AEMC Spring Loaded 25 A Kelvin Probes User Manual
- June 12, 2024
- AEMC
Table of Contents
- INTRODUCTION
- Symbols and Definitions
- Definition of Measurement Categories (CAT)
- PRECAUTIONS FOR USE
- ORDERING INFORMATION
- PRODUCT FEATURES
- OPERATION
- SPECIFICATIONS
- MECHANICAL SPECIFICATIONS
- ELECTRICAL SAFETY
- MAINTENANCE
- REPAIR
- TECHNICAL AND SALES ASSISTANCE
- LIMITED WARRANTY
- References
- Read User Manual Online (PDF format)
- Download This Manual (PDF format)
AEMC Spring Loaded 25 A Kelvin Probes User Manual
Copyright© Chauvin Arnoux®, Inc. d.b.a. AEMC® Instruments. All rights reserved.
No part of this documentation may be reproduced in any form or by any means (including electronic storage and retrieval or translation into any other language) without prior agreement and written consent from Chauvin Arnoux®, Inc., as governed by United States and International copyright laws. Chauvin Arnoux®, Inc. d.b.a. AEMC® Instruments
15 Faraday Drive • Dover, NH 03820 USA
Tel: 800-945-2362 or
603-749-6434 • Fax:
603-742-2346
This documentation is provided “as is,” without warranty of any kind, express,
implied, or otherwise. Chauvin Arnoux®, Inc. has made every reasonable effort
to ensure that this documentation is accurate; but does not warrant the
accuracy or completeness of the text, graphics, or other information contained
in this documentation. Chauvin Arnoux®, Inc. shall not be liable for any
damages, special, indirect, incidental, or inconsequential; including (but not
limited to) physical, emotional or monetary damages due to lost revenues or
lost profits that may result from the use of this documentation, whether or
not the user of the documentation has been advised of the possibility of such
damages.
INTRODUCTION
Thank you for purchasing AEMC® Instruments Spring Loaded 25 A Kelvin Probes. This test probe is designed for use with AEMC® Instruments Micro-ohmmeters Model 6240 and 6255 or equivalent.
For best results from your instrument and for your safety, read the enclosed operating instructions carefully and comply with the precautions for use. Only qualified and trained operators should use this product.
Symbols and Definitions
| CAUTION – Risk of Danger! Indicates a WARNING. Whenever this symbol is
present, the operator must refer to the user manual before operation
---|---
| Indicates Important information to acknowledge
| This product complies with the Low Voltage & Electromagnetic Compatibility
European directives (73/23/CEE & 89/336/CEE)
|
The UKCA marking certifies that the product is compliant with the requirements
that apply in the United Kingdom, specifically regarding Low-Voltage Safety,
Electromagnetic Compatibility, and the Restriction of Hazardous Substances
| In the European Union, this product is subject to a separate collection
system for recycling electrical and electronic components in accordance with
directive WEEE 2002/96/EC
Definition of Measurement Categories (CAT)
CAT IV:
Corresponds to measurements performed at the primary electrical supply (< 1000
V). Example: primary overcurrent protection devices, ripple control units, and
meters.
CAT III:
Corresponds to measurements performed in the building installation at the
distribution level. Example: hardwired equipment in fixed installation and
circuit breakers.
CAT II:
Corresponds to measurements performed on circuits directly connected to the
electrical distribution system. Example: measurements on household appliances
and portable tools.
PRECAUTIONS FOR USE
The protection offered by this accessory is compromised if it is used in unapproved, unspecified, or improper ways. These safety warnings are provided to ensure the safety of personnel and proper operation of the instrument.
- Observe the maximum voltage, intensity, and measurement category. Do not use the accessory on networks where the voltage or category exceeds those stated in this manual.
- Observe the conditions of use (e.g. temperature, humidity, altitude, degree of pollution and location of use).
- Do not use the accessory if it is open, damaged, or incorrectly reassembled.
- Any repairs must be carried out by trained and qualified personnel using appropriate safety precautions.
ORDERING INFORMATION
Kelvin Probes Pistol Grip, 10 ft, 25 A Spring Loaded: Cat. #2118.84
Kelvin Probes Pistol Grip, 20 ft, 25 A Spring Loaded : Cat. #2118.85
PRODUCT FEATURES
The pistol-grip spring loaded 25 A test probes can measure low resistances very accurately using micro-ohmmeters with (.157 to .236) in (4 to 6) mm diameter binding posts or .157 in (4 mm) diameter safety terminals.
-
The pistol-grip spring loaded test probes ensure good contact at the measurement point even if the surface is uneven.
-
Each 25 A test probe has very low resistance.
-
The measurement is made in 4-wire mode, so lead resistance compensation is not necessary.
NOTE: At the other end of the leads, there are .157 in (4 mm) diameter safety terminals that spaded lugs .236 in (6 mm) diameter screw into. These probes can be used with micro-ohmmeters having either binding posts from (.157 to .236) in (4 to 6) mm in diameter or safety terminals .157 in (4 mm) in diameter.
SPRING LOADED PISTOL 25 A KELVIN PROBE
OPERATION
-
Connect the two spring loaded 25 A test probes to the four terminals of the instrument.
-
Loosen the nuts on the binding posts. Insert the spaded lugs while carefully connecting the red leads to the voltage terminals (P). Then, tighten the nut.
