AtlasScientific B07VDMNB92 Gen 2 Consumer Grade pH Probe Double Junction Silver User Manual

June 6, 2024
AtlasScientific

Atlas Scientific B07VDMNB92 Gen 2 Consumer Grade pH Probe Double Junction Silver User Manual

AtlasScientific B07VDMNB92 Gen 2 Consumer Grade pH Probe Double Junction
Silver User Manual

AtlasScientific B07VDMNB92 Gen 2 Consumer Grade pH Probe Double Junction
Silver User Manual - Consumer Grade

Measurements

AtlasScientific B07VDMNB92 Gen 2 Consumer Grade pH Probe Double Junction
Silver User Manual - Measurements

Specifications

Reference electrode Silver / silver chloride
Double junction Yes
Max depth 45m (148 ft.)
Cable length  1 meter
Weight 49 grams
Speed of response 95% in 4 second
Isopotential point pH 7.00 (0 mV)
Dimensions 12mm x 150.6mm (0.5″ x 6″)
SMA connector Female
Sterilization  Chemical only
Food Safe Yes
Included with probe  SMA to BNC adapter

AtlasScientific B07VDMNB92 Gen 2 Consumer Grade pH Probe Double Junction
Silver User Manual - Do not Boil & Freeze

Materials

AtlasScientific B07VDMNB92 Gen 2 Consumer Grade pH Probe Double Junction
Silver User Manual - Materials

This pH probe can be fully submerged in fresh or salt water, up to the SMA connector indefinitely.

Typical applications

  • Standard lab use
  • Field use
  • Soil
  • Samples containing heavy metals
  • Hydroponics / aquaponics
  • Beer, wine, alcohol, and food production

NSF/ANSI 51 Compliant

Atlas Scientific LLC, hereby certifies that,
Consumer Grade pH Probe Part # ENV-30-pH

Complies with NSF/ANSI Standard 51

AtlasScientific B07VDMNB92 Gen 2 Consumer Grade pH Probe Double Junction
Silver User Manual - NSF ANSI 51 Compliant

Operating principle

A pH (potential of Hydrogen) probe measures the hydrogen ion activity in a liquid. At the tip of a pH probe is a glass membrane. This glass membrane permits hydrogen ions from the liquid being measured to defuse into the outer layer of the glass, while larger ions remain in the solution. The difference in the concentration of hydrogen ions (outside the probe vs. inside the probe) creates a VERY small current. This current is proportional to the concentration of hydrogen ions in the liquid being measured.

AtlasScientific B07VDMNB92 Gen 2 Consumer Grade pH Probe Double Junction
Silver User Manual - Operating principle

A pH electrode is a passive device that detects a current generated from hydrogen ion activity. This current (which can be positive or negative) is very weak and cannot be detected with a multimeter, or an analog to digital converter. This weak electrical signal can easily be disrupted and care should be taken to only use proper connectors and cables.
AtlasScientific B07VDMNB92 Gen 2 Consumer Grade pH Probe Double Junction
Silver User Manual - Result will always read zero
The current that is generated from the hydrogen ion activity is the reciprocal of that activity and can be predicted using this equation:

AtlasScientific B07VDMNB92 Gen 2 Consumer Grade pH Probe Double Junction
Silver User Manual - Equations

Where R is the ideal gas constant. T is the temperature in Kelvin. F is the Faraday constant.

Because a pH probe is a passive device it can pick up voltages that are transmitted through the solution being measured. This will result in incorrect readings and will slowly damage the pH probe over time. In this instance, proper isolation is required.

AtlasScientific B07VDMNB92 Gen 2 Consumer Grade pH Probe Double Junction
Silver User Manual - NEVER EXTEND THE CABLE

DO NOT CUT THE PROBE CABLE WITHOUT REFERRING TO THIS DOCUMENT!

AtlasScientific B07VDMNB92 Gen 2 Consumer Grade pH Probe Double Junction
Silver User Manual - DO NOT MAKE YOUR OWN UNSHIELDED
CABLES!

