HACH SP-510 Hardness Analyzer User Manual

June 12, 2024
HACH

HACH SP-510 Hardness Analyzer

HACH-SP-510-Hardness-Analyzer-product

Product Information

Product Name: SP-510

Model Number: DOC023.53.80422

Edition: 13

User Manual Date: 03/2022

Specifications

Specifications are subject to change without notice.

General Specifications

  • Dimensions (W x D x H): 42 x 31.5 x 18 cm (16.5 x 12.5 x 7 in.)

Product Usage Instructions

Section 3 Installation

3.1 Install the instrument

Follow the provided installation instructions to properly install the instrument.

3.2 Plumb the instrument

Refer to the plumb instructions in the user manual to correctly plumb the instrument.

Section 5 Startup

5.1 Start the instrument

To start the instrument, follow these steps:

  1. Ensure the instrument is properly connected to a power source.
  2. Press the power button to turn on the instrument.

5.2 Prime the reagents (optional)

If required, follow the reagent priming instructions provided in the user manual to prime the reagents before use.

Section 6 Calibration

6.1 Complete a 2-point calibration

Perform a 2-point calibration using the instructions provided in the user manual to ensure accurate measurements.

Section 8 Maintenance

8.11 Replace the fuse

If the fuse needs to be replaced, refer to the instructions in the user manual for proper replacement steps.

Section 9 Troubleshooting

9.1 Troubleshooting for a hard reading

If you encounter difficulties in obtaining a reading, follow the troubleshooting steps outlined in the user manual to resolve the issue.

Section 10 Replacement parts and consumables

Refer to Section 10 of the user manual for information on replacement parts and consumables for the SP-510.

Product Specifications

Specifications are subject to change without notice.

General specifications

Specification Details
Dimensions (W x D x H) 42 x 31.5 x 18 cm (16.5 x 12.5 x 7 in.)
Enclosure rating IP62
Weight 11.3 kg (25 lbs)
Installation environment Indoor
Mount Wall
Pollution degree/installation category 2/II
Protection class I
Altitude Maximum 2000 m (6562 ft)
Light source Class 1 LED (light emitting diode) with peak wavelength of 610

nm. Estimated 50,000 hours minimum life; hard water and soft water indicator.
Detector| Silicon photo detector
Optical path length| 8.89 mm (0.35 in.)
Time delay| One reading above the trip point is necessary to trigger a soft- to-hard water alarm transition. One reading below the trip point is necessary to trigger a hard-to-soft water alarm transition.
Trip point| 0.3, 1, 2, 5, 10, 20, 50 and 100 mg/L. Refer to Trip point specifications on page 4.
Reagents| Buffer and indicator solution. 500 mL of each reagent every 2 months.
Reagent shelf life| 1 year
Power requirements| 115/230 VAC, 70 VA, 50/60 Hz, 1.25 A Fuse
Electrical connections| Two 3-wire barrier terminal blocks. Wire range: 18–12 AWG (0.75–1.0 mm2)
Hard water alarm relay| SPDT relay, set to on when the hard water indicator is on. The alarm can be disabled.
Contact rating| 5 A resistive at 100–240 VAC
Operating temperature| 5 to 40 °C (32 to 104 °F)
Storage temperature| -40 to 60 °C (-40 to 140 °F)
Operating humidity| Relative humidity: 5–95% at different temperatures, non- condensing
Sampling rate| New sample: every 1.9 minutes ±5% at 60 Hz; 2.3 minutes

±5% at 50 Hz

Specification Details
Sample flow rate to sample conditioning 50 to 500 mL/minute flow rate

necessary (250 mL/minute recommended)
Inlet pressure to instrument| 1 to 5 psig (0.07 to 0.34 bar), 1.5 psig (0.1 bar) is optimum, > 5 psig (0.34 bar) can cause sample tubing failure
Inlet pressure to sample conditioning| 1.5 psig to 75 psig (0.1 to 5.2 bar)
Inlet fitting| At instrument, 6.35 mm (0.25 in.) OD polyethylene tubing with quick-disconnect fitting
Drain fitting| Hose barb for 12.7 mm (0.5 in.) ID flexible tubing
Sample temperature range| 5 to 40 °C (41 to 104 °F)
Certifications| CE, cETLus
Warranty| 1 year (EU: 2 years)

