SchureMed 800-0050 Multi Axis Arm Positioner Instructions
- June 11, 2024
- SchureMed
Table of Contents
- SchureMed 800-0050 Multi Axis Arm Positioner
- Product Information
- Product Usage Instructions
- Instructions for Use
- Set Up and Use of Multi-Axis Arm Positioner
- GENERAL SPECIFICATIONS
- COMPONENT OVERVIEW
- GENERAL INFORMATION
- CLEANING RECOMMENDATION
- Symbol Glossary
- References
- Read User Manual Online (PDF format)
- Download This Manual (PDF format)
SchureMed 800-0050 Multi Axis Arm Positioner
Product Information
- The Multi-Axis Arm Positioner 800-0050 is a surgical table accessory designed to provide support for non-operative arms during surgical procedures. It is intended for use by medical professionals within hospitals and surgery centers. The device can be used in various positions such as seated, lateral, prone, neuro, and park bench positions.
- The unique pivoting design allows for infinite positioning.
- The device attaches to the surgical table rail with a clamp and comes with a replacement pad (508-0108 Multi-Axis Arm Positioner Concave Pad, 16 x 6 x 2) and requires Schure Socket XL (800-0134) which is sold separately. The device has a maximum load capacity of 500 lbs (227 kg).
Product Usage Instructions
Before using the Multi-Axis Arm Positioner with a patient, it is essential to become familiar with its features. Always practice use on a nurse, physician, or appropriate volunteer before using it clinically.
- Step 1: Attach the device to the surgical table rail with the clamp.
- Step 2: Insert the Schure Socket XL (800-0134) into the device.
- Step 3: Place the non-operative arm of the patient on the concave pad of the device.
- Step 4: Use the positioning handle to pivot and position the arm as required.
Cleaning and Disinfection Instructions
It is recommended to follow current Association of periOperative Registered Nurses Journal Guidelines for proper cleaning and disinfection procedures. Always adhere to standards for blood-borne pathogens from the Occupational Safety and Health Administration. Use recommended protective clothing, gloves, masks, and eye protection to clean the accessory.
- Step 1: Remove major contaminants from the accessory with disposable materials and follow appropriate bio-hazard waste disposal procedures.
- Step 2: Apply cleaning fluid liberally to the entire accessory and wipe with a clean, lint-free cloth until all moisture and cleaning fluid are removed from the accessory.
- Step 3: Let the accessory dry before using it again.
Note: Strictly read and follow the manufacturer’s directions for cleaning
fluids. DO NOT use cleaners containing phenolics.
If there are obvious signs of damage or functional issues, do not reuse the
device. Consult the manufacturer before reusing it. Any serious incident that
has occurred in relation to the device should be reported to the manufacturer
and the competent authority of the Member State in which the user and/or
patient is established.
Instructions for Use
Multi-Axis Arm Positioner 800-0050
Replacement Pad
Multi-Axis Arm Positioner Concave Pad 508-0108
INTENDED USE
Multi-Axis Arm Positioner provides support for nonoperative arm during
surgical procedures. The intended users of this device are medical
professionals within hospitals and surgery centers.
INSTRUCTIONS
Become familiar with the features of patient positioning device before use
with a patient. Always practice use on a nurse, physician or appropriate
volunteer prior to using clinically.
Set Up and Use of Multi-Axis Arm Positioner
- Attach Schure Socket XL P/N 800-0134 (sold separately) onto side rail
- Insert mounting post into Schure Socket XL
- Turn handle clockwise to loosely tighten Schure Socket XL
- Loosen Positioning Handle, move to desired position
- Adjust Schure Socket XL to position in lateral, prone, neuro and park bench positions
- Tighten and lock Positioning Handle and Schure Socket XL
Detaching Multi-Axis Arm Positioner from Surgical Table
- Turn clamp handle counterclockwise to loosen
- Remove from clamp
GENERAL SPECIFICATIONS
- Maximum Height: 21” +/- 0.5” (53 cm +/- 1 cm)
- Minimum Height: 7.5” +/- 0.5” (19 cm +/- 1 cm)
- Armboard Length: 16” +/- 0.5” (41 cm +/- 1 cm)
- Armboard Width: 6” +/- 0.5” (15 cm +/- 1 cm)
- Maximum Pad Depth: 2.25” +/- 0.5” (6 cm +/- 1 cm)
- Middle of Pad Depth: 1.5” +/- 0.5” (4 cm +/- 1 cm)
- Net weight: 5.1 lbs. +/- 0.5 lbs. (2.3 +/- .22 kg)
- Gross Weight: 8.2 lbs. +/- 0.5 lbs. (3.7 +/- .22 kg)
- Single-person installation
COMPONENT OVERVIEW
Multi-Axis Arm Positioner is a surgical table accessory providing support for non-operative arm during surgical procedures. Device attaches to surgical table rail with clamp, and used in seated, lateral, prone, neuro and park bench positions. Unique pivoting design allows for infinite positioning.
