Baxtran ABD Measuring Scales User Manual

June 12, 2024
Baxtran

ABD COMPACT BALANCE WEIGHT CONTROL
User Manual

Baxtran ABD Measuring Scales

The manufacturer reserves the right to modify the specifications of its products in order to make technical improvements or comply with new regulations.

EXCITATION

ABD

Input 230~240V
Output 10V 600mA
Rechargeable Battery 6V/4Ah

BEFORE ITS USE

  1. Use an independent electric source to prevent electronic disturbances.
  2. Don’t place any object on the platform when switching on the indicator.
  3. Please, warm-up the scale during 2-3 minutes before using it.
  4. Avoid sudden changes in temperature and draughts.
  5. Don’t overload the scale; do not exceed its maximum capacity.

CONSUMPTION

ABD

Battery life: Without back illumination, approx., 160 hours.

With back illumination, approx., 120 hours.

KEYBOARD DESCRIPTION

Baxtran ABD Measuring Scales - KEYBOARD DESCRIPTION

| First function; Press this key to switch on the scale.
---|---
| First function; Press and hold the key during 3 seconds to switch of the indicator.
| First function; to choose the unit of weight.
Second function; to exit from the programming mode.
| First function; : to remove (tare) the weight of a container.
Second function; to move to the right in the programming mode.
| First function; 0.
Second function; to move to the left.
| First function; to view the brute weight after that to do the tare (weight) the container.
Second function; To increase the values inside the programming.
| First function; to accumulate in memory the value of the weight that appears in the screen.
Second function; manual transmission of data through RS-232 port to a PC or printer. (No function ABD)
Third function; confirmation key in the programming mode..

ABD APPLICATIONS

5.1 NORMAL WEIGHING MODE
5.1.1 EQUIPMENT CONFIGURATION
5.1.2 FIRST CALIBRATION
5.1.3 USE
5.2 PIECE COUNTING MODE
5.2.1 EQUIPMENT CONFIGURATION
5.2.2 FIRST CALIBRATION
5.2.3 USE
5.3 LIMITS AND ALARM
5.4 UNIT RANGE/ MULTI RANGE/ MULTI INTERVAL
5.1 NORMAL WEIGHING MODE
5.1.1 CONFIGURATION OF THE EQUIPMENT
See section LF2 of the technical parameters
5.1.2 FIRST CALIBRATION
See section LF1 of the technical parameters
5.1.3 USE
Switch on the equipment when all the parameters have been correctly configured and the equipment has been calibrated

  • Make sure than the value of the indicator, without load on the platform, is 0. If this is not the case, press .
  • Place the weight on the platform and the platform will show the weight.
  • The accumulation and sending of data will depend on the mode chosen in the section UF-6

5.2 PIECE COUNTING MODE
5.2.1 CONFIGURATION OF THE EQUIPMENT
See section LF2 of the technical parameters
5.2.2 FIRST CALIBRATION
See section LF1 of the technical parameters
5.2.3 USE
Switch on the equipment when all the parameters have been correctly configured. Make sure that the value on the visor, with no load on the platform, is 0. If this is not the case press the key .
STEPS TO FOLLOW;

  1. Press the key until the symbol PCS appears on the screen.
  2. Press the key successively to choose the quantity of pieces of the sample. On the screen will appear, successively, C10/C20/ C50/C100/C200.
  3. Place the sample on the platform, and wait until the sign of stability and press the key .
  4. Place the product on the platform and the screen will show the number of pieces.

The accumulation and the sending of data will depend on the mode chosen in the section UF-6.

  • To turn to the normal weighing mode, press the key
  • If the user wants to go back to the piece counting mode, using the same sample of reference, press the key  again.
  • If the user want to change the sample of reference, the user must repeat the steps described above.