Trip point specifications

Alarm trip point| Minimum trip value| Maximum trip value| Temperature influence on trip point
---|---|---|---
0.3 mg/L| 0.22 mg/L| 0.38 mg/L| –0.03 mg/L per °C
1 mg/L| 0.75 mg/L| 1.25 mg/L| –0.03 mg/L per °C
2 mg/L| 1.5 mg/L| 2.5 mg/L| –0.03 mg/L per °C
5 mg/L| 3.75 mg/L| 6.25 mg/L| –0.06 mg/L per °C
10 mg/L| 7.5 mg/L| 12.5 mg/L| –0.08 mg/L per °C
20 mg/L| 15.0 mg/L| 25.0 mg/L| –0.09 mg/L per °C
50 mg/L| 37.5 mg/L| 62.5 mg/L| –0.29 mg/L per °C
100 mg/L| 75.0 mg/L| 125.0 mg/L| –0.60 mg/L per °C

General information

In no event will the manufacturer be liable for direct, indirect, special, incidental or consequential damages resulting from any defect or omission in this manual. The manufacturer reserves the right to make changes in this manual and the products it describes at any time, without notice or obligation. Revised editions are found on the manufacturer’s website.

Safety information
The manufacturer is not responsible for any damages due to misapplication or misuse of this product including, without limitation, direct, incidental and consequential damages, and disclaims such damages to the full extent permitted under applicable law. The user is soley responsible to identify critical application risks and install appropriate mechanisms to protect processes during a possible equipment malfunction.
Please read this entire manual before unpacking, setting up or operating this equipment. Pay attention to all danger and caution statements. Failure to do so could result in serious injury to the operator or damage to the equipment.
Make sure that the protection provided by this equipment is not impaired. Do not use or install this equipment in any manner other than that specified in this manual.

Use of hazard information

D A N G E R
Indicates a potentially or imminently hazardous situation which, if not avoided, will result in death or serious injury.

W A R N I N G
Indicates a potentially or imminently hazardous situation which, if not avoided, could result in death or serious injury.

C A U T I O N
Indicates a potentially hazardous situation that may result in minor or moderate injury.

NOT I C E
Indicates a situation which, if not avoided, may cause damage to the instrument. Information that requires special emphasis.

Precautionary labels
Read all labels and tags attached to the instrument. Personal injury or damage to the instrument could occur if not observed. A symbol, if noted on the instrument, will be included with a danger or caution statement in the manual.

| This is the safety alert symbol. Obey all safety messages that follow this symbol to avoid potential injury. If on the instrument, refer to the instruction manual for operation or safety information.
---|---
| Electrical equipment marked with this symbol may not be disposed of in European domestic or public disposal systems. Return old or end-of-life equipment to the manufacturer for disposal at no charge to the user.
| This symbol indicates the need for protective eye wear.
| This symbol identifies a risk of chemical harm and indicates that only individuals qualified and trained to work with chemicals should handle chemicals or perform maintenance on chemical delivery systems associated with the equipment.
| This symbol indicates that a risk of electrical shock and/or electrocution exists.
| This symbol indicates that a risk of fire is present.
| This symbol, when noted on the product, identifies the location of a fuse or current limiting device.

Compliance and certification

CAUTION
This equipment is not intended for use in residential environments and may not provide adequate protection to radio reception in such environments.

Canadian Radio Interference-Causing Equipment Regulation, ICES-003, Class A:
Supporting test records reside with the manufacturer.
This Class A digital apparatus meets all requirements of the Canadian Interference-Causing Equipment Regulations.
Cet appareil numérique de classe A répond à toutes les exigences de la réglementation canadienne sur les équipements provoquant des interférences.

FCC Part 15, Class “A” Limits
Supporting test records reside with the manufacturer. The device complies with Part 15 of the FCC Rules. Operation is subject to the following conditions:

  1. The equipment may not cause harmful interference.
  2. The equipment must accept any interference received, including interference that may cause undesired operation.

