HESAI DM200 Drone Mounted Remote Methane Leak Detector User Manual
- June 13, 2024
- HESAI
Table of Contents
- Notice
- Safety Notice
- Introduction
- Specifications
- Installation Guide
- UART Serial Port Data and Parameter Configuration
- Appendix I Warranty and Maintenance
- Appendix II EC-Declaration of Conformity
- Appendix III FCC Statement
- Customer Support
- Read User Manual Online (PDF format)
- Download This Manual (PDF format)
HESAI DM200 Drone Mounted Remote Methane Leak Detector
Notice
Please read this manual carefully before using the equipment.
CAUTION
To avoid violating the warranty and to minimize the chances of getting electrically shocked, please do not disassemble the device on your own accord. The device must not be tampered with and must not be changed in any way. There are no user-serviceable parts inside the device. For repairs and maintenance inquiries, please contact an authorized Hesai Technologies service personnel.
DISCLAIMER
The information contained within this user’s manual and the functions offered is intended to provide information about products. All reasonable efforts have been made to ensure the accuracy of the information. However, Hesai cannot be held responsible for any errors. Hesai does not warrant the accuracy and reserves the right to make changes to the catalog and its functions at any time without notice.
Safety Notice
Prior to using this product, please read the safety notices carefully to ensure the safety and to avoid any improper usage.
Warning
Be sure to observe the following points:
Avoid direct eye exposure. The lasers may harm your eyes.
Do not point the lasers at people when the product is powered on.
Do not allow children to handle the product.
The blue working light will turn on when there is an infrared laser emission.
Infrared detection lasers and green indicator lasers have outlet labels next
to them, and Class 3R warning labels are attached to the side of the device.
CAUTION
Use of controls or adjustments or performance of procedures other than those specified herein may result i hazardous radiation exposure
Laser Safety
Infrared Measurement Laser :
Class 3R, Wavelength: 1.65 μm, <25 mW
Green Alignment Laser: Class 3R, Wavelength: 520 nm, <5 mW
The device satisfies the requirements of:
- IEC/EN 60825-1: 2014;
- Complies with 21 CFR 1040.10 and 1040.11 except for conformance with IEC 60825-1 Ed.3., as described in Laser Notice No.56, dated May 8, 2019.
Introduction
Hesai Drone-mounted Remote Methane Leak Detector is an ultralight and highly sensitive module that utilizes an infrared laser to remotely detect methane leaks. It can be mounted on the drones, vehicles, and ships for applications such as pipeline survey, residential area detection, etc.
The detector measures the methane concentration by emitting an infrared laser and receiving the laser reflection. The reading is expressed by a methane column density (ppm*m). It is the integral of the methane concentration (ppm) and the distance (m) on the optical path.
Specifications
Table 1. Main Specifications
Target Gas | Methane(CH4)and methane-containing gases |
---|---|
Working Principle | Tunable Diode Laser Absorption Spectroscopy (TDLAS) |
Sensitivity | 5 (Actual) ppm*m |
Detection Range (CH4) | 0~99999 ppm*m |
Measurement Accuracy | ±10% (100~50000 ppm*m) |
Response Time | 10 ms (configurable) |
Detection Distance | 300 m (@75% reflectivity) (Actual distance may vary due to |
background type and conditions)
Weight| < 600 g
Size| 70 mm87 mm120 mm
Operating Voltage| DC 7~12 V
Rated Current| <1 A 0.3 A (typical value) under 8 V power supply
Data Output| Methane concentration, light intensity, temperature, etc
Laser Classes| Infrared Measurement Laser: Class 3R, Wavelength: 1.65 μm, <25
mW Green Alignment Laser: Class 3R, Wavelength: 520 nm, <5 mW
Operating Temperature| -20℃~+45℃
Enclosure Level| IP54
Beam Size| Infrared light divergence angle: 3 mrad (6 cm at 20 m)
- Specifications are subject to change without notice
Installation Guide
Installation Notice
Prior to installing the detector, please read the notes carefully to ensure safe and appropriate usage:
- Supply the detector as required (DC 7~12 V).
- Provide good heat dissipation to the laser.
- Pay attention to static electricity. Please release the body static electricity or wear antistatic bracelet before the installation.
- Do not weld/replace any component unless it is recognized by Hesai.
- Do not unplug the laser fiber. Do not bend the fiber even at a small angle.
- Do not attempt to remove any circuit boards unless it is recognized by Hesai.
- Do not use the PCB of detector A on detector B.
- Do not disassemble or repair the device on your own accord. For repairs and maintenance inquiries, please contact an authorized Hesai Technologies service personnel.
