AZ INSTRUMENT 84131 Long Cable Dissolved Oxygen Pen
Product Information
The Long Cable Dissolved Oxygen Pen is a high-quality instrument designed for
measuring dissolved oxygen levels in various applications. It features a
patented design with a 3-meter long cable, providing flexibility and
convenience during measurements.
The pen comes in two models: 84131 with a floating probe and 8413 with an
anti-collision cap probe.
This pen utilizes a high accuracy double precious metal D.O. probe, ensuring
precise and reliable readings. It is equipped with automatic temperature
compensation, manual salinity and barometric pressure compensation, and a
high-sensitivity membrane set for
improved accuracy.
The product has a built-in rechargeable battery that can be easily charged using the included Type C charging cable. With just 1.5 hours of charging, it provides approximately 60 hours of continuous use.
It comes with a full package that includes the D.O. Meter, D.O. Probe with a 3-meter cable, spare dissolved oxygen membrane set, electrolyte, charging cable, manual, and a hard black carrying case for convenient storage and transportation.
Product Usage Instructions
- No need to remove the anti-collision cover of the dissolved oxygen head before use.
- The product comes pre-installed with electrolyte.
- To turn on the pen, simply short press the power button. Customizable parameters can be set to improve accuracy.
- Allow the pen to undergo 5 minutes of polarization time after the first boot for better stability.
- Prior to each measurement, place the probe in the air for 100% saturated oxygen calibration. This calibration ensures optimalautomatic temperature compensation.
- Regularly check the reading at the same spot and time to obtain the best comparison results.
- If you are a frequent user, consider replacing the membrane set and injecting electrolyte if there is a significant difference between the measured value and the previous measurement, even if it is less than half a year.
- When using the membrane cover, make sure to screw it tightly until the membrane and the metal electrode are in close contact. Avoid trapping air bubbles between the membrane and metal.
- For power supply, unscrew the battery cover and use the included charging cable to charge the built-in rechargeable battery. A 1.5-hour charge provides approximately 60 hours of continuous use.
- The pen features a LCD display that shows various information such as battery level, calibration status, reading lock, dissolved oxygen value, atmospheric pressure unit, salinity unit, dissolved oxygen unit, water temperature, and units.
- The keypad operation includes short and long presses for different functions such as turning on/off, locking the reading, switching display units, entering calibration mode, adjusting values, and confirming settings.
QUICK GUIDE
- No need to remove the anti-collision cover of the dissolved oxygen head before use.
- This product has been pre-installed with electrolyte.
- Short press the power button to turn on. Setting parameters can be customized to improve accuracy.
- Please allows 5 minutes polarization timeafter the first boot, the stability of the product will be better.
- Before each measurement, place the probe in the air for 100% saturated oxygen calibration. The calibration done at this point allows automatic temperature compensation to work best.
- Check the reading at regular time and regular spot can provide you the best comparison result.
- Maintenance after a period of use is simple. Shake the probe slightly and visually check the amount of electrolyte. If there is almost no liquid, unscrew the membrane set and drop sufficient electrolyte. Provided two spare membrane sets are enough for one year. Generally speaking, add liquid and replace the membrane set once every six months is fine, frequently unscrewing the membrane set and adding electrolyte will not help the performance. While drop electrolyte, fill about 3/4 full at a time, If having spillage after installation, just wipe it dry.
- If you are a frequent user, even if it is less than half a year, when the difference between the measured value and the previous time is too large, please replace the membrane set and inject electrolyte.
- Put on the membrane cover and screw it tightly until the membrane and the metal electrode are in close contact. No air bubbles trapped between the membrane and metal
INTRODUCTION
Thank you for purchasing this long cable dissolved oxygen pen Please read the manual in detail and keep it for future use. Proper use and maintenance of this product can result in accurate measurements and extended service life.
Features:
- 84131 equipped with floating probe to ensure your measure result is from water not from sand
- Large screen with adjustable backlight brightness.
- The pen type is easy to carry, suitable for all places.
- Accurate dissolved oxygen and temperature readings.
- Low battery indicator, charged through USB type cable. 4% and mg/L and °C/°F switchable.