Changes or modifications to this equipment not expressly approved by the party responsible for compliance could void the user’s authority to operate the equipment. This equipment has been tested and found to comply with the limits for a Class A digital device, pursuant to Part 15 of the FCC rules. These limits are designed to provide reasonable protection against harmful interference when the equipment is operated in a commercial environment. This equipment generates, uses and can radiate radio frequency energy and, if not installed and used in accordance with the instruction manual, may cause harmful interference to radio communications. Operation of this equipment in a residential area is likely to cause harmful interference, in which case the user will be required to correct the interference at their expense. The following techniques can be used to reduce interference problems:

  1. Disconnect the equipment from its power source to verify that it is or is not the source of the interference.
  2. If the equipment is connected to the same outlet as the device experiencing interference, connect the equipment to a different outlet.
  3. Move the equipment away from the device receiving the interference.
  4. Reposition the receiving antenna for the device receiving the interference.
  5. Try combinations of the above.

Product overview

D A N G E R

Chemical or biological hazards. If this instrument is used to monitor a treatment process and/or chemical feed system for which there are regulatory limits and monitoring requirements related to public health, public safety, food or beverage manufacture or processing, it is the responsibility of the user of this instrument to know and abide by any applicable regulation and to have sufficient and appropriate mechanisms in place for compliance with applicable regulations in the event of malfunction of the instrument.

The SP-510™ Hardness Monitor continuously measures water softener levels to find hardness breakthroughs based on softener exhaustion. The monitor is used in commercial and industrial water applications. The monitor has an automatic control system that starts regeneration sequences with the alarm circuit.
Hardness is measured at different levels as CaCO3 with the applicable indicator and reagent solutions. A relay closes when the instrument measures a hardness value that is more than the trip point. Refer to Figure 1 for the product overview.

Figure 1 Product overviewHACH-SP-510-Hardness-Analyzer-fig
8

1 Power switch 7 Keypad
2 Power access port 8 Pump/valve module
3 Relay and alarm contact access port 9 Indicator and reagent

bottles
4 Air purge (optional)| 10 Sample inlet
5 Access cover| 11 Enclosure drain
6 Colorimeter| 12 Sample drain

Product components
Make sure that all components have been received. Refer to Figure 2. If any items are missing or damaged, contact the manufacturer or a sales representative immediately.

Figure 2 Product componentsHACH-SP-510-Hardness-Analyzer-fig
9

1 SP-510 Hardness Monitor 5 Indicator solution
2 Installation kit1 6 Magnesium sulfate solution
3 Maintenance kit2 7 TitraVer® (EDTA) hardness titrant

4 Buffer solution

Installation

D A N G E R
Electrocution hazard. Always remove power to the instrument before making electrical connections.
D A N G E R
Multiple hazards. Only qualified personnel must conduct the tasks described in this section of the document.

Install the instrument
Install this instrument on an indoor wall, away from direct sunlight.

  • For a complete purge of the sample line during each cycle, install the instrument as close to the sample point as possible.
  • Leave sufficient clearance at the bottom and sides for tubing and cable connections. Refer to Figure 3 and Figure 4.
  1. Refer to the installation kit documentation for more information.
  2. Refer to the maintenance kit documentation for more information.

Figure 3 Dimensions for wall installationHACH-SP-510-Hardness-Analyzer-
fig 10

Figure 4 Door clearance

Plumb the instrument

C A U T I O N
Fire hazard. This product is not designed for use with flammable liquids.
C A U T I O N
Chemical hazard. If there is a leak in the fluid system, hazardous substances may leak out of the lower enclosure. Put the supplied reagent bottle tray or a bucket under the drain to catch any spills.
C A U T I O N
Chemical exposure hazard. Dispose of chemicals and wastes in accordance with local, regional and national regulations.

Use quick-connect type connectors for ¼-inch OD tubes to install the sample inlet and drain connections.

  1. Install 3 feet of tubing in the ½-inch enclosure drain to keep dust out of the analyzer. Refer to Figure 1 on page 7, item 11.
    Note: Make sure that the enclosure drain is open. The drain must be open to remove sample water if leaks occur.

  2. Push the ¼-in OD poly tube on the sample line connector. The tube will “stop” two times as it is pushed on the connector. For more information, refer to the installation kit documentation.

  3. Make sure that the tubes are pushed completely on the connector. Incorrect installation can cause the tube to come off of the connector when water pressure is applied.
    Note: The sample drain connector will install on ½-inch ID flexible tubes (customer-supplied).

Connect the air purge (optional)
An air purge is necessary in an environment with high humidity or caustic vapors.
Note: Use only dry, oil-free instrument air at 0.42 m3 (15 standard cubic feet) per hour (scfh).

