TGL16M Benchtop High Speed Refrigerated Centrifuge Instruction Manual
- June 12, 2024
- TGL
Table of Contents
TGL16M Benchtop High Speed Refrigerated Centrifuge
Safety alerts and precautions:
Safety alerts:
- Before use, please read the operation instruction handbook.
- Do not use in wet, hot or dusty locations to avoid instrument failure.
- When the instrument is running, it must not move the instrument and must not open the door cover of the instrument.
- The countertop of the installation instrument should be solid and flat, there should be no shaking phenomenon, all four rubber machine feet should be in full contact with the countertop and uniform force.
- Centrifuge tube added liquid as evenly as possible, if the difference between the addition will produce large vibration, at this time should be stopped to check, so that the addition meets the requirements, centrifugal test tube must be placed in even symmetry, not singular into or asymmetrical into the rotor.
- After each use, open the instrument door cover, allow the water in the centrifugal cavity to evaporate, and clean the inner cavity of the instrument. The instrument must have a reliable ground wire when using a power supply.
Precautions:
- Do not strongly collide, vibrate or re-throw the instrument to avoid non-serviceable consequences of damage to the mechanical or electrical parts of the instrument.
- Do not use the instrument in wet or hot places to avoid extensive damage to the electrical part of the instrument.
【Product Name】 Centrifuge
【Model】TGL16M
【Product performance structure and composition】 It consists of a control
system, centrifugal cavity, drive system, rotor, refrigeration system (If it
is a refrigerated centrifuge) and safety protection.
【Scope of application】 This product is suitable for use by trained
professional operators and technicians.
【 Taboo 】 It is taboo to use without professional operators or personnel
without relevant operational qualifications.
【Precautions】 See above.
【Alerts and suggestive content】 Stop prompts
【Operating instructions】 See below
【 Label, packaging identification sample 】Please read the instructions
【Storage conditions/methods】
Rooms with ventilation/drying/light avoidance/non-corrosive gases should be
stored.
Overview
This manual adapts to TGL16M benchtop high-speed refrigerated centrifuges can be widely used in biochemistry, genetic engineering, radioimmunology and other fields, is an ideal experimental instrument for research institutes. This machine uses DC variable frequency motor, microcomputer control, liquid crystal display, lift speed, low noise, small vibration, electronic door lock safe and reliable. (See Table 1/Table 2 for major technical performance)
Table 1: Technical parameters
Model | TGL16M |
---|---|
Max. Speed | 16500 rpm |
Max. RCF | 20653 xg |
Max. Capacity | 4 x 100 ml |
Speed Accuracy | ± 10 rpm |
Temperature Range | -20~ +40 °C |
Temperature Accuracy | ± 1°C |
Presentation | LCD |
Timer Range | 1sec ~ 99min59sec |
Motor | AC frequency conversion |
Power Supply | AC110 V 60 Hz 16A |
Compressed units | R134a (No fluorine refrigerant ) |
Full Power | 1 kW |
Noise | ≤65 db |
Dimensions(LxDxH) | 20.5 x 27.6 x 14.6 in / 520 x 700 x 370 mm |
Weight(w/o rotor) | 152.1 lb / 69 kg |
Table 2: Instrument rotor parameters
Nu mber| Name| Max. Speed| Max. RCF| Capacity| Tube
Size(mm)
---|---|---|---|---|---
NO.1
| Fixed angle rotor| 16500 r/min| 20653 xg| 12 x 1.5/2 ml| Φ10 x 42
NO.2| 13500 r/min| 18760 xg| 24 x 1.5/2 ml|
Φ10 x 42
NO.3
| 12000 r/min| 16843 xg| 48 x 1.5/2 ml| Φ10 x 42
NO.4| 13000 r/min| 15635 xg| 48 x 0.2/0.5 ml|
Φ8 x 33
NO.5
| 15000 r/min| 13856 xg| 3 x 8 x 0.2 ml (PCR)| Φ6 x 22
NO.6| 14000 r/min| 12826 xg| 6 x 8 x 0.2 ml (PCR)|
Φ6 x 22
NO.7
| 15000 r/min| 19350 xg| 10 x 5 ml| Φ14 x 54
NO.8| 12000 r/min| 14856 xg| 12 x 10 ml|
Φ16 x 87
NO.9
| 12000 r/min| 14510 xg| 8 x 15 ml| Φ16 x 118
NO.10
| 12000 r/min| 14510 xg| 8 x 20 ml| Φ22 x 104
NO.11| 12000 r/min| 14856 xg| 6 x 30 ml|
Φ25 x 94
NO.12
| 14000 r/min| 15635 xg| 4 x 50 ml| Φ29 x 106
NO.13| 12000 r/min| 14856 xg| 6 x 50 ml|
Φ29 x 106
NO.14
| 12000 r/min| 15635 xg| 4 x 100 ml| Φ37 x 124
NO.15| 12000 r/min| 15869 xg| 24 capillary tubes|
Φ4 x 100
NO.16
| Microplate swinging bucket rotor| 4000 r/min| 2310 xg| 2 x 48 holes|
Installed
A. The native power supply is a :su1ga:-p11a:se AC l 10V60Hz 15A.
B. The power supply should have a safe and reliable wire, strictly
prohibit the ground wire instead of zero line, not to pull the machine wire
indiscri1ninately.
