EL’ | OT’ co 2’
Carbon Dioxide WiFi Cloud-Connected Data Logger
Quick Start Guide
Set up a Cloud Account
To begin setting up your EL-I0T-0O2,you first need an Easy Log Cloud account.
- Visit easylogcloud.com and click Sign Up Now
- Download the Easy Log Cloud App onto your phone or tablet
https://www.lascarelectronics.com/software/easylog-software /easylog-cloud
Preparing the power supply
Select the correct plug type for your local region and fit it to the power supply. Note you e tofititata Slgmpmgleym?:\ Shush Enid rotate to lock it into place.
Plugging in the unit
Plug in your device and it will start automatically.
Install the batteries
Your EL-I0T-CO2 is designed to operate from the power supply provided, but can work for a short time from battery power, if mains power is Lost. If you wish to fit batteries, insert 4 x AA batteries into the battery compartment, taking care to place the batteries in the correct orientation. The sounder will beep when the batteries are first inserted.
Getting to know your EL-I0T-C02 data logger
|
---|---
1. Warning notification| 12. CO2, Units
---|---
2. Calibration reminder| 13. Humidity reading
3. CO, alarm status| 14. Humidity units
4. Temperature reading| 15. Humidity alarm status
5. Temperature alarm status| 16. MIN / MAX – CLEAR M/M button
6. Temperature units| 17. AUDIT – CONFIG button
7. Minimum reading| 18. RESET ALARM – CALIBRATION button
8. Maximum reading| 19. Alarm light and mute button
9. WiFi signal strength| 20. Battery compartment
10. Battery level| 21. Reset button
11. CO2, reading|
Connect to and configure in the Cloud
Sign in to the Easy Log Cloud App on your mobile device. Select “Setup Device”
from the burger menu and follow the on-screen instructions to configure your
EL-I0T-CO2.
Once your EL-IOT-CO2 is connected to your WiFi network and Easy Log account,
replace the battery cover and wall mount bracket. Setup is now complete.
Install your device in the location you wish to monitor.
You can now view the EL-I0T-C02’s data and change settings either in the
Syslog Cloud App or by visiting your account at:
www.easylogcloud.com
Display Status Indicator
- On startup, the LCD runs through a test sequence in which all elements are activated, the alarm LEDs light up and the sounder beeps.
- The CO2 sensor may take a few minutes to self-calibrate after startup, before readings are given. During this time keep in fresh air.
Display | Logger Status | Explanation |
---|---|---|
Startup | The logger is starting up, the number displayed is the firmware |
version
| Not configured| The logger has not yet been connected to the EasyLog Cloud,
readings will be displayed but not uploaded
| Setup Mode| The device is in setup mode, ready to connect to the EasyLog
Cloud
| Connected to App| The device has connected to the EasyLog App as part of the
setup process
| Not logging| The device has been turned on but can’t connect to the Cloud,
readings will be displayed but not logged. As soon as the device connects to
the Cloud it will start logging
| Setup Fail| The device has been unable to complete the setup process, check
the availability of your WiFi network and try again
| Firmware Update| The device is loading new firmware (you can check for
updates on the EasyLog Cloud)
| Update Failed| The device has failed to load new firmware and will
automatically revert to the previous version
| CO2 Calibration Due| You should activate the CO2 Fresh Air Calibration
process
| CO2 Fresh Air Calibration| The device is recalibrating the CO2 sensor, leave
it exposed to fresh air during this process
| CO2 Calibration Failed| The device was unable to calibrate the CO2 sensor,
you should expose it to fresh air and try again
****| Device warning| Press MIN/MAX to cycle through readings screens to see
warning screens
Button Functions
Button | Press | Function |
---|---|---|
Short Press | Cycle between current, minimum and maximum readings, and any |
warning screens
Long Press| Reset the maximum and minimum values for all parameters
| Short Press| Create an audit mark in the data record and initiate data
synchronization with the Cloud
Long Press| Enter Setup mode
| Short Press| Reset all active alarms (they will immediately re-trigger if an
alarm threshold is breached)
Long Press| Enter CO2 Fresh Air Calibration mode
Important Safety Information
WARNING: Failure to follow these safety instructions could result in
fire, electrical shock, other injury or damage.
Repairing or modifying
Never attempt to repair or modify this product. Dismantling, may cause damage
that is not covered under the warranty. Servicing should only be provided by
an approved supplier. If the product has been punctured, or seriously damaged
do not use it and return it to an approved supplier.
Power supply
Only use 1.5V AA alkaline batteries or a genuine EL-I0T-CO2 power supply to
power your EL-IOT data logger.
Disposal and recycling
You must dispose of this product and batteries according to relevant laws and
regulations. This product contains electronic components and therefore must be
disposed of separately from household waste.
Caution: Do not leave product in direct sunlight. Risk of explosion if
battery i replaced by an incorrect type. Dispose of used batteries according
to the instructions.
Lascar Electronics UK
Telephone: +44 (0) 1794 884 567
Email: sates@lascar.co.uk
Lascar Electronics US
Telephone: +1(814) 835-0621
Email: us-sales@lascarelectronics.com
Lascar Electronics HK
Telephone: +852 2389 6502
Email: sateshk@lascar.com.hk
www.lascarelectronics/data-loggers
Manufactured and supplied by:
Lascar Electronics Ltd
Module House Whit parish,
Wiltshire SP5 2S1 United Kingdom Issue 4 11/2022
Read User Manual Online (PDF format)
Read User Manual Online (PDF format) >>