- It is recommended that the length of the external cable should not exceed 3m.
Mechanical Installation
There are two mechanical mounting holes on the top of the detector for the
horizontal installation, and four mechanical mounting holes at the back of the
detector for the vertical installation. The mechanical drawing of the detector
is shown below.
Interface Introduction
The interfaces are at the back of the detector. Descriptions of the interfaces
are shown below.
Table 2. Detector Interfaces
Interface | Description |
---|---|
1 | Positive power supply, DC 7-12V, Rated Current <1A |
2 | Negative power supply, GND |
3 | Data output, TX |
4 | Data input, RX |
UART Serial Port Data and Parameter Configuration
UART Serial Port Data Output
Serial Port Configuration
Baud Rate: 57600; Data: 8 bytes; Stop: 1 byte; Parity: None
Data Format
After power-on, the detector will use around 3 seconds to complete the self- test. If there is an issue, the detector will repeat self-testing until the problem is solved. After booting up, the serial port will send out data in the following format:
Table 3. Outputs Data Format
Header| Methane Concentration| Light Intensity| | Relevancy 1000| Environment
Temperature| | | | SNR| Reserved Bytes| Version Number| Error Code| Checksum
---|---|---|---|---|---|---|---|---|---|---|---|---|---
| 5 bytes| 5 bytes| 5 bytes| 5 bytes| 5 bytes| 5 bytes| 5 bytes| 5 bytes| 3
bytes| 3 bytes| 3 bytes| 3 bytes|
$| 00318,| 02147,| 00735,| 00999,| 00036,| 00024,| -0078,| 00000,| 003,| 000,|
010,| 000,| 3C
Table 4. Data Description
Data | Description |
---|---|
Methane Concentration | the unit is ppm*m |
Light Intensity| the light intensity of infrared detection laser
Relevancy| reference value of whether the signal is distorted, <00900, no outputs
Environment Temperature| measured from PCB board thermistor
SNR| signal to noise ratio reference value
Version Number| used to record the software version number, e.g. 011
Error Code| 000
001
002| normal operation; temperature control failure; laser diode error;| 004
008
016| circuit board temperature exceeds -20~80℃; laser case temperature exceeds
-20~80 °C; reference channel error.
Checksum| Starting character is ‘’, and the Checksum value is obtained by the
data difference or operation between ‘$’ and’ ‘. Checksum calculation
example: unsigned char CalcChecksum(unsigned char* str, int len) { unsigned
char cs; int i; cs = 0; for (i=0; i<len; i++) { cs ^= str[i]; } return cs; }
Error Code Proper Handling Method
When the error message appears, try restarting the device. If error code 001, 004, or 008 appears, please try to restart the device after it is completely cooling down. If the error still exists, please contact Hesai for the technical support.
NOTE Do not try to repair the device on your own accord.
Parameter Configuration
Users can use the serial port debugging tool to configure parameters for the detector. The specific configuration parameters are as follows:
Table 5. Parameter Configuration
Functions | Command | DM100 Return | Description |
---|---|---|---|
Enter configuration mode | “+++\r\n” | “WELCOME\r\n” | Serial port configuration |
is only valid in configuration mode
Set measurement time| “SSPD,%f\r\n”| “AOK\r\n”| Default setting: 0.1 s.
Options: 0.1 s~10 s
Set save power| “SD,Savepower\r\n”| “AOK\r\n”| –
Set wakeup| “SD,Wakeup\r\n”| “AOK\r\n”| –
Set SNR| “SSNR,%f\r\n”| “AOK\r\n”| Default setting: 2. Options: 0.1~10
Set light intensity threshold| “SITH,%d\r\n”| “AOK\r\n”| Default setting:
100,Options: 50~1000 (integer)
Get device information| “DSP\r\n”| “DeviceSN:%s\r\n LaserSN:%s\r\n
SetTld:%x\r\n IS1651:%d\r\n SNRThreshX10:%d\r\n IntensityThresh:%d\r\n”|
Example: “DeviceSN:HESAI194 LaserSN:ABCD1234
SetTld:825 IS1651:1
SNRThreshX10:10 IntensityThresh:100”
Set green alignment laser| “SGL,ON\r\n” “SGL,OFF\r\n”| “AOK\r\n”| Default
setting: ON
Reset| “SF,1\r\n”| “AOK\r\n”| Restore system default setting
Exit| “—\r\n”| “END\r\n”| Enter test mode after exiting configuration mode
- After sending the configuration command, the system returns the value “AOK\r\n”; if the command is incorrect, it returns “ERR\r\n”. End a command with “\r\n”.