C. The placement of the site indoor environment non-corrosive gas
relative humidity shall not be greater than 85%, there should be no other
strong vibration source. The table surface of the unit should be rigid and
hard enough to support the instrument, the countertop should be flat without
any slope, so that the four rubber soles of the machine and the table to
maintain smooth contact.
Introduction to the interface and instructions for use:
-
The centrifuge placed on a solid countertop, four rubber machine feet should be in good contact with the table surface, shake the centrifuge with your hand, check whether the centrifuge is placed smoothly.
-
Open the door cover, the centrifuge tube into the rotor test tube hole, centrifuge tube must be placed in even symmetry (high-speed centrifuge centrifuge test fluid should be weighed to join), pay attention to the rotor cover tight, and re-check whether the test tube is symmetrically placed, screws are tight.
-
Close the door cover, pay attention to make the door cover locked, finished by hand to check whether the door cover is closed.
Plug in the power supply and press the power switch. -
Set speed, temperature, time, rotor number, program, speed-up gear, speed-down gear: ( operator panel diagram below)
(unscrew the rotor cover, with a handle into the rotor center hole
Introduction to keystrokes:
Set: Switch adjustment parameters, this key is set for switching adjustment
speed, time, speed etc.
Up: Increase hypothesis parameter value; This pressed key adjustment
rotational speed value and time value establishment.
Down: Reduce the set parameter value, this button adjusts the speed value and
time value setting.
Enter: Exit the set state and save the setting parameters.
Short: Press this key centrifuge to start centrifugation, relax and stop
working.
Start: Start the motor, (press this key to start the centrifuge after setting
all parameters).
Stop/Open: Eliminate alarm when alarm, operating status press can stop midway
standby state open cover.
Parameter set:
Press the setting key to cycle through the set parameters, the parameters set
the cycle order (number of keys): 1: speed; 2: temperature; 3: time minutes;
4: time seconds; 5: rotor number; 6: program number; 7: speed upshift (Larger
the number, Faster the speed), 8: the lower speed (Larger the number, Faster
the descent)
according to the actual situation. Press the stop button to eliminate When
there is only one rotor, the rotor number adjustment is skipped and the room
temperature interface skips the temperature adjustment. Press the up and down
key to adjust the parameter value, adjust the parameter to the expected value,
press the confirmation key to exit the set state and save the modified
parameters.
When the above steps are complete, press the “Confirm” key to confirm the
speed, temperature and time set above, and then press the “Start” key to start
the centrifuge (when pressing the “Start” key, if the instrument cannot start,
check that the door cover is closed).
Precautions:
- The table of the centrifuge should be flat, four rubber machine feet should be in contact with the table and uniform force, so as not to produce vibration.
- Centrifuge tube liquid should weigh the balance, if the difference between the addition of liquid too large operation will produce large vibration, at this time should be stopped to check, so that the addition meets the requirements, centrifugal test tube must be placed in even symmetry.
- If there is a centrifugal test tube rupture, will cause large vibration should be immediately stopped treatment.
- After each shutdown and then power on the interval should not be less than 5 minutes, so as not to block the compressor to turn damage.
- After each centrifuge is completed, open the instrument door cover, let the water evaporate, and regularly take out the rotor, can be raised) otherwise long time on the shaft, may rust to death, the rotor will not be taken out and cause the centrifuge machine scrap.
- The life of the native host 6 years, rotor service life of 5years, expired should replace the rotor. All rotors must not be use dat more than their maximum speed.
Troubleshooting/warranty coverage
Common troubleshooting methods:
Common failures | Possible reason | Elimination method |
---|---|---|
Power on, the display window is not lit | No 110V power supply | Check power |
supply
Fuse is broken| Replacement
Power plug-in is loose| Plug in
Press the start button centrifuge does not operate
| Door cover is not closed| Close the door
cover
Wiring plug-in is loose| Plug in
Control board is damaged| Replacement
Motor has electricity does not turn, the motor is damaged| Replace the
motor
Centrifuges vibrate a lot
| Centrifuge tube is not placed symmetrically| Check
Centrifuge tube is broken| Replacement
Rotor is not rotating| Check
When the centrifuge fails, check the fault condition corresponding to the fault number and resolve the failure of the centrifuge the fault display.
- E-1:Imbalance E-2: Overspeed E-3: Door
- E-4:Speed-hall error E-5 :Overvoltage (during shutdown)
- E-6: Overcurrent E-7: No speed measured
- E-8: Communication
- E-9: The external input power supply fluctuates too much
- E-10: Temperature sensor
Wiring diagram
Benchtop high-speed refrigerated centrifuges
Packing list
No | Name | Specification | Quantity | Note |
---|
1
| Host| TGL16M| 1|
Wooden box
2
| Rotor| 24 x 1.5 ml| 1|
3| Rotor| |
|
4
| Rotor| | |
5| Test tube| 1.5 ml|
24
|
6
| Test tube| | |
7
| Test tube| | |
8|
Accessories/tools
| | |
9| Operation instruction|
|
1
|
10
| Power cord| | 1|
11|
Packing list
| |
1
|
Product form picture
Read User Manual Online (PDF format)
Read User Manual Online (PDF format) >>