EU-45802-HPS Pressure
Injectable Arrowg+ard Blue Plus Two-Lumen CVC
Instructions
EU-45802-HPS Pressure Injectable Arrowg+ard Blue Plus Two-Lumen CVC
Pressure Injectable Arrowg+ard Blue Plus® Two-Lumen CVC
Contents
1: Arrowg+ard Blue Plus®1 Two-Lumen Catheter: 8 Fr. (2.8 mm OD) x 20 cm,
Pressure Injectable
1: Spring-Wire Guide, Nitinol, Marked: .032″ (0.81 mm) dia. x 23-5/8″ (60 cm)
(Straight Soft Tip on One End – “J” Tip on Other) with Arrow Advancer with ECG
Mark
1: Introducer Needle: Echogenic 18 Ga. x 2-1/2″ (6.35 cm) XTW
1: Syringe: 5 mL Luer-Slip
1: Tissue Dilator: 9 Fr. x 10.2 cm
2: Dust Cap: Non-Vented
1: SecondSite TPA Adjustable Hub: Fastener
1: SecondSite IPA Adjustable Hub: Catheter Clamp
1: SharpsAway® II Locking Disposal Cup
1: Safety Scalpel: #11 1Licensed under US Patent No. 7,329,412.
All components are CE 2797 unless otherwise noted.
Warning: Read all package insert warnings, precautions, and instructions
prior to use. Failure to do so may result in severe patient injury or death.
www.teleflex.com/IFU Not made with natural rubber latex.
Store below 25°C (77°F). Avoid excessive heat above 40°C (104°F).
Fluid path components are non-pyrogenic.
Contraindications: The Arrowg+ard Blue Plus antimicrobial catheter is
contraindicated for patients with known hypersensitivity to chlorhexidine,
silver sulfadiazine and/or sulfa drugs.
Lumen| Priming Volume* (mL)| Gravity Flow
Rate† (mL/hr)| MAX Pressure Injection Flow
Rate** (mL/sec)
---|---|---|---
Distal ( 14 G a.)
Proximal( 14 Ga.)| 0.73
0.75| 5354
3984| 10
10
- Priming volumes are approximate and are done without accessories.
Flow rate values are approximate and are determined using deionized water at 100 cm head height.
** Pressure injection flow rates are determined using media of 11.8 centipoise (cP) viscosity, with 152 cm administration set tubing.
EU Authorized
Representative and Importer:
Teleflex Medical IDA
Business and Technology Park
Dublin Road, Athlone
Co.
Westmeath, Ireland
Arrow International LLC
Subsidiary of Teleflex
Incorporated 3015
Carrington Mill Blvd. Morrisville, NC 27560 USA
www.teleflex.com
LBL062496 ROO (2021-10)
Pressure Injectable Arrowg+ard Blue Plus® Two-Lumen CVC
References
Read User Manual Online (PDF format)
Read User Manual Online (PDF format) >>