Dental Direkt Ti-Base 2CUT noLock Cooking Hob
Product Information
The DD Ti-Base 2CUT noLock is a titanium bonding base that can be shortened and is non-engaging. It is made from Ti6Al4V (medical grade 5) ASTM 136 ELI. The package includes 1 x DD Lab Screw (blue) and 1 x DD Implant Screw (silver).
Purpose:
The DD Ti-Base 2CUT noLock is designed for use in dental implant procedures.
Indication:
This product is indicated for use in implant-supported prosthetic
restorations.
Contraindication:
Please refer to the contraindications provided by the original implant
manufacturer.
Intended Users:
This product is intended for use by dental professionals.
Intended Patient Group:
This product is intended for use with patients requiring implant-supported
prosthetic restorations.
Instructions for Laboratory Use:
The Lab Screw should not be used for the final restoration of the implant
components in the patient.
Instructions for Practice Use:
The Implant Screw should be tightened with the specific torque recommended
by the system (refer to the table provided). It is recommended to tighten the
Implant Screw 10 minutes after placement.
Cleaning, Disinfection, and Sterilization:
Cleaning/Disinfection: Follow the manual cleaning procedure provided or
use an appropriate cleaning solution.
Sterilization: After cleaning, sterilize the product and store it in a
sterile packaging. The recommended drying time is 20 minutes.
Product Usage Instructions
- Prepare the DD Ti-Base 2CUT noLock for use in the dental implant procedure.
- Ensure that the Lab Screw is not used for the final restoration of the implant components in the patient.
- Tighten the Implant Screw using the specific torque recommended by the system. Refer to the provided table for the recommended torque values.
- It is recommended to tighten the Implant Screw 10 minutes after placement.
- Clean and disinfect the DD Ti-Base 2CUT noLock using the provided manual cleaning procedure or an appropriate cleaning solution.
- Sterilize the product and store it in a sterile packaging. The recommended drying time is 20 minutes.
Contents
- 1 x DD Ti-Base 2CUT noLock (titanium bonding base, can be shortened, non-engaging)
- 1 x DD Lab Screw, blue
- 1 x DD Implant Screw, silver
Product description
The Dental Direkt Ti-Base 2CUT noLock is used as a titanium bonding base for
customized bridge and bar restorations. The titanium bonding bases are
available for various implant systems and consist of a prefab-ricated implant
interface, a screw channel, a circumferential platform for shaping the
emergence profile and a cylinder that can be shortened for customized patient
care. Dental Direkt titanium bonding bases are made of Ti6AI4V (medical grade
5) for single use and are supplied non-sterile, each with a separately
packaged Implant and Lab Screw.
Caution: Dental Direkt titanium bonding bases must only be used in combination
with the DD Implant Library.
Intended purpose:
DD Solid Connect® implant components are used to support prosthetic
restorations in combination with endosseous dental implants in the upper
and/or lower jaw.
Indication:
For attachment of customized bridge and bar restorations.
Contraindication:
- Intolerance to the constituents
- Restorations with an angulation correction of more than 25° to the implant axis
Please note: Observe the contraindications specified by the original implant manufacturer used in each case.
Intended users
DD Solid Connect® implant components should only be used by dentists and dental technicians in compliance with the procedures in the instruc-tions for use.
Intended patient group:
Fixed restorations using Dental Direkt Solid Connect® implant compo-nents are
suitable for the permanent dentition in adult patients of any gender and
nationality.
Instructions for use in the laboratory
In the laboratory, the technician selects the DD Ti-Base 2CUT noLock that
corresponds to the implant to be placed (see table). DD Ti-Base 2CUT no-Lock
is supplied with a chimney height of 6 mm. The chimney height can be shortened
at predefined grooves to a height of 5 mm or 4 mm by using a 0.3 mm thick
fabric-reinforced cutting disc. The geometry of the tita-nium bonding base
must not be altered in any way except for shortening the chimney height and
removing the burr from the edges. Always use an extraction system when cutting
and sandblasting. Metal dusts are harmful to health, so always wear
respiratory protection and safety goggles. During subsequent fabrication of
prosthetic restorations, the minimum wall thickness specified by the material
manufacturer must be observed. The individual restoration is constructed using
dental design software. It is possible to veneer a ceramic abutment using
suitable veneering materials and following the manufacturer‘s instructions.
Veneering must always be completed before bonding. The bonding surfaces of the
titanium bonding base must be treated with Al₂O₃ abrasives 110 μm and 2 bar
max. before bonding. The surfaces are bonded with a suitable luting material
(e.g. Multilink® Hybrid Abutment) in accordance with the manufacturer’s
instructions. After curing, all adhesive residues on basal surfaces and in the
screw channel must be carefully removed. The titanium bonding base must not
display any sign of damage. An undamaged and correct implant interface is
essential for correct placement of the prosthetic restoration.
Caution: On the model, only use the Lab Screw (max. torque: 10 Ncm).
Instructions for use in the dental practice
The customized prosthetic restoration is placed in the patient’s mouth using the silver Implant Screw. In the case of long-span bridge restora-tions, care must be taken to ensure that the screw connections are free of stress. To avoid any stresses, the implant screws are tightened diagonally.
- Caution: The Lab Screw must not be used for placement of the implant components in the patient’s mouth.
- Caution: The Implant Screw must be tightened to the torque specified for the system (see table).
- Please note: It is recommended to tighten the Implant Screw 10 minutes after placement.
- Please note: When DD Solid Connect® implant components are used for restorations with reduced-diameter implants, limited indications apply. These can be found in the relevant instructions for use from the implant manufacturer.
- Please note: DD Solid Connect® implant components from a particular series can only be combined with the corresponding compatible implant system.This also applies to the corresponding platform diameters.