5.3 LIMITS AND ALARM
The user can configure the superior and inferior limits of the sample placed on the platform.
The display will show if the sample is lower to the inferior limit Lo, above the superior limit Hi or in the zone between the two limits OK The user can configure when he wants to make the alarm ring and the stability needed to make it happen.
All the procedure is described in the section UF-2.
If the user wants to define the limits in the normal weighing mode and wants to use the limits in piece counting mode, he must define the new limits for this mode, when it changes to weight mode again, the user will recover the limits he already has. The same  happens otherwise.
5.4 UNIT RANGE / MULTI-RANGE / MULTI-INTERVAL
The indicator can be configured with an only range, a maximum weight and a value of step. It can also be configured as multi range or multi interval, in such cases there is a maximum weight.
From 0 to the medium weight of these maximum weight it is used the value of the chosen step ( step 1) and from the half to the maximum weight it is used the next value in the step ( step 2).
The screen indicators R1 and R2 point out the range which the user is using at every moment.
In the multi interval mode the weight increases, in the range use it used step 1, when the user goes to range 2 the step 2 is used.
When the weight decreases and the user go back to range 1, the step 1 is used again. On the contrary, in mode multi range, when the weight decreases and the device go back to range 1, the device continues using step 2 until it reaches 0.
In the section LF2 of the technical parameters the user can choose the range mode.

SETTINGS

PARAMETER DESCRIPTION
Internal calculation (A/D)

Limit Configuration of weight ( superior and inferior)
auto auto off
| Back illumination of display
Four modes of hold
RS-232 Output (PC/PRINT) (No function ABD)
| Configuration of the speed of the converter (A/D)
blind
Configuration of gravity
| Speed of filter
Stabilization speeds

PARAMETERS CONFIGURATION

7.1 INTERNAL COUNTING (A/D)
7.2 CONFIGURATION OF THE WEIGHT LIMITS (SUPERIOR AND INFERIOR)
7.3 AUTO SWITCH OFF
7.4 CONFIGURATION OF THE ILLUMINATION OF THE DISPLAY
7.5 HOLD FUNCTION
7.6 RS-232 DATA EXIT (NO FUNCTION ABD)
7.7 CONFIGURATION OF THE SPEED OF THE AD CONVERTER
7.8 BLIND
7.9 CONFIGURATION OF THE GRAVITY
7.10 SPEED OF FILTER
7.11 STABILIZATION SPEEDS

To access the configuration of parameters when the screen is in zero, the user must press at the same time the keys and . Press the key or  to choose the desired character (UF-1 ~ UF-11) To go back to the previous mode press the key .
7.1 INTERNAL COUNTING (A/D) |

  1. Press the key to view the internal sums of the scale.
  2. To go to the next parameter, press the key the screen will show the value of the voltage of the battery.
  3. To exit this mode and go back to the normal weighing, the user must press the key or .

7.2 CONFIGURATION OF THE LIMITS OF WEIGHT ( SUPERIOR AND LOWER) |

  1. Press the key to access the parameter.

  2. The display will show the message “ ” ( inferior limit)

  3. Use the keys and to move the cursor and the key  to choose the desired number.

  4. Press the key to confirm.

  5. The display will show the message “ ” (Superior limit= Hi)

  6. Use the keys and to move the cursor and press the key  to choose the desired number.
    Note: You can can stay over the last digit to modify if you wish.

  7. Press the Key to confirm.

  8. El display shows the value.
    A B C
    0 0 0
    (configuration of the alarm) DISPLAY| VALUE| ESTABILITY
    ---|---|---
    A| 0| There is no need to stabilize the alarm to make it ring
    1| The alarm must be stabilized to ring
    B| 0| Always 0
    C| 0
    1
    2| Alarm switched off
    The alarm ring if it is place on the band OK
    The alarm rings if it is situated below the Hi

  9. Press the keys and to move the cursor and the key to choose the desired number.

  10. Press the key to confirm.

7.3 AUTO SWITCH OFF |

MODES:

  • AoFF 00 – Auto switch off deactivated
  • AoFF 01 – Auto switch off activated in a minute. The scale is going to switch off automatically after 1 minute of not being used.
  • You can configure the value wished from 1 to 99 minutes.
  1. Press the key to have an access to the parameter.
  2.  Press the keys and to move the cursor and the key to choose the desired number.
  3. Press the key to confirm.

7.4 DISPLAY BACKLIGHTING |
MODES:

  • A: Automatic.
  • ON: Illumination Activated.
  • OFF: Illumination Deactivated
  1. Press the key to have an access to the parameter.
  2. Press the key to select the desired mode.
  3. Press the key to confirm.

7.5 HOLD FUNCTION |
(Once the object is retired from the plate, the display maintains the weight fixed until press the Key ,This function is very useful for the weighing of animals). Note; if you activate the hold mode, not can use the function limits and accumulation.

  1. Press the key to access to the parameter.
  2. Press the key to choose the desired mode.
  3. Press the key to confirm.

