Product Information

  • Product Name: Stirrup Clamp 800-0396
  • Model Number: 800-0396
  • Copyright: All rights reserved. Any duplication, adaption or translation without prior written consent is prohibited unless otherwise provided for in relevant copyright laws.
  • Initial Release Date: 6/6/2023
  • Manufacturer: SchureMed (081001460)
  • Intended Use: The Stirrup Clamp is designed to hold operating room accessories on a surgical table. It is primarily used together with the Great White Stirrups.
  • Device Dimensions (maximum): Not specified

Side Rail Dimensions:

  • US: 0.374 x 1.122 (9.5 mm x 28.5 mm) – PN# 800-0396
  • Europe: 0.394 x 0.984 (10 mm x 25 mm) – PN# 800-0396-EU

Product Usage Instructions

  1. Become familiar with the features of the Stirrup Clamp before using it with a patient. Practice using it on a nurse, physician, or appropriate volunteer prior to using it clinically.
  2. To attach the clamp to the side rail:
  3. Place the clamp on the rail with the flat bottom in the down position.
  4. Push the clamp down on the rail until it attaches securely.
  5. Turn the handle clockwise to lock the clamp in position.
  6. To detach the clamp from the side rail:
  7. Unlock the clamp by turning the handle counterclockwise.
  8. Slide the clamp away from the standoffs.
  9. Push the top of the clamp down with your thumb until it releases from the rail.
  10. Grab the handle and lift the clamp from the side rail.
  11. User Notice: Any serious incident related to the device should be reported to the manufacturer and the competent authority of the Member State where the user and/or patient is established.
  12. Cleaning and Maintenance:
  13. Follow proper cleaning and disinfection procedures according to hospital protocols and guidelines.
  14. Apply cleaning fluid liberally to the entire accessory and wipe it with a clean, lint-free cloth until all moisture and cleaning fluid are removed.
  15. Allow the accessory to dry before storage or reuse.

Instructions for Use

Stirrup Clamp
800-0396
800-0396-EU

Copyright Notice
All rights reserved. Any duplication, adaption or translation without prior written consent is prohibited unless otherwise provided for in relevant copyright laws.
Initial Release: 6/6/2023

SchureMed (081001460)

GENERAL SPECIFICATIONS

Device Dimensions (maximum)

  • Length:2.75” +/- 0.5” (7 cm +/- 1 cm)
  • Width: 1.94” +/- 0.5” (4.9 cm +/- 1 cm)
  • Depth: 2.98”+/- 0.5” (6.8 cm +/- 1 cm)
  • Handle Collapsed: 2.75” +/- 0.5” (7 cm +/- 1 cm)
  • Handle Extended: 3.0” +/- 0.5” (7.62 cm +/- 1 cm)
  • Device Weight: 0.5 +/- 0.5 lbs. (.22 kg +/- .22 kg)
  • Attaches to surgical table anywhere on rail
  • Single-person installation

INTENDED USE
Intended use is to hold operating room accessories on surgical table. Clamp is primarily used together with our Great White Stirrups.

INSTRUCTIONS
Become familiar with the features of patient positioning device before use with a patient. Always practice use on a nurse, physician or appropriate volunteer prior to using clinically.
Stirrup Clamp is spring-loaded and attaches anywhere along surgical table’s side rail. Accepts 1” x 1/4” (2.5 cm x .64 cm) flat bar mounting post. Holds all accessories in vertical position.

Attaching to Side Rail

  1. Place clamp on rail so flat bottom is in down position
  2. Push clamp down on rail until it attaches to the rail.
  3. Secure clamp by turning handle clockwise to locked position

Detaching from Side Rail

  1. Unlock clamp by turning handle counterclockwise and remove accessory
  2. Slide clamp away from standoffs, then push top of clamp down with thumb until it releases from rail, then grab handle and lift clamp from side rail

COMPONENT OVERVIEW
Stirrup Clamp is a surgical table clamp that attaches to surgical table side rails. Clamp is primarily used in conjunction with our Great White Stirrups.
See side rail dimensions below:
US: 0.374” x 1.122” (9.5 mm x 28.5 mm) PN# 800-0396
Europe: 0.394” x 0.984” (10 mm x 25 mm ) PN# 800-0396-EU

GENERAL INFORMATION

  • Product not made with Natural Rubber Latex
  • Device supports 1,000 lb. (454 kg) proportional patient load (6’4” (193 cm) tall patient per 99%human body model)
  • Product warranty covers product from manufacturing defects for period of 2 years
  • Product is maintenance-free, check product condition before next use
  • Life of device is 5 years under normal use
  • Store device between –4°F to +86°F (-20°C to 30°C)

DISPOSAL

  • General – Prevent infection by cleaning and disinfecting product before disposal
  • Packaging – Dispose packaging material via household waste according to national requirements
  • SchureMed accepts back used or retired products – or dispose of product in accordance with national requirements

Product Use Warnings

WARNING!
Maximum load should not exceed appropriate proportion of a patient weighing 1,000 lbs. (454 kg). Use care with low-maximum load capacity surgical tables that accessory rails are not overloaded.

Cleaning Recommendation
Follow current Association of periOperative Registered Nurses (AORN) Journal Guidelines for proper cleaning and disinfection procedures.

CAUTION
Strictly read/follow manufacturer’s directions for cleaning fluids. DO NOT use cleaners containing phenolics.

  1. Remove major contaminants from accessory with disposable materials. Follow appropriate bio-hazard waste disposal procedures.
  2. Apply cleaning fluid liberally to entire accessory and wipe with clean, lint-free cloth until all moisture and cleaning fluid is removed from accessory
  3. Let accessory dry

USER NOTICE
Any serious incident that has occurred in relation to the device should be reported to the manufacturer and the competent authority of the Member State in which the user and/or patient is established.

IFU-800-0396 Rev 3.00