PeakTech 6140/6150 Laboratory Power Supply User Manual

June 4, 2024
PeakTech

PeakTech logo.

Unser Wert ist messbar….PeakTech 6140 6150 Laboratory Power
Supply

PeakTech® 6140 / 6150
Operation manual
Laboratory Power Supply

Safety Precautions

This product complies with the requirements of the following directives of the European Union for CE conformity: 2014/30/EU (electromagnetic compatibility), 2014/35/EU  (low voltage), and 2011/65/EU (RoHS).
To ensure the safe operation of the equipment and eliminate the danger of serious injury due to short circuits (arcing), the following safety precautions must be observed.
Damages resulting from failure to observe these safety precautions are exempt from any legal claims.

  • Prior to the connection of the equipment to the mains outlet, check that the available mains voltage corresponds to the voltage setting of the equipment.
  • Connect the mains plug of the equipment only to a mains outlet with an earth connection.
  • do not place the equipment on damp or wet surfaces.
  • do not subject the equipment to direct sunlight or extreme temperatures.
  • do not subject the equipment to extreme humidity or dampness
  • replace a defective fuse only with a fuse of the original rating. Never short circuit fuse or fuse housing
  • not exceed the maximum permissible input rating.
  • Conduct measuring works only in dry clothing and in rubber shoes, i. e. on isolating mats.
  • comply with the warning labels and other info on the equipment
  • do not cover the ventilation slots of the cabinet to ensure that air is able to circulate freely inside.
  • do not insert metal objects into the equipment by way of the ventilation slots
  • do not place water-filled containers on the equipment (danger of short-circuiting in case of knock over of the container)
  • do not operate the equipment near strong magnetic fields (motors, transformer etc.)
  • do not subject the equipment to shocks or strong vibrations
  • keep hot soldering iron or guns away from the equipment
  • allow the equipment to stabilize at room temperature before taking up measurement (important for exact measurements)
  • do not modify the equipment in any way
  • do not place the equipment face-down on any table or workbench to prevent damaging the controls at the front.
  • Do not operate the meter before the cabinet has been closed and screwed safely as the terminal can carry voltage.
  • The measurement instrument is not to be operated unattended.
  • Periodically wipe the cabinet with a damp cloth and mid detergent. Do not use abrasives or solvents.
  • The meter is suitable for indoor use only.
  • Do not store the meter in a place of explosive, inflammable substance.
  • Opening the equipment and any service- and repair work must be performed by qualified service personnel. Repair work should be performed in the presence of a second person trained to administer first aid, if needed.
  • Power Supplies do not belong to children’s hands.
  • In order to extend the service life of the devices, let the devices cool down and switched off after 8 hours of use with full load.

Cleaning the cabinet
Prior to cleaning the cabinet, withdraw the mains plug from the power outlet. Clean only with a damp, soft cloth and a commercially available mild household cleaner. Ensure that no water gets inside the equipment to prevent possible shorts and damage to the equipment.

Technical Data

Input voltage: 115/230 V AC, 50/60 Hz 10% (switchable)
Fuse: 115 V – 4 AT / 230 V – 2 AT
Output voltage: P 6140: 0 – 30 V DC

P 6150: 0 – 30 V DC
5 V / 12 V (fixed outputs)
Output current:| P 6140: 0 – 5 A DC
P 6150: 0 – 5 A DC
0,5 A DC (fixed output)
Source regulation:| adjustable output: ≤ 1 x 10 -4 + 4 mV (I ≤  3 A)
≤ 1 x 10 -4 + 7 mV (I ≤  3 A)
≤  2 x 10 -3 + 5 mA (I ≤  3 A)
≤  2 x 10 -3 + 7 mA (I≤ 3 A) fixed outputs: ≤ 1 %
Ripple and noise: Adjustable output| CV ≤ 1 mV rms (I ≤3 A)
CV ≤ 2 mV rms (I ≤ 3 A)
CC ≤ 3 mA rms
fixed outputs: ≤ 10 mVrms
Protection:| current limit
Indication accuracy| Volt-indication: LED 1% + 2 digits
Amp-Indication: LED 2% + 2 digits
Dimensions (W x H x D)| 135 x 155 x 285 mm (P 6140)
170 x 155 x 285 mm (P 6150)
Weight| 6 kg (P 6140)
7 kg (P 6150)
Accessories| Power cord, operation manual

