PROTEOR MATIK Polycentric Prosthetic Knee Joint Instructions

June 12, 2024
PROTEOR

PROTEOR-LOGO

PROTEOR MATIK Polycentric Prosthetic Knee Joint

PROTEOR-MATIK-Polycentric-Prosthetic-Knee-Joint-PRODUCT-
IMG

COMPONENTS INCLUDED

Designation Ref. Included/Sold separately
Knee 1P200(-KD) Included
One-part cosmetic cover 1G18 Sold separately

DESCRIPTION, PROPERTIES AND MECHANISM OF ACTION

PROTEOR-MATIK-Polycentric-Prosthetic-Knee-Joint-FIG-1

Description

  • A4-axis pneumatic knee available in two different versions:
  • 1P200: Male pyramid proximal adapter, suitable for patients who have had a transfemoral amputation or hip disarticulation.
  • 1P200-KD: Threaded proximal adapter (M36×1.5), suitable for patients with a long amputation stump or who have had a knee disarticulation.
  • For both versions, the distal connection is made via a male pyramid connector.

Properties

Ref. 1P200 1P200-KD
Weight 743 g
Maximum flexion 170°

Maximum patient weight

(Including carried load)

| 125 kg

PROTEOR-MATIK-Polycentric-Prosthetic-Knee-Joint-FIG-2

This device has undergone a three-million-cycle test, in accordance with standard NF EN ISO 10328 at a P6 load level (i.e. 125 kg), which corresponds to a service life of 4 to 5 years, depending on the patient’s mobility level.

Mechanism of action

  • When in extension, the system of linkages ensures that the instantaneous centre of rotation is to the rear of the load line, which keeps the knee stable during the stance phase. The speed of every flexion and extension movement can be adjusted independently.
  • The extension assist is not adjustable.

INTENDED USERS/INDICATIONS

This medical device is supplied to healthcare professionals (prosthetists/orthotists) who then instruct the patient in how to use it. The prescription is drawn up by a doctor. The doctor works with a prosthetist/orthotist to assess whether the patient is suited to using the device. This device should be used on ONE PATIENT ONLY. It must not be reused on other patients.

  • This device is exclusively intended to be used as prosthetic equipment for patients who have undergone transfemoral amputation, hip disarticulation or knee disarticulation. It is specifically recommended for active or slightly active patients (L2/L3) and allows them to perform everyday activities.
  • Maximum weight (including carried load): 125 kg.

Not suitable for children

The maximum knee flexion is 170°. This flexion may, however, be limited by the volume of the socket or by the cosmetic covering.

CLINICAL BENEFITS

The device allows:

  • Everyday mobility with minimum maintenance
  • The patient to choose a lighter and smaller knee.
  • Flexion of up to 170° for sitting or kneeling.
  • Good clearance between the foot and the ground in the swing phase, to avoid contact.
  • High levels of stability and safety in the stance phase thanks to the 4-axis joint geometry.

ACCESSORIES AND COMPATIBILITIES


Ref.

| ****

1P200

| ****

1P200-KD

---|---|---



Proximal adapter

| ****

Anchor

| 1K179+ 1K183

1K173 or 1K176

1K177 (max. weight 100 kg)

| ****

1K179


Connectors

| ****

1K172 or 1K207-HD

| ****

1K184


Distal adapter

| Tube and connector| 1D41-HD and 1G01-HD

1D52-P6

ASSEMBLY AND FITTING TO THE PATIENT

Aligning

Static alignment:

  • In the sagittal plane, the load line passes through the greater trochanter and between 0 and 10 mm in front of the axis of the knee.
  • Align the foot in compliance with the manufacturer’s instructions.
  • In the frontal plane, the load line passes through the middle of the socket, through the middle of the knee and between the first and second toes.

INSTRUCTIONS FOR USE

PROTEOR-MATIK-Polycentric-Prosthetic-Knee-Joint-FIG-5

Dynamic alignment

PROTEOR-MATIK-Polycentric-Prosthetic-Knee-Joint-FIG-5

  1. To increase the stability, incline the knee as indicated (to move the instantaneous centre of rotation to the rear) by adjusting the proximal and distal connectors. To reduce the stability, incline the knee in the opposite direction.
  2. Then reposition the foot, in accordance with the recommended alignment, by adjusting the connectors between the knee and the foot.

The threaded M36×1.5 adapter for the KD version does not accommodate any alignment adjustment. The socket must therefore be made accordingly.
Check the alignment using suitable equipment (line laser, plumb line, etc.)

