensions (L x W x H): 215 x 74 x 43mm ; 8 x 3 x 2 in
  • Weight (including battery): 285g (10 ounces)
  • Max conductor size: 30mm (1in)

Supply

  • Battery type: 1 battery 9V NEDA 1604 IEC 6F22 JIS 006P.
  • Low battery indication: “ ” is displayed when the battery level is low
  • Battery life: about 200 hours
  • Auto Power Off: after 15 minutes of idleness (not disabled)

Disp l ay

  • Characteristics: 4 LCD (max 6000 counts), decimal point, unit symbol indication, bargraph and backlight
  • Sample rate: 2 times/sec
  • Conversion mode: TRMS
ENVIRONMENTAL CONDITIONS

Climatic conditions

  • Reference temperature: 18°C ÷ 28°C (65°F ÷ 83°F)
  • Operating temperature: 5 ÷ 40 °C (41°F ÷ 104°F)
  • Operating humidity: <80%RH
  • Storage temperature: -20 ÷ 60 °C (-4°F ÷ 140°F)
  • Storage humidity: <80% RH

This product conforms to the prescriptions of the European directive on low voltage 2006/95/EEC and to EMC directive 2004/108/EEC.

STANDARD ACCESSORIES
  • Test leads – Code 4413-2
  • Adapter + type K bead probe
  • Carrying bag
  • Battery
  • User manual
OPTIONAL ACCESSORIES

The herewith type K probes are available:

Model| Description| Temperature range| Accuracy (at 100°C)| Probe length (mm)| Probe diameter (mm)
---|---|---|---|---|---
TK107| Air and gas temperature| -40 ¸ 800 °C| ± 2.2rdg| 200| 1.5
TK108| Inner temperature of semisolid substances| -40 ¸ 800 °C| ± 2.2rdg| 200| 3
TK109| Inner temperature of fluids| -40 ¸ 800 °C| ± 2.2rdg| 200| 4
TK110| Surface temperature| -40 ¸ 400 °C| ± 2.2rdg| 200| 5
TK111| Surface temperature with fixed tip at 90°| -40 ¸ 400 °C| ± 2.2rdg| 260| 5

SERVICE

WARRANTY CONDITIONS

This equipment is guaranteed against material faults or production defects, in accordance with the general sales conditions. During the warranty period (one year), faulty parts may be replaced. The manufacturer reserves the right to decide either to repair or replace the product.

In case of returning of the instrument, all transport charges must be paid by the customer. The instrument must be accompanied by a delivery note indicating the faults or reasons of returning. The returned tester must be packed in its original box. Any damage occurred in transit because of lack of original packaging will be debited to the customer.
The manufacturer is not responsible for any damage against persons or things. Accessories and batteries are not covered by warranty.

The warranty won’t be applied to the following cases:

  • Faults due to improper use of the equipment
  • Faults due to combination of the tester with incompatible equipment.
  • Faults due to improper packaging.
  • Faults due to servicing carried out by a person not authorized by the company.
  • Faults due to modifications made without explicit authorization of our technical department.
  • Faults due to adaptation to a particular application not provided for by the definition of the equipment or by the instruction manual.

The contents of this manual cannot be reproduced in any form without our authorization.

Our products are patented. Our logotypes are registered. We reserve the right to modify characteristics and prices further to technological developments.

SERVICE

If the equipment doesn’t work properly, before contacting the SERVICE, test the batteries, the test leads, etc., and change them if necessary.
If the equipment still doesn’t work, make sure that your operating procedure complies with the one described in this manual.
In case of returning of the instrument, all transport charges must be paid by the customer. The instrument must be accompanied by a delivery note indicating the faults or reasons of returning. The returned tester must be packed in its original box. Any damage occurred in transit because of lack of original packaging will be debited to the customer.

Read User Manual Online (PDF format)

Read User Manual Online (PDF format)  >>

Download This Manual (PDF format)

Download this manual  >>

Related Manuals