-
If the instrument has safety terminals, unscrew the spaded lugs to make the connection
For details of the measurement configuration, technical specifications and error messages, refer to the user manuals for the meters being used.
SPECIFICATIONS
ENVIRONMENTAL CONDITIONS
Indoor use
- Operating Range: (14 to 131) °F (-10 to 55) °C and (10 to 85) % RH
- Storage Range: (-40 to 158) °F (-40 to 70) °C and (10 to 90) % RH
- Degree of Pollution: 2
- Altitude: < 6,561.68 ft (2000 m)
- Altitude Storage: < 32,808.4 ft (10,000 m)
MECHANICAL SPECIFICATIONS
Pistol-grip Kelvin Probe
Dimensions (L x H x D):
- Grip (4.25 x 1.57) in (108 x 40) mm
- probe (6.06 x 1.18) in (154 x 30) mm
- thickness 1.1 in (28 mm)
- Diameter of the probes: .079 in (2 mm)
- Center-to-center distance between the probes: .138 in (3.5 mm)
- Travel of the probes: .276 in (7 mm)
- Lead: 10.33 ft (3.15 m) long
- Spaded lugs: for diameters from (.157 to .236) in (4 to 6) mm
- Safety terminals: diameter of .157 in (4 mm)
- Weight: 10 ft: Set of two: (2 x 14.82) oz (2 x 420) g
- 20 ft: Set of two: (2 x 29.64) oz (2 x 840) g
- Resistance: approximately (2 x 25) mΩ
- Protection index: IP 40 according to IEC 60 529
- IK 04 according to IEC 50102
- Drop test: According to IEC 61010-1
ELECTRICAL SAFETY
Electrical safety according to IEC 61010-031, IEC 61557-4, IEC 61557-14
- Maximum current: 25 A for 1 min
- Maximum voltage: 300 V CAT III
MAINTENANCE
Use only factory specified replacement parts. AEMC® Instruments will not be held responsible for any accident, incident, or malfunction following a repair completed other than by its service center or by an approved repair center.
CLEANING
WARNING: Disconnect the accessory completely
- Use a soft cloth lightly dampened with soapy water.
- Rinse with a wet cloth and then dry with a dry cloth.
- Do not submerge the instrument in water.
- Do not use alcohol, solvents, or hydrocarbons.
REPAIR
For instrument repair:
You must contact our Service Center for a Customer Service Authorization
Number (CSA#). This will ensure that when your instrument arrives, it will be
tracked and processed promptly. Please write the CSA# on the outside of the
shipping container.
Ship To: Chauvin Arnoux®, Inc. d.b.a. AEMC® Instruments 15 Faraday Drive
Dover, NH 03820 USA
Phone: 800-945-2362 (Ext. 360)
603-749-6434 (Ext. 360)
Fax: 603-742-2346 or
603-749-6309
E-mail: repair@aemc.com
(Or contact your authorized distributor)
NOTE: You must obtain a CSA# before returning any instrument.
TECHNICAL AND SALES ASSISTANCE
If you are experiencing any technical problems or require any assistance with the proper operation or application of your instrument, please call, mail, fax, or e-mail our technical support team:
Chauvin Arnoux®, Inc. d.b.a. AEMC® Instruments
Phone: 800-343-1391 (Ext. 351)
Fax: 603-742-2346
E-mail: techsupport@aemc.com
www.aemc.com
LIMITED WARRANTY
The instrument is warrantied to the owner for a period of two years from the date of original purchase against defects in manufacture. This limited warranty is given by AEMC® Instruments, not by the distributor from whom it was purchased. This warranty is void if the unit has been tampered with, abused, or if the defect is related to service not performed by AEMC® Instruments.
Full warranty coverage and product registration is available on our website at www.aemc.com/warranty.html
Please print the online Warranty Coverage Information for your records.
What AEMC® Instruments will do:
If a malfunction occurs within the warranty period, you may return the
instrument to us for repair, provided we have your warranty registration
information on file or a proof of purchase. AEMC® Instruments will repair or
replace the faulty material at our discretion.
REGISTER ONLINE AT: www.aemc.com/warranty.html
Warranty Repairs
What you must do to return an Instrument for Warranty Repair:
First, send an email to repair@aemc.com requesting a Customer Service Authorization Number (CSA#) from our Service Department. You will be provided a CSA Form and other required paperwork along with the next steps to complete the request. Then return the instrument along with the signed CSA Form. Please write the CSA# on the outside of the shipping container. Return the instrument, postage or shipment pre-paid to:
Chauvin Arnoux®, Inc. d.b.a. AEMC® Instruments
15 Faraday Drive, Dover, NH 03820 USA
Phone: 800-945-2362 (Ext. 360)
603-749-6434 (Ext. 360)
Fax: 603-742-2346
E-mail: repair@aemc.com
Caution: To protect yourself against in-transit loss, we recommend that you insure your returned material.
NOTE: You must obtain a CSA# before returning any instrument.
Costumer Support
AEMC® Instruments
15 Faraday Drive • Dover, NH 03820 USA
Phone: 603-749-6434 •
800-343-1391 • Fax:
603-742-2346
www.aemc.com
References
- Get Electrical Testers – Power Analyzers | AEMC Instruments
- AEMC Warranty Registration
- Get Electrical Testers – Power Analyzers | AEMC Instruments
- AEMC Warranty Registration
Read User Manual Online (PDF format)
Read User Manual Online (PDF format) >>