MODES:

  • HOLD 0 : Deactivated.
  • HOLD 1 : Animal in movement
  • HOLD 2 : Value of peak
  • HOLD 3 : Hold steady
  • HOLD 4 : Hold steady with self cancelling at zero.

HOLD 1.

  • When the user can access this parameter, the screen shows the message PCt,
  • Use the keys and to move the cursor and the key to choose the desired value of the range of HOLD, you can choose a number from 001 to 100.
  • Press the key to confirm.
  • It will appear on the screen the message time 8, use the keys and to move the cursor and the key to choose the number of times you want to repeat during the range of hold.
  • Press the key to confirm. Example: PCt small and big time means more accuracy and longer stabilization..

7.6 ACCUMULATION| (NO FUNCTION ABD)

FORMAT OF THE DATA

232 1
232 2
232 3| NO FUNCTION ABD
---|---
232 4
232 5
232 6| NO FUNCTION ABD
232 7| Connection sending and manual accumulation, passing by zero and pressing the key . When the display is set to 0, press the key twice and removed from weighing memory.
232 8| Connection sending and manual accumulation, passing by zero and pressing the key . When the display is set to 0, press the key twice and removed from weighing memory.
232 9| Connection sending and manual accumulation, passing by zero and pressing the key . When the display is set to 0, press the key twice and removed from weighing memory.
232 10| Connection sending and manual accumulation, passing by zero and pressing the key . When the display is set to 0, press the key twice and removed from weighing memory.

7.6.1 TRANSMISSION SPEED

  1. Press the key to access the parameter
  2. Press the key to choose the data exit mode wanted.
  3. Press the key to confirm.
    The screen shows the different possibility:** SPEED OF TRANSMISSIOIN**

    b1200| 1200 baudios
    b2400| 2400 baudios
    b4800| 4800 baudios
    b9600| 9600 baudios
    b19200| 19200 baudios
    b38400| 38400 baudios

  4. Press the key to choose the speed of transmission needed.
  5. Press the key to confirm.

7.7 CONFIGURATION OF THE SPEED OF THE CONVERTER|

  1. Press the key to access the parameter.

  2. Press the key to choose the desired mode:
    • Mode 1 : Normal
    • Mode 2: Fast
    • Mode 3: Slow

  3. Press the key to confirm.

7.8 BLIND|

It appears 0 on screen until the next division is selected. It starts to show values from that division.
Example:
Scale with e =2g
Blind in 5 divisions
It will show 0 until it reaches 25= 10 g, the first value it will show will be 12g

  1. Press the key to access the parameter
  2. Press the key to select.
  3. Press the key to confirm.

7.9 CONFIGURATION OF GRAVITY|

  1. Press the key to view the value of the actual gravity.
  2. To change the value, press the key , next you must use the key and to move the cursor and the key to select the desired number.
  3. Press the key to confirm.

CONFIGURATION OF THE READJUSTMENTS IN CALIBRATION

  1. When the user is in the normal mode of weighing, he must press the keys and , the message ECF-1 is going to appear on the display.
  2. Press the keys and to select the desired function: ECF-1, ECF-2 or ECF-3

* ECF-1 CALIBRATION OF ZERO + WEIGHT

Press the key , the display will show CALZ.
Press the key , to put the reading of the display to zero.
Press the keys and to move the cursor.
Press the key to introduce the value of the weight of calibration..
Place the weight of calibration on the platform and press the key to do the calibration once the reading is steady.
* ECF-2 CALIBRATION OF ZERO*
Press the key  , the display will show CALZ.
Press the key , to put the reading of the display to zero.
Press the key , to calibration.
ECF-3 CALIBRATION OF WEIGHT (SPAN)**
Press the key , the display will show the value of the weight of calibration.
Press the keys and to move the cursor.
Press the key to modify the value of the weight of calibration.
Press the key to confirm.
Place the weight of calibration on the platform and press the key to do the calibration once the reading is stable.

TECHNICAL PARAMETERS

DO NOT MODIFY THE TECHNICAL PARAMETERS IF IT IS NOT STRICTLY NEEDED. A BAD CONFIGURATION OF THIS SECTION CAN CAUSE A WRONG FUNCTIONING OF THE SCALE.
ENTRANCE AND EXIT OF THE CALIBRATION

DISPLAY DESCRIPTION AND SEQUENCE OF USE
With the visor switched off, press and hold the key and until the message

**** , appears on the screen, then you can release the keys
With the keys , and introduce the code .
Press the key to start or the key to exit the menu and the indicator will begin again auto- magically.

