HT ITALIA HTA103 Thermometer for Measuring Temperature User Manual
- June 4, 2024
- HT ITALIA
Table of Contents
HT ITALIA HTA103 Thermometer for Measuring Temperature
SAFETY PRECAUTIONS AND PROCEDURES
This instrument conforms to safety guidelines relating to electronic measuring instruments. For your own safety and that of the apparatus, you must follow the procedures described in this instruction manual and specially read all the notes preceded by the symbol carefully. Strictly keep to the following instructions before and during measurements:
- Do not perform measurements in environments with explosive gas, fuels or dust.
- Do not perform any measurement in case of unusual conditions of the instrument such as deformation, breakage, leakage of substances, absence of display reading etc.
The following symbols are used on the instrument and in this manual:
Caution:
refer to the instructions reported on this manual; improper use may damage the
apparatus or its components.
Ground reference
PRELIMINARY INSTRUCTIONS
- Use the instrument only as specified in this manual otherwise, the protection provided by the instrument may be impaired.
- Only the accessories supplied with the instrument guarantee compliance with the safety standards. They must be in good conditions and must be replaced, if necessary, with identical models.
- Do not effect measurements beyond the limits specified in this manual.
- Check that batteries have been placed correctly.
DURING USE
Carefully read the following recommendations and instructions:
CAUTION
No compliance with the Warnings and/or Instructions may damage the apparatus
and/or its components or injure the operator.
- Use the instrument and the probes only within the ranges specified in this manual and in the probes’ ones.
- Do not perform any measurement on materials under voltage. This could damage the instrument.
- If during a measurement the read value or sign never changes, please verify if the HOLD function is active.
AFTER USE
- When the measurements are completed switch off the instrument.
- Remove batteries when the apparatus remains unused for long periods.
GENERAL DESCRIPTION
The instrument can performs the following measurements:
- Temperature measurement in °C/°F/°K by means of K, J and T type probes
- Maximum, minimum values measurements
- Data HOLD
- Offset adjustment for compensation of external probes
- Backlight
- Auto Power OFF
Each function could be selected via the relative key. The selected quantity appears on a high-contrast display with indication of measurement unit and of the active functions. There are also the function keys, for their use please see § 4.2.
PREPARATION FOR USE
INITIAL
The instrument has been checked from every point of view before shipment. Every care has been taken to make sure that the instrument reaches you in perfect conditions. However, it’s advisable to make a rapid check in order to detect eventual damages, which may have occurred in transit. Should this be the case, enter immediately the usual claims with the carrier. Make sure that all the accessories listed in § 7.3 are contained in the package. In case of discrepancies contact the dealer. In case of returning of the tester, please keep to the instructions given in § 8.
POWER SUPPLY
The instrument is supplied by 1x9V alkaline battery type IEC 6F22 included in the package. Battery life: about 200 hours. When battery is low, the symbol ” ” appears on the display. Replace it immediately, following the instructions given in § 6.2.
STORAGE
In order to guarantee the accuracy of the measurements, after a period of storage in extreme environmental condition, wait for the necessary time so that the instrument returns to normal measuring conditions (see § 7.2.1).
NOMENCLATURE
INSTRUMENT DESCRIPTION
- Instrument description
CAPTION:
- Input terminal for JKT probes
- LCD display
- Arrow key
- Arrow key
- . HOLD/ENTER key
- MIN key
- MAX key
- UNIT/ key
- CAL key
- JKT key
- . ON/OFF key
- Tripod hole
- Battery cover
DESCRIPTION OF DISPLAY
- Display description
CAPTION:
-
Probe type indication
-
Auto Power OFF
(APO) symbol -
Low battery indication
-
Main display
-
Measurement unit
-
MAX function active
-
MIN function active
-
HOLD function active
-
Secondary display
FUNCTION KEY DESCRIPTION
ON/OFF key
By pushing ON/OFF key permits to turn on and off the instrument. After turn
on, the backlight is active and it automatically disabled after approx. 20s.
UNIT/ key
By pushing UNIT/ key permits the temperature measurement unit selection among
the option: “°C” (Celsius), “°F” (Fahrenheit) and “K” (Kelvin). The UNIT/ key
is disabled if the function “HOLD” is activated..
The formulas that links Celsius and Fahrenheit degrees are:
T [°F] = T [°C] * 1.8 + 32
so: 0°C are equal to 32°F, 100°C are equal to 212°F
By pushing long (>2s) of UNIT/ key permits to activate/disable the backlight.
