PROTEOR RUSH Foot EVAQ8 Collection Vacuum Suspension Instruction Manual

June 4, 2024
PROTEOR

Vacuum Suspension
Instruction Manual

SAFETY

  • This product is only to be used in lower limb prosthetic devices.
  • This system is intended to produce elevated vacuum suspension for lower limb amputees already fit with an appropriate vacuum suspension socket. The socket should have a good fit with no voids. This is to help prevent blisters and to prevent tissue from being drawn into the voids or poorly fit areas.
  • This product should only be fit by trained prosthetic professionals.
  • The vacuum pump heel, socket valve, and associated barbs & tubing should be inspected often for signs of damage or excessive wear.
  • Ensure that the tubing is properly secured at all times.

CONTENTS AT ARRIVAL

  1. Vacuum pump heel
  2. Straight barb
  3. Exhaust filter
  4. Tubing
  5. Inline filter
  6. Socket right angle barb and housing
  7. Tube securing hook and loop tape

FUNCTION

The RUSH FOOT EVAQ8 is a product designed to provide the user with an elevated vacuum suspension for a lower limb prosthesis. The vacuum pump compresses and expands throughout the gait cycle. The cycling of the vacuum pump pulls air from the socket, and after multiple steps, the system can achieve vacuum levels of 18 to 22 inches. The number of steps needed to achieve maximum vacuum levels depends on the free air in the system. This is determined by the fit of the socket, the number of socks used, and the length of tubing required.

PROTEOR RUSH Foot EVAQ8 Collection Vacuum Suspension -
Dimension

ASSEMBLY AND INSTALLATION

The RUSH FOOT EVAQ8 system comes preassembled inside the heel of the RUSH Foot®. To connect the vacuum system to the socket:

  1. Connect a short piece of the tubing (4) to the socket right angle barb (6).
  2. Install the inline filter (5) into the distal end of the tubing (the inline filter may be placed anywhere in the tubing run between the socket and vacuum pump heel).
  3. Connect the second piece of tubing (4) to the distal end of the inline filter (5) and route the tube to the medial side of the pylon or wrap the tubing around the pylon (to prevent damage to the tube or snagging while walking).
  4. Cut the tubing (4) to the desired length and connect it to the straight barb (2) located in the recessed area of the vacuum pump heel (1).
  5. Secure the tubing to the pylon using the included hook and loop tape (7) or another appropriate tape.

TROUBLESHOOTING

If a vacuum cannot be generated or maintained by the system, complete the following troubleshooting guide:

  • Inspect tubing (4) for cracks, holes, or other potential leaks.
  • Reseat tubing at each connection point to ensure a good seal.
  • Inspect the inline filter. Remove the inline filter (5) from the tubing and hold up to a light to look through it. If light can be seen through it, the filter is clean. If the light is blocked, blast air from a syringe through the inline filter (5) from the distal to the proximal end (reverse of normal flow) to attempt to clear the blockage. If the blockage persists, the filter needs to be replaced.
  • Inspect socket right angle valve (6) and installation area for leaks.
  • Inspect socket expulsion valve (if used) and installation area for leaks.
  • Inspect socket and suspension sleeve for potential leaks.

MAINTENANCE

The RUSH FOOT EVAQ8 Collection components may need periodic cleaning or replacement during the life cycle of the system and are not replaceable under the warranty as it is considered normal wear:

  • Tubing (4)
  • Inline filter (5)
  • One-way valves housed inside the vacuum heel

Periodic Inspection of the system:

  • Visually inspect the tubing (4) for kinks, cracks, or wear that may leak air into the system. Replace tubing if any of these conditions exist.

  • Remove the inline filter (5) from the tubing and hold up to a light to look through it. If light can be seen through it, the filter is clean. If light is blocked, blast air from a syringe through the inline filter (5) from distal to the proximal end (reverse of normal flow) to attempt to clear the blockage. If blockage persists, the filter needs to be replaced.

  • If the amount of vacuum pulled by the system decreases, the one-way valves contained in the vacuum heel may occasionally need to be cleaned and flushed with distilled water or isopropyl alcohol to ensure proper function. This procedure should be done only by a qualified professional.

  • To flush the one-way valves and vacuum heel:
    1. Disconnect tubing from distal end of the inline filter.
    2. Remove the exhaust filter (3) and take care to not lose the one-way valve. The one-way valve may be inside the bottom of the exhaust filter, or may stay inside the vacuum heel until step 4 of this procedure which should dislodge the one-way valve from the housing.
    3. Place tubing still connected to heel in cup of distilled water or isopropyl alcohol, or connect syringe filled with distilled water or isopropyl alcohol to tubing.
    4. Slowly cycle the vacuum pump by gently compressing the heel to draw the distilled water or isopropyl alcohol into the system and cycle the water or alcohol through the heel and out the hole where the exhaust filter is installed.
    5. Once the distilled water or isopropyl alcohol exiting the heel is clear or 5-10 heel compressions have been performed (whichever is longer) remove the tubing from the water or alcohol source. Cycle the heel a few more times to push any remaining water or alcohol out of the heel. It may be beneficial to invert the foot and allow gravity to aid in removing the last bit of liquid out of the system.
    6. Remove the one-way valve from the exhaust filter (3) or retrieve the dislodged one-way valve from the exhaust side of the vacuum heel. Flush the one-way valve with distilled water or isopropyl alcohol using a syringe or with compressed air.
    7. Rinse the exhaust filter (3) with distilled water or isopropyl alcohol.
    8. Reinstall the one-way valve into the exhaust filter (3) as shown in Figure 1. 9. Reinstall the exhaust filter (3) into the vacuum heel and torque to 14 in-lbs.
    10. Reconnect the tubing (4) to the inline filter (5) and socket.

  • If the preceding procedure hasn’t resolved the decreased vacuum pull, the one-way valves and tubing should be replaced.

WARRANTY

The RUSH FOOT EVAQ8 has a warranty period of 36 months from the date the shipment is received by the customer, excluding the following normal wear parts: one-way valves, tubing, and inline filter.

**| PROTECTOR USA
1236 West Southern Ave, Suite 101
Tempe. AZ 85282 USA
P 001.855.450.7300
F 001.480.361.4526
www.proteonssa.com| **

---|---|---
| PROTECTOR SAS
6 rue de la Redoute
21850 Saint Apollinaire
France|

Call toll-free at855-450-7300 or visit us at proteorusa.com.
Proudly handcrafted in the U.S.A.
For detailed information on the RUSH Foot® collection, visit rushfoot.com
V1EVAQ8NOV18PROTEOR RUSH Foot EVAQ8 Collection Vacuum Suspension - Logo
1

Documents / Resources

| PROTEOR RUSH Foot EVAQ8 Collection Vacuum Suspension [pdf] Instruction Manual
RUSH Foot EVAQ8 Collection, RUSH Foot, EVAQ8 Collection, RUSH Foot EVAQ8 Collection Vacuum Suspension, Vacuum Suspension
---|---

References

Read User Manual Online (PDF format)

Read User Manual Online (PDF format)  >>

Download This Manual (PDF format)

Download this manual  >>

Related Manuals