SIMPSON 5002, 5003, 5005, 5006 Digital Portable Multimeter Owner’s Manual
- June 12, 2024
- Simpson
Table of Contents
- Special Features
- Application
- Model Wise Functional Overview
- Technical Specification
- Reference Condition for Accuracy
- Influence Quantity
- Applicable Regulations & Standards
- Display
- Accessories For Operation at a PC
- Scope of Supply
- References
- Read User Manual Online (PDF format)
- Download This Manual (PDF format)
5002, 5003, 5005, 5006 Digital Portable Multimeter
Owner’s Manual
Simpson 5002 5003 5005 5006 series of new multimeters is made for professional use that offers safety, high resolution, large range count, reliability, ruggedness, a complete tool for test automation and is equipped with more than 30 different measuring functions.
Special Features
- True RMS Digital Multimeter
- Data logger & View function (up to 32000 readings)
- Plug and Play USB connectivity with PC
- 100kHz bandwidth for voltage measurement
- 1kHz Low Pass Filter mode
- GO-NO-GO function
- VAC with 1MOhm impedance 4-20mA/0-20mA scale type measurement
- Single fuse for mA & A
- Adjustable square wave output
- Temperature measurement with J, K, Pt100 & Pt1000 sensors
- External power adapter for long hours of measurements
- Selectable clamp ratio for current measurement
- Conductance Measurement
- Frequency/Time Period Measurement
Application
Low input impedance (Ri = 1MΩ)
Trouble shooting a branch circuit with dead or disconnected circuit is made
easy with VAC1M. Low impedance VAC1M measurement helps eliminating error
readings resulting from ghost voltages caused by long wires that share a
common conduit
Single fuse(16A)
Instrument contains a single fuse of 16A common for all the ranges of current
from 600μA to 10A AC/DC as compared to the two fuses in traditional DMMs. This
eliminates the accidental blowing of 1.6A fuse due to operator’s error when
higher current is applied in lower ranges
Tool for automation, USB 2.0 Interface
With ready to use communication protocol and plug and play USB 2.0 add-on
device, one can easily automate his test system. The extensive data capturing
and analysis is possible with DMM software. With vast functionality and
editable report settings DMM software is a real help for easy report
generation and analysis of a device under test.
Square Wave Output
A square wave output can be generated from the DMM with the user selectable
frequency and adjustable duty cycle. This can be used as baud rate generator,
to check flow meters, to test frequency counters, accelerometer and frequency
transmitter. It can also be used as audio signal in audio signal testing.
Current measurement with clamp sensor
Measurement with various clamp sensors is possible, which helps in accurate
measurement of current from 60mA to 6000A without interrupting the circuit.
The measured current is automatically calculated from the selected clamp
ratio.
Low pass filter(LPF) in VAC10MΩ & VAC1MΩ
A selectable 1kHz low pass filter offers advanced variable frequency drive
filtering to help you accurately analyze nontraditional sine waves and noisy
signals.
In LPF mode DMM rejects all high frequency noise making it suitable for making
measurements on inverters and high frequency drives.
True RMS measurement with high crest factors
Accurate true RMS measurement of distorted waveform with crest factor CF
between 1 to 10.
Data Logging
DMM 5005/5006 offers continuous data logging of up to 32000 readings with real
time stamping. Log rate is adjustable from as low as 0.1 sec to as high as
1hr.
Adjustable Beep Level
With Beep level setting, the limit for continuity can be adjusted from 10Ω to
90Ω depending upon application.
Separate fuse compartment
Easier access to fuse when replacing the blown fuse.
Auto Power OFF with adjustable timing
Flexibility to adjust “Auto off ” period from 5 minutes to 60 minutes.
60mv & 600mV DC & ACDC
This helps in accurate measurement of low output voltages <600mV from sensors
& transmitters. High frequency low voltage signal from RF transmitters can
also be measured. Signal as low as 0.001mV can be measured accurately.