Material:
DD Ti-Base 2CUT noLock, DD Implant Screw and DD Lab Screw are all made of
Ti6Al4V (medical grade 5) ASTM 136 ELI.
Cleaning, disinfection and sterilization
DD Solid Connect® implant components and accompanying screws are cleaned at Dental Direkt, but not sterilized, before being packaged. After customizing, and before being used on the patient, the implant compo-nents must be cleaned, disinfected and sterilized following the procedure described by Dental Direkt.
Cleaning/disinfection:
Automatic cleaning in a suitable washer-disinfector with an appropriate
cleaning program for medical instruments (e.g. “Vario TD Program”, Miele).
OR
The following manual cleaning procedure:
- Rinse with cold tap water for 1 minute
- Rinse the screw channel using a disposable syringe (min. capacity 10 ml)
- Clean in the ultrasonic cleaner at 60°C (140°F) +/- 2°C, e.g. using Dr. Weigert neodisher® LM 2 for 10 minutes at a concentration of 5%
- Neutralize with cold distilled water for 2 minutes
- Rinse in the ultrasonic cleaner using distilled water at 60°C ( 140°F) for 10 minutes
Sterilization
Double-seal products in sterilization film conforming to EN ISO 11607-1. Steam
sterilize in a suitable sterilizer with 3x fractionated prevacuum: 134°C
(273°F) for 5 minutes,
drying time: 20 minutes. Then store in sterile packaging.
Caution: DD Ti-Base 2CUT noLock, DD Implant Screw and DD Lab Screw are
suitable for single use only. Any reuse may lead to hazards due to infection.
Possible side effects and interactions:
Allergic reactions or skin irritations are possible in very rare cases. Diffe-
rent alloys in the oral cavity can lead to galvanic reactions when in direct
contact.
Disposal:
In compliance with local regulations.
Warranty
Our products undergo strict quality control and conform to the latest technical standards. Please refer to our current warranty terms for further details.
Please note: Any serious incident that occurs in connection with the pro- duct should be reported to the manufacturer and the competent authority of the Member State in which the user and/or patient is located
Please note: The safety and clinical performance summary report can be
requested at info@dentaldirekt.de.
We are continuously developing and enhancing our devices, and therefore
reserve the right to make changes.
The latest version of the instructions for use can also be found on our
website at: www.dentaldirekt.de/en
This version replaces all previous versions.
Explanation of symbols:
- Manufacturer
- Date of manufacture
- Batch code
- Catalog number
- Keep dry
- Non-sterile
- Do not use if package is damaged
- Torque
- Consult electronic instructions for use www.dentaldirekt.de/IFU
- Caution: Under U.S. federal law, the device may only be sold by or on behalf of a dentist.
- Do not re-use
- Medical device
- Unique device identification
Dental Direkt Series| Compatible with: manufacturer|
System| Size (mm)| Torque (Ncm)
---|---|---|---|---
DD AAS-Serie| Dentsply Implants®| Astra Tech®| 3,5 / 4,0| 25
4,5 / 5,0| 25
DD BSK-Serie| Bredent| Sky®| 3,5 NP| 25
4,0 / 4,5 RP| 25
DD CCA-Serie| Camlog®| Camlog®| 3,3| 20
3,8| 20
4,3| 20
5,0| 20
DD CCO-Serie| Camlog®| Conelog®| 3,3| 20
3,8| 20
4,3| 20
5,0| 20
DD DEV-Serie| Dentsply Implants®| Astra EV| 3,0| 25
3,6| 25
4,2| 25
4,8| 25
5,4| 25
DD DFX-Serie| Dentsply Implants®| Frialit®/Xive®| 3,4| 25
3,8| 25
4,5| 25
5,5| 25
DD MAR-Serie| MegaGen| AnyRidge®| 3,5 – 6,0| 30
DD MMI-Serie| Medentis Medical®| ICX| 4,1| 30
DD MSE-Serie| MIS Implants®| Seven®| 3,3 NP| 30
4,2 SP| 30
5,0 WP| 30
DD NBR-Serie| Nobel Biocare™| Brånemark| 3,5 NP| 35
4,1 RP| 35
DD NNA-Serie| Nobel Biocare™| Nobel Active™| 3,5 NP| 25
4,3 / 5,0 RP| 35
DD NRE-Serie| Nobel Biocare™| Replace Select®| 3,5 NP| 35
4,3 RP| 35
5,0 WP| 35
DD OTS-Serie| Osstem Implant| TS| 3,5 M| 20
4,0 – 6,0 R| 20
DD SBO-Serie| Straumann| Bone Level®| 3,3 NC| 35
4,1 / 4,8 RC| 35
DD SSY-Serie| Straumann| synOcta®| 4,8 RN| 35
6,5 WN| 35
DD TSP-Serie| Thommen Medical| SPI®| 3,5| 15
4,0| 25
4,5| 25
5,0| 25
6,0| 25
DD ZBC-Serie| Zimmer Biomet 3i| Certain®| 3,4| 30
4,1| 30
5,0| 30
DD ZTA-Serie| Zimmer Dental| Tapered Screw- Vent®| 3,5| 30
4,5| 30
5,7| 30
Dental Direkt GmbH · Industriezentrum 106 – 108
32139 Spenge · Germany
Tel.: +49 5225 863190
eIFU_DD Ti-Base 2CUT noLock_Rev.06_20230124
References
- Dental Direkt - close to you
- Dental Direkt - close to you
- Your search results | Dental Direkt
- Downloads | Dental Direkt
- Validated Milling Centers - Dental Direkt
- eIFU - electronic Instructions for Use | Dental Direkt
Read User Manual Online (PDF format)
Read User Manual Online (PDF format) >>