TGL20M Benchtop High Speed Refrigerated Centrifuge Instruction Manual

June 12, 2024
TGL

Benchtop high-speed refrigerated centrifuge

TGL20M

operation instruction

Before use, please read the operation instruction handbook

Safety alerts and precautions:

1. Safety alerts
  1. Before use, please read the operation instruction handbook.
  2. Do not use in wet, hot or dusty locations to avoid instrument failure.
  3. When the instrument is running, it must not move the instrument and must not open the door cover of the instrument.
  4. The countertop of the installation instrument should be solid and flat, there should be no shaking phenomenon, all four rubber machine feet should be in full contact with the countertop and uniform force.
  5. Centrifuge tube added liquid as evenly as possible, if the difference between the addition will produce large vibration, at this time should be stopped to check, so that the addition meets the requirements, centrifugal test tube must be placed in even symmetry, not singular into or asymmetrical into the rotor.
  6. After each use, open the instrument door cover, allow the water in the centrifugal cavity to evaporate, and clean the inner cavity of the instrument. The instrument must have a reliable ground wire when using a power supply.
2. Precautions
  1. Do not strongly collide, vibrate or re-throw the instrument to avoid non-serviceable consequences of damage to the mechanical or electrical parts of the instrument.
  2. Do not use the instrument in wet or hot places to avoid extensive damage to the electrical part of the instrument.

[ Product Name ] Centrifuge         [ Model ] TGL20M [ Product performance structure and composition ] It consists of a control system, centrifugal cavity, drive system, rotor, refrigeration system (If it is a refrigerated centrifuge) and safety protection.
[ Scope of application ] This product is suitable for use by trained professional operators and technicians.
[ Taboo ]  It is taboo to use without professional operators or personnel without relevant operational qualifications.
[ Precautions ] See above.
[ Alerts and suggestive content ] Stop prompts
[ Operating instructions ] See below
[ Label, packaging identification sample ] Please read the instructions

[ Storage conditions/methods ] Rooms with ventilation/drying/light avoidance /non-corrosive gases should be stored.

1. Overview

This manual adapts to TGL20M benchtop high-speed refrigerated centrifuges can be widely used in biochemistry, genetic engineering, Radioimmunology and other fields, is an ideal experimental instrument for research institutes. This machine uses DC variable frequency motor, microcomputer control, liquid crystal display, lift speed, low noise, small vibration, electronic door lock safe and reliable. (See Table 1/Table 2 for major technical performance)

Table 1: Technical parameters

Model TGL20M
Max. Speed 20000 rpm
Max. RCF 28948 xg
Max. Capacity 4 x 100 ml
Speed Accuracy ± 10 rpm
Temperature Range -20~ +40 °C
Temperature Accuracy ± 1°C
Presentation LCD
Timer Range 1sec ~ 99min 59sec
Motor AC frequency conversion
Power Supply AC110 V 60 Hz 16 A
Compressed units R134a (No fluorine refrigerant )
Full Power 1 kW
Noise ≤65 db
Dimensions(LxDxH) 20.5 x 27.6 x 14.6 in / 520 x 700 x 370 mm
Weight(w/o rotor) 152.1 lb / 69 kg

Table 2: Instrument rotor parameters

Number Name Max. Speed Max. RCF Capacity TubeSize(mm)
NO.1 Fixed angle rotor 20000 r/min 28948 xg 12 x 1.5/2 ml Φ10×42
NO.2 15000 r/min 22600 xg 24 x 1.5/2 ml Φ10×42
NO.3 13000 r/min 17930 xg 48 x 1.5/2 ml Φ10×42
NO.4 14500 r/min 21807 xg 48 x 0.2/0.5 ml Φ8×33
NO.5 15000 r/min 13856 xg 3 x 8 x 0.2 ml (PCR) Φ6×22
NO.6 14000 r/min 12826 xg 6 x 8 x 0.2 ml (PCR) Φ6×22
NO.7 15000 r/min 19350 xg 10 x 5 ml Φ14×54
NO.8 15000 r/min 22680 xg 12 x 10 ml Φ16×87
NO.9 15000 r/min 22680 xg 8 x 15 ml Φ16×118
NO.10 15000 r/min 22680 xg 8 x 20 ml Φ22×104
NO.11 14000 r/min 19060 xg 6 x 30 ml Φ25×94
NO.12 13500 r/min 18950 xg 4 x 50 ml Φ29×106
NO.13 13000 r/min 18840 xg 6 x 50 ml Φ29×106
NO.14 12000 r/min 15635 xg 4 x 100 ml Φ37×124
NO.15 12000 r/min 15869 xg 24 capillary tubes Φ4×100
NO.16 Microplate swinging bucket rotor 4000 r/min 2310 xg 2 x 48 holes
2. Installed

A. The native power supply is a single-phase AC110V 60Hz 15A.
B. The power supply should have a safe and reliable wire, strictly prohibit the ground wire instead of zero line, not to pull the machine wire indiscriminately.
C. The placement of the site indoor environment non-corrosive gas relative humidity shall not be greater than 85%, there should be no other strong vibration source. the table surface of the unit should be rigid and hard enough to support the instrument, the countertop should be flat without any slope, so that the four rubber soles of the machine and the table to maintain smooth contact.