Min / Max / Avg measurement
Min/Max/Avg function records the minimum, maximum and average of all the
readings applied since its activation. With dual display it makes it even
flexible for the user to keep the trace of the applied readings while viewing
Min/Max/Avg readings. The average reading is useful for smoothing out unstable
inputs,& verifying circuit performance.
Dedicated keys for easy navigation
Dedicated navigation keys makes scrolling through menu and setting of
parameters easy & comfortable.
External Power Adapter (DC Jack)
The external power supply adapter helps in conserving battery while performing
long hours of measurements. When DC jack is connected batteries inside DMM are
electronically disconnected, and reconnected in absence of mains, hence there
is no need of removing the battery when using the power adapter.
100kHz Bandwidth
Alternating voltages with frequencies up to 100kHz can be measured accurately.
This is useful while analyzing high frequency analog signals.
Self battery voltage measurement
Capable of measuring self battery voltage.
Room temperature measurement
Room temperature can be sensed and measured without any external sensor. The
same is used as internal reference temperature in thermocouple based
temperature measurements
Fully programmable GO NO-GO
The Go – NoGo function gives an indication through a buzzer for the applied
input lying inside or outside the set band. The values for low limit, high
limit and buzzer condition can be easily set through NoGo function in menu
settings. Once the NoGo function is set, user can get busy doing other
activities in the vicinity of the meter, whenever the condition is met it will
be indicated by a buzzer. It eliminates the need of operator to continuously
monitor the display.
View Function
Data logged on meter can be viewed directly on the meter itself, hence the
data analysis is also possible without a PC based software. However for
graphical and large data analysis PC based software can be used.
Dangerous Contact Voltage Indication
Presence of hazardous voltage (>35Vrms 50/60Hz and 50Vdc) at the contact
terminal are indicated on display. This is very useful while performing
measurements in the circuit which takes longer time to discharge its
capacitors, or where unexpected danger voltage are present.
Model Wise Functional Overview
Functions/Features | 5002 | 5003 | 5005 | 5006 |
---|---|---|---|---|
Voltage VDC (Ri>9MΩ) | • | • | • | • |
Voltage VAC TRMS (Ri>9MΩ ) | • | • | • | • |
Voltage LoZ VAC TRMS (Ri=1MΩ ) | • | • | • | |
Voltage VAC TRMS (Ri>9MΩ) LPF 1kHz | • | • | • | |
Voltage LoZ VAC TRMS (Ri=1MΩ) LPF 1kHz | • | • | • | |
Voltage VACDC (Ri>9MΩ) | • | • | • | • |
High impedance, high bandwidth mV measurement | 600mV | 60mV/600mV | 60mV/600mV |
60mV/600mV
Bandwidth VAC & mV ACDC| 10kHz| 10kHz| 10kHz| 100 kHz
Frequency Measurement| | | •| •
Duty cycle %
Voltage level measurement dB, dBu, dBm| | •| •| •
Resistance| •| •| •| •
Conductance measurement| •| •| •| •
Continuity test (I const = 1 mA)| •| •| •| •
Diode measurement (I const = 1 mA)| •| •| •| •
Temperature measurement (TYP J,TYP K)| | •| •| •
Temperature measurement (PT100,PT1000)| •| | •| •
Capacitance measurement| | | •| •