  1. Find the air purge connection on the left side of the instrument enclosure. Refer to Figure 1 on page 7.
  2. Remove the plug in the quick-connect fitting.
  3. Push the ¼-inch poly tubing into the fitting. Make sure to have two different “stops” so that the tubing cannot disconnect when air pressure is applied.

Sample line guidelines
Select a good, representative sampling point for the best instrument performance. The sample must be representative of the entire system.
To prevent erratic readings:

  • Collect samples from locations that are sufficiently distant from points of chemical additions to the process stream.
  • Make sure that the samples are sufficiently mixed.
  • Make sure that all chemical reactions are complete.

Connect the sample stream
Install the sample line into the center of a larger process pipe to minimize interference from air bubbles or bottom sediment. Figure 5 shows examples of good and bad installation.
Keep the sample lines as short as possible to prevent the accumulation of bottom sediment. The sediment can absorb some of the analyte from the sample and cause low readings. The sediment can later release the analyte and cause high readings. This exchange with the sediment also causes a delayed response when the analyte concentration in the sample increases or decreases.

Figure 5 Sampling methodsHACH-SP-510-Hardness-Analyzer-fig
12

1 Air 2 Sample flow

Install the pump valve pressure plate
The pressure plate and screws are supplied with the installation kit. Remove the tape from the pump tubes before installation.
Make sure to turn the screws in small increments and move from one screw to the other so that the plate is pulled down equally. Do not overtighten. To install the pressure plate, refer to Figure 6.

Figure 6 Install the pressure plateHACH-SP-510-Hardness-Analyzer-fig
13

Install the sample conditioning kit
The sample conditioning kit is supplied with the instrument. The kit removes large particles from the sample stream with a 40-mesh strainer. The ball valve on the raw sample inlet controls the flow to the filter. The ball valve on the instrument supply line controls the flow rate of filtered sample to the instrument. For complete sample conditioning installation instructions, refer to the installation kit documentation. Refer to Figure 7.

Figure 7 Sample flow through the conditioning kitHACH-SP-510-Hardness-
Analyzer-fig 14

1| Sample flow| 4 Bypass tee, unfiltered sample| 7 Unfiltered- sample ball valve (shown open)
---|---|---|---
2| Bypass flow| 5 Flow observation point| 8| Low-flow valve option
3| Drainage flow| 6 Filtered-sample bypass ball valve (shown open)| 9 High-flow valve option

Electrical installation

D A N G E R

Electrocution hazard. Always remove power to the instrument before making electrical connections.

Remove the access cover
Remove the access cover to connect to the wiring terminals. Refer to Figure 8.

Figure 8 Access cover removalHACH-SP-510-Hardness-Analyzer-fig
15

Electrical connections for conduit

D A N G E R
Electrocution hazard. Use only fittings that have the specified environmental enclosure rating. Obey the requirements in the Specifications section.
D A N G E R
Electrocution hazard. If this equipment is used outdoors or in potentially wet locations, a Ground Fault Circuit Interrupt (GFCI/GFI) device must be used for connecting the equipment to its main power source.
W A R N I N G
Electrocution hazard. Install a 10 A circuit breaker for mains power. Identify the circuit breaker with a label, as a local disconnect for this equipment.

The instrument has plugs in all of the conduit holes. To keep the IP62 environmental rating, use liquid-tight sealing-type conduit fittings and cord strain reliefs. Connect the equipment in accordance with local, state or national electrical codes. If connections are not made through a conduit opening, install a liquid-tight seal in place of the plugs. Refer to Figure 9.

Figure 9 Power connectionsHACH-SP-510-Hardness-Analyzer-fig
16

1 Conduit (preferred) 2 Power cord (optional)

Wiring for power
To connect the instrument to power, refer to Figure 10 and Table 1.

Figure 10 Power connectionHACH-SP-510-Hardness-Analyzer-fig
17

1 Voltage switch (in 115 V position) 3 AC power connector
2 Voltage switch (in 230 V position) 4 Fuses (F1 and F2)

Table 1 Terminal wiring

Wire color code| Protective earth ground| Hot or Ø1| Neutral or Ø2
---|---|---|---
North America| Green| Black| White
International electrotechnical commission (IEC)| Green with yellow line| Brown| Blue

Select the voltage
NOTICE
To prevent serious damage to the instrument, make sure that the line voltage is set correctly. Refer to Figure 10 on page 14.
The instrument is set to 115-volt operation at the factory. To change the line voltage to 230-volt operation, slide the voltage switch to the 230 V position. Refer to Figure 10 on page 14.
Note: The fuses are approved for North American and European use and do not need to be changed with the voltage.