JKT key
By pushing JKT key permits to set the type of external probe (K, J or T) to be
connected to the instrument. Cyclically the K, J or T settings are displayed.
MAX and MIN key
By pushing MAX or MIN key permits to detect the Maximum or Minimum values of
measured temperature. The secondary display shows the symbol linked to the
selected function: The symbol “MAX” for maximum value and “MIN” for minimum
value. The MAX or MIN values are dynamically updated while the real time value
of temperature is displayed on the main display
HOLD/ENTER key
By pushing HOLD/ENTER key the instrument’s measured value is frozen on the
display and the symbol “HOLD” appears on it. The HOLD function does not
influence the secondary display.
keys
Use the and arrow keys to set the offset values during the calibration
operation of the probe fitted in the terminal input of the instrument (see §
4.3.7).
CAL key
The CAL key permits to perform compensation on the temperature measurement due
to a possible error of the input probe. Follow the below steps:
-
By pushing long (>3s) CAL key to enter in the calibration section. The below screen is shown at display:
- Compensation of temperature probe
-
On the main display the real time value of the temperature is displayed while the measurement offset value is displayed on the secondary display. The default value is “0.0”. Do not modify this default value if any compensation is needed
-
Push the or arrow keys respectively to increase or decrease the offset value. The maximum set values are: 5°C or 9°F with a 0.1°C / 0.1°F resolution
-
. By pushing long (>3s) again CAL key to save the settings and quit this mode. The instrument keeps the set value after each shutdown
Disabling Auto Power OFF
The instrument is provided with Auto Power OFF (APO) feature which
automatically permits to switch it off after 15 minutes of idleness in order
to preserve the internal battery. To disable this feature follows the below
steps:
-
With instrument on, push and keep the HOLD/ENTER key
-
By pushing long (>1s) the ON/OFF key. The “APO OFF” message is shown at display for a while
- Disable/enable Auto Power OFF
and the “ “ symbol disappears from display
- Disable/enable Auto Power OFF
-
Repeat the same operation of the point 1 and 2 for enable again the function with “APO ON” message shown at display
- Disable/enable Auto Power OFF
or the feature is automatically restored after each shutdown
OPERATING INSTRUCTIONS
TEMPERATURE MEASUREMENT
CAUTION
Maximum input voltage is 24V AC or 60V DC. Do not attempt to apply any voltage
that exceeds the limits indicates in this manual to avoid electrical shock
hazard or damage the instrument.
-
Turn on the instrument by pushing the ON/OFF key. If no probe is connected to the instrument, the “- – – -“ symbol is shown at two displays
-
Select the type of probe (see § 4.3.3) and the measurement unit (see § 4.3.2)
-
Insert the connectors of the probe on input terminal respecting the polarity (+) and (-)(see Fig. 1 – part 1) shown on the instrument and on the probe connector
-
The real time value of the temperature is shown on the main display
- Use of the instrument for temperature measurement
-
Press HOLD/ENTER key to freeze result at display (see § 4.3.5)
-
Press the MAX or MIN keys to read the Maximum or Minimum values on secondary display
-
Switch off the instrument after measurement
MAINTENANCE
GENERAL
- Whether in use or in storage, please do not exceed the specification requirements to avoid any possible damage or danger during use
- Do not place this instrument in high temperature or expose to direct sunlight
- Be sure to turn the instrument off after use. For long time storage, remove the battery to avoid leakage of battery liquid that would damage the interior parts
BATTERY REPLACEMENT
When the symbol ” ” is displayed replace battery.
CAUTION
Only expert and trained technicians must perform this operation. Remove all
the probes from the devices under test before replacing the battery.
- Switch off the instrument
- Remove the temperature probe
- Remove the battery cover (see Fig. 1 – part 13)
- Remove the battery from the battery fastener
- Set the new battery into battery fastener, and return it to the battery case
- Replace the battery cover
- Use the appropriate battery disposal methods for Your area
CLEANING
For cleaning the instrument use a soft dry cloth. Never use a wet cloth, solvents or water, etc.
END OF LIFE
CAUTION:
this symbol indicates that equipment, its accessories and battery shall be
subject to a separate collection and proper disposal.