Current ADC| 600mA| 6 A/16 A
(20 A)| 600 μ A/ 6 m A
60mA/600mA
6A/10A(16 A)| 600 μ A/ 6 m A
60mA/600mA
6A/10A(16 A)
Current AAC+DC TRMS
Current AAC TRMS| |
Bandwidth @AAC+DC or AAC 10 kHz| •| •| •| •
Measurement with Clamp Sensor| •| •| •| •
Data Logging / Viewing Function| | | •| •
Protective rubber holster| •| •| •| •
Fuse 16A/ 1000V| 1.6A| | •| •
0-20mA/ 4-20mApercentage scale| | | •| •
Square wave Out| | | •| •
Self battery voltage measurement| •| •| •| •
MIN/MAX/AVG and Auto Hold Functions| •| •| •| •
Dangerous contact voltage indication| •| •| •| •
REL/Zero function| •| •| •| •
USB IR-interface| Optional
External power supply adapter
External power supply adapter| 1000 V CAT III 600 V CAT IV| 1000 V CAT I
600V CAT II| 1000 V CAT III
600 V CAT IV| 1000 V CAT III
600 V CAT IV
Environmental Condition
Operating temperature | -10 to +50°C, 14°F to 122°F |
---|---|
Storage temperature | – 25 to +70°C, -13°F to 158°F |
Relative humidity | < 75% non condensing. |
IP | I. P 50 for Housing, IP20 for terminals. |
Altitude | Up to 2000 m |
Technical Specification
Voltage
Measurement Function
| Measuring Range| Resolution| Input Impe dance| Intrinsic Uncertainty under Reference Condition ±(…% of the rdg.+…Digits)|
Overload Capacity 2)
---|---|---|---|---|---
DC 7| AC 1) 3)| ACDC 1) 3)| Value| Time
V| 6V| 100µV| >9 MΩ| 0 .05 + 5| 0 .5 + 9| 1 + 30| 1000 V DC/ AC RMS Sine|
Continuous
60 V| 1mV| 0 .05 + 5
600 V| 10 mV| 0 .05 + 9
1000 V| 100 mV| 0 .09 + 10
mV| 60 mV| 1 µV| >10 MΩ| 0 .09 + 15| –| 1 + 30| Max10 s
600 mV| 10 µV| 0 .09 + 15
Influence Quantity| Range of Influence| Range| Accuracy|
Simpson 5006| Others 2)
Frequency 6) 9)| >15 Hz… 45 Hz| 60 mV~5),
600 mV~| 3 +30
65 Hz… 100 kHz
15 Hz… 45 Hz| 6 V, 60V,
600 V~| 2 +9| 3 +9
65 Hz… 1 kHz| 1 +9| 3 +9
1 kHz…..20 kHz| 3 +9| 4 +9 10)
20 kHz….100kHz 8)| 3 .5 +30|
15 Hz… 45 Hz| 1000 V~| 2 +9| 3 +9
65 Hz… 1 kHz| 2 +9| 3 +9
1 kHz…..10 kHz| 3 +30|
1 ) Specified Accuracy is valid as of 3% of the measuring range. With Short- circuited test probes: residual value of 1 to 30 d at zero point due to the TRMS converter.
2 ) At 0°C to 40°C (Accuracy Range)
3 ) In VAC measurement, Frequency will be shown above 10% of the present range, except for 1000V & 60mV range i.e. 25% & 50% respectively.
4 ) Frequency Influence upto 10kHz.
5 ) Frequency response up to 50 kHz
6 ) Frequency response is valid from 10% to 100% of range
7 ) With Zero Balancing
8 ) Frequency response up to 100 kHz, for greater than 50 kHz plus 2.5%
9 ) Overload capacity of the voltage measurement input: power Limiting: Frequency x Voltage Max : 6×10 V xHz for V>100V
10 ) Frequency response greater than 2 kHz plus 2.5%
Frequency, Duty Cycle
Measurement Function| Measuring Range| Frequency| Intrinsic
Uncertainty ±(% of the rdg. + .Digits)| Overload Capacity 1)
---|---|---|---|---
Value| Time
Hz5)| 600Hz, 6kHz, 60kHz, 600kHz, 114-1z| fmin2): 6Hz| 0.05 +5| 1000 V DC/AC
RMS Sine| Max 10 s
Hz(V)3)| 10Hz…..100kHz| 0.1 +54)
Duty Cycle (%)| 2.0…98%| 151-1z …. 1 kHz| 0.1 R +5 d
5.0…98%| …. 10 kHz| 0.2 R per kHz+ 5d
10…90%| …. 50kHz| 0.5 R per kHz+ 5d
- At 0°C to 40°C (Accuracy Range)
- Lowest measurable frequency for square measuring signals symmetrical to the zero point ( ±5V).