CALIBRATION OF THE WEIGHT

DISPLAY DESCRIPTION and SEQUENCE OF USE

Baxtran ABD Measuring Scales - icon89| The calibration can be done with any weight, but the weight ca not be inferior to 1/100 of the maximum capacity and it must also never be exceeded.
• Press the key to start the calibration of zero ( press  to exit the calibration and go back to the menu LF1)
• Use , and , 0 ~ 9 and then pto introduce the weight with which the calibration will be done. ( press the key to exit the calibration and go back to the menu LF1)
• Place the required weight on the scale as it is indicated in the display.
• Once everything is steady, press the key to calibrate it ( press the key ESC to exit the calibration and go back to the menu LF1).
THE CALIBRATION IS GOING TO FINISH AND THE SCALE WILL GO VACK TO THE WEIGHING MODE AUTOMATICALLY.

CONFIGURATION

FIRST STEP

DISPLAY DESCRIPTION AND SEQUENCE OF USE

FIRST STEP

| • DISPLAY OF THE INTERNAL COUNTING
SECOND STEP| • POSSIBLE VALUES OF THE PARAMETERS

A B C D E F

| | A: Metric system| 0:NO| 1: kg| 2:T| 3:g
---|---|---|---|---
B: American system| 0:NO| 1: lb| 2:lb oz| 3:VISS
C: other unities| 0:NO| 1: TW kg| 2:HK kg|
D: PCS| 0:OFF| 1: ON| |
E: double range| 0:OFF| 1: multi interval| 2:multi range|
F: units of calibration| | 1:use metric| 2:use American unities|

lb oz cannot be selected as unities of calibration.
The scale won’t let us continue up to the next step if there is a mistake during the programming.

| • Use , and , 0 ~ 9 and then press to introduce THE MAXIMUM CAPACITY.
•Use the Keys , to change the position of the DECIMAL POINT.
Baxtran ABD Measuring Scales - icon36
• Use to select the DIVISION: Baxtran ABD Measuring Scales -
icon38
| After introducing the parameters LF2, the indicator will show the last configuration saved. All the steps to follow must be completed, if they are not done the indicator will continue with the previous configuration.
Proceed with the calibration of weight after LF2.
• Press the key to start and display the internal value. Press to exit the menu and the scale will automatically restart.
• Press the key To continue with the configuration. (press the key to exit from the configuration and go back to the menu LF2).
Use the keys , and , 0 ~ 9 and then press the key to choose the UNITS OF WEIGHT . (press the key to exit from the configuration and go back to the menu LF2).
| • Use the keys , and , 0 ~ 9 and then press the key to choose the MAXI-MUM WEIGHT. press the key to exit from the configuration and go back to the menu LF2).
• Use the keys , and then To move the decimal point ( press the key To exit from the configuration and go back to the menu LF2).
| • Use and then the change the step.(Press the key to exit from the configuration and go back to the menu LF2).
• Use ,  and To continue with the other configuration or press the key to exit from the menu and the scale will begin again automatically).

LINEAL CALIBRATION
Up to six steps of calibration W0 ~W6
Make sure that the plate of the scale is empty before starting the calibration.
Press the key or the first point of calibration.
The previous points of calibration will be removed and the message CLEAR is going to appear on the screen.
Press the key to proceed with the next point of calibration.
Press the key to go back to the previous point of calibration.
Press the key to save everything.
Press the key to finish the calibration and go back to the menu LF3.

DISPLAY DESCRIPTION AND SEQUENCE OF USE
• Press the Key to start or the key to exit the menu and the scale will

start again auto-magically.

• Press the Key to calibrate the zero. ( press the key  to exit from the calibration and go back to the menu LF3)
• Place 1/3 of the maximum weight on the plate and press to calibrate 2/3 th the capacity. (press the key to exit from the calibration and go back to the menu LF3).