- One key to do 100% D.O. calibration in the air, chemicals is not needed.
- Easy to maintain and high-sensitivity membrane set.
- Built-in automatic temperature compensation
- Manual salinity and barometric pressure compensation
- High accuracy double precious metal D.O. probe
POWER SUPPLY
This product has a built-in rechargeable battery. Please unscrew the battery cover, use the included charging cable to charge. About 60 hours of continuous use after 1.5 hours of charging.
This enclosed cable is for charging, no data transter unction.
MATERIAL SUPPLIED
The full package contains:
- D.O. Meter (built-in battery) × 1
- D.O. Probe (w/ 3m cable,84131: floating
- probe, 8413: anti-collision cap probe)
- Spare dissolved oxygen membrane set × 2
- Electrolyte ×1
- Type C charging cable ×1
- Manual x1
- Hard black carrying case x 1
LCD DISPLAY
KEYPAD OPERATION
Note: Green text keypad means press longer time
- Short press to turn on, when no key operation for 20 minutes, it will automatically turn off
- In the power-on state, short press to lock the reading
- In power-on state, long press to power off
- If you don’t want to have the function of automatic shutdown function, press and hold the power button when turning it on until the word “n” appears on LCD
- In the setting state, press this key to confirm setting
- In the power-on state, short press to switch the display unit of dissolved oxygen. It can be mg/L or %
- In the power-on state, long press to enter calibration
- In the setting state, press to adjust value upward
- In the power-on state, short press to turn on the backlight of LCD, and short press to turn backlight off
- In the power-on state, long press to enter the setting mode for customized settings
- In the setting state, press to adjust value downward
OPERATION
Step 1: Charge and check probe status
This product has a built-in rechargeable battery. Unscrew the battery cover on
the head, and use the included charging cable to charge. About 60 hours of
continuous use after 1.5 hours charging time. On 8413, the anti-collision
cover of the D.O. probe head has protective function and no need to remove
before use. This product is pre-filled with electrolyte. After the product is
charged, it is ready to use. Before each regular use, shake the probe gently
and visually inspect the electrolyte inside of the probe is enough.
Step 2: Power on the meter
Short press the “Power” button to turn on the LCD; if you need to turn it
off,long press the “Power” button.
Step 3: Calibration
The instrument has been well calibrated at the factory, but calibration is
still necessary before each use to get best result. Simply do calibration in
air, chemical powder is not required. See the Calibration section for
calibration details.
Step 4: Get D.O. Value
Immerse the electrode in the test solution and shake gently. You will see the
dissolved oxygen and temperature from the LCD. For 84131, the patented design
floating probe can ensure you get measurement from water, probe won’t sink
into sand. When there is no key operation for 20 minutes, it will
automatically shut down. The probe cable is 3 meters long, so it can be used
even in deep pools.
Step 5: switch D.O. Unit
The display is preset to %, you can short press the key to switch to mg/L
Step 6: Backlight for dark places
To make it easier to see readings in dim areas, press key.When the backlight
is not needed, press again to turn it off.
Step 7: Freeze the reading (Hold)
In order to make the hand recording work is easier, press the key to
temporarily lock the reading value. Onte¢Re handwriting job is done, press
again to dismiss.
D.0 100% SATURATION CALIBRATION IN AIR
In order to get accurate readings, we recommend that before each operation and
after replacing the membrane module, saturated dissolved oxygen calibration is
required. No need to remove the anti-collision cover.
Step 1: In normal mode, convert dissolved oxygen units to % and place the
probe in the air, wait a few minutes till the reading value is stable.
Step 2: Long press key to do 100% saturation calibration. “CAL” will
flash on the screen
Step 3: Wait for a few seconds to 1 minute, see that the reading is
stable and CAL no longer flashes, short press to complete the calibration.
Step 4: After calibration, the displayed value should be 99.0%~101.0%. If an error occurs during calibration, this may cause by low electrolyte or a bad probe condition; see next section to learn how to maintain a dissolved oxygen probe.
PROBE MAINTENANCE
Visual Check
Check the probe when you receive a new probe or see strange readings or if the
probe has been in use for six months.
- There is no problem if the electrolyte is not full. While it is less than 1/2 full, fill up the electrolyte.
