MEAN WELL MPD-65A Series 65W Dual Output Medical Type Owner’s Manual
- June 12, 2024
- MEAN WELL
Table of Contents
MPD-65A Series 65W Dual Output Medical Type
Owner’s Manual 65W Dual Output Medical Type
MPD-65 series
Features :
- Universal AC input / Full range
- Low leakage current <250M
- Protections: Short circuit / Overload / Over voltage
- Cooling by free air convection
- Medical safety approved (2 x MOPP between primary to secondary)
- 100% full load burn-in test
- Fixed switching frequency at 45KHz ‘ 3 years warranty
SPECIFICATION
MODEL | MPD-65A | MPD-65B |
---|
OUTPUT
| OUTPUT NUMBER| CH1| CH2| CH1| CH2
DC VOLTAGE| 5V| 12V| 5V| 24V
RATED CURRENT| 5.5A| 2.8A| 3.5A| 2A
CURRENT RANGE| 0.4 ~ 7A| 0.2 ~ 3.2A| 0.4 ~ 6A| 0.2 ~ 2.6A
RATED POWER| 61.1W| 65.5W
OUTPUT POWER (max.)| 72W with 18CFM min. Forced air convection
RIPPLE & NOISE (max.) Note.2| 60mVp-p| 150mVp-p| 60mVp-p| 150mVp-p
VOLTAGE ADJ. RANGE| CH1:4.5 ~ 5.5V| CH1:4.5 ~ 5.5V
VOLTAGE TOLERANCE Note.3| ±4.0%| ±7.0%| ±4.0%| ±7.0%
LINE REGULATION| ±1.0%| ±2.0%| ±1.0%| ±2.0%
LOAD REGULATION| ±3.0%| ±4.0%| ±3.0%| ±4.0%
SETUP, RISE TIME| 800ms, 20ms/230VAC 800ms, 20ms/115VAC at full
load
HOLD UP TIME (Typ.)| 80ms/230VAC 12ms/115VAC at full load
INPUT
| VOLTAGE RANGE| 90 ~ 264VAC 127 ~ 370VDC
FREQUENCY RANGE| 47 ~ 63Hz
EFFICIENCY(Typ.)| 75%| 78%
AC CURRENT (Typ.)| 1.6A/115VAC 1A/230VAC
INRUSH CURRENT (Typ.)| COLD START 20A/115VAC 40A/230VAC
LEAKAGE CURRENT **Note.7**| Earth leakage current < 250μ A/264VAC ,
Touch current < 60μ A/264VAC
PROTECTION
|
OVERLOAD
| 73 ~ 105W rated output power
Protection type : Hiccup mode, recovers automatically after fault condition is
removed
OVER VOLTAGE
| 5.75 ~ 6.75VDC on CH1
Protection type : Hiccup mode, recovers automatically after fault condition is
removed
ENVIRONMENT
| WORKING TEMP.| -10 ~ +55℃ (Refer to “Derating Curve”)
WORKING HUMIDITY| 20 ~ 90% RH non-condensing
STORAGE TEMP., HUMIDITY| -20 ~ +85℃, 10 ~ 95% RH
TEMP. COEFFICIENT| ±0.04%/℃ (0 ~ 50℃) on +5V output
VIBRATION| 10 ~ 500Hz, 2G 10min./1cycle, period for 60min. each along X,
Y, Z axes
SAFETY & **EMC**
(Note 4)
| SAFETY STANDARDS| ANSI/AAMI ES60601-1, TUV EN60601-1, IEC60601-1
approved
ISOLATION LEVEL| Primary-Secondary: 2xMOPP, Primary-Earth:1xMOPP
WITHSTAND VOLTAGE| I/P-O/P:4KVAC I/P-FG:2KVAC O/P-FG:0.5KVAC
ISOLATION RESISTANCE| I/P-O/P, I/P-FG:100M Ohms / 500VDC / 25℃/ 70% RH
EMC EMISSION| Compliance to EN55011 (CISPR11) Class B, EN61000-3-2,-3
EMC IMMUNITY| Compliance to EN61000-4-2,3,4,5,6,8,11, EN60601-1-2,
medical level, criteria A
OTHERS
| MTBF| 291.3Khrs min. MIL-HDBK-217F (25℃)
DIMENSION| 1277642mm (LWH)
PACKING| 0.25Kg; 54pcs/16Kg/1.28CUFT
NOTE| 1. All parameters NOT specially mentioned are measured at 230VAC
input, rated load and 25℃ of ambient temperature.
2. Ripple & noise are measured at 20MHz of bandwidth by using a 12″ twisted
pair-wire terminated with a 0.1uf & 47uf parallel capacitor.
3. Tolerance : includes set up tolerance, line regulation and load
regulation.
4. The power supply is considered a component which will be installed into a
final equipment. All the EMC tests are been executed by mounting the unit on a
360mm*360mm metal plate with 1mm of thickness. The final equipment must be re-
confirmed that it still meets EMC directives. For guidance on how to perform
these EMC tests, please refer to “EMI testing of component power supplies.”
(as available on http://www.meanwell.com)
5. Mounting holes M1 and M2 should be grounded for EMI purposes.
6. Heat Sink HS1,HS2 can not be shorted.
7. Touch current was measured from primary input to DC output.
※ Product Liability Disclaimer: For detailed information, please refer to
https://www.meanwell.com/serviceDisclaimer.aspx
https://www.meanwell.com/Upload/PDF/Medical_EN.pdf
Mechanical Specification
AC Input Connector (CN1) : Molex 5277-02 or equivalent
Pin No. | Assignment | Mating Housing | Terminal |
---|---|---|---|
1 | AC/L | Molex 5195 or equivalent | Molex 5194 or equivalent |
2 | AC/N |
DC Output Connector (CN2) : Molex 5273-06 or equivalent
Pin No. | Assignment | Mating Housing | Terminal |
---|---|---|---|
1 | +V |
Molex 5195 or equivalent
|
Molex 5194 or equivalent
2,3| +5V
4,5| COM
6| NC
DC Output Connector (CN2) : Molex 5273-06 or equivalent
: Grounding Required
1.HS1,HS2 cannot be shorted
2.M1 is safety ground
Block Diagram
Derating Curve
Static Characteristics