AtlasScientific EZO Complete-TMP USB Temperature Meter User Guide

June 12, 2024
AtlasScientific

AtlasScientific EZO Complete-TMP USB Temperature Meter

Product Information

The EZO Complete-TMP is a USB Temperature meter designed and manufactured by AtlasScientific. It is a dustproof and splash-resistant device with a USB connector and a probe connector. The readings provided by the EZO Complete-TMP are interference-free and unaffected by other sensors in the same water. The device can read temperatures from any type and brand of PT-100 or PT-1000 temperature sensors with single-point accuracy. The readings are displayed at a rate of 1 reading per second.

Product Usage Instructions

To use the EZO Complete-TMP, follow these instructions:

  1. Ensure that the USB connector and the probe connector are free from water or any other liquid. Water entering the USB connector can cause an electrical short and permanently damage the device. Water entering the probe connector can cause a temporary connector to short, affecting temperature readings and potentially damaging the temperature probe.
  2. Connect the EZO Complete-TMP to your computer using the USB connector.
  3. Install the AtlasDesktop monitoring software on your computer.
  4. Once the software is installed, open it and click on the gear icon to access the setup options.
  5. Follow the on-screen calibration steps to calibrate the device. Make sure to issue calibration commands only when the readings have stabilized to avoid drifting readings.
  6. If you want to link the temperature readings from the EZO Complete-TMP to other EZO-Complete sensors for temperature compensation, navigate to the advanced setup options in the software. You can link the temperature readings to other sensors’ readings, such as conductivity, for temperature-compensated measurements.

Note: For detailed instructions on specific setup procedures or troubleshooting, refer to the product manual or check for updates on the manufacturer’s website.

OVERVIEW

Atlas-Scientific-EZO-Complete-TMP-USB-Temperature-Meter-FIG-1
\(1\)

  • Reads Temperature
  • Range -126.000 °C to 1254 °C
  • Accuracy +/– (0.1 + 0.0017 x °C)
  • Temp reading time 1 reading /sec
  • Supported probes Any type & brandB PT-100 or PT-1000
  • Calibration Single point

Interference-free

Atlas-Scientific-EZO-Complete-TMP-USB-Temperature-
Meter-FIG-1 \(2\)

The EZO complete readings are unaffected by other sensors in the same water.

Ingress protection IP62

The EZO Complete-TMP™ is dust proof and resistant to splashing water. Two areas of concern are the USB connector and the probe connector.Atlas-
Scientific-EZO-Complete-TMP-USB-Temperature-Meter-FIG-1
\(3\)

Ingress protection IP62
An electrical short can occur if water enters the USB connector. A USB short could permanently damage the EZO-Complete. A USB short is not covered under warranty.Atlas-Scientific-EZO-Complete-TMP-USB-Temperature-Meter-FIG-1
\(4\)

A connector short can occur if water enters the SMA connector. A connector short will cause the temperature readings to pin to -1023, or the probe will respond slowly to changes in temperature. A connector short is reversible and will not damage the EZO-Complete. However, frequent shorts will eventually damage the temperature probe .Atlas-Scientific-EZO-Complete-TMP-USB-
Temperature-Meter-FIG-1 \(5\)

The SMA connector is part of your probe; Nothing should be in contact with this part.

Do not use this as a mounting point!

Setup

Atlas-Scientific-EZO-Complete-TMP-USB-Temperature-Meter-FIG-1
\(7\)

  • Temperature probe with SMA Connector
  • Temperature probe with BNC Connector
  • Micro Temperature probe with SMA Connector
  • Tinned leads with the SMA to terminal block connector

Atlas-Scientific-EZO-Complete-TMP-USB-Temperature-Meter-FIG-1
\(8\)

Setup
Once you have installed the AtlasDesktop monitoring software, you can begin monitoring and logging your readings.

Best practices for calibration

Always watch the readings throughout the calibration process. Issue calibration commands once the readings have stabilized.

Never do a blind calibration! Issuing a calibration command before the readings stabilize will result in drifting readings

Calibration procedure

Atlas-Scientific-EZO-Complete-TMP-USB-Temperature-
Meter-FIG-1 \(12\)Atlas-
Scientific-EZO-Complete-TMP-USB-Temperature-Meter-FIG-1
\(13\)

Advanced setup

The EZO-Complete-TMP readings can be linked to other EZO-Complete sensors to provide temperature compensation to the other sensors readings.
Example: We setup an EZO-complete monitoring system to read conductivity and temperature.Atlas-Scientific-EZO-Complete-TMP-USB-Temperature-Meter-
FIG-1 \(14\)

Entering the conductivity setup page, you can link the readings from your EZO- Complete temperature sensor to your conductivity readings, this will give you temperature compensated conductivity readings.Atlas-Scientific-EZO-Complete-
TMP-USB-Temperature-Meter-FIG-1 \(15\)

The connected EZO-complete-TMP is the device with address 129. If multiple temperature sensors are connected, multiple address will be listed. (The temperature sensor address is listed in the temp sensor setup page) Copyright © Atlas Scientific LLC
PATENT PROTECTED: This is an evolving document, check back for updates. Written by Jordan Press Designed by Noah Press

Read User Manual Online (PDF format)

Loading......

Download This Manual (PDF format)

Download this manual  >>

Related Manuals