- Overload capacity of the voltage measurement input : Power limiting: Frequency x voltage max : 6×10 Vx Hz for U> 100V.
- Input sensitivity, sinusoidal signal, 10% to 100% of the measuring range
- At input ±5Vrms, Square wave, Bipolar inputs.
R= Range d= digit
Current
Measurement
Function| Measuring
Range| Resolution| Voltage Drop
Approx.| Intrinsic Uncertainty under Reference Condition ± (…% of the
rdg.+…Digits)| Overload 32) Capacity
---|---|---|---|---|---
| DC 4)| AC 1)| ACDC 1)| Value| Time
mA| 600 μA| 10 nA| 60 mV| 05 + 15| 1 +10| 15 + 10| 0.7A| Continuous
6 mA| 100 nA| 60 mV| 0.5 + 5| 1+10| 15 + 10
60 mA| 1 µA| 60 mV| 01 + 5| 1 + 10| 1.5 + 10
600 mA| 10 μA| 60 mV| 02 + 5| 1 + 10| 15 + 10
A| 6A| 100µA| 60 mV| 0.9 + 10| 1 +10| 15 + 10| 10 A =5 min 3)
10A| 1 mA| 300 mV| 0.9 + 10| 1+10| 15 + 10
Influence Quantity| Range of Influence| Range| Accuracy|
Simpson 5006 ±(…% of the rdg +…Digits)| Others
Frequency 5)| >15 Hz ….45 Hz| 600µA. 10A| 3+10
16 Hz….10 kHz
- Specified Accuracy is valid as of 3% of the measuring range. With Short- circuited test probes: residual value of 1 to 30 d at zero point due to the TRMS converter.
- At 0°C to 40°C (Accuracy Range)
- Off time 30 min and TA= 40°C
- With Zero Balancing
- Frequency response is valid from 10% to 100% of range
Resistance, Diode, Continuity
Measurement Function| Measuring Range 4)| Resolution| Open
Ckt. Voltage| Meas. curr. @ range limit| Intrinsic Uncertainty ±(…%
of the rdg +…Digits)| Overload Capacity 2)
---|---|---|---|---|---|---
Value| Time
Ω 1)| 600 Ω| 10mΩ| <1.4V| Approx. 300 μ A| 0.1 + 10| 1000 V DC/ AC RMS Sine|
Max 10 s
6 k Ω| 100mΩ| Approx. 250 μ A| 0.1 + 10
60k Ω| 1Ω| Approx. 100 μ A| 0.1 + 10
600kΩ| 10 Ω| Approx. 12 μ A| 0.5 + 10
6MΩ| 100 Ω| Approx. 1.2 μ A| 1 + 10
60MΩ| 10k Ω| Approx. 125 nA| 5 + 10
Continuity| 600 Ω| –| Appx. 8V| Approx. 1 mA| 3 + 5
Diode 1)| 6.0V 3)| –| Appx. 8V| Approx. 1 mA| 0.5 + 5
- Measurement of Resistance, Diode will be more accurate after removal from device under test
- At 0°C to 40°C (AccuracyRange)
- Displays up to max6.0 V, “OL” in excess of 6.0V.