Connect the alarm relays

C A U T I O N
Fire hazard. Relay loads must be resistive. Always limit current to the relays with an external fuse or breaker. Obey the relay ratings in the Specifications section.
The current to the relay contacts must be 5 A or less. Make sure to have a second switch available to remove power from the relays locally in case of an emergency or for maintenance. Power can be removed with an external switch and a 5 A fuse or with a switched 5 A circuit breaker.

Figure 11 shows the alarm relay contacts connected to the terminal strip with normally open and normally closed terminations. Terminals are unpowered and rated for 5 A at 100–240 VAC resistive load.
The relay connector accepts 18–12 AWG (0.75–1.0 mm2) wire. Select the necessary wire gage that operates with the application. A wire gauge less than 18 AWG (0.75 mm2) is not recommended.

Figure 11 Alarm connectionsHACH-SP-510-Hardness-Analyzer-fig
18

Table 2 Relay wiring

Terminal block Terminal 1 Terminal 2 Terminal 3
J7 COM Normally open (NO) Normally closed (NC)

Install the buffer solution
Buffer solutions are formulated at the factory and are ready to install without preparation. Put the bottle in the instrument as shown in Figure 12 on page 16. Refer to Buffer and indicator solution for the applicable trip point on page 24.

Install the indicator solution
Indicator solutions are formulated at the factory and are ready to install without preparation. Put the bottle in the instrument as shown in Figure 12. Refer to Buffer and indicator solution for the applicable trip point on page 24.

Figure 12 Install the buffer and indicator reagentsHACH-SP-510-Hardness-
Analyzer-fig 19

1 Indicator reagent tube label and bottle| 2 Buffer tube label and bottle
---|---

Install the stir bar
A stir bar is included in the installation kit. Install the stir bar in the colorimeter sample cell for correct operation. Refer to Figure 13.

Figure 13 Install the stir barHACH-SP-510-Hardness-Analyzer-fig
20

User interface and navigation

Keypad description
Refer to Figure 14 and to Table 3 for the keypad description and navigation information.

Figure 14 SP-510 keypad

Table 3 Keypad description

Key Function Description
1 HARD LED The monitor found hardness in the water. The status indicator

light is red.
2| SOFT LED| The monitor is in operation and no hardness shows. The status indicator light is green.
3| ALARM OFF| Sets the alarm to off. When more cycles are necessary to examine the water, set the alarm to off. The status indicator light is yellow.
4| HARD CAL| Completes a hard calibration. Refer to Calibration on page 18. The status indicator light is yellow.
5| ZERO CAL| Completes a zero-point calibration. Refer to Calibration on page 18. The status indicator light is yellow.

Startup

Start the instrument
Note: Install the pressure plate correctly to prevent sample flow directly into the reagents. Refer to Install the pump valve pressure plate on page 11.

  1. Open the supply valve to the instrument.

  2. Let the pressure in the tubes stabilize.
    Note: If leaks occur under pressure, examine all connections. Secure connections until all leaks have stopped.

  3. Set the power to on.

  4. Operate the instrument for approximately 2 hours with the sample and reagents.
    Note: The sample cell will have bubbles on the surface until it is fully wetted. The bubbles can cause irregular readings until the sample flow stabilizes.

Prime the reagents (optional)
Prime the reagents to decrease the start up time to 15 minutes.

  1. Set the sample flow to the instrument to off.
  2. Disconnect the reagent feed line after the “Y” fitting at the colorimeter.
  3. Connect a syringe to the tubing that comes from the “Y” fitting.
  4. Loosen the two screws on the pump pinch plate so that the reagents can be pulled through the reagent lines.
  5. Use the syringe to remove the air and fill the lines with reagent.
  6. Tighten the screws on the pump pinch plate. Do not overtighten.
  7. Remove the syringe from the tubing and connect the fitting on the colorimeter.

Calibration

C A U T I O N

Chemical exposure hazard. Obey laboratory safety procedures and wear all of the personal protective equipment appropriate to the chemicals that are handled. Refer to the current safety data sheets (MSDS/SDS) for safety protocols.