ONLY USE SHIELDED CABLES.

Extending the probe cable length

You can extend the cable to greater than 100 meters with no loss of signal. Atlas Scientific has tested up to 300 meters without a problem, however you run the risk of turning your pH probe into an antennae, picking up noise along the length of your cable.

If you want to extend your cable, we recommend that you use proper isolation, such as the Basic EZO TM Inline Voltage Isolator, or Tentacle Shield. Be sure to calibrate your probe with the extended cable.

Extending a probe cable can be easily done with our SMA Extension Cables. Simply connect the SMA end of the probe to the Extension cable, and you are all set.

AtlasScientific B07VDMNB92 Gen 2 Consumer Grade pH Probe Double Junction
Silver User Manual - Extending a probe cable can be easily
done

If you need to water proof a SMA connection, we highly recommend using a product like Coax-Seal to safely cover and prevent any water damage that may occur.

AtlasScientific B07VDMNB92 Gen 2 Consumer Grade pH Probe Double Junction
Silver User Manual - If you need to water proof a SMA
connection

Helpful operating tips

pH probes must stay wet and cannot be allowed to dry out, this is why every pH probe is shipped in a plastic soaker bottle containing pH probe storage solution. The probe should remain in the bottle until it is used. If the probe is used infrequently, the bottle and its solution should be saved and the probe stored inside.

  1. To remove the soaker bottle from the probe, hold the soaker bottle by the cap and turn the bottle until it separates from the cap; Then slide the cap off.
    Ready to use

AtlasScientific B07VDMNB92 Gen 2 Consumer Grade pH Probe Double Junction
Silver User Manual - To remove the soaker bottle from the
probe

  1. During shipment the air bubble in the probes stem may move into the bulb area. If bubbles are seen in the bulb area, hold the probe by its top cap and shake downward as done with a clinical thermometer. Bubbles

AtlasScientific B07VDMNB92 Gen 2 Consumer Grade pH Probe Double Junction
Silver User Manual - During shipment the air bubble in the probes stem may
move into the bulb area

  1. Vigorously stir the probe in the sample, calibration solution, or rinse solution. This action will bring solution to the probes surface quicker and improve the speed of response.

AtlasScientific B07VDMNB92 Gen 2 Consumer Grade pH Probe Double Junction
Silver User Manual - Vigorously stir the probe in the
sample

Probe cleaning

Coating of the pH bulb can lead to erroneous readings including shortened span (slope). The type of coating will determine the cleaning technique. Soft coatings can be removed by vigorous stirring or by the use of a squirt bottle. Organic chemical, or hard coatings, should be chemically removed. A light bleach solution or even a 5 ­ 10% hydrochloric acid (HCl) soak for a few minutes, often removes many coatings. If cleaning does not restore performance, reconditioning may be tried. Do not use a brush or abrasive materials on the pH probe.

AtlasScientific B07VDMNB92 Gen 2 Consumer Grade pH Probe Double Junction
Silver User Manual - Probe cleaning

How often do you need to recalibrate a pH probe?

Because every use case is different, there is no set schedule for recalibration.

If you are using your probe in a fish tank, a hydroponic system or any environment that has generally weak levels of acids and bases you will only need to recalibrate your probe once per year for the first two years. After that every ~six months.

If you are using the pH probe in batch chemical manufacturing, industrial process, or in a solution that is known to have strong acids and bases, then calibration should be done monthly or in extreme cases after each batch.

Probe reconditioning

When reconditioning your pH probe is required due to aging, we recommend you use the Atlas Scientific pH probe reconditioning kit.

AtlasScientific B07VDMNB92 Gen 2 Consumer Grade pH Probe Double Junction
Silver User Manual - Probe reconditioning

Copyright © Atlas Scientific LLC AtlasScientific B07VDMNB92 Gen 2 Consumer
Grade pH Probe Double Junction Silver User Manual

Read User Manual Online (PDF format)

Read User Manual Online (PDF format)  >>

Download This Manual (PDF format)

Download this manual  >>