TECHNICAL SPECIFICATIONS
CHARACTERISTICS
Accuracy is calculated as [%rdg + degrees] at 23°C ± 5°C <70%HR
Temperature measurement with K probe
Range | Resolution | Accuracy | Overload protection |
---|---|---|---|
-250°C ÷ 1372°C | 0.1 °C |
±(1%rdg+ 1°C) (T<-99.9°C)
±(1%rdg+0.5°C) (T≥ -99.9°C)
|
60V DC
24V AC rms
-418°F ÷ 2502°F| 0.1 °F|
±(1%rdg+ 1.8°F) (T<-148°F)
±(1%rdg+0.9°F) (T≥ -148°F)
Temperature measurement with J probe
Range | Resolution | Accuracy | Overload protection |
---|---|---|---|
-200°C ÷ 1000°C | 0.1 °C |
±(1%rdg+ 1°C) (T<-99.9°C)
±(1%rdg+0.5°C) (T≥ -99.9°C)
|
60V DC
24V AC rms
-328°F ÷ 1832°F| 0.1 °F|
±(1%rdg+ 1.8°F) (T<-148°F)
±(1%rdg+0.9°F) (T≥ -148°F)
Temperature measurement with T probe
Range | Resolution | Accuracy | Overload protection |
---|---|---|---|
-250°C ÷ 400°C | 0.1 °C |
±(1%rdg+ 1°C) (T<-99.9°C)
±(1%rdg+0.5°C) (T≥ -99.9°C)
|
60V DC
24V AC rms
-418°F ÷ 752°F| 0.1 °F|
±(1%rdg+ 1.8°F) (T<-148°F)
±(1%rdg+0.9°F) (T≥ -148°F)
General data
Mechanical characteristics
Dimensions (Lx W x H): 190 x 65 x 45mm; (7 x 3 x 2in)
Weight (including battery): 235g (8 ounces)
Mechanical protection: IP40
Supply
Battery type: 1x9V alkaline NEDA 1604 IEC 6F22 JIS 006P
Low battery indication: “ ” symbol is displayed
Battery life: ca 50h (backlight ON), ca 210h
(backlight OFF)
Auto Power Off: after 15 minutes of idleness (disabled)
Display
Characteristics: double 4 LCD, sign, decimal point and
backlight
Sample rate: 3times/s
ENVIRONMENTAL CONDITIONS
Climatic conditions
Reference temperature: 25°C ; (77°F)
Operating temperature: 0°C ÷ 50 °C ; (32°F ÷ 122°F)
Operating humidity: <70%RH
Storage temperature: -10°C ÷ 60°C ; (14°F ÷ 140°F)
Storage humidity: <70%RH
Max height of use: 2000m (6562ft)
ACCESSORIES
Accessories provided
- K-type wire probe (Code: TK101)
- Battery
- Carrying bag
- User manual
Optional accessories
- Type K probe for air and gas Cod. TK107
- Type K probe for semi solid substances Cod. TK108
- Type K probe for liquid Cod. TK109
- Type K probe for surfaces Cod. TK110
- Type K probe for surfaces with a 90° lead Cod. TK111
SERVICE
WARRANTY CONDITIONS
This equipment is guaranteed against material faults or production defects, in
accordance with the general sales conditions. During the warranty period (one
year), faulty parts may be replaced. The manufacturer reserves the right to
decide either to repair or replace the product. In case of returning of the
instrument, all transport charges must be paid by the customer. The instrument
must be accompanied by a delivery note indicating the faults or reasons of
returning. The returned tester must be packed in its original box. Any damage
occurred in transit because of lack of original packaging will be debited to
the customer. The manufacturer is not responsible for any damage against
persons or things. Accessories and batteries are not covered by warranty.
The warranty won’t be applied to the following cases:
- Faults due to improper use of the equipment.
- Faults due to combination of the tester with incompatible equipment.
- Faults due to improper packaging.
- Faults due to servicing carried out by a person not approved by the company.
- Faults due to modifications made without explicit authorization of our technical department.
- Faults due to adaptation to a particular application not provided for by the definition of the equipment or by the instruction manual.
The contents of this manual cannot be reproduced in any form without our
authorization.
Our products are patented. Our logotypes are registered. We reserve the right
to modify characteristics and prices further to technological developments.
SERVICE
If the equipment doesn’t work properly, before contacting the SERVICE, test the batteries, the probes and change them if necessary. If the equipment still doesn’t work, make sure that your operating procedure complies with the one described in this manual. In case of returning of the instrument, all transport charges must be paid by the customer. The instrument must be accompanied by a delivery note indicating the faults or reasons of returning. The returned tester must be packed in its original box. Any damage occurred in transit because of lack of original packaging will be debited to the customer.