Complete a 2-point calibration
Complete a calibration after reagent installation or when an optical system component is replaced. The standardization procedure makes the sample first hard, then soft.

  1. Remove the plug from the top port of the colorimeter. Refer to Figure 13 on page 16.

  2. Push HARD CAL. The hard cal LED is on continuously.

  3. When the hard cal LED flashes, add two drops of Magnesium Standard Solution into the colorimeter.

  4. When the LED stops flashing and is on continuously, wait for the cycle to complete.
    At the end of the cycle, the LED sets to off to show a successful calibration.

  5. HARD CAL failure: When the LED starts to flash, push HARD CAL to confirm a hard cal failure. Do steps 2–4 again.

  6. Push ZERO CAL. The zero cal LED is on continuously.

  7. When the zero cal LED flashes, add two drops of EDTA Solution, 10 g/L, into the colorimeter.

  8. When the LED stops flashing and is on continuously, wait for the cycle to complete.
    At the end of the cycle, the LED is set to off to show a successful calibration.

  9. ZERO CAL failure: When the LED starts to flash, push ZERO CAL to confirm a zero cal failure. Do steps 1–7 again.

  10. Replace the plug in the top port of the colorimeter.

Operation

After the instrument power is set to on, the SOFT LED flashes until the first reading cycle is completed. The cycle takes 2 minutes. Then, the SOFT LED shows a stable green light.
The HARD LED comes on with a red light when the sample reagent mixture measurement is higher than the trip point for one measurement. Push ALARM OFF to set the alarm relay to off. The ALARM OFF light will show a yellow light.
The instrument operates maintenance-free for a minimum of 2 months. Periodically, do a visual check to see if bubbles are in the tubing. If there are bubbles, look for an air leak. Refer to Maintenance schedule on page 19 for maintenance tasks.

Maintenance

D A N G E R

Multiple hazards. Only qualified personnel must conduct the tasks described in this section of the document.

Maintenance schedule
Table 4 shows the recommended schedule of maintenance tasks. Facility requirements and operating conditions can increase the frequency of some tasks.

Table 4 Maintenance schedule

Task| 2 months| 3 months| 6 months| 1 year| As necessary
---|---|---|---|---|---
Replace the reagent on page 19| x| | | |
Replace the pump tubes on page 20| | x3| x4| |
Replace the tubing on page 20| | | | x|
Clean the colorimeter on page 20| | | | x|
Replace the sample conditioning filter on page 20| | | | x|
Replace the fuse on page 21| | | | | x

Clean the instrument
Clean the exterior of the instrument with a moist cloth and a mild soap solution and then wipe the instrument dry as necessary.

Clean the instrument interior

C A U T I O N

Chemical exposure hazard. Obey laboratory safety procedures and wear all of the personal protective equipment appropriate to the chemicals that are handled. Refer to the current safety data sheets (MSDS/SDS) for safety protocols.

Refer to the MSDS for instructions to safely clean reagent spills and leaks. Obey all local and federal regulations to dispose of cleaning materials.

Clean spills

C A U T I O N

Chemical exposure hazard. Dispose of chemicals and wastes in accordance with local, regional and national regulations.

  1. Obey all facility safety protocols for spill control.
  2. Discard the waste according to applicable regulations.

Replace the reagent

C A U T I O N

Chemical exposure hazard. Dispose of chemicals and wastes in accordance with local, regional and national regulations.

Note: Do not pour the remaining reagent into new bottles.
One 500-mL bottle of each buffer and indicator solution can be used for approximately 2 months.

Ambient operating temperatures more than 27 °C (80.6 °F)
Ambient operating temperatures less than 27 °C (80.6 °F)

  1. Discard the old containers with remaining contents in compliance with MSDS and regulatory requirements.
  2. Install the new bottles. Refer to Install the buffer solution on page 15 and Install the indicator solution on page 15.

Change the alarm trip point
Install the applicable buffer and indicator solution to change the alarm trip point to a different hardness level. Let the instrument operate for 2 hours to purge the used reagents. Calibrate the instrument again. Refer to Trip point specifications on page 4 and to Calibration on page 18.

Replace the pump tubes
Over a period of time, the clamping action of the pump/valve module will soften the tubing. This causes the tubing to break and prevents liquid flow. Leaks can occur. The tubing wear is increased at high temperatures. Therefore, the pump tube replacement schedules are based on the ambient operating temperature. Refer to Maintenance schedule on page 19 for the maintenance interval. Refer to the documentation supplied with the maintenance kit.