 Operation

3.1. Controls and description P 6140

PeakTech 6140 6150 Laboratory Power Supply - Controls

  1. Amp and Volt display: indicating output current and voltage

  2. Voltage fine-adjustment: fine-adjusting output voltage
    2a. Voltage coarse-adjustment: coarse-adjusting output voltage

  3. Current adjustment: adjusting the current-limit protection point

  4. Power switch: the unit is „ON“ when this button switch is depressed

  5. The output terminal (+): connecting the positive terminal of the load

  6. Case ground

  7. The output terminal (-): connecting the negative terminal of the load

  8. Output-switch: output voltage switchable

  9.  Constant-current indicator: the LED illuminates when the units are in the current regulated state

  10. Constant-voltage indicator: the LED illuminates when the units are in voltage regulated state

  11. 3.2. Controls and description P 6015 APeakTech 6140 6150 Laboratory Power Supply - output

  12. Amp and Volt display: indicating output current and voltage

  13. Voltage coarse-adjustment: coarse-adjusting output voltage

  14. Current adjustment: adjusting the current-limit protection point

  15. Power switch: the unit is „ON“ when this button switch is depressed

  16. The output terminal (+): connecting the positive terminal of the load

  17.  Case ground

  18. The output terminal (-): connecting the negative terminal of the load

  19. Output-switch: output voltage switchable

  20. Constant-current indicator: the LED illuminates when the units are in the current regulated state

  21. Constant-voltage indicator: the LED illuminates when the units are in voltage regulated state

  22. Output terminals: 12 V/0,5 A

  23. Output terminals: 5 V/ 0,5 A

3.3. Operating method

  1. When the adjustable output is used as CV output, first should rotate clockwise the CC adjustment (3.) to maximum, then turn on the power switch (4.), adjust CV adjustment (2.) till output voltage reach the required voltage value, at this time, the CC state indicator (9.) go out the CV state indicator (10.) light on.
  2. Used as CC output, after turning on the power switch (4.), first rotate clockwise the CV adjustment (2.) to maximum, while rotating counterclockwise the CC adjustment (3.) to a minimum, connect the required load, again adjust clockwise adjustment (3.) till output current reach the required current value. At this time, the CV state indicator (9.) goes out and the CC state indicator (10.) light on.
  3. Used as the CV output, in general, the CC adjustment (3.) should be set to maximum, but for these units, the current-limiting protection point can also be set arbitrarily. Setting procedure:
  • turn on the power, rotate counterclockwise the CC adjustment (3.) to a minimum
  • then make the positive and negative output terminal in short connection and rotate clockwise the CC adjustment (3.) till output current equal to the required current- limiting protection point, so the current-limiting protection point is well set, disconnect the short connection. 4. It is possible to turn off the output voltage without switching off the power supply with help of push-button “output”. So you can easily turn off the outputs and the test circuits are without tension.

Caution!

These units have excellent protection functions. The adjustable output has current-limit protection. As there is a controlling circuit for regulating the transistor’s power loss in the circuit, when short-circuit occurs, the power loss on large power transistors is not very high, it can’t cause any damage to the unit. But there is still power loss when short-circuit, in order to reduce aging and energy consumption, this situation should be found as soon as possible and turn off the power, then exclude the faults.
When operating is finished, put it in a dry place of good ventilation, and keep it clean. If it is not in use for a long period, pull off the power supply plug for storage.
For maintenance, the input voltage must be cut off.
All rights, also for translation, reprinting and copy of this manual or parts are reserved. Reproductions of all kinds (photocopy, microfilm or other) only by written permission of the publisher. This manual is according the latest technical knowledge. Technical changings which are in the interest of progress, are reserved. We herewith confirm that the units are calibrated by the factory according to the specifications as per the technical specifications.
We recommend calibrating the unit again, after 1 year. © PeakTech ® 09/2021 MP/Ehr.

PeakTech Prüf- und Messtechnik GmbH
– Gerstenstieg 4 – DE-22926 Ahrensburg / Germany
+49-(0) 4102-97398 80 +49-(0) 4102-97398 99
 info@peaktech.de www.peaktech.de