My Weigh iBalance i3100 Table Top Precision Scale Instruction Manual
- June 12, 2024
- My Weigh
Table of Contents
- iBalance i3100 Table Top Precision Scale
- PRECAUTIONS BEFORE USING THE BALANCE
- SCALE FEATURES
- DISPLAY SYMBOLS
- KEYPAD FUNCTIONS
- POWER SUPPLY
- OPERATION INSTRUCTIONS
- ERROR MESSAGES
- CALIBRATION
- COUNTING FUNCTION
- WEIGHT RESPONSE SPEED
- COMMUNICATION AND BACKLIGHT
- SELECTING DESIRED WEIGHING MODES
- DATA TRANSMISSION SERIES RS-232 INTERFACE (only for communication)
- PRESET UNITS
- ADDITIONAL UNITS
- SPECIFICATIONS
- References
- Read User Manual Online (PDF format)
- Download This Manual (PDF format)
iBALANCE i3100™
USER MANUAL
Capacity
3100g x 0.01g, 6.8lb x 0.0002oz
iBalance i3100 Table Top Precision Scale
Thank you for purchasing the My Weigh® iBalance® 3100TM digital scale. Please
read all operating instructions carefully before use. This electronic scale is
a precision instrument. With normal care and proper treatment, it will provide
years of reliable service. For more information please visit
www.myweigh.com
Never load the scale with more than the maximum capacity. Although the
iBalance® 3100TM is designed to be extremely durable with extra overload
protection built into the case, overloading will permanently damage it! Avoid
any exposure to extreme heat or cold, your scale works better when operated at
normal room temperature. Keep your scale in a clean environment. Dust, dirt,
moisture, vibration, air currents and/or a close proximity to other electronic
equipment can all cause an adverse effect on the reliability and accuracy of
your scale. Handle with care. Gently apply all items to be weighed onto tray
top. Avoid shaking, dropping or otherwise shocking the scale. Scales are
delicate instruments and unlike cellular phones, scales have delicate sensors
that determine how much an item weighs. If you drop or shock your scale, these
sensors “feel” the shock and are sometimes destroyed. This happens with all
digital scales. We design our scales to be as resistant to shock or drops as
possible, however there is no way for us to protect 100% against load cell or
sensor damage. Failure to follow these instructions will void your warranty.
Always allow the scale to acclimate to normal room temperature for at least
one hour before use. Give your scale sufficient warm up time. Usually 30-60
seconds before calibration to give the internal components a chance to
stabilize.
PRECAUTIONS BEFORE USING THE BALANCE
- Matter charged with static electricity can affect accuracy. Discharge all static electricity. For example, one method is to use Static-Guard spray, and spray it on both sides of the weighing platform.
- The balance must be in an exactly horizontal position in order to achieve accurate measurement results. In order to bring the balance into a horizontal position, the adjustable feet are turned either clockwise or counter-clockwise until the air bubble on the front panel is in the center of the marked circle.
- Please use an independent power outlet to avoid interference from other electrical appliances.
- Don’t place any objects on the platform before powering on.
- When possible please allow the scale to warm up for several minutes before operation.
- Items should always be placed on the center of the platform when being weighed.
SCALE FEATURES
DISPLAY SYMBOLS
| Scale is in ZERO mode.
---|---
| Scale is in TARE mode.
| BATTERY needs recharging.
| The display reading is STABLE.
KEYPAD FUNCTIONS
| ON/OFF power switch.
---|---
| SAMPLING & COUNTING.
| UNIT selection
| TARE is used to deduct the weight of an item or container. The symbol will
appear and reading will go to zero. Press it again to exit the tare mode (when
empty), the tare indication will disappear.
| CAL is used for entering calibration and other function settings
POWER SUPPLY
The i3100TM is powered by DC 6 V rechargeable battery or directly by power adapter.
OPERATION INSTRUCTIONS
Weighing Procedures
-
Press to turn on the scale.
When the power is turned on, the scale will countdown for a few seconds and “0” will appear on the display. -
Select the weighing unit with key.
Press to select a weighing unit.
Once the unit has been selected, the selected unit will be displayed next to the weight value. -
Start weighing Verify the reading is “0”. Place objects on the weighing platform to weigh.
When the reading becomes stable, the stable indicator is displayed.
Tare
Tare can be used for eliminating the weight value of an empty container. Place
an empty container on the scale and press . Then place the items to be weighed
in the container.
NOTE: When all weight is removed from the weighing tray, the tared value
of a container will be displayed as a negative number. Press again to return
the scale to zero.
ERROR MESSAGES
When an applied load exceeds the capacity, “——” will appear on the display. Remove excessive load immediately. The unit may return to normal operation. Remember: You can permanently damage the scale and void your warranty by overloading it! The LCD will display”ERR “if the weight placed on the platform is incorrect during calibration.
CALIBRATION
When to calibrate – calibration is RARELY required. Calibration may be required when the scale is first set up for use, or if the scale is moved to a different altitude or new location. This is necessary because the weight of a mass in one location is not necessarily the same in another location. Also, with time and use, mechanical deviations can occur.
How to calibrate: you must have an accurate 1kg weight or combination of weights in order to calibrate
- Press and hold , release when the LCD shows “CAL”.