Replace the tubing
The manufacturer recommends that one tube is replaced at a time. Refer to the documentation supplied with the maintenance kit.

Clean the colorimeter

W A R N I N G

Chemical exposure hazard. Obey laboratory safety procedures and wear all of the personal protective equipment appropriate to the chemicals that are handled. Refer to the current safety data sheets (MSDS/SDS) for safety protocols.

The colorimeter measuring cell can collect sediment or cause layers to grow on the inside walls. Clean the cell with an acid solution and a cotton swab on a monthly basis.
Items to collect:

  • Sulfuric acid standard solution, 19.2 N, 100 mL MDB
  • Wooden or paper cotton-tipped swab
  • Bent steel paper clip
  • Stir bar (optional)
  1. Remove the plug from the top port of the colorimeter. Refer to Figure 13 on page 16.

  2. Remove the magnetic stir bar with a bent steel paper clip as a retrieval tool.

  3. Use the pipette to add 2–3 drops of 19.2 N sulfuric acid standard solution in the colorimeter. Note: Lower normality sulfuric acid is not sufficient to clean the colorimeter.

  4. Let the sulfuric acid stay in the colorimeter for 15 minutes.

  5. Use a wooden or cotton-tipped swab to clean the interior surfaces of the colorimeter cell. Move the cotton-tipped swap gently up and down.
    Note: Do not use a plastic swab to clean the colorimeter with sulfuric acid. The acid will dissolve the plastic.

  6. Make sure that the measurement cell is dry.

  7. Clean the stir bar with water or alcohol or replace the stir bar.

  8. Replace the plug in the top port of the colorimeter.

Replace the sample conditioning filter
Examine the sample conditioning filter regularly when lots of solids are in the sample. Replace the sample conditioning filter if necessary. Refer to Maintenance schedule on page 19 for the maintenance interval. Refer to the documentation supplied with the installation kit.

Replace the fuse

D A N G E R
Electrocution hazard. Remove power from the instrument before doing maintenance or service activities.

D A N G E R
Fire hazard. Use the same type and current rating to replace fuses.

Remove the two fuses (F1 and F2) and replace them with two new fuses with the same specifications, T, 1.25 A, 250 V. The same fuse rating is used for the 115 V and for the 230 V operation. Refer to Figure 10 on page 14.

Troubleshooting

Problem Possible cause Solution
The SOFT LED light is not on and the pump motor does not operate. No power

Make sure that the power switch is set to on.

•   Connect the power cord.

•   Replace the fuses if necessary.

The SOFT LED light is not on and the pump motor does operate.| Problem with the power supply| Replace the main circuit board. Contact technical support.
The SOFT LED light is on but the pump motor does not operate.| Operation power is low| Make sure that the line voltage is within specification.
Voltage selector setting is not correct| Set the line voltage selector switch to the correct voltage.
Motor cable is not connected| Connect the motor cable connection.
Motor does not operate| Replace the motor.
The reading is continually high. The HARD LED is on.| •   No stir bar

•   LED is not connected

•   No sample flow

•   Out of reagents

| Refer to Troubleshooting for a hard reading on page 22. If the problem continues, cycle power to the instrument, examine all power connections and fuses, replace the circuit board or contact technical support.
Problem| Possible cause| Solution
---|---|---
The HARD LED is flashing.| The instrument could not save the calibration information.| Contact technical support.
The instrument could not save the disabled alarm status.
The SOFT LED is flashing continuously for more than

5 minutes after the power is set to on.

| The optical path has a blockage or the optical switch does not work.| • Let the instrument complete the cycle.

•   Make sure that the motor operates.

•   Set the power to off and then to on and wait for 5 minutes.

•   Contact technical support.

Troubleshooting for a hard reading
When the red LED is on continuously and the instrument measures hard water, do the steps that follow.

Cause Solution

If the flow is too low, the sample cell does not flush completely all the color out of the colorimeter. This causes a zero reading. If the flow is too high, some of the water will bypass the pinch block and cause the color to be diluted.| Make sure to set the flow rate to 200 mL/minute.
The sample line is pinched off in the pinch block.| •   Pull the sample line from the colorimeter. Water that is already in the line could leak out. Examine and adjust the flow rate when a steady water stream comes out while the line is pinched. Set the flow rate to 200 mL/minute.