- With Zero Balancing
Temperature
Measurement
Function| Measuring Range| Intrinsic Uncertainty
±(…% of the rdg +-Digits)| Overload Capacity 1)
---|---|---|---
Value| Time
Temperature °C/ °F| Pt 100| -200°C. +850°C/-328°F to 1562°F| 2)
0.3 + 15| 1000 V DC/AC R MS Sine| Max 10s
Pt1000| -150°C..+850°C
-238°F to 1562°F| 0.3 + 15 2)
TC K| -200°C..+1372°C
-328°F to 2501.6°1–| 1 % +20 2)
TC J| -210°C..+1200°C
-346°F to 2192°F| 1 % +20 2)
1 ) At 0°C to 4 0°C (Accuracy Range)
2) Plus Sensor Deviation
Capacitance
Measurement Function| Measuring Range| Resolution| V O
Max| Intrinsic Uncertainty ± (% of the rdg +Digits)| Overload
Capacity 1) 2)
---|---|---|---|---|---
Value| Time
F 3) 4)| 1 0 n F| 1 0 p F| 0 .7 V| 1 +1 0 2)| 1000VDC/AC RMS
Sine
| Max 10 s
1 0 0 n F| 1 0 0 p F| 1 + 6 2)
1 µ F| 1 n F| 1 + 6 2)
1 0 µ F| 1 0 n F| 1 + 6 2)
1 0 0 µ F| 1 0 0 n F| 5 + 6 2)
1 0 0 0 µ F| 1 µ F| 5 + 6 2)
1 ) At 0 °C to 4 0 °C (Accuracy Range)
2 ) Applies to measurements at film capacitors and battery operated :
3 ) Measurement of Capacitance will be more accurate after removal from device
undertest
4 ) With Zero Balancing
Square Wave Out
Output | Range | Accuracy |
---|---|---|
Frequency | 30Hz – 10kHz | 0.1% x output frequency + 2 counts of DMM display |
Duty Cycle | 10% – 100% 2) | 0.2% of Full scale 1) |
Amplitude | Fixed -3.15 to 3.15V | ±0.4V |
- For signal greater than 1kHz, add 0.2% per kHz to the accuracy
- In Multiple of 10
Influence Error
Influence Quantity| Range of Influence| Measured Quantity/
Measuring Range 1)| Variation ± (….%of rdg +. digits)/10k
---|---|---|---
Temperature| -10ºC to 21ºC & +25 °C to 50°C| VDC| 0.2 + 20
V~, VACDC| 0.4 + 10
600Ω to 600 kΩ| 0.5 + 10
600 kΩ| 1 + 10
mA/ADC| 0.6 + 10
mA/AAC, ACDC| 0.8 + 10
10nF…10µF| 1 + 5
100µF…1000µF| 1.5+10
Hz,%| 0.2 + 10
ºC/ºFpt100/pt1000| 0.5 + 10
ºC/ºF thermocouple K/J| 0.2 + 10
Relative humidity| 75% 3Days Meter off| V, A, Hz, %, Diode F, Ω| 1 × intrinsic error
Battery voltage| 1.8 to 3.6V| V, A, Hz, %, Diode,
F, Ω
| 1 × intrinsic error
- With Zero Balancing
Reference Condition for Accuracy
Reference Temperature | 23°C ± 1K, 73.4°F ± 1K |
---|---|
Relative Humidity | 45%…55% RH |
Waveform of measured quantity | Sinusoidal |
Input frequency | 45…65 Hz |
Battery Voltage | 3 V ± 0.1 V |
Influence Quantity
Influence Quantity| Range of Influence| Measuring Ranges|
Attenuation
---|---|---|---
Common Mode interference voltage| Noise quantity max. 1000 V dc| V dc| > 120
dB
Noise quantity max. 1000 V ~ 50-60 HZ sinusoidal| 6.0 V~, 60 V~| >80 dB
600 V~| > 70 dB
1000 V~| > 60 dB
Normal Mode interference ratio| Noise quantity V ~ Value of the measuring
range at a time
Max. 1000V~, 50Hz, 60Hz Sinusoidal| V dc| > 50dB
Noise quantity max. 1000 V dc| V~| >110dB
Applicable Regulations & Standards
EMC Immunity | IEC 61326-1:2012, Table A.1 |
---|---|
Immunity | IEC 61000-4-2 : 8 KV atmosphere discharge, 4 KV contact discharge |
IEC 61000-4-3 : 3 V/m
Safety| IEC 61010-1-2010
IP for water & dust| IEC 60529 : IP 50 For Instrument and IP20 for socket