- The display now shows “1000.00”. This is where you input the calibration weight you are going to use to calibrate (1000 grams is recommended). Use to cycle through the numbers shown on the screen and use on the selected digit to adjust selected digit. To calibrate using the recommended 1000 grams, the screen must read “1000.00”. Once this is set, place the 1000 gram calibration weight on the tray.
- The display will show “OK” and return to normal weighing mode. Calibration is complete.
COUNTING FUNCTION
- Press to cycle through the weighing modes until the display will show” CA 10 PCS”(means sample size is 10 pcs)
- Press to toggle sample size, “10z l20z l50z l100” , “200”, “500”, “1000”pcs will appear in. Stop at the one you want to use.
- Put the exact quantity of samples desired on the platform and press , the set sample size will appear.
- Keep adding objects to be counted on the pan, the total number of the objects will be displayed. If the unit weight is too small for the counting resolution, the display will show “Err”
- To return to normal weighing press
WEIGHT RESPONSE SPEED
The i3100 allows you to adjust the scale reaction time and division selction.
- Press and hold (do not release) and power the scale on with . Wait until the display shows “nb0 – nb4” and release
- Press to select your response speed. (nb0: fastest, nb4 slowest) Press to confirm.
COMMUNICATION AND BACKLIGHT
-
Press and hold and press to power the scale on. The scale will power on and go through the self test.
COMMUNICATION -
The scale should now read “bAUd” followed by “12”, “24”, “48” or “96”. These values represent the baud rate settings: 1200, 2400, 4800, 9600.
-
Use to scroll through the settings and press to confirm.
-
The display will now read “Pr”(Press Sending),”Co”(Stable Sending”or”St”(Continuous Sending)
-
Use to scroll through the settings and then press to confirm
BACKLIGHT -
The display should now read “bL On” (Backlight ON), “bL OFF” (Backlight OFF) or “bL AUTO” (Automatic Backlight).
-
Use to scroll through the settings and press to confirm
-
Once is pressed the scale will return to normal weighing.
SELECTING DESIRED WEIGHING MODES
The 3100TM comes with a preset 9 units with the option to add another 9 units if required. To enable or disable each unit weighing modes follow these steps:
- Press and hold and together, then press and release , the scale will power on. Release all the keys when the display shows “PASS”.
- Press 4 times and the display will show “YES” and the corresponding unit. Press again to enable (yes) or disable (no) the selected unit.
- Press to move to the next unit. The scale will return to normal weighing mode when all the unit settings are finished.
DATA TRANSMISSION SERIES RS-232 INTERFACE (only for communication)
- iBalance 0232C,s -RS232 C’s UART signal
- Format
(1) Baud rate: 1200 bps 2400 bps 4800bps 9600 bps
(2) Data bits: 8 bits
(3) Parity bit: none
(4) Stop bit: 1 bit
(5) Code ASCII
DATA FORMAT
HEAD1 | HEAD2 | DATA | UNIT | CR |
---|---|---|---|---|
12 3 | 4 56 | 7 8 9 10 11 12 13 14 15 16 17 | 18 19 20 | 21 22 23 24 25 |
HEAD1 (2BYTES)
OL – overload
ST – stable| HEAD2 (2BYTES)
NT – net weight mode
US – unstable
---|---
DATA(8BYTE)
2D (HEX) = “- ”(negative sign)
2E (HEX) = “. ”(decimal point)| 20 (HEX) = “ ”(blank)
---|---
UNIT (4 byte)
g = 20 (HEX); 20 (HEX);
kg = 20 (HEX); 20 (HEX);
ct = 20 (HEX); 20 (HEX);
ozt = 20 (HEX); 6F (HEX);
CR = OD (HEX); OA (HEX)| 20 (HEX); 67 (HEX)
6B (HEX); 67 (HEX)
63 (HEX); 74 (HEX)
7A (HEX); 74 (HEX)
---|---
TRANSMISSION EXAMPLE
stable net + 0.168 g
HEAD, HEAD, DATA UNIT CR
ST, NT + 0.168 g OA, OD
PRESET UNITS
• g
• ct
• oz
• ozt
• lb
• dr
• gn
• dwt
• gsm| GRAM
CARAT
OUNCE
TROY OUNCE
POUND
DRAM
GRAIN (UK)
PENNY WEIGHT
GRAMS PER METRIC SQUARED
---|---
ADDITIONAL UNITS
• MM
• tl.J
• tl.T
• tl.H
• t
• 1/8
• 1/4
• TMr
• TAr| MOMME (JPN)
JEWELRY TAEL (HONG KONG)
TAEL (TWN)
TAEL (HONG KONG)
TOLA (INDIA)
1/8 OUNCE
1/4 OUNCE
TOLA, MASHA, RATTI
TOLA, ANNIE, RATTI
---|---
SPECIFICATIONS
Capacity| 3100g x 0.01g| Units| g, ct, oz, ozt, lb, dr, gn, dwt,
gsm, PCS , MM, tl.J, tl.T. tl.H, t, 1/8, 1/4, TMr, TAR
---|---|---|---
Scale dimension| 267mm x 191mm x 70mm
Tray dimension| 127mm x 146mm
Scale Weight| 1100g
Operating temperature| Optimum 10-40°C (50-104°f)
Power Source| DC 6 V Rechargeable Battery or power adaptor
Tare range| Up to scale’s maximum capacity