•   Make sure that the pressure plate of the pinch block is not overtightened.

•   Examine the back of the pinch blocks and make sure that they do not have grooves.

No stir bar is installed or more than one stir bar is installed.| •   Make sure that a stir bar is installed.

•   Use a paper clip to retrieve the stir bar from the colorimeter and make sure that multiple stir bars are not installed.

•   If the stir bar does not operate, mix the sample manually after the reagents have been added. Use the end of a wooden Q-tip and mix the reagents approximately for 10 seconds. If the instrument gets a reading, the mixing is not sufficient. Replace the stir coil. Refer to Replacement parts and consumables on page 23.

Cause Solution
The instrument does not receive any buffer or indicator solution. Remove the

two reagent lines from the Y connector. Only one drop of reagent comes out per cycle. If no reagent comes out, examine the pressure plate. Make sure that the pinch block is not overtightened and the reagent lines are not pinched. Examine the back of the pinch block for grooves. Replace the pinch block if necessary.
The chemistry does not operate correctly.| Mix 1-mL of each reagent and 80-mL of sample. A color change must show. If no color change shows, replace the reagents.
The instrument does not operate correctly.| Pinch the sample inlet line to stop the sample flow. The next cycle reads soft. If it does not read soft, examine the LED if it shows an orange light. Examine the photocell and clean if necessary. After this cycle, the instrument reads hard again.

Replacement parts and consumables

W A R N I N G
Personal injury hazard. Use of non-approved parts may cause personal injury, damage to the instrument or equipment malfunction. The replacement parts in this section are approved by the manufacturer.
Note: Product and Article numbers may vary for some selling regions. Contact the appropriate distributor or refer to the company website for contact information.
Refer to the installation kit and maintenance kit documentation for more information.

Replacement parts

Description Item no.
Cam/coupler assembly 5410500
Circuit board assembly 5694800
Colorimeter flow-cell assembly 5516200
Colorimeter drain plug 5103600
Follower block, reagent 4274100
Follower block, sample 4274200
Fuse (T, 1.25 A, 250 V) UL/CSA/CE accepted 5516700
LED 4350800
LED housing, lower 5412700
LED housing, upper 5412800
Maintenance kit 5516500
Installation kit 5516400
Motor assembly 5411900
Photo detector 5411300
Pressure plate 5411800
Stir bar, micro, 3 x 8 mm 5412900
Description Item no.
--- ---
Stir coil 5411100
Thumb screw to hold pressure plate (2x) 5410100
Tubing, 0.16 cm (0.0625 in.) ID, white 4271700
Tubing, 0.11 cm (0.043 in.) ID, brown 5412100
Y fitting/strainer, 40 mesh 4661600

Accessories

Description Item no.
Seal, oil-tight 4221000
Power cord kit for 115 V North American operation 4630600
Power cord kit for 230 V European operation 4630800

Consumables

Description Quantity Item no.
EDTA solution, dropping bottle 10 g/L, 1 oz. 102133
Magnesium standard solution, dropping bottle 10 g/L, 1 oz. 102233
SP510 Hardness Monitor indicator solution for 25 ppm 500 mL 2502749
SP510 Hardness Monitor indicator solution for 50 ppm 500 mL 2502849
SP510 Hardness Monitor indicator solution for 100 ppm 500 mL 2502949
Sulfuric acid standard solution, 19.2 N, MDB 100 mL 203832

Buffer and indicator solution for the applicable trip point

Trip point Buffer 5 item no. Indicator item no.
0.3 mg/L 2768549 2794649
1 mg/L 2768549 2769049
2 mg/L 2768549 2769149
5 mg/L 2768549 2769249
10 mg/L 2768649 2769249
20 mg/L 2768749 2769249
50 mg/L 2768849 2769249
100 mg/L 2768949 2769249

HACH COMPANY World Headquarters

HACH LANGE GMBH

HACH LANGE Sàrl

  • 6, route de Compois
  • 1222 Vésenaz
  • SWITZERLAND
  • Tel. +41 22 594 6400
  • Fax +41 22 594 6499

© Hach Company/Hach Lange GmbH, 2013, 2014, 2016–2018, 2020–2022. All rights reserved.

References

Read User Manual Online (PDF format)

Loading......

Download This Manual (PDF format)

Download this manual  >>

Related Manuals