Replacement Pad
508-0108 Multi-Axis Arm Positioner Concave Pad, 16″ x 6″ x 2″ (41 cm x 15 cm x 5 cm)
Other required products for use: 800-0134 Schure Socket XL (sold separately)
- US: 0.374” x 1.122” (9.5 mm x 28.5 mm) PN# 800-0134
- Denyer: 0.236” x 1.496” (6 mm x 38 mm) PN# 800-0134-DEN
- Europe: 0.394” x 0.984” (10 mm x 25 mm ) PN# 800-0134-EU
- Eschmann (UK): 0.236” x 1.260” (6 mm x 32 mm) PN# 800-0134-UK
- Japan: 0.354” x 1.260” (9 mm x 32 mm ) PN# 800-0134-JPN
- Swiss: 0.394” x 1.181” (10 mm x 30 mm ) PN# 800-0134-SWISS
GENERAL INFORMATION
- Product not made with Natural Rubber Latex
- Device supports 500 lb. (227 kg) proportional patient load (6’4″ (193 cm) tall patient per 99% humanbody model)
- Product warranty covers product from manufacturing defects for period of 2 years
- If damaged in shipping, call Customer Service at 888-724-8763 or 781-982-7000 to obtain ReturnMaterial Authorization number. For product warranty issues, contact
Customer Service.
- CE marked medical device according to MDR (EU) 2017/745
- Product is maintenance-free, check product condition before next use
- Life of device is 5 years under normal use
- Store device between -4°F to +86°F (-20°C to 30°C)
DISPOSAL
- General – Prevent infection by cleaning and disinfecting product before disposal
- Packaging – Dispose packaging material via household waste according to national requirements
- SchureMed accepts back used or retired products – or dispose of product in accordance with nationalrequirements
PRODUCT USE WARNINGS
WARNING! Maximum load should not exceed appropriate proportion of a
patient weighing 500 lbs. (227 kg). Use care with low-maximum load capacity
surgical tables that accessory rails are not overloaded.
WARNING! Hazards result from incorrect use. Strictly follow Instructions
for Use with your Operating Table system.
WARNING! Do not reuse device if there are obvious signs of damage or
functional issues. Consult manufacturer before reusing
CLEANING RECOMMENDATION
Follow current Association of periOperative Registered Nurses Journal Guidelines for proper cleaning and disinfection procedures
WARNING!
Adhere to standards for blood-borne pathogens from the Occupational Safety and
Health Administration. Use recommended protective clothing, gloves, masks and
eye protection to clean accessory.
CAUTION
Strictly read/follow manufacturer’s directions for cleaning fluids. DO NOT use
cleaners containing phenolics.
- Remove major contaminants from accessory with disposable materials. Follow appropriate biohazard waste disposal procedures.
- Apply cleaning fluid liberally to entire accessory and wipe with clean, lint-free cloth until all moisture and cleaning fluid is removed from accessory
- Let accessory dry
USER NOTICE
Any serious incident that has occurred in relation to the device should be
reported to the manufacturer and the competent authority of the Member State
in which the user and/or patient is established.
Basic UDI-DI: 081001460F0038E9
Symbol Glossary
Manufacturer
SchureMed ( 081001460)
Authorized Representative
Emergo Europe, Prinsessegracht 20, 2514 AP The Hague, The Netherlands
IFU-800-0050 REV 3.07
Latest Revision: 2022-01
References
Read User Manual Online (PDF format)
Read User Manual Online (PDF format) >>