Adjusting the walking pace

On delivery, the knee is set up for walking at medium pace. For patient safety reasons, the prosthetist/orthotist must begin the adjustment process with the knee set to its factory settings and with the patient walking between parallel bars.

  • The model of foot used may have an effect on the adjustment of the knee. The knee should be readjusted whenever the foot is changed.
  • Prioritise patient safety in all circumstances, particularly to ensure that slopes and stairs can be negotiated safely.
  • After performing any maintenance the knee must be readjusted.
  • To adapt the knee to the patient’s pace of walking, adjust hex set screws E and F using an Allen (hex) key.
  • It is advisable to perform these adjustments in the following order:

Adjusting the Flexion (F)

  • Turn hex set screw F: Screw in the set screw to reduce the angle of flexion, and unscrew to increase it.

Adjusting the Extension (E)

  • Turn hex set screw E: Screw in to increase the cushioning of the impact when the knee reaches full extension.

Once these adjustments have been made, ensure that the knee returns to full extension when walking slowly to ensure that there is no risk of the patient falling.

Finishing
It is advisable to reinforce the cosmetic cover in front of the joint to prolong its service life.

Maximum flexion stop
When the patient is crouching or kneeling, the socket must come into contact with the rear flexion limit on the rear linkage. It must not touch the rear linkage, or the plate, since this might damage or destroy the knee.

The cosmetic cover may modify the initial knee settings and limit certain movements. It is advisable to always check the operation of the knee, once the cosmetic cover has been fitted.

DETECTING ANY MALFUNCTIONING
If you notice that the device is behaving abnormally or if its characteristics feel different, or if it has been badly knocked, consult your prosthetist/orthotist.

WARNINGS, CONTRAINDICATIONS, SIDE EFFECTS

Warnings

  • Caution, the patient may need to take time to get used to a new knee.
  • To reduce the risk of falling when going down stairs, it is advisable to hold on to handrail.
  • So as not to damage the knee, do not apply talcum powder to eliminate rubbing noises – use silicon spray instead. Talc degrades the mechanical components, which might cause a malfunction and possibly a fall.
  • PROTEOR disclaims all liability if talcum powder is used.
  • It is possible to trap a finger or pinch clothing in the knee joint. To avoid the any injury caused by the movement of the joint, make sure that no-one’s hand or fingers is near the mechanism.
  • The knee is resistant to bad weather, but needs to be dried immediately if it does get wet. Any water that gets into the pneumatic mechanism may lead to the deterioration of the extension assist spring and thus the destruction of the knee. It must, therefore, be protected against any ingress of water

Contraindications

  • The following are strictly prohibited: removing the covers, screwing in or unscrewing any screws on the knee with the exception of the two adjustment screws E and F.
  • Never lubricate the knee pins; this can cause them to degrade rapidly.
  • The knee has been designed for patients with a maximum weight of 125 kg (including carried load).
  • The warranty does not cover damage or degradation caused by misuse, unsuitable alignment, use without suitable protection in a very dusty environment or any other inappropriate use.
  • The knee must not be exposed to environments which might induce corrosion of the metal components (fresh water, sea water, chlorinated water, acids, etc.).
  • Showering or taking a bath while wearing the prosthesis is prohibited; this might impair its properties and its correct operation.
  • Never leave this device near a source of heat since it might burn or release toxic fumes.
  • The use of solvents is prohibited.

Side effects

  • There are no side effects directly associated with the device.
  • Any serious incident that occurs which relates to the device must be reported to the manufacturer and to the competent authority of the member state in which the incident occurred.

MAINTENANCE, STORAGE, DISPOSAL AND SERVICE LIFE

Maintenance/cleaning

  • You can clean the knee using a damp sponge
  • Do not immerse the knee or hold it under running water
  • Dry you knee immediately if it is exposed to bad weather or accidental splashing.
  • The stops may need to be replaced:
  • Spare stop: EE014 (2 are required)

Storage

  • Service and storage temperature: -10°C to +40°C
  • Relative humidity of the air: no restrictions

Disposal

  • The various components of this device are considered as special waste: elastomer, plastic, aluminium, titanium, steel and brass. They must be treated in accordance with current, applicable legislation.

Service life

  • It is advisable to arrange for a prosthetist/orthotist to check the product once a year.

DESCRIPTION OF THE SYMBOLS

PROTEOR-MATIK-Polycentric-Prosthetic-Knee-Joint-FIG-9 REGULATORY INFORMATION

  • This product is a CE-marked medical device that is certified as conforming with regulation (EU) 2017/745.