- Be sure to check the middle part of the probe surface,
The work area should be kept clean as shown on the right photo
The best condition of the probe is the electrode reaction surface is clean and there is no air bubbles attached.
How to replace membrane set
Replace the membrane pack when any of the following occurs:
- a. The probe has been used for more than half a year.
- b. There is an error code E1/E3, but the central working area of the probe is still clean.
- c. The membrane is damaged or wrinkled.
- d. The measurement speed is much slower than before.
Steps:
1. Prepare a new membrane set. The structure of this membrane set is sensitive, not recommendec to reuse the membrane set.
2. Rotate to remove the membrane set from the probe
3. Slowly drop the electrolyte into the new membrane set until it is 3/4 full.
4. Slowly drop the electrolyte into the new membrane set until it Screw the new membrane set onto the probe.
5. After installing the membrane set, it is fine to found small air bubbles in the electrolyte. But, the work area in the middle should be free from bubbles.
6. 100% saturated calibration in air after membrane replacement
Note: Under normal use, the membrane set included with the goods should be enough for the life time of probe. It is not necessary to replace membrane set too frequently.
SETUP
This meter has an advanced setting mode that allows you customize the
parameters. If want to change the parameters, press key >2 seconds to enter
the setting mode.
The first setting you see is the salinity value. If you want to select other
parameters, you can press “ ” to select
When you see the parameter you want to set, use” ” and to adjust the value up
and down. After adjustment, press” ” to confirm.
TROUBLESHOOTING
- Already press power key but no display
- Make sure you have pressed power key more than 0.1 second
- Charging the battery first and then try again.
- Can’t turn off the meter
- Short press power key could not turn of the meter. Please press power key longer time to turn off the meter.
- Too big difference from previous measurement
Fill the electrolyte and replace the membrane according to the maintenance method of the probe. After replacement, please do 100% air saturated dissolved oxygen calibration.
D.O. value
E01: Indicates that the sensing element is damaged. Sensing element is a
consumable and cannot be repaired
E03: The signal strength is too high, it is normal to appear for few
seconds when the power is turned on, but if E03 continues, put probe in the
air to re-calibrate. If no improvement, fill the electrolyte and replace
membrane. If still invalid, sensing element is damaged.
E21: The difference between the current measured temp. and the
calibration temp. is more than 15 degreesC, please do calibration again
Temp. value
E01: Indicates that the sensing element is damaged, the sensing element
is a consumable and cannot be repaired
E02: The temperature measured at the moment is too low, put it in the
environment of 25 °C for 5 minutes, if there is no improvement, the sensing
element is damaged, the# sensing element is consumable and cannot be repaired
E03: The temperature measured at the moment is too high, put it in the
environment of 25 °C for 5 minutes, if it does not improve, the sensing
element is damaged, the sensing element is consumable and cannot be repaired.
SPECIFICATION
WARRANTY
The meter is warranted to be free from defects in material and workmanship for a period of one year from the date of purchase. This warranty covers normal operation but does not cover battery, misuse, abuse, alteration, tampering, neglect, improper maintenance, or damage resulting from leaking batteries. Proof of purchase is required for warranty repairs. Warranty is void if the meter used to be taken apart.
RETURN AUTHORIZATION
Authorization must be obtained from the supplier before returning items for any reason. When requiring a RA (Return Authorization), please include data regarding the defective reason, the meters are to be returned along with good packing to prevent any damage in shipment and insured against possible damage or loss.
Accuracy, the Zenith of Measuring / Testing Instruments !
- Hygrometer/Psychrometer
- Thermometer
- Anemometer
- Sound Level Meter
- Air Flow meter
- Infrared Thermometer
- K type Thermometer
- K.J.T. type Thermometer
- K.J.T.R.S.E. type Thermometer
- pH Meter
- Conductivity Meter
- T.D.S. Meter
- D.O. Meter
- Saccharimeter
- Manometer
- Tacho Meter
- Lux / Light Meter
- Moisture Meter
- Data logger
- Temp./RH transmitter
- Wireless Transmitter
More products available !
Read User Manual Online (PDF format)
Read User Manual Online (PDF format) >>