Appendix I Warranty and Maintenance
Warranty
During the warranty period, Hesai will provide free maintenance service when the device can’t operate due to the problem of software or hardware. But Hesai will not provide the service if the problem is caused by rule-breaking operations. Situations that break the warranty include but are not limited to the following:
- Missing warranty document or effective purchase vouchers;
- Did not use the instrument as per instructed;
- Unauthorized modification, disassembly, or repair;
- Intentional damages;
- Stolen, lost or discarded device;
- Damaged device due to use with unauthorized accessories and services;
- Those damages caused by natural disasters, such as fire, lightning, flood, earthquake, etc.
Maintenance
For maintenance requirement, please follow the following steps:
- Report the problem to Hesai for the preliminary problem diagnosis.
- Return the product to Hesai.
- Hesai conducts the maintenance test report, calculates maintenance fee (if any), and provides the maintenance agreement.
- Sign the maintenance agreement and pay the maintenance fee if needed.
- Receive the repaired instrument and finish the delivery inspection.
Normal Maintenance Time (after receiving the product): 7 days. The specific maintenance time shall be determined according to the actual situation.
Appendix II EC-Declaration of Conformity
EC-Declaration of Conformity
Manufacturer Name: Hesai Photonics Technology Co., Ltd.
Manufacturer Address: Room J385, Building 6, No. 1288 Yecheng Rd., Jiading
District, Shanghai City, China
Tel: 00+86-21-31588240; Fax: 00+86-21-31588240; E-mail:
lijinyu@hesaitech.com
We hereby declare that our products below described electrical equipment manufactured by us fulfills the requirements :
- Electromagnetic Compatibility Directive (EMC) 2014/30/EU
- Restriction of Hazardous Substance Directive (RoHS) 2011/65/EU last amended by (EU)2015/863
- Waste Electric and Electronic Equipment (WEEE) Directive 2012/19/EU
- Screening of SVHCs subject to authorization, according to (EU) No 143/2011, (EU) No 125/2012, (EU) No 348/2013 and (EU) No 895/2014 and (EU) No 2017/999 (Annex XIV of EC No 1907/2006) and SVHCs In candidate list by European Chemical Agency (ECHA), and the EU Court of Justice rules on SVHCs In articles
Declare that product
Product Name: Drone-mounted Remote Methane Leak Detector
Product Model: DM100
Is In conformity with the following standards:
EN 61326-1:2013; IEC 61326-1:2012; EN 60825-1:2014; IEC 60825-1:2014; GB 7247.1-2012 21 CFR 1040.10 and 1040.11 except for deviations pursuant to Laser Notice No.SO, Dated June 24,2017 Article 11 Recovery and Recycling, Calculation of Theoretical Recovery and Rec.Y£!i!!g_ Rate
Mr. Li Jinyu I Certification Manager
Shanghai, P. R. China
25 June, 2019
Hesai Photonics Technology Co., ltd Building ~2. Hongqiao World Centre,
Shanghai 021- 31588240
www.hesaitech.com
Appendix III FCC Statement
FCC ID :2ASO2DM100
This device complies with part 15 of the FCC Rules. Operation is subject to the following two conditions:
- This device may not cause harmful interference, and (2) this device must accept any interference received, including interference that may cause undesired operation.
Please note that changes or modifications not expressly approved by the party responsible for compliance could void the user’s authority to operate the equipment.
Note: This equipment has been tested and found to comply with the limits for a Class B digital device, pursuant to part 15 of the FCC Rules. These limits are designed to provide reasonable protection against harmful interference in a residential installation. This equipment generates, uses and can radiate radio frequency energy and, if not installed and used in accordance with the instructions, may cause harmful interference to radio communications. However, there is no guarantee that interference will not occur in a particular installation. If this equipment does cause harmful interference to radio or television reception, which can be determined by turning the equipment off and on, the user is encouraged to try to correct the interference by one or more of the following measures:
- Reorient or relocate the receiving antenna.
- Increase the separation between the equipment and receiver.
- Connect the equipment into an outlet on a circuit different from that to which the receiver is connected.
- Consult the dealer or an experienced radio/TV technician for help.
Customer Support
HESAI WeChat
Hesai Technology Co., Ltd.
Phone: 400-805-1233
Website : www.hesaitech.com
Address : Building L2, Hongqiao World Centre, Shanghai, China
Business Email : info@hesaitech.com
Service Email : service@hesaitech.com
Read User Manual Online (PDF format)