Pollution degree:| 2
Installation category:| 1000 V CATIII / 600 V CATIV, 600V CATII for DMM 5003
High Voltage Test| 7.4 kV (IEC 61010-1-2010), 3.5kV For DMM 5003
Test & Procedure| IS 13875
Battery
Battery Voltage | 2 X 1.5 V Cells (LR6 Battery) |
---|---|
Battery type | Alkaline manganese cells. |
Battery Life | Appx. 100 Hrs. (Backlight off) |
Battery test | Automatic display of symbol when battery voltage drops below |
approx. 2.4V
Mechanical Design
Housing | PC ABS |
---|---|
Dimension | 200 x 91 x 54 mm |
Weight | Approx. 0.5 kg with batteries |
Crest Factor
Additional error caused by signal’s crest factor: 1 < CF < 3: 1% R+ 30D
3 < CF < 10: 3% R
Curve 1 : Range from 0.06V to 60V, 0.6mA to 60mA, 6A
Curve 2 : Range 600V/600mA
Curve 3 : Range 1000V/10A
Note: With Unknown Waveform (CF >2), measurement should be made with
manual range selection.
R = Reading
D = Digit
Internal Clock
Time Format | dd. MM. yy hh. mm. ss |
---|---|
Resolution | 1 s |
Accuracy | + 1min. per month |
Temperature Influence | 50 ppm/K |
Display
LCD display field 67 mm X 54 mm with digital display, analog scale and with
display of measurement unit, and Various special functions.
Analog
Display:| LCD scale with bar graph or pointer, depending on the selected
parameter setting
---|---
Scaling:| 2 bar/pointer corresponds to 2500 counts at the digital display
Over range Display (Digital):| By triangle “ ”
Polarity Display:| With automatic switching
Sample rate (Digital):| 10 measurements/sec and display refresh
Digital
Display: | 7-segment characters |
---|---|
Character Height: | Main Display – 12.88mm |
Sub Display – 7.37mm
Resolution:| 60,000 counts
Overflow Display:| “OL” is displayed
Polarity Display:| “-” (minus) is displayed if plus pole is connected to “ ⊥”
Measuring Rate:| 10 measurement/sec with the Min-Max function except for the
capacitance, frequency and duty cycle measuring Function
Refresh Rate:| 4 times/sec
Number of Digits:| 5
Fuse
Fuse | FF (UR) 16 A/ 1000 V AC/DC; 10 mm x 38 mm ( Simpson 5005 & 5006) |
---|
FF (UR) 1.6 A/ 1000 V AC/DC ; 6.3 mm x 32 mm ( Simpson 5002)
Switching Capacity| 30 kA at 1000 V AC/DC (Simpson 5005 & 5006)
10 kA at 1000 V AC/DC (Simpson 5002)
Accessories For Operation at a PC
Interface Adapter For USB Communication
Communication: | Bi-Directional |
---|---|
Baud Rate: | 9600 |
Data Bit: | 8 |
Stop Bit: | 1 |
Flow Control: | None |
A CD ROM is included which contains current drivers for Windows operating systems, Installation Guide, Datalogger User Manual and Datalogger Setup File.
Scope of Supply
Model Name | Scope of Supply |
---|
Simpson 5002
Simpson 5003
Simpson 5005
Simpson 5006| 1. Digital Meter
2. Cable Set
3. Protective Case
4. Battery
5. Operating Manual
6. Test Certificate
OPTIONAL ACCESSORIES
1. External Power Supply Adapter
2. USB Interface Adapter + Software CD
Simpson Electric Company
520 Simpson Avenue
Lac Du Flambeau, WI 54538-0099 (USA)
Web: www.simpsonelectric.com
Contact: +1 715-588-3311
Revision 2022/10