3. Introduction to the interface and instructions for use:
  1. The centrifuge placed on a solid countertop, four rubber machine feet should be in good contact with the table surface, shake the centrifuge with your hand, check whether the centrifuge is placed smoothly.
  2. Open the door cover, the centrifuge tube into the rotor test tube hole, centrifuge tube must be placed in even symmetry (high-speed centrifuge centrifuge test fluid should be weighed to join), pay attention to the rotor cover tight, and re-check whether the test tube is symmetrically placed, screws are tight.
  3. Close the door cover, pay attention to make the door cover locked, finished by hand to check whether the door cover is closed. Plug in the power supply and press the power switch.
  4. Set speed, temperature, time, rotor number, program, speed-up gear, speed-down gear: (operator panel diagram below)

(unscrew the rotor cover, with a handle into the rotor center hole

Introduction to keystrokes :
Set: Switch adjustment parameters, this key is set for switching adjustment speed, time, speed etc.
Up: Increase hypothesis parameter value; This pressed key adjustment rotational speed value and time value establishment.
Down: Reduce the set parameter value, this button adjusts the speed value and time value setting.
Enter: Exit the set state and save the setting parameters.
Short: Press this key centrifuge to start centrifugation, relax and stop working.
Start: Start the motor, (press this key to start the centrifuge after setting all parameters).
Stop/Open: Eliminate alarm when alarm, operating status press can stop midway standby state open cover.
Parameter set:
Press the setting key to cycle through the set parameters, the parameters set the cycle order (number of keys): 1: speed; 2: temperature; 3: time minutes; 4: time seconds; 5: rotor number; 6: program number; 7: speed upshift (Larger the number, Faster the speed), 8: the lower speed (Larger the number, Faster the descent) according to the actual situation. Press the stop button to eliminate When there is only one rotor, the rotor number adjustment is skipped and the room temperature interface skips the temperature adjustment. Press the up and down key to adjust the parameter value, adjust the parameter to the expected value, press the confirmation key to exit the set state and save the modified parameters.
When the above steps are complete, press the “Confirm” key to confirm the speed, temperature and time set above, and then press the “Start” key to start the centrifuge (when pressing the “Start” key, if the instrument cannot start, check that the door cover is closed).

4. Precautions
  1. The table of the centrifuge should be flat, four rubber machine feet should be in contact with the table and uniform force, so as not to produce vibration.
  2. Centrifuge tube liquid should weigh the balance, if the difference between the addition of liquid too large operation will produce large vibration, at this time should be stopped to check, so that the addition meets the requirements, centrifugal test tube must be placed in even symmetry.
  3. If there is a centrifugal test tube rupture, will cause large vibration should be immediately stopped treatment.
  4. After each shutdown and then power on the interval should not be less than 5 minutes, so as not to block the compressor to turn damage.
  5. After each centrifuge is completed, open the instrument door cover, let the water evaporate, and regularly take out the rotor, can be raised) otherwise long time on the shaft, may rust to death, the rotor will not be taken out and cause the centrifuge machine scrap.
  6. The life of the native host 6 years, rotor service life of 5 years, expired should replace the rotor. all rotors must not be used at more than their maximum speed.
5. Troubleshooting/warranty coverage

Common troubleshooting methods:

Common failures Possible reason Elimination method
Power on, the display window is not lit No 110V power supply Check power

supply
Fuse is broken| Replacement
Power plug-in is loose| Plugin
Press the start button centrifuge does not operate| Door cover is not closed| Close thedoorcover
Wiring plug-in is loose| Plugin
Control board is damaged| Replacement
Motor has electricity does not turn, the motor is damaged| Replacethemotor
Centrifuges vibrate a lot| Centrifuge tube is not placed symmetrically| Check
Centrifuge tube is broken| Replacement
Rotor is not rotating| Check

When the centrifuge fails, check the fault condition corresponding to the fault number and resolve the failure of the centrifuge the fault display.

  • E-1: Imbalance   E-2: Overspeed   E-3: Door
  • E-4: Speed-hall error   E-5: Overvoltage (during shutdown)
  • E-6: Overcurrent   E-7: No speed measured
  • E-8: Communication
  • E-9: The external input power supply fluctuates too much
    E-10: Temperature sensor
Wiring diagram

TGL20M Benchtop High Speed Refrigerated Centrifuge A1

Benchtop high-speed refrigerated centrifuges
Packing list

No

| Name| Specification| Quantity|

Note

---|---|---|---|---
1| Host| TGL20M| 1| Wooden box
2| Rotor| 48x 1.5 ml| 1|
3| Rotor| | |
4| Rotor| | |
5| Test tube| 1.5 ml| 48|
6| Test tube| | |
7| Test tube| | |
8| Accessories/tools| | |
9| Operation instruction| | 1|
10| Power cord| | 1|
11| Packing list| | 1|
| | | |
| | | |
| | | |

Product form picture

Read User Manual Online (PDF format)

Read User Manual Online (PDF format)  >>

Download This Manual (PDF format)

Download this manual  >>

Related Manuals