Baxtran ABD Measuring Scales - icon94| • Place 2/3 of the maximum weight on the plate and press to calibrate 2/3 th the capacity. (press the key to exit from the calibration and go back to the menu LF3).
• Place the maximum weight on the plate and press the key to calibrate the full capacity. (press the key to exit from the calibration and go back to the menu LF3).
| • Press the key to complete the lineal calibration. (press the key to exit from the calibration and go back to the menu LF3)
• Use the keys ,    and then the key to continue with other adjustments and press the key to exit from the menu and the scale will begin again automatically.

VELOCIDAD DEL CONVERTIDOR AD

SPEED 1 standard speed 15Hz.
SPEED 2 high speed 30Hz.
SPEED 3 low speed 7.5Hz
This function stays blocked when UF-5 is in mode HOLD 1.
The value of factory is 1

DISPLAY DESCRIPTION AND SEQUENCE OF USE
• Press the key to start or the key to exit from the menu and the scale is

going to begin again automatically.
• Use the key and then the key to select the speed of the AD converter ( press the key to exit from the configuration and go back to the menu LF4)
• Use the keys , and then . To continue with the other adjustment and press the key to exit from the menu and the scale will start again automatically.

BLIND

DISPLAY DESCRIPTION AND SEQUENCE OF USE
ZP 0 OFF

ZP 1 One division will not show being at zero
ZP 2 Two division will not show being at zero
ZP 3 Three divisions will not show being at zero
ZP 4 Four divisions will not show being at zero
ZP 5 Five divisions will not show being at zero
This function is going to be blocked when UF-5 is in mode HOLD 1
The value of factory is ZP 0
| • Press the key to start or the key to exit from the menu and the scale will start again automatically.
• Use the key and the key to select how many divisions will show when in zero ( press ESC to exit from the configuration and go back to the menu LF4)
• Use the keys , and then the key to continue with other adjustment or press the key to exit from the menu and the scale is going to start automatically.

APPROVAL CONFIRMITY

DISPLAY DESCRIPTION AND SEQUENCE OF USE
nonE Version not approved

DON’T MODIFY, IN NO WAY, THIS PARAMETER. IT MUST ALWAYS BE CONFIGURED AT nonE.
The change of this parameter implies the blocking of some functionalities.

GRAVITY
Introduce the gravity of your zone before doing the first calibration.
Introduce the gravity of destination after doing the calibration.
*The value of the gravity will be denied if it is bigger than 9.83217 ( gravity of the pole) or inferior to 9.78031 ( gravity of the equator).
Value of factory: 9.8035

DISPLAY DESCRIPTION AND SEQUENCE OF USE
• Press the key to continue or the key to exit from the menu and the scale

will start again automatically.
• The screen is going to show the number of pre-calibration during one second.
• Press the key to continue.
• Use the keys , and , 0 ~ 9 and then the key to introduce the value of gravity ( press ESC to exit from the configuration and go back to the menu LF7)

INITIAL ZERO

SETZY resets of the point of zero every time that the scale is begun again.
SETZN resets of the point zero OFF

DISPLAY DESCRIPTION AND SEQUENCE OF USE
• Press the key to start or the key to exit from the menu and the scale

will start again automatically.
• Use the key and then press the key  to choose the initial zero mode. (Press the  Key to exit of the configuration and go back to menu LF8)
• Use the keys , and then the key to continue with other adjustments  0 press the key to exit from the menu and the scale will start again automatically.

ERROR MESSAGES

MESSAGE DESCRIPTION
lnitial zero too high (over 10% of max. cap).

Baxtran ABD Measuring Scales - icon68| lnitial zero too high (over 10% of max. cap).
| Initial zero too high (approved model).
| Initial zero too low (less than 10% of max. cap).
| Initial zero too low (approved model).
| Unstable internal value.
| Internal code is not stable.
| Overloading
| Overloading
| Overloading
| Linearity correcting not well or linearity correcting cancel.
| Weight is too low.
| Internal code is too low.
| E2ROM abnormal.
| Total price is more than 999999.
| Battery volume is too low.
| Calibration failure, check the loadcell.

GUARANTEE

This scale has a warranty against all manufacture and material defects, for a period of a year starting with the delivery date.
During this period, GIROPES, will be in charge of the repairing of the scale.
This warranty does not include the damages done by overload or wrong use.
The warranty does not cover the delivery expenses necessary for the repair of the scale.

trade mark propriety of:
Pol. Empordà Internacional C/ Molló, 3
17469 VILAMALLA – (Girona) SPAIN
T. (34) 972 527 212 – F. (34) 972 527 211 Hosteleria10.com